Multi-center clinical field study; CLIA-waived point-of-care clinical sites
The clinical field study evaluated the performance of the Sofia Strep A+ FIA in a real-world point-of-care setting by comparing test results from symptomatic patients against standard bacterial culture and an FDA-cleared molecular assay.
Point-of-care; CLIA-waived; Clinical field study; Diagnostic performance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Multi-center clinical field study; Prospective clinical field trial; Study Period: 2014
851 patients with symptoms suggestive of bacterial pharyngitis; Sample Size: 851; Number of Sites: 7
Standard bacterial culture and FDA-cleared Group A Streptococcus RT-PCR assay
Sensitivity, Specificity, PPV, NPV
Indications for Use
The Sofia® Strep A+ FIA detects Group A Streptococcal antigens from throat swabs from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection.
Device Story
Sofia® Strep A+ FIA is an immunofluorescence-based lateral flow assay for qualitative detection of Group A Streptococcus (GAS) antigens. Input: throat swab specimen applied to a 2-D barcoded cassette. Operation: specimen is chemically lysed and wicked through a nitrocellulose strip containing europium chelate-impregnated microparticles coated with anti-GAS antibodies; GAS antigens bind to these particles and are captured by secondary antibodies on the strip. The Sofia® analyzer scans the strip using a 365 nm LED and filtered photodiode to detect fluorescent signals (Stokes shift 365 to 618 nm). Output: analyzer processes signals via an assay-specific algorithm to provide a positive, negative, or invalid result on the screen. Used in professional/laboratory settings by healthcare personnel. The device provides objective, automated interpretation of results, aiding in clinical diagnosis of GAS infection. Benefits include rapid, objective detection of GAS antigens, reducing reliance on subjective visual interpretation of lateral flow tests.
Clinical Evidence
Prospective clinical study (n=851) across 7 U.S. sites compared Sofia® Strep A+ FIA results to bacterial culture. Sensitivity was 93.7% (95% CI: 89.1%-96.5%) and specificity was 94.4% (95% CI: 92.4%-95.9%). Discordant analysis performed using an FDA-cleared molecular device. Analytical studies included precision, reproducibility, inclusivity (21 strains), and cross-reactivity/interference (88 organisms). Bench testing confirmed performance stability across temperatures and inter-analyzer consistency.
Technological Characteristics
Lateral-flow immunofluorescence assay. Uses polyclonal rabbit antibodies and Europium chelate-dyed polystyrene microparticles. Reagent solution contains sodium nitrite and hydrochloric acid. Analyzed by Sofia instrument using method-specific algorithms. Qualitative output. Room temperature storage. Built-in procedural control zone and automated cartridge expiration/insertion checks.
Indications for Use
Indicated for symptomatic patients (e.g., sore throat) suspected of having pharyngitis to detect Group A Streptococcal antigens in throat swab specimens.
Regulatory Classification
Identification
Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identify Streptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.
Predicate Devices
Sofia® Strep A+ FIA with Sofia® (k123793)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
K141775
B. Purpose for Submission:
To obtain a substantial equivalence determination for the Sofia® Strep A+ FIA.
C. Measurand:
Group A β-hemolytic Streptococcus (GAS; Streptococcus pyogenes) antigens in throat swab specimens.
D. Type of Test:
The Sofia® Strep A+ FIA is an immunofluorescence-based lateral flow *in vitro* diagnostic test for the qualitative detection of GAS antigens isolated from throat swab specimens obtained from symptomatic patients.
E. Applicant:
Quidel Corporation
F. Proprietary and Established Names:
Sofia® Strep A+ FIA and Sofia®
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3740 - Streptococcus spp. Serological Reagents
2. Classification:
Class I
3. Product code:
GTY – Antigens, Streptococcus spp., All Groups
{2}
KHO – Fluorometer, for Clinical Use
4. Panel:
83- Microbiology
H. Intended Use:
1. Intended use(s):
The Sofia® Strep A+ FIA detects Group A Streptococcal antigens from throat swabs from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection.
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
Sofia®
I. Device Description:
The Sofia® Strep A+ FIA is a lateral flow immunofluorescence-based assay encased in a 2-D barcoded cassette. Antigen-extracted specimen is applied to the sample pad through a sample port, which is wicked through photoresistant europium chelate-impregnated polystyrene microparticles coated with polyclonal primary antibodies to Streptococcus pyogenes antigens, then through a localized stripe of secondary antibodies (Strep A Line), followed by a reference stripe that acts as a procedural control zone and reference line. The test is designed to reduce autofluorescence background by using a large Stokes shift from 365 to 618 nm. Once the barcode is scanned with the external barcode analyzer, the cassette is uni-directionally inserted into an analyzer drawer in the front of the apparatus. The Sofia® analyzer scans the region of the test strip containing the Strep A and Reference lines in the result window to yield three possible results: 1) Positive for Strep A, 2) Negative for Strep A, 3) Invalid. Europium chelate is detected by a scanning optics unit, exciting by a 365 nm filtered LED (Light Emitting Diode) and captured by a filtered photodiode, converted from analog RFU (Relative Fluorescence Units) data to digital.
