Medline ReNewal Reprocessed Stryker SERF AS Energy Probes
K172608 · Surgical Instrument Service and Savings, Inc. · NUJ · Oct 18, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172608
Device Name
Medline ReNewal Reprocessed Stryker SERF AS Energy Probes
Applicant
Surgical Instrument Service and Savings, Inc.
Product Code
NUJ · General, Plastic Surgery
Decision Date
Oct 18, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medline ReNewal Reprocessed Stryker SERFAS Energy Probe is a disposable, radio frequency probe used in electrosurgical procedures for resection, ablation, and coagulation of soft tissue, as well as the hemostasis of blood vessels in patients undergoing arthroscopic surgery of the knee, shoulder, ankle, hip, elbow, and wrist.
Device Story
Disposable radio frequency (RF) probe for electrosurgical procedures; used in arthroscopic surgery (knee, shoulder, ankle, hip, elbow, wrist). Device inputs: RF energy from compatible generator; manual control via integrated finger switches or foot switches. Device transforms electrical energy to perform soft tissue resection, ablation, coagulation, and hemostasis; includes suction line for removal of tissue/fluids. Operated by surgeons in clinical/OR settings. Output: thermal energy for tissue effect; suction for site clearance. Benefits: enables minimally invasive arthroscopic tissue management. Reprocessed device undergoes cleaning, disinfection, inspection, refurbishment, testing, and sterilization.
Radio frequency electrosurgical probe; disposable. Features integrated finger switches and suction line. Compatible with Stryker SERFAS and Crossfire energy systems. Standards: IEC 60601-1 (general safety), IEC 60601-2-2 (electrosurgical safety), IEC 60601-1-2 (EMC). Sterilization: validated process for reprocessed single-use devices.
Indications for Use
Indicated for patients undergoing arthroscopic surgery of the knee, shoulder, ankle, hip, elbow, and wrist requiring soft tissue resection, ablation, coagulation, or hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health and Human Services logo, and on the right, there is the FDA logo. The FDA logo is in blue and consists of the letters "FDA" followed by the words "U.S. Food & Drug Administration".
October 18, 2017
Surgical Instrument Service and Savings (dba Medline ReNewal) Ms. Stephanie Boyle Mays Regulatory Specialist, Quality Assurance and Regulatory Affairs 1500 NE Hemlock Ave. Redmond, Oregon 97756
Re: K172608
Trade/Device Name: Medline ReNewal Reprocessed Stryker SERF AS Energy Probes (Models:279-351-100, 279-351-230, 279-351-250, 279-351-300,279-351-400, and 279-401-100) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NUJ Dated: August 30, 2017 Received: August 31, 2017
Dear Ms. Mays:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{2}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172608
#### Device Name
Medline ReNewal Reprocessed Stryker SERFAS Energy Probes (models: 279-351-230, 279-351-250, 279-351-300, 279-351-400, and 279-401-100)
#### Indications for Use (Describe)
The Medline ReNewal Reprocessed Stryker SERFAS Energy Probe is a disposable, radio frequency probe used in electrosurgical procedures for resection, ablation of soft tissue, as well as the hemostasis of blood vessels in patients undergoing arthroscopic surgery of the knee, shoulder, ankle, hip, elbow, and wrist.
Type of Use (Select one or both, as applicable)
| <div> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|--------------------------------------------------------------------------------|
| <div> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Device<br>Model | Device Name | Original<br>Manufacturer |
|-----------------|--------------------------------------------------------|--------------------------|
| 279-351-100 | 90-S SERFAS Energy Suction Probe<br>3.5 mm | Stryker |
| 279-351-230 | 30-S SERFAS Energy Suction Probe<br>3.5 mm | Stryker |
| 279-351-250 | 50-S SERFAS Energy Suction Probe<br>3.5 mm | Stryker |
| 279-351-300 | Super 90-S SERFAS Energy Suction Probe<br>3.5 mm | Stryker |
| 279-351-400 | 90-S Accelerator SERFAS Energy Suction Probe<br>3.5 mm | Stryker |
| 279-401-100 | 90-S MAX SERFAS Energy Suction Probe<br>4.0 mm | Stryker |
Reprocessed Single-Use Device Models Included in Clearance:
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Medline Renewal logo. The word "Medline" is in blue, with a blue symbol to the left of it. To the right of "Medline" is a vertical line, followed by the word "Renewal" in green and blue. Below "Renewal" is the phrase "Full Circle Reprocessing" in blue.
