K171681 · Hermes Medical Solutions AB · KPS · Nov 21, 2017 · Radiology
Device Facts
Record ID
K171681
Device Name
Hermes Medical Imaging Suite v5.7
Applicant
Hermes Medical Solutions AB
Product Code
KPS · Radiology
Decision Date
Nov 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Hermes Medical Imaging suite that provides software applications used to process, display, analyze and manage nuclear medicine and other medical imaging data transferred from other workstation or acquisition stations.
Device Story
HERMES Medical Imaging Suite v5.7 is a software platform for processing, displaying, analyzing, and managing nuclear medicine and medical imaging data. It receives data transferred from external workstations or acquisition stations. The suite includes various modules for clinical analysis, such as Colonic Transit, Remnant Liver, Parathyroid, Dosimetry, Classic DMSA, Oesophageal Transit/Reflux, HIDA, Salivary Gland, Bone 3Phase Analysis, and Hybrid Recon. It is used by healthcare professionals in clinical settings to assist in diagnostic workflows. By providing automated or semi-automated analysis tools, the device aids clinicians in interpreting imaging data, potentially improving diagnostic accuracy and efficiency compared to manual calculations.
Clinical Evidence
Bench testing and comparative analysis performed. Hybrid Recon compared to v5.6 using 10 subjects (myocardial SPECT) showed Pearson correlation coefficients of 0.96-0.99 for QPS parameters. Phantom studies (Jaszczak, IEC) demonstrated reconstructed activity concentration errors of -1.1% (v5.7) vs -3.5% (v5.6). Additional modules (DMSA, Colonic Transit, Oesophageal, 3Phase Bone, HIDA, Salivary Gland) were validated against predicate software or manual calculations, with differences generally <5% (max <11% for Bone 3Phase).
Technological Characteristics
Software-based medical imaging suite. Operates as a processing and analysis workstation for nuclear medicine data. Utilizes established image processing algorithms for reconstruction and quantification. Connectivity via data transfer from external acquisition stations. Verification and validation followed design control procedures with risk mitigation implemented.
Indications for Use
Indicated for use by clinicians to process, display, analyze, and manage nuclear medicine and other medical imaging data transferred from other workstations or acquisition stations.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Xeleris 3.1 processing and review workstation (K130884)
Submission Summary (Full Text)
{0}------------------------------------------------
November 21, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Hermes Medical Solutions AB % Joakim Arwidson Quality Manager Skeppsbron 44 Stockholm. 11130 SWEDEN
Re: K171681
Trade/Device Name: HERMES Medical Imaging Suite v5.7 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: September 29, 2017 Received: October 26, 2017
Dear Joakim Arwidson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) FBD K171681
Device Name
HERMES Medical Imaging Suite v5.7
Indications for Use (Describe)
HERMES Medical Imaging suite that provides software applications used to process, display, analyze and manage nuclear medicine and other medical imaging data transferred from other workstation or acquisition stations.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
{3}------------------------------------------------
# 5.0 510 (k) SUMMARY
### A. Submitted by:
- Submitters name and address: Hermes Medical Solutions AB Skeppsbron 44 111 30 Stockholm Sweden
#### . Submitters telephone number
| Phone: | +46 8 19 03 25 |
|---------|-----------------------------------|
| Cell: | +46 708 19 03 08 |
| Fax: | +46 8 18 43 54 |
| E-mail: | joakim.arwidson@hermesmedical.com |
#### ● Contact person
Joakim Arwidson Quality Manager Hermes Medical Solutions AB Skeppsbron 44 111 30 Stockholm Sweden
- . Reqistration number 9710645
### B. Preparation date:
2017-04-18
### C. Proprietary/Trade name, Common name, Classification name:
- Proprietary/Trade name . Hermes Medical Imaging Suite v5.7
- . Common name Image processing systems
- . Classification name Emission Computer Tomography System, Class II, 21CFR892.1200
### D. Legally marketed device (predicate device):
Following legally marketed device has been used for comparison.