2
{3}
A positive result for the analyte is determined by detection and analysis of the fluorescent signal at the test and reference lines, which are processed by an assay-specific algorithm. The algorithm employs a smoothing algorithm to the data, identifies peak maxima, minima and width, then calculates the RFU value based on peak height for the Strep A test line. The Limit of Blank (LoB) is 1948 RFU. When the test line value is $\geq 1948$ RFU, the test result is positive; when the value is $< 1948$ RFU, the test results is negative. The procedural control zone cutoff for a valid versus invalid result is 30,000 RFU. If internal controls fail at any point, the result is invalid and an error code is presented. Results are presented on a screen and can be printed on an integrated printer (optional). The Sofia® analyzer has the capacity to operate in Read-Now Mode and Walk-Away Mode; however Read-Now Mode functionality has been disabled for this the Sofia® Strep A+ FIA.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Sofia® Strep A+ FIA with Sofia® (k123793)
2. Predicate 510(k) number(s):
k123793
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Sofia® Strep A+ FIA with Sofia® (k141775) | Sofia® Strep A FIA with Sofia® (k123793) |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Sofia® Strep A+ FIA with Sofia® (k141775) | Sofia® Strep A FIA with Sofia® (k123793) |
| **Intended Use** | The Sofia® Strep A+ FIA detects Group A Streptococcal antigens from throat swabs from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection. | The Sofia® Strep A FIA employs immunofluorescence technology to detect Group A Streptococcal antigens from throat swabs of symptomatic patients. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection. |
| **Instrument** | Sofia® | Same |
| **Analyte** | Group A Streptococcal antigen | Same |
| **Qualitative** | Yes | Same |
| **Automated Analysis** | Yes | Same |
| **Read Results** | Read results on instrument screen or print with optional printer | Same |
| **Calibrator** | Yes – Calibration Cassette and QC Card provided | Same |
| **Read Result Time** | 5 Minutes | Same |
| **Specimen Type** | Throat swab | Same |
| **Test Principle** | Immunofluorescence Device | Same |
| **Format** | Lateral-flow Test Cassette | Same |
| **Antibodies Used** | Polyclonal rabbit antibodies that are specific to Group A Streptococcus carbohydrate antigen | Same |
| **Detection Particle** | Polystyrene microparticles dyed with Europium chelate | Same |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Sofia® Strep A+ FIA with Sofia® (k141775) | Sofia® Strep A FIA with Sofia® (k123793) |
| Transfer Device | Fixed volume pipette used to transfer patient sample mixed with reagent into Test Cassette | Same |
| External Controls | Test kit contains Positive and Negative Control Swabs | Same |
| Quality Control Features | Built-in features include: • Built-in procedural control zone scanned by the analyzer to determine whether adequate flow occurred • Analyzer prevents used or expired cartridge from being read by the reader • Cassette properly inserted | Same |
| Storage | Room Temperature | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Sofia® Strep A+ FIA with Sofia® (k141775) | Sofia® Strep A FIA with Sofia® (k123793) |
| Operational Modes | Walk-Away Mode only | Read-Now and Walk-Away Modes |
| Reagent | One reagent bottle containing sodium nitrite and hydrochloric acid in glass ampoule | One reagent bottle containing sodium nitrite and acetic acid in glass ampoule |
{6}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Sofia® Strep A+ FIA with Sofia® (k141775) | Sofia® Strep A FIA with Sofia® (k123793) |
| Performance Characteristics | Walk-Away Mode:
Sensitivity:
93.7%[95% CI: 89.1% - 96.5%]
Specificity:
94.4%[95% CI: 92.4% - 95.9%] | Read Now + Walk-Away Mode:
Sensitivity:
90.6%[95% CI: 84.3% - 94.6%]
Specificity:
96.1%[95% CI: 94.2% - 97.3%]
Read-Now Mode:
Sensitivity:
89.3%[95% CI: 82.2% - 93.8%]
Specificity:
96.0%[95% CI: 94.0% - 97.3%]
Walk-Away Mode:
Sensitivity:
100.0%[95% CI: 80.6% - 100.0%]
Specificity:
97.2%[95% CI: 85.8% - 99.5%] |
K. Standard/Guidance Document Referenced (if applicable):
Not applicable.
