K171244 · Nimbleheart, Inc. · DRX · Aug 23, 2017 · Cardiovascular
Device Facts
Record ID
K171244
Device Name
Physiotrace
Applicant
Nimbleheart, Inc.
Product Code
DRX · Cardiovascular
Decision Date
Aug 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2360
Device Class
Class 2
Indications for Use
The Physiotrace™Smart is a telemetry device intended for physiological monitoring of men and women above 18 years of age at home, workplace, exercise facilities and alternate care settings. The Physiotrace™ Smart records single lead ECG data for up to 60 minutes during resting and exercise activities; and transmits it to a server for review by healthcare team. The Physiotrace™ Smart is indicated for use as a general patient monitor to provide physiological information as part of an occupational welfare monitoring system, for general research and performance measurement purposes, or where prescribed by a healthcare professional. The Physiotrace™ Smart is not indicated for use on critical care patients. The Physiotrace™ Smart is not indicated for diagnosis of cardiac conditions.
Device Story
Wearable torso-worn device; utilizes dry gold-plated brass electrodes; acquires single-lead ECG signals; integrates Bluetooth Low Energy (BLE) transmitter. Mobile application controls data acquisition; displays device status, heart rate, and ECG waveform; stores session data; relays to cloud server for healthcare team review. Used in home, workplace, and exercise settings; operated by patient. Output provides physiological data for occupational welfare, research, or performance measurement. 60-minute session limit prevents off-label critical care use. Benefits include non-invasive monitoring without electrolytic gels or adhesives.
Clinical Evidence
Clinical study performed during exercise scenarios. Device performance, safety, and effectiveness found equivalent or superior to predicate. Bench testing confirmed compliance with medical device safety, biocompatibility, and EMC standards.
Technological Characteristics
Wearable torso-strap form factor; gold-plated brass dry electrodes; Bluetooth Low Energy (BT 4.0, 4 dBm) connectivity; 60-minute session limit; lead-off detection for both electrodes. Complies with IEC 60601-2-47. Biocompatible materials.
Indications for Use
Indicated for physiological monitoring of adults (18+) in home, workplace, exercise, and alternate care settings. Records single-lead ECG for up to 60 minutes during rest or exercise. Not for critical care patients or cardiac diagnosis.
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
Predicate Devices
BioHarness 3.0 by Zephyr Technology Corporation (K113045)
Submission Summary (Full Text)
{1}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 23, 2017
NimbleHeart, Inc. Sonal Tambe Chief Executive Officer 1300 White Oaks Rd., Suite #201 Campbell, California 95008
Re: K171244
Trade/Device Name: Physiotrace™ Smart Regulation Number: 21 CFR 870.2360 Regulation Name: Electrocardiograph Electrode Regulatory Class: Class II Product Code: DRX, MWI, MHX Dated: July 15, 2017 Received: July 17, 2017
Dear Sonal Tambe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jessica E. Paulsen -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Over-The-Counter Use (21 CFR 801 Subpart C)
510(k) Number (if known) K171244
Device Name
Physiotrace™ Smart
#### Indications for Use (Describe)
The Physiotrace™Smart is a telemetry device intended for physiological monitoring of men and women above 18 years of age at home, workplace, exercise facilities and alternate care settings. The Physiotrace™ Smart records single lead ECG data for up to 60 minutes during resting and exercise activities; and transmits it to a server for review by healthcare team. The Physiotrace™ Smart is indicated for use as a general patient monitor to provide physiological information as part of an occupational welfare monitoring system, for general research and performance measurement purposes, or where prescribed by a healthcare professional.
The Physiotrace™ Smart is not indicated for use on critical care patients. The Physiotrace™ Smart is not indicated for diagnosis of cardiac conditions.
#### Contra-Indications
The Physiotrace™ Smart is contra-indicated for surgical procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1of 1
Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
# K171244 510(k) Summary
## Preparation Date
August 17, 2017
#### Company Information
Sonal Tambe NimbleHeart, Inc. 1300 White Oaks Rd.. #201 Campbell CA 95008 USA 408 475 7630 regulatory@nimbleheart.com
### Identification of Product and Classification
| Proprietary Name: | Physiotrace Smart |
|-------------------------|------------------------------------------------------------------------|
| Common Name: | ECG Monitor |
| Classification Name: | Monitor, Physiological, Patient(Without Arrhythmia Detection Or Alarm) |
| Regulation Description: | Cardiac monitor (including cardiotachometer and rate alarm). |
| Classification Panel: | Cardiovascular |
| Device Class: | II |
| Product Code: | DRX, MWI |
| CFR Section: | 870.2360, 870.2300 |
| Predicate Device: | BioHarness 3.0 by Zephyr Technology Corporation K113045 |
### Device Description
The Physiotrace™ Smart has a wearable design that can be wrapped and fastened around the torso. The device uses dry ECG electrodes and embeds an ECG acquisition unit with a Bluetooth Low Energy transmitter. A mobile application controls the data acquisition, displays the status of the device, heart rate and optionally the ECG waveform during a recording session. The mobile App also stores the ECG and the exercise session information and relays it to a cloud server for permanent storage and review by healthcare staff. The Physiotrace™ Smart is designed to be used without electrolytic gels and without adhesives that necessitate skin preparation.
