K171299 · Aju Pharm Co., Ltd. · MAI · Aug 24, 2017 · Orthopedic
Device Facts
Record ID
K171299
Device Name
Fixone Biocomposite Anchor
Applicant
Aju Pharm Co., Ltd.
Product Code
MAI · Orthopedic
Decision Date
Aug 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fixone Biocomposite Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures: Shoulder: Rotor Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction; Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair; Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis; Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
Device Story
Fixone Biocomposite Anchor is a manual orthopedic fixation device used to reattach soft tissue to bone. The system consists of a biodegradable anchor (PLGA copolymer and β-TCP) and a non-resorbable suture (cobraided UHMWPE and PET fibers) delivered via a dedicated driver. The device is used by surgeons in clinical settings (OR) to perform procedures such as rotator cuff repair, Bankart repair, and ligament reconstructions. The surgeon manually implants the anchor into the bone to secure the suture, which then holds the soft tissue in place. This mechanical fixation facilitates tissue healing. The device is provided non-sterile and requires autoclave sterilization by the user before use.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993 series: cytotoxicity, systemic toxicity, pyrogenicity, sensitization, genotoxicity, implantation), mechanical testing (fixation strength per ASTM F2502), and chemical characterization (EO residues, heavy metals, pH, UV spectrum). All tests passed, demonstrating performance equivalent to predicate devices.
Technological Characteristics
Materials: PLGA (70%) and β-TCP (30%) anchor; UHMWPE/PET suture. Principle: Manual screw-in bone fixation. Dimensions: 4.5mm to 6.5mm diameters. Sterilization: User-performed autoclave (134°C/20 min). Standards: ASTM F2502, ISO 10993, ISO 11135.
Indications for Use
Indicated for soft tissue-to-bone fixation in shoulder, foot/ankle, knee, and elbow orthopedic procedures, including ligament and tendon repairs.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
AJU Pharm Co., Ltd. % Mr. Peter Chung President Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
August 24, 2017
Re: K171299
Trade/Device Name: Fixone Biocomposite Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: July 10, 2017 Received: July 17, 2017
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{2}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K171299
Device Name Fixone Biocomposite Anchor
The Fixone Biocomposite Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures:
Shoulder: Rotor Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction;
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, lliotibial Band Tenodesis;
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
## 510(k) Summary
### [as required by 807.92(c)]
#### 1. Applicant
- Company : AJU Pharm Co.,Ltd. 1)
- 2) Address : A-207, 697, Pangyo-ro, Seongnam-si, Gyeonggi-do, Korea
- 3) Tel : 82-31-765-4420
- Fax : 82-31-602-7818 4)
- 5) Prepared date : Jul. 4, 2016
- 6) Contact person : Peter Chung, 412-512-8802
- 7) Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
- 8) Submission date : Jul. 11, 2016
#### 2. Device Information
- Trade name : Fixone Biocomposite Anchor 1)
- 2) Common name : Biodegradable Orthopedic Bone Screw
- Regulation name : Fastener, Screw, Fixation, Bone, Suture 3)
- 4) Product code : MAI
- Regulation number : 888.3030 5)
- ୧) Class of device : Class II
- 7) Panel : Orthopedic
#### 3. The legally marketed device to which we are claiming equivalence
K043337, Arthrex, Inc. Bio-Corkscrew Suture Anchor K101679, Arthrex, Inc. PushLock Anchors K073412, Depuy Mitek, Healix BR Anchor
#### 4. Device description
The Fixone Biocomposite Anchor is intended for reattaching soft tissue to bone with sutures. The anchor is manufactured from biodegradable materials (PLGA copolymer and β-ΤCP). A nonresorbable suture manufactured from cobraided UHMWPE and PET fibers is inserted into the anchor is implanted using a provided driver.