L. Test Principle:
Analyte detection from the chemically lysed swab specimen with the lateral flow device begins with the pipetting of the extracted clinical sample into the test cassette’s sample port and onto the sample pad, after which the sample is drawn by capillary action into and through the label pad, through the nitrocellulose strip and into the absorbent pad. The specimen interacts with distinct chemical environments as the fluid migrates along the course of the lateral flow device. Within the label pad, the specimen comes into contact with antibodies that have been coupled to the europium chelate-impregnated microbeads. During this interaction, the beads coated with anti-GAS polyclonal antibodies bind corresponding GAS carbohydrate antigens that are present in the specimen. The bead-coupled antigen-antibody complexes then begin to flow through the test strip. As they migrate on, if GAS antigens are present in the sample, they and the fluorescent beads to which they are bound will be subsequently captured on the surface of the nitrocellulose by the respective location-fixed, anti-GAS-specific capture antibodies. The flow and capture of the fluorescent microbeads coated with GAS antigens allows the accumulation of a fluorescent signal at the specific analyte line location on the test strip. Upon completion of the test, the Sofia® analyzer scans the test strip and objectively interprets the assay result. There are three possible results: (1) positive for Strep A; (2) negative for Strep A; and (3) invalid. It is
{7}
important to point out that the fluorescence signals obtained with this assay are invisible to the unaided eye. The test results can only be obtained with the proper use of the Sofia® analyzer. This ensures fully objective interpretation of the test result.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
All analytical studies were conducted in Read-Now Mode with the exception of the reproducibility studies, which were conducted with the Walk-Away mode. Although the Read-Now Mode is not available in this version of the device, this feature was unlocked in order to simplify testing. The primary difference between these modes is that in Read-Now Mode, the operator is responsible for the incubation time; in the Walk-Away Mode, the instrument will wait to read the strip until the specified development time has passed. A validation study was conducted to demonstrate that the Read-Now Mode does not impact the results of the analytical studies in the conditions tested. See section M.3.c below for more information and the results of this study.
#### a. Precision/Reproducibility:
Specimens for precision and reproducibility studies were prepared by re-suspending each cultured strain in a contrived negative matrix, freezing until testing then diluting the strain to the indicated concentration in the table below. The rayon swab that is provided with the assay was inoculated with a fixed volume of the stock and run with the Sofia® Strep A+ FIA. The LoD values were based on the values obtained in the LoD study. See section M.2.b below for details on the matrix equivalency and freeze-thaw study.
The reproducibility of the Sofia® Strep A+ FIA was evaluated at three (3) laboratory sites (two external, one in-house) on six (6) different analyzers. Reproducibility was assessed using a panel of four (4) simulated samples in negative clinical matrix that include negative, high negative, low positive and moderate positive GAS. The panels and controls were processed and tested on the Sofia® Strep A+ FIA at each site by 2 operators for 5 non-consecutive days over 2 weeks on 6 different instruments in Walk-Away Mode (2 operators x 3 replicates x 5 days x 3 sites = 90 results per concentration). The reproducibility study results are acceptable. The results are shown in the Table I below.
7
{8}
8
| Table I: Reproducibility | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Panel ID | | Negative (C_{0}) | | High Negative (C_{5}) | | Low Positive (C_{95}) | | Moderate Positive (C_{100}) | |
| Site | Operator | Detected Pos/Total | %Pos | Detected Pos/Total | %Pos | Detected Pos/Total | %Pos | Detected Pos/Total | %Pos |
| Site 1 | 1 | 0/15 | 0% | 2/15 | 13.3% | 14/15 | 93.3% | 15/15 | 100% |
| | 2 | 0/15 | 0% | 1/15 | 6.7% | 13/15 | 86.7% | 15/15 | 100% |
| Site 2 | 1 | 0/15 | 0% | 0/15 | 0% | 14/15 | 93.3% | 15/15 | 100% |
| | 2 | 0/15 | 0% | 1/15 | 6.7% | 9/15 | 60.0% | 15/15 | 100% |
| Site 3 | 1 | 0/15 | 0% | 3/15 | 20.0% | 14/15 | 93.3% | 15/15 | 100% |
| | 2 | 0/15 | 0% | 2/15 | 13.3% | 14/15 | 93.3% | 15/15 | 100% |
| Combined | | 9/90 | 0/90 | 0% | 9/90 | 10.0% | 78/90 | 86.7% | 100% |
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The Sofia® Strep A+ FIA has an internal a procedural control zone used to monitor sample processing and evaluate the presence of inhibitory substances or inadequate specimen preparation. External controls are used to confirm the integrity of assay reagents and cassette detection. The external quality controls used in these studies were positive control swabs spiked with heat inactivated, GAS cells and negative control swabs spiked with Group C Streptococcal antigen. The positive control serves as both a processing and extraction control. The positive and negative external control isolates were tested each day during the clinical studies.
Specimen stability studies demonstrated that contrived specimens were stable when stored at 2-8°C or 23°C for up to 96 hours. These studies were conducted with negative matrix and positive contrived specimens at 2.4 x LoD in replicates of six (6) per condition on a single instrument for 0, 6, 24, 48, 72 and 96 hours.
Additional temperature studies were conducted with negative matrix and positive contrived specimens at 1.6 x LoD in replicates of ten (10) per condition on ten (10) different instruments. These studies were conducted for 30 minutes at 2-8°C, 24°C and at 37°C.
