K171391 · Stryker · GEI · Jan 26, 2018 · General, Plastic Surgery
Device Facts
Record ID
K171391
Device Name
Stryker 90-S Max SERFAS Energy Probe
Applicant
Stryker
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 26, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker SERFAS 90-S Max electrosurgical probe is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic and arthroscopic procedures of joints such as the knee, shoulder, elbow, hip, ankle, and wrist.
Device Story
Disposable, single-use electrosurgical RF probe; accessory to SERFAS Energy System and Crossfire Arthroscopy System. Used in arthroscopic procedures for soft tissue resection, ablation, coagulation, and hemostasis. Operated by surgeons in clinical settings. Device delivers RF energy to tissue via probe tip; enables removal of tissue and coagulation of bleeding vessels. Benefits include precise tissue management during minimally invasive orthopedic surgery.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included aggressive use, tip cantilever, torsion, bending moment, impact, electrode pull, heat resistance, captured tip strength, leak testing, heat shrink strength, mechanical force, shaft compression, and thermal damage assessment. All tests passed.
Technological Characteristics
Disposable RF electrosurgical probe; sterile (Ethylene Oxide). Features modified probe tip (electrode and ceramic) and outer lumen heat shrink. Operates as accessory to RF energy systems for soft tissue ablation/coagulation. No software or active electronic components within the probe itself.
Indications for Use
Indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic and arthroscopic procedures of joints including knee, shoulder, elbow, hip, ankle, and wrist.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 26, 2018
Stryker Benjamin Krebs Quality Engineer 5900 Optical Court San Jose, California 95138
Re: K171391
Trade/Device Name: Stryker 90-S Max SERFAS Energy Probe Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: May 9, 2017 Received: May 11, 2017
Dear Benjamin Krebs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
X171391 510(k) Number (if known)
Stryker SERFAS 90-S Max Electrosurgical Probe Device Name
hip, ankle, and wrist. super in as into is a your studiose in a first success in a since success and the sizes show. of the for notising to be noticess in the is in is is in it is in the in the more the more of the sett Indications for Use (Describe)
| | | Over-The-Counter Use ()Se (Frange | | Subgat C) ( ) the grass ( Use Hart 21 PFR 80 Suppated
Type of Use (Select one or both, as applicable)
. See to A notisuber drowteds and to stremeritypes of your selligge notices and CONTINUE OF A SEPARATE PAGE IF NEEDED.
#### A WOJE SERICA JIAMS FRATS ARA FHT OT MAOR CETTINGS AUOY CHES TON OO*
of this information collection, including suggestions for reducting to: to the securities nebrud sint pribrished stremmos bres. notitemori to noitemorini to noitest on the see of the beben a be be niemiem bus not of projects of the projective do really ent
PRASItaff@fdda.hhs.gov Paperwork Reduction Act (PRA) Stat Office of Chief Information Officer Food and Drug Administration Destment of Health and Human Services
information unless it displays a currently voile OMB number. to noitsellos 6, ot broqzer of beniuper in nozneq s bris 10, 2011 yem van van van van variene and
See PRA Statement below. Expiration Date: 06/30/2020
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
# 1. General Information
| 510(k) Sponsor | Stryker Corporation |
|-------------------------|------------------------------------------------------------------------------|
| Address | 5900 Optical Court<br>San Jose, CA 95138 |
| FDA Registration Number | 2936485 |
| Correspondence Person | Mr. Benjamin Krebs<br>Quality Engineer<br>Stryker Corporation |
| Contact Information | Email: ben.krebs@stryker.com<br>Phone: (408) 754-2897<br>Fax: (408)-754-2969 |
| Date Prepared | January 25, 2018 |
# 2. Device Identification
| Proposed Device: | |
|---------------------|----------------------------------------------------------------|
| Proprietary Name | Stryker® SERFAS 90-S Max Electrosurgical Probe |
| Common Name | RF Probe |
| Classification Name | Electrosurgical Cutting and Coagulation Device and Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Product Code | GEI |
| Regulatory Class | II |
# Primary Predicate Device:
| Proprietary Name | Stryker® SERFAS 90-S Electrosurgical Probe |
|------------------------|----------------------------------------------------------------|
