BioMonitor 2-AF, BioMonitor 2-S, FIT - Fast Insert Tool 1 + 2, Set
K171514 · Biotronik, Inc. · MXD · Jun 29, 2017 · Cardiovascular
Device Facts
Record ID
K171514
Device Name
BioMonitor 2-AF, BioMonitor 2-S, FIT - Fast Insert Tool 1 + 2, Set
Applicant
Biotronik, Inc.
Product Code
MXD · Cardiovascular
Decision Date
Jun 29, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The BioMonitor 2 is indicated to detect the following cardiac arrhythmias: - Atrial fibrillation, - Bradycardia, - Sudden rate drop. - High ventricular rate (HVR), - Asystole. The BioMonitor 2 is indicated for: - Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias - Patients who experience transient symptoms that may suggest a cardiac arrhythmia - The device has not been tested for and it is not intended for pediatric use
Device Story
Insertable cardiac monitor; senses subcutaneous electrocardiograms (SECG) via two integrated electrodes. Detects arrhythmias (atrial fibrillation, bradycardia, sudden rate drop, HVR, asystole); triggers automatic SECG recording. Patients use Remote Assistant for manual recording. Device stores SECG episodes; transmits data via BIOTRONIK Home Monitoring to Service Center for physician review. Used in clinical/home settings; physician-implanted. Output aids diagnosis of otherwise undetected arrhythmias; informs clinical decision-making regarding cardiac management.
Clinical Evidence
No clinical data. Bench testing only, including mechanical system verification, usability testing, sterilization validation, biocompatibility, and packaging validation.
Technological Characteristics
Insertable cardiac monitor; subcutaneous ECG sensing via two integrated electrodes. Includes FIT1/FIT2 insertion accessories. Data transmission via BIOTRONIK Home Monitoring. Sterilization validation performed. No changes to core sensing/detection technology compared to predicate.
Indications for Use
Indicated for adult patients with clinical syndromes/situations at increased risk of cardiac arrhythmias or experiencing transient symptoms suggesting cardiac arrhythmia. Detects atrial fibrillation, bradycardia, sudden rate drop, high ventricular rate (HVR), and asystole. Not for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
BIOTRONIK Home Monitoring Service Center (P950037/S66)
BIOTRONIK Home Monitoring Service Center (P950037/S69)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 29, 2017
BIOTRONIK, Inc. Jon Brumbaugh Vice President, Regulatory Affairs and Compliance 6024 Jean Road Lake Oswego, Oregon 97035
Re: K171514
Trade/Device Name: BioMonitor 2 Insertable Cardiac Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: May 22, 2017 Received: May 24, 2017
Dear Jon Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
M.A. Hillebrand
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K171514
#### Device Name
BioMonitor 2 Insertable Cardiac Monitor
Indications for Use (Describe)
The BioMonitor 2 is indicated to detect the following cardiac arrhythmias:
- · Atrial fibrillation,
- · Bradycardia,
- · Sudden rate drop.
- · High ventricular rate (HVR),
- · Asystole.
The BioMonitor 2 is indicated for:
- · Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- Patients who experience transient symptoms that may suggest a cardiac arrhythmia
- · The device has not been tested for and it is not intended for pediatric use
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1. Submitter
BIOTRONIK 6024 SW Jean Road Lake Oswego, OR 97035 Phone: (888) 345-0374 Fax: 503-451-8519
Contact Person: Jon Brumbaugh
Date Prepared: June 29, 2017
#### 2. Device
| Name of Device | BioMonitor 2 Insertable Cardiac Monitor |
|----------------------|-------------------------------------------------------------------------------|
| Common or Usual Name | Insertable Cardiac Monitor |
| Classification Name | Arrhythmia detector and alarm (including ST-segment measurement<br>and alarm) |
| Classification | Class II (21 CFR 870.1025) |
| Product Code | MXD |
# 3. Predicate Devices
BIOTRONIK BioMonitor 2 (K152995, cleared April 11, 2016)
# 4. Device Description
The BioMonitor 2 senses subcutaneous electrocardiograms (SECG) using two integrated electrodes and has the capability of detecting a number of arrhythmias. The BioMonitor 2 uses BIOTRONIK's Home Monitoring to send recorded SECG and statistics to the Home Monitoring Service Center (P950037/S66, dated November 21, 2008; and P950037/S69, dated March 23, 2009) for further analysis. BioMonitor 2 is intended to aid in the diagnosis of cardiac arrhythmias that might otherwise go undetected.
BioMonitor 2 is designed to detect and trigger automatic recording of certain cardiac arrhythmias; these arrhythmias can be classified as follows:
- atrial fibrillation
- bradycardia ●
- sudden rate drop
- high ventricular rate (HVR)
- asystole. .
Patients can also manually trigger recording of the cardiac information by use of the Remote Assistant. The memory capacity of BioMonitor 2 is such that the device can record at least 66 minutes of subcutaneous electrocardiograms (SECG). The device automatically stores a maximum of 55 separately recorded SECG-episodes of 40 seconds each (60 seconds maximum), and a maximum of 4 patient triggered SECG-episodes of 7.5 minutes. The BioMonitor 2 reports the recorded episodes and related statistical data through the physician's programmer and BIOTRONIK's Home Monitoring®
{4}------------------------------------------------
# 5. Indications for Use
The BioMonitor 2 Insertable Cardiac Monitor is indicated to detect the following cardiac arrhythmias:
- Atrial fibrillation, .
- Bradycardia, ●
- Sudden rate drop,
- High ventricular rate (HVR), ●
- . Asystole.
The BioMonitor 2 Insertable Cardiac Monitoris indicated for:
- . Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- . Patients who experience transient symptoms that may suggest a cardiac arrhythmia
- The device has not been tested for and it is not intended for pediatric use.
#### 6. Comparison of Technological Characteristics with the Predicate Device
The technological principles of the subject and predicate device are the same. The differences represent minor modifications to the currently marketed BioMonitor 2 as follows:
- . Dimensional changes to the FIT accessories:
- FIT1 assists to create the device pocket for insertion. Dimensions at the tip and handle of ● this accessory were modified for improved handling.
- . FIT2 supports the device during insertion. Dimensions of this accessory were modified for improved support during insertion.
- Packaging changes to the FIT accessories:
- FIT1 and FIT2 were provided separately in double sterile pouches, and may now be ● provided together in a sterile blister package.
- A visual inspection was added to the receiving inspection of batteries. Subsequent testing . and inspections, including acceptance criteria for device performance, are unchanged.
# 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
To demonstrate that the modified BioMonitor 2 meets the same performance criteria, the following tests were conducted using the same test methods and acceptance criteria as for the predicate device.
- Mechanical system verification of FIT1 and FIT2
- . Visual inspection and handling of FIT1 and FIT2
- Usability testing: system verification and validation
- Visual inspection of labeling ●
- Compatibility testing of BioMonitor 2 with FIT 2 ●
- . Procedure of implantation: Product Validation
- Bioburden
- Pvrogenicitv
- Sterilization validation ●
- . Biocompatibility
- Packaging validation, including seal integrity, environmental conditioning, sealing process, and label integrity
{5}------------------------------------------------
No clinical testing was deemed necessary or completed in the premarket notification submission for a determination of substantial equivalence.
# 8. Conclusions
The subject device results from minor modifications to the predicate device. The performance testing demonstrates that the subject device meet the same functional acceptance criteria for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.