Extracted specimens were demonstrated to be stable for up to 24 hours at 24°C. Studies were conducted with negative matrix and positive contrived specimens at 1.6 x LoD in replicates of 5 (five) per condition for 0, 1 and 30 minutes, 1, 2, 6 and 24 hours on six (6) different instruments.
{9}
Data was provided to demonstrate that the Sofia Strep $\mathrm{A + }$ FIA can be operated at temperatures between $15^{\circ}\mathrm{C} - 31^{\circ}\mathrm{C}$ . These studies were conducted with negative matrix and positive contrived specimens at $1.6\mathrm{x}$ LoD in replicates of ten (10) per condition on a single instrument at $15^{\circ}\mathrm{C}$ , $23^{\circ}\mathrm{C}$ and $31^{\circ}\mathrm{C}$ .
An inter-analyzer study demonstrated that results did not vary between instruments. These studies were conducted with negative matrix and positive contrived specimens at $1.6 \times \mathrm{LoD}$ using ten (10) analyzers in replicates of 20 for positive specimens and 10 for negative specimens.
The studies above meet or exceed the requirements in the labeling.
All controls performed as expected for all stability studies.
These study results are acceptable.
# d. Detection limit:
The limit of detection (LoD) of the Sofia® Strep A+ FIA was determined using contrived stocks of three (3) strains of GAS which were prepared per the protocol in M.1.a. The LoD was defined as the point at which at least $95\%$ of all replicates tested positive $(C_{95})$ . Preliminary testing was conducted to identify the approximate threshold for negative and positive detection for each strain. Low-positive $(C_{50})$ testing was then conducted with 60 replicates per strain. Dose-response curves were then generated to interpolate the expected LoD concentration for each strain. For each strain, the LoD was validated with 20 replicates. The LoD study results are shown in Table II below.
| Table II: LoD for Group A β-hemolytic Streptococcus | | | |
| --- | --- | --- | --- |
| Strain | Strain ID | CFU/ml | RFU |
| Group A Streptococcal strain 1 (Streptococcus pyogenes) | ATCC 19615 | 1.3 x 105 | 3357 |
| Group A Streptococcal strain 2 (Streptococcus pyogenes) | ATCC 700942 | 2.7 x 105 | 3599 |
| Group A Streptococcal strain 3 (Streptococcus pyogenes) | ATCC 700952 | 9.2 x 104 | 2706 |
These study results are acceptable.
# e. Analytical Sensitivity:
Inclusivity studies were conducted with twenty-one (21) GAS strains in replicates of twenty (20) per strain at approximately $2 \times$ LoD ( $5.8 \times 10^{5} \mathrm{CFU/ml}$ ) on with five (5) different instruments against three (3) different reagent lots. Each cultured strain was
{10}
prepared per the protocol described above in M.1.a above. The inclusivity study results and the final organism concentrations tested are shown in Table III below. All GAS strains were correctly detected by the assay (100% detected).
| Table III: Group A β-hemolytic Streptococcus Inclusivity | |
| --- | --- |
| Strain | Strain ID |
| Group A Streptococcal strain 1 | ATCC 19615 |
| Group A Streptococcal strain 2 | ATCC 700942 |
| Group A Streptococcal strain 3 | ATCC 700952 |
| Group A Streptococcal strain 4 | CI-52123 |
| Group A Streptococcal strain 5 | CI-52120 |
| Group A Streptococcal strain 6 | CI-62055 |
| Group A Streptococcal strain 7 | CI-52152 |
| Group A Streptococcal strain 8 | CI-62092 |
| Group A Streptococcal strain 9 | CI-52151 |
| Group A Streptococcal strain 10 | ATCC 700482 |
| Group A Streptococcal strain 11 | ATCC BAA 1315 |
| Group A Streptococcal strain 12 | ATCC 700459 |
| Group A Streptococcal strain 13 | ATCC 12203 |
| Group A Streptococcal strain 14 | ATCC 700944 |
| Group A Streptococcal strain 15 | CI-52154 |
| Group A Streptococcal strain 16 | CI-5036 |
| Group A Streptococcal strain 17 | CI-5095 |
| Group A Streptococcal strain 18 | CI-5017 |
| Group A Streptococcal strain 19 | CI-5060 |
| Group A Streptococcal strain 20 | CI-5112 |
| Group A Streptococcal strain 21 | CI-5008 |
These study results are acceptable.
f. Analytical specificity:
i. Microbial Cross-reactivity:
A study was performed to evaluate the performance of the Sofia® Strep A+ FIA in the presence of eighty-eight (88) other microorganisms commonly found in throat specimens. Each potentially cross-reacting microorganism was tested in replicates of five (5) or more with clinically relevant levels of viruses (1.0 x 10 $^5$ pfu/ml or the highest achievable concentration) and bacteria (1.0 x 10 $^6$ cfu/mL). All strain combinations were spiked into contrived negative matrix, prepared per the
{11}
protocol described above in M.1.a above. The strains included in the cross-reactivity study are shown in Table IV below.