This device is could used with instrument that manufactured by Aju Pharm Co.,Ltd. It is consist of 18 models. It provide non-sterile (user must sterilization before use).
| Product name | Instrument / 18 model codes including Fixone.I.B-A/T450a |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model name | Fixone.I.B-A/T450a, Fixone.I.B-A/T475a, Fixone.I.B-A/T500a, Fixone.I.B-A/T525a, Fixone.I.B-A/T550a, Fixone.I.B-A/T575a, Fixone.I.B-A/T600a, Fixone.I.B-A/T625a, Fixone.I.B-A/T650a, Fixone.I.B-Awl450a, Fixone.I.B-Awl475a, Fixone.I.B-Awl500a, Fixone.I.B-Awl525a, Fixone.I.B-Awl550a, Fixone.I.B-Awl575a, Fixone.I.B-Awl600a, Fixone.I.B-Awl625a, Fixone.I.B-Awl650a |
| Intended use | The instrument of make the hole in the bone. |
| Manufacturer | Ajumedical / Korea |
| Sterilization | Non-sterile |
| Sterilization method | Autoclave / 134°C/ 20 min |
#### 5. Intended Use :
#### The Fixone Biocomposite Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures:
Shoulder: Rotor Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction;
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;
Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis;
{5}------------------------------------------------
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
#### 6. Performance data:
- Bench test were performed. Bench testing included biocompatibility, mechanical testing 1) including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
| Test item | Requirements | Results |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| External surface | | |
| Measurement | | |
| Insertion torque | | |
| Fixation strength | ASTM F2502 and USP<28> | Pass |
| Tensile strength | | |
| Extractable color | | |
| Extraction test | | |
| pH | The difference should be 1.5 and less. | |
| Potassium permanganate<br>reducing substances | The difference of the consumption of potassium permanganate<br>should be 2.0 mL and less. | |
| Residue after evaporation | Record the weight of the residue should be 1.0mg and less. | |
| Heavy metals | Any brown color produced within 10 minutes in the tube containing<br>the extract of the prepared sample does not exceed that in the tube<br>containing the standard lead solution | Pass |
| UV spectrum(250nm~350nm) | Maximum absorbance between 250 to 350 nm should be 0.1 and less. | |
| Property | When observing it with the naked eye, test solution should be clear<br>and have no forign particles. | |
#### 2) Biocompatibility
#### Anchor
| # | Test item | Test method / Test criteria | Test result |
|----|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------|
| 1 | Cytotoxicity | ISO 10993-5(2009) Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | Pass |
| 2 | Acute systemic toxicity test | ISO 10993-11(2009) Biological evaluation of medical devices - Part 11: Tests for systemic toxicity | Pass |
| 3 | Pyrogen Test | ISO 10993-11 Test for systemic toxicity, pyrogen test | Pass |
| 4 | Intracutaneous(intradermal) reactivity test | ISO 10993-10(2013) Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Pass |
| 5 | Maximization test for delayed hypersensicivity | ISO 10993-10(2013) Test for irritation and skin sensitization, Maximization test for delayed hypersensitivity | Pass |
| 6 | Bacterial revers mutation test | ISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation test | Pass |
| 7 | Mammalian erythrocyte micronucleus test | ISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation test | Pass |
| 8 | Implantation test | ISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bone | Pass |
| 9 | Bioabsorbable screws test | ASTM F2502 Standard specification and test methods for bioabsorbable plates and screws for internal fixation implants | Pass |
| 10 | Subchronic toxicity test | ISO 10993-11 Biological Evaluation of Medical Devices Part 11- Test for systemic toxicity | Pass |
#### Suture
| # | Test item | Test method / Test criteria | Test result |
|---|---------------------------|----------------------------------------------------------------------------------------------------|-------------|
| 1 | Cytotoxicity | ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in<br>vitro cytotoxicity | Pass |
| 2 | Systemic toxicity<br>test | ISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for<br>systemic toxicity | Pass |
{6}------------------------------------------------
| ന | Pyrogen Test | ISO 10993-11 Test for systemic toxicity, pyrogen test | | |
|---|-------------------|------------------------------------------------------------------------------|------|--|
| प | Intracutaneous | ISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for | Pass | |
| | reactivity test | irritation and skin sensitization | | |
| 5 | Maximization | ISO 10993-10, Test for irritation and skin sensitization, Maximization test | Pass | |
| | sensitization | for delayed hypersensitivity | | |
| 6 | Genotoxicity test | ISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation test | | |
| 8 | Implantation test | ISO 10993-6, Tests for local effects after implantation, Annex D test | Pass | |
| | | methods for implantation in bone | | |
| g | Hemolysis test | ISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of | Pass | |
| | | tests for interactions with blood | | |
The performance tests demonstrated that Fixone Biocomposite Anchor performs in a substantially equivalent manner to the predicate device.
