K171492 · Spiration, Inc. · EOQ · Nov 20, 2017 · Ear, Nose, Throat
Device Facts
Record ID
K171492
Device Name
B7 2C Occlusion Balloon
Applicant
Spiration, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Nov 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The B7-2C Occlusion Balloon is to be used for selective endoscopic bronchography, saline injection associated with bronchoalveolar lavage, bronchial hemostasis and airway occlusion to localize air leaks.
Device Story
B7-2C Occlusion Balloon is a single-use, sterile catheter accessory for flexible bronchoscopes. Device consists of an inflatable latex balloon, catheter, bifurcation, irrigation port, air feed cap, and stopcock. Operated by physicians in clinical settings; inserted through bronchoscope instrument channel. Balloon is inflated to apply pressure against airway walls to achieve occlusion. Primary clinical utility includes selective bronchography, saline delivery for lavage, hemostasis, and localization of air leaks. Device provides mechanical airway occlusion; output is the physical state of the airway (occluded vs. patent) allowing clinicians to identify air leak sources. Benefits include improved diagnostic capability for air leaks and therapeutic support for hemostasis.
Clinical Evidence
Bench testing only. Performance testing (N=30) demonstrated complete airway occlusion with 90% conformance at 95% confidence. Biocompatibility testing (cytotoxicity, intracutaneous irritation, materials-mediated pyrogen, sensitization, and systemic toxicity) successfully passed. Sterilization validated to SAL 10^-6.
Technological Characteristics
Materials: Latex rubber, polyethylene impregnated with barium sulfate. Dimensions: 1050 mm length, 2.55 mm pre-inflation diameter, 13 mm post-inflation diameter, 1.7 cc max volume. Operation: Mechanical balloon inflation via stopcock/air feed. Connectivity: None. Sterilization: Ethylene Oxide. Single-use.
Indications for Use
Indicated for patients requiring selective endoscopic bronchography, saline injection for bronchoalveolar lavage, bronchial hemostasis, or airway occlusion for air leak localization.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2017
Spiration, Inc. Cheryl Frederick Executive Director, Regulatory Affairs 6675 185th Avenue N.E. Redmond, Washington 98052
Re: K171492
Trade/Device Name: B7 2C Occlusion Balloon Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOO Dated: August 22, 2017 Received: August 24, 2017
Dear Cheryl Frederick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K171492
Device Name
B7-2C Occlusion Balloon
#### Indications for Use (Describe)
The B7-2C Occlusion Balloon is to be used for selective endoscopic bronchography, saline injection associated with bronchoalveolar lavage, bronchial hemostasis and airway occlusion to localize air leaks.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5 510(k) Summary – K171492
#### Submitter Information
| Date of 510(k) Summary Preparation: | May 19, 2017 |
|--------------------------------------|-----------------------------------------------------------------------------------------------|
| Name and Address of Private Labeler: | Spiration, Inc.<br>6675 185th Avenue NE<br>Redmond, WA 98052 |
| Name and Address of Manufacturer: | Aomori Olympus Co., LTD<br>2-248-1 Okkonoki<br>Kuroishi-shi Aomori<br>Japan 036-0357 |
| Contact Person: | Cheryl Frederick<br>Executive Director, RA/QA<br>Phone: (425) 636-5470<br>Fax: (425) 497-8802 |
| Subject Device | |
Device Trade Name: B7 2C Occlusion Balloon Common Name: Bronchoscope and Accessories
Classification Regulation:
Product Code: Review Panel:
#### Predicate Device
Trade Name: 510(k) Number: Manufacturer:
BP-25 Occlusion Balloon
Bronchoscope and Accessories
Bronchoscope (flexible or rigid) and Accessories 21 CFR 874.4680 EOQ 77- Ear, Nose and Throat
Olympus B7-2C K962901 Aomori Olympus Co., LTD
#### Device Description
Olympus Respiratory America (ORA) dba Spiration, Inc. is submitting this Traditional 510(k) premarket notification to add a modified name and modified Indication for use to for the already cleared Olympus B7-2C Balloon Catheter (K962901, August 20, 1996). The current cleared Indication for Use for the B7-2C Balloon Catheter is as follows (the 510(k) was submitted with 2 model #'s with identical indications):
Olympus B5-2C/B7-2C Balloon Catheters are to be used for selective endoscopic bronchography, medicine injection to bronchi, saline injection
{4}------------------------------------------------
associated with bronchoalveolar lavage, foreign body removal from bronchi and bronchial hemostasis.
The device design and materials are completely unchanged. it will be provided packaged and sterilized to ORA, and we will private label it. The only difference will be the package labeling and Instructions for Use indicating that the device is the B7-2C Occlusion Balloon, and the modified Indication for Use.
In the modified indication for use, the B7-2C Occlusion Balloon will also be available for localizing air leaks.
We proposed the following modified indication:
The B7-2C Occlusion Balloons is to be used for selective endoscopic bronchography. saline injection associated with bronchoalveolar lavage, bronchial hemostasis and airway occlusion to localize air leaks.