- Hermes Medical Imaging Suite v5.6 (K153056)
- . Xeleris 3.1 processing and review workstation (K130884)
### E. Description of the device that is subject of this premarket notification:
The base product design of Hermes Medical Imaging Suite v5.7 is the same as for the Hermes Medical Imaging Suite v5.6 (K153056). The following modules have been added in this applicatio to Hybrid Viewer NM Processing: Colonic Transit, Remnant Liver, Parathyroid, Dosimetry, Classic DMSA, Oesophageal Transit/Reflux, HIDA, Salivary Gland, Bone3Phase Analysis and Uniformity.
{4}------------------------------------------------
# F. Intended use:
Hermes Medical Imaging suite that provides software applications used to process, display, analyze and manage nuclear medicine and other medical imaging data transferred from other workstation or acquisition stations.
# G. Technological characteristics:
The proposed device Hermes Medical Imaging Suite has the same technological characteristics as the original device and the same indication for use.
# H. Testing:
The tests for verification and validation followed Hermes Medical Solutions AB design controlled procedures. The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria.
## l. Substantially Equivalent/Conclusions:
The proposed device Hermes Medical Imaging Suite v5.7 and the predicate devices HERMES Medical Imaging Suite v5.6 (K153056) have the same indication for use.
The proposed device will use similar technology and fundamental concepts and operation are also the same, as described in the 510(k) submission.
The following comparative tests have been made.
- Hybrid Recon in Hermes Medical Imaging Suite v5.7 has been compared to Hybrid Recon in Medical Imaging Suite v5.6 (K153056). The comparison included 10 subjects, where each subject went through stress&rest myocardial SPECT scan with Tc99m tetrofosmin. The QPS parameters SSS, Stress volume, Stress Area, Stress Defect Area, SRS, Rest Volume, Rest Area, Rest Defect Area were calculated with the Cedar Sinai Application and compared between the applications (Pearson correlation coefficient r between 0.96 to 0.99).
- -Hybrid Recon in Hermes Medical Imaging Suite v5.7 has been compared to Hybrid Recon in Medical Imaging Suite v5.6 (K153056). The comparison included studies on Jaszczak phantom and IEC Phantom. The results on the Jaszak Phantom showed an average error between the reconstructed and true activity concentration of -3.5 % with Hybrid Recon in Medical Imaging Suite v5.6 and -1.1 % with Hybrid Recon in Medical Imaging Suite v5.7. The results on the IEC Phantom show that the error of the reconstructed activity concentration is around 5% when the target is large enough. In small targets the partial volume effect lowers the accuracy as expected.
- Classic DMSA in HV NM Processing/Hermes Medical Imaging Suite v5.7 has been compared to the classic DMSA application in Medical Imaging Suite v5.3 (K153056). For relative function values, the maximum% difference between Classic and HybridViewer is < 5%, which represents an absolute % difference of < 2%.
- -Colonic Transit in HV NM Processing/Hermes Medical Imaging Suite v5.7 has been compared to GE Xeleris 3.1 review and workstation (K130884). The results obtained from both applications were similar.
- -Oesophageal Transit/Reflux Analysis in HV NM Processing/Hermes Medical Imaging Suite v5.7 has been compared to manual calculations. For the Oesophageal Transit all residual % results are in complete agreement and all Transit time values are in agreement within
{5}------------------------------------------------
0.5s. For the Oesophageal Reflux all % results agree completely and most values in counts/second agree within 5%.
- 3Phase Bone Analysis in HV NM Processing/Hermes Medical Imaging Suite v5.7 has been compared to manual calculations. The biggest difference for any value was <11%.
- -HIDA in HV NM Processing/Hermes Medical Imaging Suite v5.7 was compared to manual calculations from first principles. The maximum difference between the results for all calculated values for all studies was 5%.
- -Salivary Gland Analysis in HV NM Processing/Hermes Medical Imaging Suite v5.7 has been compared to manual calculations from first principles. The maximum difference between the results for all calculated values for all studies is < 0.5%
The results from the comparisons show good results and are acceptable.
In summary, the Hermes Medical Imaging Suite v5.7 described in this submission is in our opinion substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.