Clinical study data (30 patients); Post Market Observation (PMO) data (15 patients)
Clinical data from an Australian clinical study and a European Post Market Observation were used to demonstrate the implantation procedure and sensing quality of the BioMonitor 2.
Implantable Cardiac Monitor; Clinical study; Post Market Observation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
BioMonitor 2 Clinical Study; Clinical study of implantation procedure and sensing quality; Study Period: December 18, 2014 through July 06, 2015
Patients requiring an implantable cardiac monitor; Sample Size: 30; Number of Sites: 5
Not applicable for this study
Implantation procedure and sensing quality
Post Market Observation (PMO); Post Market Observation; Study Period: As of October 9, 2015
The BioMonitor 2 indicated to detect the following cardiac arrhythmias: - atrial fibrillation - bradycardia - sudden rate drop - high ventricular rate (HVR) - asystole The BioMonitor 2 is indicated for use in: - Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias - Patients who experience transient symptoms that may suggest a cardiac arrhythmia - The device has not been tested for and is not intended for pediatric use
Device Story
BioMonitor 2 is a small, leadless, implantable cardiac monitor (ICM) used for continuous subcutaneous ECG monitoring. It detects arrhythmias including atrial fibrillation, bradycardia, sudden rate drop, high ventricular rate, and asystole. The device automatically records arrhythmia episodes and supports patient-triggered recordings via a Remote Assistant. It is implanted subcutaneously; physicians manage the device and review transmitted data via the BIOTRONIK Home Monitoring system. The device provides clinicians with diagnostic data to identify cardiac arrhythmias, facilitating clinical decision-making and patient management. It features a smaller form factor and weight compared to its predecessor and is MR conditional for 1.5T and 3.0T full-body scans.
Clinical Evidence
Clinical study included 30 patients across 5 Australian sites (Dec 2014–July 2015) to evaluate implantation procedure and sensing quality. Additional data from 15 European post-market observations provided. No specific sensitivity/specificity metrics reported in the summary; evidence focused on successful implantation and device performance.
Technological Characteristics
Leadless implantable device; 88.4 x 15.2 x 6.2 mm; 10.1 g; 5 cc volume. Sensing via two electrodes on device body. MR conditional (1.5T and 3.0T). Connectivity via Home Monitoring system. Tested per ISO 14708-1, ISO/TS 10974, and EN 45502 standards. Sterilization per ISO 11135-1.
Indications for Use
Indicated for patients with clinical syndromes/situations at increased risk of cardiac arrhythmias or experiencing transient symptoms suggesting arrhythmia. Not for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is presented in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 11, 2016
Biotronik, Inc. Jon Brumbaugh Vice President, Regulatory Affairs and Compliance 6024 Jean Road Lake Oswego, Oregon 97035
Re: K152995
Trade/Device Name: BioMonitor 2 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MXD Dated: March 10. 2016 Received: March 11, 2016
Dear Jon Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152995 - Page 1 of 1
Device Name BioMonitor 2
Indications for Use (Describe)
The BioMonitor 2 indicated to detect the following cardiac arrhythmias:
- atrial fibrillation
- bradycardia
- sudden rate drop
- high ventricular rate (HVR)
- asystole
The BioMonitor 2 is indicated for use in:
- Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- Patients who experience transient symptoms that may suggest a cardiac arrhythmia
- The device has not been tested for and is not intended for pediatric use
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# BioMonitor 2 Implantable Cardiac Monitor
Traditional 510(k) Premarket Notification
# 1. SUBMISSION INFORMATION
| Date 510(k) Summary Prepared: | October 9, 2015 |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name and Address of Sponsor: | BIOTRONIK, Inc.<br>6024 Jean Road<br>Lake Oswego, OR 97035 |
| Establishment Registration Number: | 1028232 |
| Name and Address of Manufacturer: | BIOTRONIK SE & Co. KG (reg. no. 9610139)<br>Woermannkehre 1,<br>12359 Berlin, Germany<br>011-49-30-689-05-1210<br><br>Sterigenics Germany GmbH<br>Kasteler Straße 45<br>D- 65203 Wiesbaden, Germany |
| Contact Person(s) and Phone Number: | Jon Brumbaugh<br>VP, Regulatory Affairs and Compliance<br>Phone (888) 345-0374<br>Fax (800) 913-6993<br>jon.brumbaugh@biotronik.com |
| Device Name: | Trade Name: BioMonitor 2<br>Common Name: Implantable Cardiac Monitor<br>Classification Name: Arrhythmia detector and alarm (including ST-segment<br>measurement and alarm).<br>Classification: Class II (21 CFR 870.1025)<br>Product Code: MXD |
## General Description:
The BioMonitor 2 is a small, leadless, implantable device that uses two electrodes on the body of the device to monitor continuously the patient's subcutaneous ECG. The BioMonitor 2 is designed to record automaticallythe occurrence of arrhythmias in a patient. Recordings can also be triggered by use of the associated Remote Assistant. Arrhythmia may be classified as atrial fibrillation, bradycardia, asystole, or high ventricular rate. The device memory can automatically store a maximum of 55 separately recorded SECG-episodes of 40 seconds each, and 4 patient triggered SECG-episodes of 7.5 minutes.
