Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Skin Protectant and Tested for Use With Chemotherapy Drugs (Dawn Blue)
K171874 · Hartalega Sdn Bhd · LZA · Oct 12, 2017 · General Hospital
Device Facts
Record ID
K171874
Device Name
Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Skin Protectant and Tested for Use With Chemotherapy Drugs (Dawn Blue)
Applicant
Hartalega Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Oct 12, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
The Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Skin Protectant and Tested for Use with Chemotherapy Drugs (Dawn Blue) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
Device Story
Nitrile powder-free examination glove; coated with colloidal oatmeal USP skin protectant. Used by healthcare professionals in clinical settings to provide a protective barrier against patient-to-examiner contamination. Tested for permeation resistance against specific chemotherapy drugs (e.g., Carmustine, Cisplatin, Cyclophosphamide, etc.). Provides physical barrier protection; does not involve electronic inputs, algorithms, or software.
Clinical Evidence
Bench testing only. Device performance validated through chemotherapy drug permeation testing per ASTM standards, measuring breakthrough times for various agents including Carmustine and Thiotepa.
Technological Characteristics
Nitrile rubber material; powder-free; coated with colloidal oatmeal USP. Non-sterile, disposable, ambidextrous form factor. Barrier protection principle. No energy source, connectivity, or software components.
Indications for Use
Indicated for use as a protective barrier on the examiner's hands to prevent contamination between patient and examiner. Suitable for use during handling of chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K160562 — Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Tested for Use with Chemotherapy Drugs · Hartalega Snd. Bhd. · Oct 13, 2016
K180505 — Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP with Low-Dermatitis Potential Claim and Tested For Use with Chemotherapy Drugs (White) · Hartalega Sdn Bhd · Jun 14, 2018
K162819 — Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Tested for Use with Chemotherapy Drugs · Hartalega Sdn Bhd · Jan 9, 2017
K180645 — Powder Free Examination Gloves with Colloidal Oatmeal USP and Tested for Use with Chemotherapy Drugs - (Lemon Green) · Hartalega Sdn Bhd · Nov 16, 2018
K113100 — POWDER FREE NITRILE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS · Maxter Glove Manufacturing Sdn Bhd · Feb 6, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
October 12, 2017
Hartalega Sdn. Bhd. % Nurul Kong Quality Assurance Senior Manager Hartalega Snd. Bhd. No. 7, Kawasan Perusahaan Suria Bestari Jaya, 45600 MY
Re: K171874
Trade/Device Name: Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Skin Protectant and Tested for Use With Chemotherapy Drugs (Dawn Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: September 5. 2017 Received: September 13, 2017
Dear Nurul Kong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
### K171874
#### Device Name
Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Skin Protectant and Tested for Use with Chemotherapy Drugs (Dawn Blue)
#### Indications for Use (Describe)
The Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Skin Protectant and Tested for Use with Chemotherapy Drugs (Dawn Blue) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.
The list of Chemotherapy Drugs tested (with breakthrough times) is as below:
| Carmustine (3.3 mg/ml) | 21.9 minutes |
|--------------------------------------|---------------|
| Cisplatin (1.0 mg/ml) | > 240 minutes |
| Cyclophosphamide (20 mg/ml) | > 240 minutes |
| Dacarbazine (10.0 mg/ml) | > 240 minutes |
| Doxorubicin Hydrochloride (2.0mg/ml) | > 240 minutes |
| Etoposide (20.0 mg/ml) | > 240 minutes |
| Fluorouracil (50.0 mg/ml) | > 240 minutes |
| Methotrexate (25 mg/ml) | > 240 minutes |
| Mitomycin C (0.5 mg/ml) | > 240 minutes |
| Paclitaxel (6.0 mg/ml) | > 240 minutes |
| Thiotepa (10.0 mg/ml) | 36.0 minutes |
| Vincristine Sulfate (1.0 mg/ml) | > 240 minutes |
Please note that Carmustine and Thiotepa have extremely low permeation times of 21.9 minutes.
#### Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.