OLYMPUS B5-2C/B7-2C BALLOON CATHETER

K962901 · Olympus America, Inc. · EOQ · Aug 20, 1996 · Ear, Nose, Throat

Device Facts

Record IDK962901
Device NameOLYMPUS B5-2C/B7-2C BALLOON CATHETER
ApplicantOlympus America, Inc.
Product CodeEOQ · Ear, Nose, Throat
Decision DateAug 20, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4680
Device ClassClass 2
AttributesTherapeutic

Indications for Use

B5-2C Balloon Catheter (Color of Branch - Blue) The B5-2C Balloon Catheter is designed to be used as an accessory to Olympus endoscopes with the minimum channel size of 2.0 mm. B7-2C Balloon Catheter (Color of Branch - Yellow) The B7-2C Balloon Catheter is designed to be used as an accessory to Olympus endoscopes with the minimum channel size of 2.8 mm.

Device Story

Balloon catheter accessory for Olympus endoscopes; introduced through instrument channel to bronchial site. Components: inflatable balloon, catheter, bifurcation, irrigation port, air feed cap, stopcock. B5 model (5 French) for 2.0mm channels; B7 model (7 French) for 2.8mm channels. Used by physicians during bronchoscopy for diagnostic imaging (bronchography), therapeutic delivery (medicine/saline), foreign body retrieval, and hemostasis via tamponade. Color-coded branches (blue/yellow) for identification. 1050mm length.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Catheter-based balloon system; 5 French (B5) and 7 French (B7) diameters; 1050 mm length. Manual inflation/irrigation via stopcock and port. Color-coded bifurcation branches. Compatible with Olympus endoscopes (2.0mm or 2.8mm channels).

Indications for Use

Indicated for selective endoscopic bronchography, bronchial medicine injection, saline injection for bronchoalveolar lavage, bronchial foreign body removal, and bronchial hemostasis (tamponade) in patients undergoing bronchoscopy.

Regulatory Classification

Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

Predicate Devices

Submission Summary (Full Text)

{0} K962901 # AUG 20 1996 510 (k) SUMMARY ## OLYMPUS B5-2C/B7-2C BALLOON CATHETER | Device Name: | Olympus B5-2C/B7-2C Balloon Catheters | | --- | --- | | Common/Usual Name: | Balloon Catheter | | Classification Name: | Bronchoscope and Accessories | | Predicate Devices: | Olympus B5-2C/B7-2C Biliary Balloon Catheter (K904669) | | Submitted By: (Contact Person) | Mr. Barry Sands Olympus America Inc. Endoscope Division Two Corporate Center Drive Melville, New York 11747-3157 (516) 844-5474 | | Summary Preparation Date: | July 9, 1996 | ## Statement of Intended Use **B5-2C Balloon Catheter (Color of Branch - Blue)** The B5-2C Balloon Catheter is designed to be used as an accessory to Olympus endoscopes with the *minimum* channel size of 2.0 mm. **B7-2C Balloon Catheter (Color of Branch - Yellow)** The B7-2C Balloon Catheter is designed to be used as an accessory to Olympus endoscopes with the *minimum* channel size of 2.8 mm. ## Indications for Use: **B5-2C/B7-2C Balloon Catheters (Two Models)** Olympus B5-2C/B7-2C Balloon Catheters are to be used for selective endoscopic bronchography, medicine injection to bronchi, saline injection associated with bronchoalveoler lavage, foreign body removal from bronchi, and bronchial hemostasis (tamponade). ## Device Description The subject balloon catheters are designed to be used as an accessory to Olympus endoscopes. It is introduced to the desired anatomical site through the instrument channel of the endoscope. 29 {1} The subject device is constructed of the inflatable balloon, catheter, bifurcation, irrigation port, air feed cap, and stopcock. The B5 series balloon catheter is 5 French in diameter and designed to be used with minimum 2.0 mm channel endoscopes, while the B7 series balloon catheter is 7 French in diameter and should be used with minimum 2.8 mm channel size endoscopes. They are identified then by a color code. For example, B5 series have been identified with the blue color branch and B7 series catheters have the yellow color branch. The length of both catheters is 1050 mm. ## General Safety In conclusion, when compared with the predicate biliary balloon (K904669), the subject Olympus B5-2C/B7-2C Balloon Catheter does not incorporate any significant change in the design, specifications, or method of operation that could affect the safety or efficacy of the subject devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...