| Table IV: Strains Included in Cross-Reactivity | | |
| --- | --- | --- |
| Strain | | |
| Aracanobacterium haemolyticum | Streptococcus mitis | Streptococcus Group G Strain #3 |
| Bacteroides fragilis | Streptococcus oralis | Streptococcus Group G Strain #4 |
| Bordetella pertussis | Streptococcus pneumoniae | Streptococcus Group G Strain #5 |
| Candida albicans | Streptococcus salivaris | Adenovirus Type 4* |
| Corynebacterium diphtheriae | Streptococcus sanguinis | Adenovirus Type 5 |
| Enterococcus faecium | Streptococcus Group B Strain #1: Streptococcus agalactiae | Adenovirus Type 11 |
| Escherichia coli | Streptococcus Group B Strain #2 | Coronavirus 229E |
| Fusobacterium necrophorum | Streptococcus Group B Strain #3 | Coxsackievirus |
| Haemophilus influenzae | Streptococcus Group B Strain #4 | Cytomegalovirus |
| Haemophilus parahaemolyticus | Streptococcus Group B Strain #5 | Echovirus |
| Klebsiella pneumoniae | Streptococcus Group C Strain #1: Streptococcus dysgalactiae | Herpes Simplex Virus 1 |
| Moraxella catarrhalis | Streptococcus Group C Strain #2 | Herpes Simplex Virus 2 |
| Neisseria gonorrhoeae | Streptococcus Group C Strain #3 | Influenza A/New Jersey/8/76 H1N1 |
| Neisseria lactamica | Streptococcus Group C Strain #4 | Influenza A/Victoria/3/75 H3N2 |
| Neisseria meningitidis | Streptococcus Group C Strain #5 | Influenza B/Hong Kong/5/72 |
| Neisseria sicca | Streptococcus Group D Strain #1 | Influenza B/Panama/45/90 |
| Neisseria subflava | Streptococcus Group D Strain #2 | Influenza C/Taylor/1233/47 |
| Proteus vulgaris | Streptococcus Group D Strain #4 | Adenovirus Type 3 |
| Serratia marcescens | Streptococcus Group D Strain #5 | Coronavirus OC43 |
| Staphylococcus haemolyticus | Streptococcus Group F Strain #1 | Epstein Barr Virus* |
| Staphylococcus intermedius | Streptococcus Group F Strain #2 | Adenovirus Type 1 |
| Staphylococcus saprophyticus | Streptococcus Group F Strain #3 | Measles (Edmonston) |
| Streptococcus anginosus | Streptococcus mutans | Parainfluenza virus 1 |
| Streptococcus gordonii | Streptococcus parasanginis | Parainfluenza virus 2 |
| Corynebacterium pseudodiphtheriticum | Streptococcus Group D Strain #3 | Parainfluenza virus 3 |
| Enterococcus faecalis | Streptococcus Group F Strain #4 | Parainfluenza virus 4A |
| Pseudomonas aeruginosa | Streptococcus Group F Strain #5 | Rhinovirus Type 2 |
{12}
| Staphylococcus aureus | Streptococcus Group G Strain #1: Streptoccus dysgalactiae | Rhinovirus Type 15 |
| --- | --- | --- |
| Staphylococcus epidermidis | Streptococcus Group G Strain #2 | Mumps (Enders) |
* Adenovirus Type 4 was tested at $4.0 \times 10^{3}$ PFU/ml; The Epstein Barr Virus was tested only with genomic copies at $4.5 \times 10^{7}$ copies/ml.
None of the eighty-eight (88) organisms cross-reacted at the concentrations tested.
# ii. Microbial Interference:
Interference studies were conducted with each of the eighty-eight (88) microorganisms listed above in Table IV. Each organism was tested in replicates of five (5) at clinically relevant levels of viruses $(1.0 \times 10^{5} \mathrm{pfu/ml}$ or the highest achievable concentration) and bacteria $(1.0 \times 10^{6} \mathrm{cfu/mL})$ in the presence of $1.6 \times$ LoD Group A Streptococcus ATCC 19615. All strain combinations were spiked into contrived negative matrix, prepared per the protocol described above in M.1.a above.
Eleven (11) of the eighty-eight (88) tested microorganisms that might be found in throat specimens interfered with the assay. When interference was observed, follow-up tests were conducted to assess the level of interference at other organism concentration levels. Results are shown in Table V below:
| Table V: Interfering Strains | |
| --- | --- |
| Strains | Interference Frequency* |
| Corynebacterium pseudodiphtheriticum** | 1/5 |
| Enterococcus faecalis** | 1/5 |
| Staphylococcus aureus** | 1/5 |
| Streptococcus mutans** | 4/15 |
| Streptococcus parasanginis** | 1/5 |
| Streptococcus dysgalactiae (Group C)** | 1/25 |
| Streptococcus Group D*** | 1/25 |
| Streptococcus Group F*** | 1/25 |
| Adenovirus Type 1** | 1/10 |
| Adenovirus Type 3*** | 3/10 |
| Epstein Barr Virus** | 1/10 |
| Mumps (Enders)*** | 1/10 |
* #interference results observed/total # tests
** No interference was observed in follow-up studies conducted at higher concentration (0/20).
*** Interference was observed in follow-up studies conducted at higher concentration for Group D (2/45) and Group F (6/90).
