AT-6XX Series Infrared Thermometer

K171455 · Xiamen Acurio Instruments Co., Ltd. · FLL · Feb 9, 2018 · General Hospital

Device Facts

Record IDK171455
Device NameAT-6XX Series Infrared Thermometer
ApplicantXiamen Acurio Instruments Co., Ltd.
Product CodeFLL · General Hospital
Decision DateFeb 9, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The AT-6XX Series Infrared Thermometer (with models AT-601 and AT-606) is intended for body temperature measurement from the auditory canal for infants and adults without contact to human body. It can be used by consumers in household environment and doctor in clinic as reference.

Device Story

Electronic infrared thermometer (models AT-601, AT-606) measures body temperature via auditory canal. Device uses infrared thermopile sensor to detect radiated infrared energy from tympanic membrane; converts energy intensity into electronic signal; processes signal via internal circuitry to calculate temperature; displays result on LCD. Operated by consumers or clinicians; probe inserted into outer ear canal; activation button initiates measurement. Provides rapid, non-invasive temperature readings to assist in clinical decision-making and patient monitoring.

Clinical Evidence

Clinical accuracy evaluation conducted per clause 201.102 of ISO 80601-2-56:2009. Results demonstrated performance substantially equivalent to the predicate device.

Technological Characteristics

Infrared thermopile sensor; ABS plastic shell; stainless steel sensor tip. Powered by DC 3.0V batteries (CR2032 or AAA). Measurement range 34°C–42.2°C; accuracy ±0.2°C (35°C–42°C). Standalone device; LCD display. Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, and ISO 80601-2-56.