| Common Name | RF Probe |
| Premarket Notification | K160050 |
| Classification Name | Electrosurgical Cutting and Coagulation Device and Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Product Code | GEI |
| Regulatory Class | II |
{4}------------------------------------------------
| Reference Device: | |
|------------------------|----------------------------------------------------------------|
| Proprietary Name | SERFAS Energy System |
| Common Name | RF Probe |
| Premarket Notification | K041810 |
| Classification Name | Electrosurgical Cutting and Coagulation Device and Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Product Code | GEI |
| Regulatory Class | II |
### Device Description
The Stryker SERFAS 90-S Max electrosurgical probe (hereafter referred to as "Proposed device") is an accessory to the SERFAS Energy System, marketed through K041810 and K160050, and the Crossfire Arthroscopy System, marketed through K071859, which is intended for resection, ablation, and coagulation of soft tissue via radiofrequency (RF) ablation. RF ablation probes are the main tool used in most arthroscopic procedures for the removal of tissue and the coagulation of bleeding vessels. The Proposed device is a disposable single-use electrosurgical device provided sterile via Ethylene Oxide sterilization.
# 3. Indications for Use
The Stryker SERFAS 90-S Max electrosurgical probe is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic and arthroscopic procedures of joints such as the knee, shoulder, elbow, hip, ankle, and wrist.
# 4. Comparison of Technological Characteristics with the Predicate Device
The Stryker SERFAS 90-S Max electrosurgical probe is similar to the predicate device with respect to technological characteristics and design. The modifications of the Proposed device, listed below, are minor and testing confirms that these modifications do not raise any new questions of safety or effectiveness, therefore supporting that the SERFAS 90-S Max electrosurgical probe is substantially equivalent to the predicate device currently on the market.
The following differences exist between the Proposed device and the predicate device:
- Probe Tip (Electrode and Ceramic) Design ●
- . Outer Lumen Heat Shrink Material
{5}------------------------------------------------
## 5. Performance Testing
The Stryker SERFAS 90-S Max electrosurgical probe was tested for performance in accordance with internal design specifications and with the applicable performance standards, which established the substantial equivalence determination. The following non-clinical tests were conducted and are summarized in the table below:
| Test Name | Description | Results |
|-------------------|-----------------------------------------------------------------------------------------------------|---------|
| Aggressive Use | Determines if a probe can last its full lifetime without<br>failure | Pass |
| Tip Cantilever | Determines if the probe tip can withstand the normal and<br>side force | Pass |
| Torsion in a Slot | Determines if the probe tip can withstand torsional forces | Pass |
| Bending Moment | Verifies failure mode while probe is used in prying<br>manner | Pass |
| Impact | Determines a probe can survive an impact of a hard object | Pass |
| Electrode Pull | Determines the force at which the electrode is pulled out of<br>the probe tip assembly | Pass |
| Heat | Determines if the probe tip can withstand extreme<br>temperatures | Pass |
| Captured Tip | Determines strength of probe tip assembly | Pass |
| Leak | Determines if probe will allow leakage. | Pass |
| Heat Shrink | Determines mechanical strength of heat shrink | Pass |
| Mechanical Force | Determines if the probe can withstand representative<br>compressive force on shaft | Pass |
| Shaft Compression | Determines if the probe can withstand representative<br>compressive force on shaft | Pass |
| Thermal Damage | Determines the thermal effect on tissue of the proposed<br>device compared to the reference device. | Pass |
Bench performance testing was conducted to ensure that the device functioned as intended and met design specifications and acceptance criteria.
### 6. Conclusion
Based on the intended use, technological characteristics, performance testing, and non-clinical testing in comparison to the predicate device, the Stryker SERFAS 90-S Max electrosurgical probe does not raise any new questions of safety and effectiveness, therefore demonstrating substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.