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR § 807.92.
| Submitter/<br>Owner | Surgical Instrument Service and Savings (dba Medline ReNewal)<br>1500 NE Hemlock Ave.<br>Redmond, OR 97756 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | August 30, 2017 |
| Contact<br>Name/Prepared<br>by | Stephanie Boyle Mays<br>Regulatory Affairs Specialist, Regulatory Affairs<br>P: 541-516-4205/F: 541-923-3375<br>E: smays@medline.com |
| Device Name<br>and<br>Classification | Propriety/Trade Name:<br>Medline ReNewal Reprocessed Stryker<br>SERFAS Energy Probes, models: 279-351-100,<br>279-351-230, 279-351-250, 279-351-300, 279-<br>351-400, and 279-401-100<br>Common Name:<br>Electrosurgical cutting and coagulation device<br>and accessories<br>Classification:<br>Class II<br>Regulation Number:<br>21 CFR § 878.4400<br>Product Code:<br>NUJ<br>Panel:<br>General & Plastic Surgery |
| 510(k) Number | K041810 |
| | Propriety/Trade Name:<br>Stryker SERFAS Energy Probes, models: 279-<br>351-100, 279-351-230, 279-351-250, 279-351-<br>300, 279-351-400, and 279-401-100 |
| Predicate<br>Device and<br>Classification | Common Name:<br>Electrosurgical cutting and coagulation device<br>and accessories<br>Classification:<br>Class II<br>Regulation Number:<br>21 CFR § 878.4400<br>Product Code:<br>GEI<br>Panel:<br>General & Plastic Surgery<br>Manufacturer:<br>Stryker Endoscopy, 5900 Optical Ct., San Jose,<br>CA 95138 |
| Intended Use | The Medline ReNewal Reprocessed Stryker SERFAS Energy Probe is a<br>disposable, radio frequency probe used in electrosurgical procedures for<br>resection, ablation, and coagulation of soft tissue, as well as the<br>hemostasis of blood vessels in patients undergoing arthroscopic surgery of<br>the knee, shoulder, ankle, hip, elbow, and wrist. |
| Product<br>Description | The Medline ReNewal Reprocessed Stryker SERFAS Energy Probes are<br>single-use devices that have been cleaned, disinfected, inspected,<br>refurbished, tested, packaged, labeled, and sterilized. All models can be<br>operated with integrated finger switches on the device handle to control<br>ablation, coagulation and power level. Each model also has a suction line<br>that removes tissue and fluids from the surgical site. In addition, the models<br>can be operated with foot switches, which offer another mode with which to |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Medline Renewal logo. The Medline logo is on the left, with the word "MEDLINE" in blue and a blue star symbol. To the right of the Medline logo is the word "Renewal" in green and blue, with the words "Full Circle Reprocessing" in blue below it.