#### 7. Predicate device comparison table
#### 1) B-type and K-type
| Manufacturer | AJU Pharm Co.,Ltd. | Depuy Mitek | Remark |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| 510(k) No. | K171299 | K073412 | N/A |
| Indication for use | The Fixone Biocomposite Anchors are<br>intended for fixation of suture (soft<br>tissue) to bone in the shoulder,<br>foot/ankle, knee and elbow in the<br>following procedures:<br><b>Shoulder</b> : Rotor Cuff Repair, Bankart<br>Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation<br>Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction;<br><b>Foot/Ankle</b> : Lateral Stabilization, Medial<br>Stabilization, Achilles Tendon Repair;<br><b>Knee</b> : Medial Collateral Ligament Repair,<br>Lateral Collateral Ligament Repair,<br>Posterior Oblique Ligament Repair,<br>Iliotibial Band Tenodesis;<br><b>Elbow</b> : Biceps Tendon Reattachment,<br>Ulnar or Radial Collateral Ligament<br>Reconstruction. | <b>Shoulder</b> : Rotor Cuff Repair, Bankart<br>Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation<br>Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction;<br><b>Foot/Ankle</b> : Lateral Stabilization, Medial<br>Stabilization, Achilles Tendon Repair;<br><b>Knee</b> : Medial Collateral Ligament Repair,<br>Lateral Collateral Ligament Repair,<br>Posterior Oblique Ligament Repair,<br>Iliotibial Band Tenodesis;<br><b>Elbow</b> : Biceps Tendon Reattachment,<br>Ulnar or Radial Collateral Ligament<br>Reconstruction. | Same |
| Classification name | Fastener, Screw, Fixation, Bone, Suture | Fastener, Screw, Fixation, Bone, Suture | Same |
| Trade name | Fixone Biocomposite Anchor | Healix BR Anchor | N/A |
| Model/type | 80 model codes including BAB-45001a | 5 model codes including 4.5 Healix BR<br>Anchor | N/A |
| Appearance | Image: Two white, screw-like anchors | Image: Three white, screw-like anchors with drivers | Similar |
| Product<br>configuration | Driver<br>Anchor<br>Suture | Driver<br>Anchor<br>Suture | Same |
| Material | PLGA(70%) + β-TCP(30%) | PLGA(70%) + β-TCP(30%) | Same |
| Anchor | | | |
| Outside diameter | 4.5mm/4.75mm/5.5mm/5.75mm/6.5mm | 4.5mm/5.5mm/6.5mm | Similar |
| Length of abchor | 14.6mm/15mm/15.7mm/16mm/16.8mm | Not known | N/A |
| Suture | | | |
| Absorbable | Non-absorbable | Non-absorbable | Same |
| Suture diameter | 0.50~0.599 (USP size 2) | 0.50~0.599 (USP size 2) | Same |
{7}------------------------------------------------
| Manufacturer | AJU Pharm Co.,Ltd. | Depuy Mitek | Remark |
|---------------------------|------------------------------------------------------|------------------------------------------------------|--------|
| Material | Polyethylene | Polyethylene | Same |
| Sterilization | EO Gas sterilization<br>According to ISO 11135: 2007 | EO Gas sterilization<br>According to ISO 11135: 2007 | Same |
| Biodegradable | Yes | Yes | Same |
| Principle of<br>operation | Manual | Manual | Same |
| Shelf-life | 5 years | N/A | N/A |
| Manufacturer | AJU Pharm Co.,Ltd. | Arthrex, Inc. | Remark |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| 510(k) No. | K171299 | K043337 | N/A |
| Indication for use | The Fixone Biocomposite Anchors are<br>intended for fixation of suture (soft<br>tissue) to bone in the shoulder,<br>foot/ankle, knee and elbow in the<br>following procedures:<br>Shoulder: Rotor Cuff Repair, Bankart<br>Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation<br>Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction;<br>Foot/Ankle: Lateral Stabilization, Medial<br>Stabilization, Achilles Tendon Repair;<br>Knee: Medial Collateral Ligament Repair,<br>Lateral Collateral Ligament Repair,<br>Posterior Oblique Ligament Repair,<br>Iliotibial Band Tenodesis;<br>Elbow: Biceps Tendon Reattachment,<br>Ulnar or Radial Collateral Ligament<br>Reconstruction. | The Arthrex Bio-Corkscrew Suture Anchor<br>is intended for fixation of suture to bone<br>in the shoulder, foot/ankle, knee,<br>hand/wrist, elbow, and pelvis. | Similar |