*** No interference was observed in follow-up studies (0/5).
*** Studies conducted at reduced concentration of the interfering organism did not demonstrate interference.
{13}
# iii. Interfering Substances:
Twenty-five (25) chemical and biological substances were evaluated for potential to cross-react and interfere with the Sofia® Strep A+ FIA, including fresh whole human blood, bovine submaxillary mucin and some snack foods that contain specimen-thickening substances. Each substance was tested in replicates of five (5) or more using $1.6 \times$ LoD Group A Streptococcus ATCC 19615, prepared per the protocol described above in M.1.a. Two (2) of the substances tested demonstrated a potential to cause false negative results, as shown in Table VI below:
| Table VI: Interfering Substances | |
| --- | --- |
| Strains | Interference Frequency* |
| Human Whole Blood 100μl/swab** | 1/5 |
| Bovine Submaxillary Mucin, 28.7 mg/ml | 1/10 |
* #false negative results/total # tests at that concentration.
** Concentrations $\leq 75\mu l$ whole blood/swab did not interfere. 2/5 results were invalid.
No cross-reactivity was observed with any of the chemical or biological substances. Of the snack foods tested, Nacho Flavor Doritos at $250\mathrm{mg / ml}$ increased the viscosity of the specimens which reduced the flow and sample volume, leading to invalid results 4/10 times in cross-reactivity studies. No interference was observed at $125\mathrm{mg / ml}$ . Interferents were noted in the limitations section in labeling.
These study results are acceptable.
# g. Assay cut-off:
The Limit of Blank (LoB) is 1948 RFU is the cutoff for this assay, which is the numerical threshold that is used to distinguish negative from positive. The procedural control zone cutoff for a valid versus invalid result is 30,000 RFU.
# 2. Comparison studies:
# a. Method comparison with predicate device:
Not applicable.
# b. Matrix comparison:
The analytical performance studies above were conducted in a contrived negative matrix that included bovine serum albumin and sodium azide with protocol that included a freeze-thaw step. In order to demonstrate that the results of analytical studies conducted in the contrived negative matrix are comparable to studies
{14}
conducted in the presence of a clinical matrix, a study was conducted to compare the following two conditions:
1) fresh bacteria spiked into a negative clinical matrix (pharyngeal swabs from patients)
2) spiked specimens prepared using a contrived negative the sample preparation protocol provided in the submission, which includes a freeze-thaw step
This study was conducted by testing twenty replicates each of four (4) different concentrations of Streptococcus pyogenes (Table VII) for each of the three (3) strains used to determine the LoD. All testing was conducted in Walk-Away Mode.
| Table VII: Matrix Comparison Study | | | | | |
| --- | --- | --- | --- | --- | --- |
| Panel ID | Pooled Negative Clinical Matrix | | Contrived Negative Matrix | | |
| | | Detected | % Pos | Detected | % Pos |
| Group A Streptococcus (ATCC 19615) | 2 x LoD | 20/20 | 100.0% | 20/20 | 100.0% |
| | 1 x LoD | 20/20 | 100.0% | 20/20 | 100.0% |
| | 0.5 x LoD | 9/20 | 45.0% | 15/20 | 75.0% |
| | 0.25 x LoD | 0/20 | 0.0% | 4/20 | 20.0% |
| Group A Streptococcus (ATCC 700942) | 2 x LoD | 20/20 | 100.0% | 20/20 | 100.0% |
| | 1 x LoD | 19/20 | 95.0% | 19/20 | 95.0% |
| | 0.5 x LoD | 7/20 | 35.0% | 4/20 | 20.0% |
| | 0.25 x LoD | 0/20 | 0.0% | 0/20 | 0.0% |
| Group A Streptococcus (ATCC 700952) | 2 x LoD | 20/20 | 100.0% | 20/20 | 100.0% |
| | 1 x LoD | 17/20 | 85.0% | 20/20 | 100.0% |
| | 0.5 x LoD | 7/20 | 35.0% | 7/20 | 35.0% |
| | 0.25 x LoD | 0/20 | 0.0% | 0/20 | 0.0% |
These studies demonstrate that the contrived negative matrix and freeze-thaw step does not alter the performance of the device in comparison to studies conducted in a clinical matrix. These study results are acceptable.