| Device<br>Characteristics | Predicate | Proposed | Comparison |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| 510(k) | Stryker SERFAS Energy<br>Probes | Medline ReNewal<br>Reprocessed Stryker<br>SERFAS Energy Probes | As Stated |
| 510(k) | K041810 | TBD | Not applicable |
| Model Numbers | • 279-351-100<br>• 279-351-230<br>• 279-351-250<br>• 279-351-300<br>• 279-351-400<br>• 279-401-100 | • 279-351-100<br>• 279-351-230<br>• 279-351-250<br>• 279-351-300<br>• 279-351-400<br>• 279-401-100 | Same |
| Intended Use | The SERFAS Energy<br>Probe is a disposable,<br>radio frequency probe used<br>in electrosurgical<br>procedures for resection,<br>ablation, and coagulation<br>of soft tissue, as well as<br>the hemostasis of blood<br>vessels in patients<br>undergoing arthroscopic<br>surgery of the knee,<br>shoulder, ankle, hip, elbow,<br>and wrist.<br>The probe includes an<br>energy-transferring cable<br>as well as several different<br>tip configurations and<br>suction probes which are<br>capable of providing<br>simultaneous fluid<br>aspirations | The Medline ReNewal<br>Reprocessed SERFAS<br>Energy Probe is a<br>disposable, radio<br>frequency probe used in<br>electrosurgical procedures<br>for resection, ablation, and<br>coagulation of soft tissue,<br>as well as the hemostasis<br>of blood vessels in patients<br>undergoing arthroscopic<br>surgery of the knee,<br>shoulder, ankle, hip, elbow,<br>and wrist.<br>The probe includes an<br>energy-transferring cable<br>as well as several different<br>tip configurations and<br>suction probes which are<br>capable of providing<br>simultaneous fluid<br>aspirations. | Same |
| 510(k) Substantial Equivalence Chart (concluded) | | | |
| | Predicate | Proposed | Comparison |
| Device<br>Characteristics | Stryker SERFAS Energy<br>Probes | Medline ReNewal<br>Reprocessed Stryker<br>SERFAS Energy Probes | As Stated |
| Product<br>Description | The Stryker SERFAS<br>Energy Probes are single-<br>use devices. All models<br>can be operated with<br>integrated finger switches<br>on the device handle to<br>control ablation,<br>coagulation and power<br>level. Each model also has<br>a suction line that removes<br>tissue and fluids from the<br>surgical site. In addition,<br>the models can be<br>operated with foot<br>switches, which offer<br>another mode with which to<br>control ablation,<br>coagulation, and the power<br>setting. The devices are<br>connected to the<br>appropriate generator by a<br>connector cable. | The Medline ReNewal<br>Reprocessed Stryker<br>SERFAS Energy Probes<br>are single-use devices that<br>have been cleaned,<br>disinfected, inspected,<br>refurbished, tested,<br>packaged, labeled, and<br>sterilized. All models can<br>be operated with integrated<br>finger switches on the<br>device handle to control<br>ablation, coagulation and<br>power level. Each model<br>also has a suction line that<br>removes tissue and fluids<br>from the surgical site. In<br>addition, the models can<br>be operated with foot<br>switches, which offer<br>another mode with which to<br>control ablation,<br>coagulation, and the power<br>setting. The devices are<br>connected to the<br>appropriate generator by a<br>connector cable. The<br>generators and the foot<br>switches are not included<br>in the scope of this project,<br>and they will not be<br>reprocessed. | As stated |
| Power Platform | Stryker SERFAS<br>● Energy System<br>Stryker Crossfire<br>● System | Stryker SERFAS<br>● Energy System<br>Stryker Crossfire<br>● System | Same |
| Technological<br>Characteristics | The technological characteristics and the fundamental<br>scientific technology of the subject devices are identical<br>to the predicate device. The proposed devices are a<br>reprocessed version of the predicate K041810 Stryker<br>SERFAS Energy Probes. The predicate devices were<br>used to support intended use, technological<br>characteristics, and functional performance<br>specifications. | | Same |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medline Renewal logo. The Medline logo is on the left, and the Renewal logo is on the right. The Renewal logo is green and blue, and the words "Full Circle Reprocessing" are below it.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo for Medline Renewal Full Circle Reprocessing. The Medline logo is on the left, and the Renewal logo is on the right. The Medline logo is a blue star with the word "MEDLINE" in blue letters. The Renewal logo is in green and blue letters, and the words "Full Circle Reprocessing" are in smaller blue letters below the Renewal logo. The logo is clean and professional.
| <b>Performance<br/>Testing</b> | The functional characteristics of the proposed Medline ReNewal<br>Reprocessed Stryker SERFAS Energy Probe have been evaluated and<br>found to be equivalent to the Stryker SERFAS Energy Probe predicate<br>devices based on the following tests: |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | sterilization validation; |
| | biocompatibility; cytotoxicity, sensitization, irritation; pyrogenicity, and acute systemic toxicity; |
| | electrical testing electromagnetic compatibility (in accordance with IEC 60601-1-2); electrical safety (in accordance with IEC 60601-1 and IEC 60601-2-2); and basic safety (in accordance with IEC 60601-1 and IEC 60601-2-2) |
| | performance qualification: simulated use; critical function bending equivalence test; critical function suction equivalence test; critical function thermal tissue damage equivalence test; critical function drop equivalence test; critical function device equivalence integrity; and product stability |
| | cleaning; protein, and hemoglobin. |
| | <b>Conclusion</b> |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.