{5}------------------------------------------------
# Predicate Devices:
- BIOTRONIK BioMonitor with AF Detection (K143503, cleared March 19, 2015) ●
# Indication for Use:
The BioMonitor 2 is indicated to detect the following cardiac arrhythmias:
- atrial fibrillation ●
- bradycardia,
- . sudden rate drop,
- high ventricular rate (HVR), ●
- asystole. ●
The BioMonitor 2 is indicated for use in:
- Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias .
- Patients who experience transient symptoms that may suggest a cardiac arrhythmia ●
- The device has not been tested for and it is not intended for pediatric use ●
# Technological Characteristics and Substantial Equivalence:
The substantial equivalence claim between the subject and the predicate device is supported by the information included in this premarket notification. This includes the following information:
- Description of the subject and predicate devices -
- -Intended use of the subject and predicate devices
- -Performance of the subject and predicate devices
- -Technological characteristics of the subject and predicate devices
- -Validation testing
{6}------------------------------------------------
# Table 1: BioMonitor 2 Compared to Predicate Device (changes are highlighted)
| Technical Data | BioMonitor | BioMonitor 2 |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| FDA Clearance | K143503 | Subject |
| Dimensions (mm) | 53.3 x 42.7 x 7.1 | 88.4 x 15.2 x 6.2 |
| Length x Width x Height | | Rigid portion 55.5 x 15.2 x 6.2 |
| Volume | 12.5 cc | 5 cc |
| Weight | 26 g | 10.1 g |
| Longevity | 48 months (+ 12 months pre-implant shelf life + 2 months ERI life) | |
| Subcutaneous ECG Recording | Yes | |
| Pre and Post Event Storage | Yes | |
| sECG Storage | 35.8 min<br>22.5 min for patient<br>triggered events<br>13.3 min for auto-<br>activated events<br>Longest/oldest/newest | >66 min<br>30 min for patient triggered events<br>36.7 min for auto-activated events<br>Longest/oldest/newest |
| Patient Activation | 7.5 min per event<br>7 min prior to activation<br>0.5 min following activation | |
| Asystole<br>Brady/rate drop<br>VT-FVT | 40 s/episode<br>30 s prior auto activation<br>10 s post auto activation | |
| AF | 40 s/episode<br>30 s prior auto activation<br>10 s post auto activation | |
| Vector Mapping Required | No | |
| Sampling Rate | 128 Hz | |
| Auto Activation Triggers | Yes | |
| Manual (Patient) Activation<br>Trigger | Yes - via magnet | Yes - via Remote Assistant |
| High Rate Trigger | Yes | |
| Programmable High Rate Count | Yes | |
| Low Rate Trigger | Yes | |
| Asystole Trigger | Yes | |
| Remote Monitoring | Home Monitoring daily<br>transmissions, up to 1<br>episode per day | Home Monitoring daily transmissions, up<br>to 6 episodes (at least one per triggered<br>event type) per day |
| QRS Detection | Combination signal out of<br>3 vectors | Single vector |
| MR Conditional | 1.5T, full body scan | 1.5T and 3.0T, fully body scan |
# Performance Data:
The following performance data were provided in support of the substantial equivalence determination.
{7}------------------------------------------------
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" in a separate submission. The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
#### 3.0T MRI Scan Testing
BIOTRONIK conducted validation testing according to the Joint Working Group's International Technical Specification for ISO/TS 10974: 2012 (E). The following tests were performed:
- ·Gradient Induced Heating
- •Vibration
- ·Static Malfunction
- •Gradient Radiated Malfunction
- •Vibration Malfunction
#### Clinical Study
The objective of this study was to provide clinical data of the implantation procedure and the sensing quality of BIOTRONIK's second generation of Implantable Cardiac Monitor (ICM) BioMonitor 2. Data from 30 patients at 5 Australian clinical sites from December 18, 2014 through July 06, 2015 is included to support the substantial equivalence of BioMonitor 2 when compared to the predicate device. Additional implantations have commenced recently in Europe as part of a controlled Post Market Observation (PMO), as of October 9, 2015 there have been 15 implantations.
#### List of Applied Standards
The BioMonitor 2 was tested in accordance with the following standards:
- EN 45502-1 (1997) Section 21
- EN 45502-2-1 (2003) Section 6.1.3 ●
- . ISO 11135-1 (2007)
- ISO 14708-1 (2014) ●
- . ISO/TS 10974 (2012)
## Conclusion:
BIOTRONIK considers the BioMonitor 2 implantable cardiac monitor to be substantially equivalent to legally marketed predicate devices through the data and information presented. No safety or effectiveness issues were identified.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.