3. Clinical studies:
a. Clinical Sensitivity:
Performance characteristics of the Sofia® Strep A+ FIA were established during a prospective study conducted from April to May 2014. Eight hundred and sixty (860) fresh, throat swab specimens from female and male patients were prospectively collected and transported to each laboratory for testing with the Sofia® Strep A+ FIA
{15}
at seven (7) distinct geographical sites across the United States. All testing was done at laboratories holding a certificate of waiver. Three swab specimens were simultaneously collected were collected using the rayon throat swabs provided with the assay. One (1) was used for clinical testing, the second was used for Sofia® Strep A+ FIA testing and the third was shipped on ice in transport media to a central reference laboratory for culture and discrepant analysis using an FDA-cleared polymerase chain reaction (PCR) assay. All eight hundred and sixty (860) fresh throat specimens were cultured for Group A β-hemolytic Streptococcus (GAS) and tested with the Sofia® Strep A+ FIA. The specimen was considered positive if culture was positive for Group A β-hemolytic Streptococcus. Two (2) results were determined to be invalid in these studies. Seven (7) additional results were excluded due to the lack of valid external controls on the same day of running.
Bacterial culture was performed at the central reference laboratory where specimens were plated on GAS selective agar with 5% sheep's blood. Plates were cultured at 37°C with 5% CO₂ and examined for the presence of β-hemolytic colonies at twenty-four (24) hours. If no β-hemolytic colonies were observed, the plates were placed at room temperature and cultured aerobically for another twenty-four (24) hours. All β-hemolytic colonies positive culture plates were tested for catalase and confirmed for the presence of GAS using the FDA-cleared Streptex® Rapid Agglutination Test for identification of Lancefield groups A, B, C, D, F and G streptococci.
The clinical performance of the Sofia® Strep A+ FIA was demonstrated with eight hundred and fifty-one (851) prospectively collected fresh throat specimens at seven sites across the United States. The breakdown of performance for the Sofia® Strep A+ FIA is summarized in Table VIII below:
| Table VIII: Clinical Performance Data for the Sofia® Strep A+ FIA vs. Swab Culture for Group A β-hemolytic Streptococcus | | | |
| --- | --- | --- | --- |
| All Sites | | | |
| Sofia® Strep A+ FIA | Swab Culture | | |
| | Positive | Negative | Total |
| Positive | 164 | 38* | 202 |
| Negative | 11** | 638 | 649 |
| Total | 175 | 676 | 851 |
| Sensitivity: 93.7% (164/175) 95% CI (89.1%-96.5%)
Specificity: 94.4% (638/676) 95% CI (92.4%-95.9%) | | | |
| * Of the 38 discordant specimens, 24 of these specimens were positive for GAS when tested with an FDA-cleared molecular device, 14 were negative.
** Of the 11 discordant specimens, 3 were negative when tested with an FDA-cleared molecular device, 8 were positive. | | | |
{16}
| Site 1 | | | |
| --- | --- | --- | --- |
| Sofia® Strep A+ FIA | Swab Culture | | |
| | Positive | Negative | Total |
| Positive | 30 | 15* | 45 |
| Negative | 1** | 132 | 133 |
| Total | 31 | 147 | 178 |
| Sensitivity: 96.8% (30/31) 95% CI (83.1%-99.4%)
Specificity: 89.8% (132/147) 95% CI (83.9%-93.7%) | | | |
| * Of the 15 discordant specimens, 10 of these specimens were positive for GAS when tested with an FDA-cleared molecular device, 5 were negative.
** Of the one discordant specimen, one was positive when tested with an FDA-cleared molecular device. | | | |
| Site 2 | | | |
| Sofia® Strep A+ FIA | Swab Culture | | |
| | Positive | Negative | Total |
| Positive | 4 | 4* | 8 |
| Negative | 0 | 31 | 31 |
| Total | 4 | 35 | 39 |
| Sensitivity: 100.0% (4/4) 95% CI (51.0%-100.0%)
Specificity: 88.6% (31/35) 95% CI (74.1%-95.5%) | | | |
| * Of the 4 discordant specimens, 3 of these specimens were positive for GAS when tested with an FDA-cleared molecular device, 1 was negative. | | | |
| Site 3 | | | |
| Sofia® Strep A+ FIA | Swab Culture | | |
| | Positive | Negative | Total |
| Positive | 2 | 0 | 2 |
| Negative | 1** | 29 | 30 |
| Total | 3 | 29 | 32 |
| Sensitivity: 66.7% (2/3) 95% CI (20.8%-93.9%)
Specificity: 100.0% (29/29) 95% CI (88.3%-100.0%) | | | |
| ** Of the one discordant specimen, one was positive when tested with an FDA-cleared molecular device. | | | |
{17}
| Site 4 | | | |
| --- | --- | --- | --- |
| Sofia® Strep A+ FIA | Swab Culture | | |
| | Positive | Negative | Total |
| Positive | 55 | 3* | 58 |
| Negative | 2** | 175 | 177 |
| Total | 57 | 178 | 235 |
| Sensitivity: 96.5% (55/57) 95% CI (88.1%-99.0%)
Specificity: 98.3% (175/178) 95% CI (95.2%-99.4%) | | | |
| * Of the 3 discordant specimens, 2 of these specimens were positive for GAS when tested with an FDA-cleared molecular device, 1 was negative.