Indications for Use

Indicated for body temperature measurement via the auditory canal in infants and adults. Intended for intermittent measurement and monitoring of body temperature in household and clinical settings.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # 510(K) SUMMARY This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.87(h) and 21 CFR 807.92. # Submitter: Xiamen Acurio Instruments Co., Ltd 3rd-4th Floors, Biomedical Industry Park, No.2028, Wengjiao West Road, Haicang District, Xiamen, Fujian, 361026, CHINA #### Contact Person: ● Angel Liu General Manager Assistant Tel: 86-592-6312776 Fax: 86-592-6312776 Email: liushuzhen@acurio.cn #### ● Date Prepared: May 12, 2017 # Name of the Devices: - Device Common Name: Infrared Thermometer ● - Device Proprietary Name: AT-6XX Series Infrared Thermometer (with models AT-601 and AT-606) - Classification Name: Thermometer, electronic, clinical - Regulation Number:21CFR 880.2910 - Product Code: FLL - Review Panel: General Hospital ### Legally Marketed Predicate Device(s): K131243 Non-contact Infrared Thermometer, Model JPD-FR 100 # Indications for Use: The AT-6XX Series Infrared Thermometer (with models AT-601 and AT-606) is {2}------------------------------------------------ intended for body temperature measurement from the auditory canal for infants and adults without contact to human body. It can be used by consumers in household environment and doctor in clinic as reference. # Device Description: The AT-6XX Series Infrared Thermometer (with models AT-601 and AT-606) is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the auditory canal. It uses infrared probe to detect the radiated infrared energy emitted by the object, solid, liquid or gas. The intensity of the emitted energy depends on the temperature of the object. The infrared probe (Thermopile) can recognize it and transfer to the proper electronic signal. The electronic signal can be processed in the Infrared Thermometer to convert to the temperature reading, which is displayed on the LCD. Therefore, the subject device is able to measure the temperature of a person by the energy the person emits. The thermometer consists mainly of two parts - an IR sensor with a built-in ambient temperature sensor and the associated circuit. To measure body temperature, the probe of the thermometer is inserted into a patient's outer ear canal. Pressing the activation button to start the measurement of target's infrared radiation. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The final measured temperature will be appeared on a LCD display. The total operation takes a few seconds. # Comparison to Predicate Devices: The Substantial Equivalence Comparison Chat is provided as follows: | Comparison items | AT-601 infrared<br>thermometer | JPD-FR100 infrared<br>dual-mode thermometer | Comparison<br>results | |---------------------|--------------------------------------------------------------------|------------------------------------------------------------|-----------------------| | Name | Infrared thermometer | Infrared dual-mode<br>thermometer | Similar | | Model | AT-601 | JPD-FR100 | | | Manufacturer | SHENZHEN ACURIO<br>INSTRUMENTS CO., LTD. | Shenzhen Jumper Medical<br>Equipment Co., Ltd. | | | Management category | Ila | Ila | Identical | | 510(K) Number | / | K131243 | N/A | | CE Number | / | | | | Structure | Mini body part, probe, lens<br>filter, LCD screen, battery<br>door | Body part, probe, lens filter,<br>LCD screen, battery door | Similar | Table1: Substantial Equivalence Comparison Chat between AT-601 and JPD-FR 100 {3}------------------------------------------------ | | Comparison items | AT-601 infrared<br>thermometer | JPD-FR100 infrared<br>dual-mode thermometer | Comparison<br>results | |------------------------------------------------|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------| | | Materials | ABS plastic is used on the<br>shell, and stainless steel<br>material is used on the<br>temperature sensor. | ABS plastic is used on the<br>shell, and stainless steel<br>material is used on the<br>temperature sensor. | Similar | | | Basic operating principle | The product provides<br>noninvasive and rapid body<br>temperature measurement by<br>detecting the infrared energy<br>emitted from the tympanic<br>membrane, and shows the<br>results by the LCD. | The product provides<br>noninvasive and rapid body<br>temperature measurement by<br>detecting the infrared energy<br>emitted from the tympanic<br>membrane, or forehead and<br>shows the results by the LCD. | Same | | | Battery | D.C.3.0V, CR2032 | D.C. 3.0V, two 1.5V AAA<br>batteries | Similar | | | Display unit | 0.1 °C | 0.1 °C | Identical | | | Measuring<br>range | 34°C~42.2°C | Forehead mode: 32°C~42.2°C<br>Ear mode: 0°C~100°C | | | Primary<br>technical<br>performance<br>indexes | Accuracy | $±$ 0.2°C (35°C~42°C), $±$<br>0.3°C (outside the range of<br>35°C~42°C) | Ear mode: $±$ 0.2°C<br>Forehead mode: $±$ 0.3°C | Similar | | | Memory<br>function | 9 sets of storage data | 20 sets of storage data | | | | Dimensions<br>(LWH) | 58mm×58mm×66mm | N/A | N/A | | | Weight | Approx. 