The proposed modified Indication for Use removes medicine injection to bronchi, foreign body removal from bronchi and adds airway occlusion to localize air leaks.
Both the predicate and proposed device, which is identical, are an accessory to Olympus bronchoscopes. Spiration will be private labeling the device with a modified name and Indications for use, which fall within the current Intended Use.
The device is introduced through the instrument channel of the bronchoscope. The device is constructed of the inflatable balloon, catheter, bifurcation, irrigation port, air feed cap and stopcock. The B7-2C device is 1050 mm in working length, with a pre-inflation diameter of 2.55 mm and a post-inflation diameter of 13 mm, with a max volume of 1.7 cc.
## Device Materials (Patient Contacting)
The B7-2C Occlusion Balloon is manufactured with materials commonly used in medical device applications. The patient contacting materials are included in Table 5-1 below:
#### Table 5-1 B7-2C Patient Contacting Materials
| Patient Contacting Materials |
|----------------------------------------------|
| Latex rubber |
| Polyethylene impregnated with barium sulfate |
Biocompatibility testing was repeated on the subject device to support this 510(k) submission, to confirm that it meets current biocompatibility requirements.
#### How Provided
The B7-2C Occlusion Balloon is provided sterile and is disposable, intended for single patient use.
#### Proposed Indication for Use
The B7-2C Occlusion Balloon is to be used for selective endoscopic bronchography, saline injection associated with bronchoalveolar lavage, bronchial hemostasis and airway occlusion to localize air leaks.
{5}------------------------------------------------
## Comparison to Predicate
A comparison table, Table 5-2, is included below, which illustrates the equivalence of the subject B7-2C device to the predicate device.
## Table 5-2: Comparison of Key Characteristics of Proposed Device to Predicate Device
| Device Name→<br>Device Characteristics ↓ | B7-2C Balloon Catheter<br>(K962901) | B7-2C Occlusion Balloon<br>(Proposed Device) |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | II | II |
| Code of Federal Regulations | 878.4680 | 878.4680 |
| Prescription | Yes | Yes |
| Indication for Use | Olympus B5-2C/B7-2C Balloon<br>Catheters are to be used for<br>selective endoscopic<br>bronchography, medicine<br>injection to bronchi, saline<br>injection associated with<br>bronchoalveolar lavage,<br>foreign body removal from<br>bronchi and bronchial<br>hemostasis. | The Olympus B7-2C Occlusion<br>Balloon is to be used for selective<br>endoscopic bronchography,<br>medicine injection to bronchi,<br>saline injection associated with<br>bronchoalveolar lavage, foreign<br>body removal from bronchi,<br>bronchial hemostasis and airway<br>occlusion to localize air leaks. |
| Materials | Latex rubber Polyethylene<br>impregnated with<br>barium sulfate | Identical |
| Minimum Endoscope<br>Channel Size | 2.8 mm | Identical |
| Diameter | 7 French | Identical |
| Length | 1050 mm | Identical |
| Sterilization | Ethylene Oxide, Single Use | Identical |
{6}------------------------------------------------
The intended use of the subject device is virtually identical to that of the predicate device with respect to bronchial hemostasis. Whereas the predicate device uses pressure aqainst the airway walls to support hemostasis, the proposed device uses pressure against the airway to localize air leaks within the lung.
#### Performance Data
Performance testing has been completed on both the predicate device as well as the subject device, which is identical to the subject device. The B7-2C Occlusion Balloon was subjected to the following verification and validation tests, as applicable:
#### Biocompatibility
Repeat biocompatibility testing was conducted on the subject device to confirm that it meets current biocompatibility requirements. The tests included cytotoxity, intracutaneous irritation, materials mediated pyrogen, sensitization and systemic toxicity. All tests were successfully passed.
#### Sterilization/Shelf-Life
The device was validated to achieve a sterility assurance level of 10°.
#### Performance Testing
Testing was completed for the subject device to evaluate the ability of the B7-2C to successfully occlude airways. All tests (N=30) showed complete occlusion of the airway. Based on the statistical plan, this shows that the B7-2C balloon is capable of occluding airways at 90% conformance and 95% confidence. All acceptance criteria defined in the protocol was met.
Verification and burst testing bench was completed on the predicate device to confirm that the performance of the device conforms to mechanical and functional specifications. The testing included:
- Verification ●
- . Burst testing
## Conclusion (Statement of Equivalence)
As confirmed in the prior 510(k) for the predicate device, which is identical to the subject device, the device successfully passed all performance testing. For the subject device, biocompatibility testing was successfully passed, as well as bench testing to verify the ability of the device to successfully occlude airways.
As stated above, the intended use of the subject device is virtually identical to that of the predicate device with respect to bronchial hemostasis. Whereas the predicate device uses pressure against the airway walls to support hemostasis, the proposed device uses pressure against the airway to allow the physician to localize air leaks within the lung.
Since the modified device name and Indication for Use are the only changes, and the proposed Indication for Use is within the original Intended Use, we believe that the information provided supports a determination of substantial equivalence to the predicate device, and market clearance of the B7-2C Occlusion Balloon.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.