** Of the 2 discordant specimens, 1 was positive when tested with an FDA-cleared molecular device, 1 was negative. | | | |
| Site 5 | | | |
| Sofia® Strep A+ FIA | Swab Culture | | |
| | Positive | Negative | Total |
| Positive | 39 | 9* | 48 |
| Negative | 4** | 121 | 125 |
| Total | 43 | 130 | 173 |
| Sensitivity: 90.7% (39/43) 95% CI (78.4%-96.3%)
Specificity: 93.1% (121/130) 95% CI (87.4%-96.3%) | | | |
| * Of the 9 discordant specimens, 7 of these specimens were positive for GAS when tested with an FDA-cleared molecular device, 2 were negative.
** Of the 4 discordant specimens, 4 were positive when tested with an FDA-cleared molecular device. | | | |
| Site 6 | | | |
| Sofia® Strep A+ FIA | Swab Culture | | |
| | Positive | Negative | Total |
| Positive | 22 | 4* | 26 |
| Negative | 2** | 62 | 64 |
| Total | 24 | 66 | 90 |
| Sensitivity: 91.7% (22/24) 95% CI (74.2%-97.7%)
Specificity: 93.9% (62/66) 95% CI (84.4%-97.6%) | | | |
| * Of the 4 discordant specimens, 2 of these specimens were positive for GAS when tested with an FDA-cleared molecular device, 2 were negative.
** Of the 2 discordant specimens, 1 was negative when tested with an additional FDA-cleared molecular device, 1 was positive. | | | |
{18}
| Site 7 | | | |
| --- | --- | --- | --- |
| Sofia® Strep A+ FIA | Swab Culture | | |
| | Positive | Negative | Total |
| Positive | 12 | 3* | 15 |
| Negative | 1** | 88 | 89 |
| Total | 13 | 91 | 104 |
| Sensitivity: 92.3% (12/13) 95% CI (66.7%-98.6%)
Specificity: 96.7% (88/91) 95% CI (90.8%-98.9%) | | | |
| * Of the 3 discordant specimens, 3 of these specimens were negative for GAS when tested with an FDA-cleared molecular device.
** Of the 1 discordant specimen, 1 was negative when tested with an FDA-cleared molecular device. | | | |
The external quality controls used in these studies were: 1) Positive control consisting of swabs spiked with heat inactivated, Group A Streptococcus cells and 2) Negative control consisting of swabs spiked with Group C Streptococcal antigen. The positive control serves as both a processing and extraction control. Positive and negative external controls were tested each day during the clinical studies. Of 253 Group A Streptococcus positive controls tested, 250 (98.8%) were detected accurately. Of the three (3) incorrect results, two (2) were positive upon retesting and one (1) was not retested. Of the 253 Group C Streptococcus negative controls tested, 253 (98.8%) were detected accurately. Of the three (3) incorrect results, all three (3) were negative upon retesting. Results from clinical specimens were included for evaluation of study performance only if valid positive and negative external control results were attained on the same day.
These study results are acceptable.
b. Clinical specificity:
See table above.
c. Other clinical supportive data (when a. and b. are not applicable):
See section M.1 above for more information about why this study was conducted.
To demonstrate that performance of the Sofia® Strep A+ FIA is not influenced by operating the device in Read-Now Mode under the conditions used for analytical studies, a comparison study was conducted with ten (10) replicates of one (1) strain for each of three (3) concentrations of GAS on five (5) Sofia® analyzers (5 analyzers x 10 replicates = 50 tests/concentration).
{19}
19
| Table IX: Read-Now versus Walk Away Mode Comparison Study | | | | | |
| --- | --- | --- | --- | --- | --- |
| Panel ID | Read-Now Mode | | Walk-Away Mode | | |
| | | Detected | % Pos | Detected | % Pos |
| Group A
Streptococcus
(ATCC 19615) | 1.6 x LoD | 50/50 | 100.0% | 50/50 | 100.0% |
| | 1 x LoD | 49/50 | 98.0% | 48/50 | 96.0% |
| | Negative | 0/50 | 0.0% | 0/50 | 0.0% |
These results demonstrate that there is no notable difference in performance between Read-Now and Walk-Away Modes in the analytical testing setting.
These study results are acceptable.
4. Clinical cut-off:
Not Applicable.
5. Expected values/Reference range:
The overall prevalence of Group A $\beta$-hemolytic Streptococcus in patients tested during this study was 20.6% (177/858) based on bacterial culture. All clinical specimens collected during this study were collected between April, 2014 and May 2014.
N. Instrument Name:
Sofia®
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ______ or No ☐ X
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ______ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
{20}
Yes ☐ X or No ☐
3. Specimen Identification:
Not applicable.
4. Specimen Sampling and Handling:
One specimen is used per test. Specimens are not sampled. See section M.1.c for information on the studies provided to validate the conditions for specimen handling.
5. Calibration:
A calibration cassette is provided with the assay. Calibration is to be conducted within two weeks. See section M.1.c for more information.
6. Quality Control:
See section M.1.c for information on internal and external controls.
P. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
Q. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.