| Classification name | Fastener, Screw, Fixation, Bone, Suture | Fastener, Screw, Fixation, Bone, Suture | Same |
| Trade name | Fixone Biocomposite Anchor | Bio-Corkscrew Suture Anchor | N/A |
| Model/type | 80 model codes including BAB-45001a | N/A | N/A |
| Appearance | | | Similar |
| Product<br>configuration | Driver<br>Anchor<br>Suture | Driver<br>Anchor<br>Suture | Same |
| Material | PLGA + β-TCP | PLLA + B-TRICALCIUM PHOSPHATE | Similar |
| Anchor | | | |
| Outside diameter | 4.5mm/4.75mm/5.5mm/5.75mm/6.5mm | 4.5mm/4.75mm/5.5mm/5.75mm/6.5mm | Same |
| Length of abchor | 14.6mm/15mm/15.7mm/16mm/16.8mm | 15mm/16mm | Similar |
| Suture | | | |
| Absorbable | Non-absorbable | Non-absorbable | Same |
| Suture diameter | 0.50~0.599 (USP size 2) | 0.50~0.599 (USP size 2) | Same |
| Material | Polyethylene | Polyethylene | Same |
| Sterilization | EO Gas sterilization<br>According to ISO 11135: 2007 | EO Gas sterilization<br>According to ISO 11135: 2007 | Same |
| Biodegradable | Yes | Yes | Same |
| Principle of<br>operation | Manual | Manual | Same |
| Shelf-life | 5 years | N/A | N/A |
{8}------------------------------------------------
#### 2) KC-type
| Manufacturer | AJU Pharm Co.,Ltd. | Arthrex, Inc. | Remark |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) No. | K171299 | K101679 | N/A |
| Indication for use | The Fixone Biocomposite Anchors are<br>intended for fixation of suture (soft<br>tissue) to bone in the shoulder,<br>foot/ankle, knee and elbow in the<br>following procedures:<br><b>Shoulder</b> : Rotor Cuff Repair, Bankart<br>Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation<br>Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction;<br><b>Foot/Ankle</b> : Lateral Stabilization, Medial<br>Stabilization, Achilles Tendon Repair;<br><b>Knee</b> : Medial Collateral Ligament Repair,<br>Lateral Collateral Ligament Repair,<br>Posterior Oblique Ligament Repair,<br>Iliotibial Band Tenodesis;<br><b>Elbow</b> : Biceps Tendon Reattachment,<br>Ulnar or Radial Collateral Ligament<br>Reconstruction. | Intended to be used for sutures or tissue<br>fixation in the foot, ankle, knee, hand,<br>wrist, elbow, shoulder, hip | Similar |
| Classification name | Fastener, Screw, Fixation, Bone, Suture | Fastener, Screw, Fixation, Bone, Suture | Same |
| Trade name | Fixone Biocomposite Anchor | Arthrex PushLock Anchors | N/A |
| Model/type | 80 model codes including BAB-45001a | N/A | N/A |
| Appearance | Image: Fixone Biocomposite Anchor | Image: Arthrex PushLock Anchors | Similar |
| Product<br>configuration | Driver | Driver | Same |
| | Anchor | Anchor | Same |
| | Suture | Suture | Same |
| | Clip | | Different |
| Material | PLGA + $β$ -TCP | PLLA + B-TRICALCIUM PHOSPHATE | Similar |
| Outside diameter | 4.5mm/4.75mm/5.5mm/5.75mm<br>6.0mm/6.5mm | 3.5mm/4.5mm/4.75mm/5.5mm | Similar |
| Length of anchor | 16.7mm/17.2mm/17.7mm | 15.0mm/19.1mm/19.5mm/20.0mm<br>24.0mm/24.5mm/28.0mm | Similar |
| Sterilization | EO Gas sterilization<br>According to ISO 11135: 2007 | EO Gas sterilization<br>According to ISO 11135: 2007 | Same |
| Biodegradable | Yes | Yes | Same |
| Principle of<br>operation | Manual | Manual | Same |
| Shelf-life | 5 years | N/A | N/A |
#### 9. Conclusion
The device is investigated for function to compare the operation of function between Fixone Biocomposite Anchor and predicate devices.
Comparison results demonstrate that the specifications and performance of the device are substantially equivalent to the legally marketed predicate device.
Therefore, it is concluded that Fixone Biocomposite Anchor is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.