80g (including<br>batteries ) | N/A | N/A | | | Intended use | It is intended for the<br>intermittent measurement<br>and monitoring of body<br>temperature. | It is intended for the<br>intermittent measurement and<br>monitoring of body<br>temperature. | Same | | Measuring site | | Ear | Ear, forehead | Same<br>when<br>measuring ear<br>temperature | # Table1: Substantial Equivalence Comparison Chat between AT-606 and JPD-FR 100 | Comparison items | AT-606 infrared<br>thermometer | JPD-FR100 infrared<br>dual-mode thermometer | Comparison<br>results | | |------------------------------------------------|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------| | Name | Infrared thermometer | Infrared<br>dual-mode<br>thermometer | Same | | | Model | AT-606 | JPD-FR401 | | | | Manufacturer | ACURIO<br>SHENZHEN<br>INSTRUMENTS CO., LTD. | Shenzhen<br>Jumper<br>Medical<br>Equipment Co., Ltd. | | | | Management category | Ila | Ila | Identical | | | 510(K) Number | | K131243 | N/A | | | | Comparison items | AT-606 infrared thermometer | JPD-FR100 infrared dual-mode thermometer | Comparison results | | | CE Number | / | | | | | Structure | Body part, probe, lens filter, LCD screen, battery door | Body part, probe, lens filter, LCD screen, battery door | Similar | | | Materials | ABS plastic is used on the shell, and stainless steel material is used on the temperature sensor. | ABS plastic is used on the shell, and stainless steel material is used on the temperature sensor. | Similar | | | Basic operating principle | The product provides noninvasive and rapid body temperature measurement by detecting the infrared energy emitted from the tympanic membrane, and shows the results by the LCD. | The product provides noninvasive and rapid body temperature measurement by detecting the infrared energy emitted from the tympanic membrane or forehead, and shows the results by the LCD. | Same | | | Battery | D.C.3.0V, 2*AAA | D.C. 3.0V, two 1.5V AAA batteries | Similar | | | Display unit | 0.1 °C | 0.1 °C | Identical | | Primary<br>technical<br>performance<br>indexes | Measuring range | 34°C~42.2°C | Forehead mode: 32°C~42.2°C<br>Ear mode: 0°C~100°C | | | | Accuracy | ± 0.2°C (35°C~42°C), ± 0.3°C (outside the range of 35°C~42°C) | Ear mode: ± 0.2°C<br>Forehead mode: ± 0.3°C | Similar | | | Memory function | 9 sets of storage data | 20 sets of storage data | | | | Dimensions (LWH) | 137mmX40mmX56mm | N/A | N/A | | | Weight | Approx. 80g (including batteries) | N/A | N/A | | | Intended use | It is intended for the intermittent measurement and monitoring of body temperature. | It is intended for the intermittent measurement and monitoring of body temperature. | Identical | | Measuring site | | Ear | Ear, forehead | Similar when measuring ear temperature | {4}------------------------------------------------ ### 510(K) Summary # Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence: The Sponsor performed a risk assessment based on the differences from the predicate device and determined that the following non-clinical testing was necessary to demonstrate that the changes did not introduce any new risks to the subject device. {5}------------------------------------------------ Laboratory testing was conducted to validate and verify that the thermometers continued to meet all requirements of related international standards, including electrical safety, EMC, software, biocompatibility. Results of these tests demonstrated compliance to the requirements of the below consensus standards and FDA Guidance documents. # Electrical Safety and Performance: - 1. IEC 60601-1:2005+A1:2012; - 2. ISO 80601-2-56:2009. Medical Electrical Equipment and medical Electrical Systems Used in the Home Healthcare Environment: 1. IEC 60601-1-11:2010+CORR.1:2011. Electromagnetic Compatibility: 1. IEC 60601-1-2:2007. ## Biocompatibility: - 1. ISO 10993-5:2009 - 2. ISO 10993-10:2010 ## Software: - 1. General Principles of Software Validation Final Guidance for Industry and FDA Staff. # Clinical Electronic Thermometers Guidance: - 1. Guidance on the content of Premarket Notification[510(k)] Submission for clinical electronic thermometers Therefore, we conclude that the AT-6XX Series Infrared Thermometer (with models AT-601 and AT-606) is substantially equivalent to the predicate device. # Discussion of Clinical Tests Performed: In order to demonstrate that the subject device did not introduce any new risks in comparison to the subject device a clinical accuracy evaluation has been conducted according to clause 201.102 of ISO 80601-2-56:2009. Clinical test results demonstrated that the subject device performs substantially equivalent to the predicate device. # Conclusions: The AT-6XX Series Infrared Thermometer (with models AT-601 and AT-606) has the same intended use and similar characteristics as the predicate device. And based on performance data, we conclude that the AT-6XX Series Infrared Thermometer (with models AT-601 and AT-606) is substantially equivalent to the predicate device. {6}------------------------------------------------ # Please wait... If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document. You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download. For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader. Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries. {7}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 9, 2018 Xiamen Acurio Instruments Co., Ltd Angel Liu General Manager Assistant 3rd-4th Floors, BI Park, No.2028 Wengjiao West Road, Haicang District Xiamen, Fujian 361026 CHINA Re: K171455 Trade/Device Name: AT-6XX Series Infrared Thermometer (with models AT-601 and AT-606) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 26, 2017 Received: January 2, 2018 Dear Angel Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {8}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...