#### Indications of Use
The Physiotrace™ Smart is a telemetry device intended for physiological monitoring of men and women above 18 years of age at home, workplace, exercise facilities and alternate care settings. The Physiotrace™ Smart records single lead ECG data for up to 60 minutes during resting and exercise activities; and transmits it to a server for review by healthcare team. The Physiotrace™ Smart is indicated for use as a general patient monitor to provide physiological information as part of an occupational welfare monitoring system, for general research and performance measurement purposes, or where prescribed by a healthcare professional.
The Physiotrace™ Smart is not indicated for use on critical care patients. The Physiotrace™ Smart is not indicated for diagnosis of cardiac conditions.
#### Contra-Indications
The Physiotrace™ Smart is contra-indicated for surgical procedures.
{5}------------------------------------------------
# K171244 510(k) Summary
| Indications of use comparison | | | |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Parameter | Physiotrace Device | Predicate Device | |
| FDA 510(k) | K171244 | K113045 | |
| Number | | | |
| Applicant | NimbleHeart, Inc. | Zephyr Technology Corp. | |
| Device Name | Physiotrace Smart | BioHarness 3.0 | |
| | | | |
| Classification | 870.2360, 870.2300 | 870.2360, 870.1025 | |
| Regulation | | | |
| Product Code | DRX, MWI | DRX, MHX | |
| Intended<br>population and<br>locations | The Physiotrace Smart is a telemetry device<br>intended for physiological monitoring of<br>men and women above 18 years of age at<br>home, workplace, exercise facilities and<br>alternate care settings. | The BioHarness 3.0 is a physiological<br>monitoring telemetry device intended<br>for monitoring of adults in the home<br>workplace and alternate care settings. | |
| Indicated use<br>scenario | The Physiotrace Smart records single lead<br>ECG data for up to 60 minutes during resting<br>and exercise activities; and transmits it to a<br>server for review by healthcare team. | The BioHarness 3.0 collects and<br>transmits measurements captured<br>during both sedentary as well as<br>rigorous activity for Heart Rate,<br>Posture and Activity. Breathing rate<br>values are accurately transmitted only<br>during sedentary periods. | |
| Indications for use | The Physiotrace™ Smart is indicated for<br>use as a general patient monitor to<br>provide physiological information as part<br>of an occupational welfare monitoring<br>system, for general research and<br>performance measurement purposes, or<br>where prescribed by a healthcare<br>professional. Physiotrace Smart is not<br>indicated for use on critical care patients.<br>The Physiotrace™ Smart is not indicated for<br>diagnosis of cardiac conditions. | The BioHarness 3.0 is indicated for<br>use as a general patient monitor to<br>provide physiological information as<br>part of an occupational welfare<br>monitoring system, for general<br>research and performance<br>measurement purposes, or where<br>prescribed by a healthcare<br>professional. | |
| Product Design | The device has a wearable and reusable<br>design that can be wrapped and fastened<br>around the torso. The device uses dry ECG<br>electrodes and embeds an ECG acquisition<br>unit with a Bluetooth Low Energy<br>transmitter. A mobile application controls the<br>data acquisition, displays the status of the<br>device, heart rate and optionally the ECG<br>waveform during a recording session. The<br>mobile App also stores the ECG and the<br>exercise session information and relays it to a<br>cloud server for permanent storage and<br>review by healthcare staff. | The device consists of a chest strap<br>and an electronics module that attaches<br>to the strap. The device stores and<br>transmits vital sign data including<br>ECG, heart rate, respiration rate, body<br>orientation and activity. The<br>BioHarness 3.0 provides a facility to<br>detect and transmit single lead ECG<br>signals to be received by Bluetooth/<br>USB qualified ECG instruments. | |
{6}------------------------------------------------
# K171244 510(k) Summary
| Technological characteristics comparison | | | | |
|------------------------------------------|-------------------------------------------------------------|----------------------------------------------|--|--|
| Parameter | NimbleHeart Physiotrace | Predicate Device | | |
| ECG Electrodes | Gold plated brass electrodes | Silver fabric electrodes | | |
| Lead off detection | Lead off detection feature for Right and<br>Left electrodes | Single indication called of Strap on/<br>off | | |
| Performance comparison | | | | |
| ECG Performance<br>standards met | 60601-2-47 | EC38 | | |
| Bluetooth radio<br>version | BT 4.0 (Max Tx power 4 dBm) | BT 2.1 + EDR (Max Tx power 10<br>dBm) | | |
## Safety and Effectiveness
The Physiotrace testing has confirmed this device meets it's the performance requirements of recognized standards for Medical device safety, biocompatibility, Electromagnetic Compatibility and for home use. A clinical study was performed during the exercise scenario as per the intended use. The device performance and safety and effectiveness was found to be equivalent or better than the predicate device.
### Substantial Equivalence
The NimbleHeart Inc., Physiotrace Smart is a Class II device that is substantially equivalent to the designated predicate device.
The indications of use scenario where Physiotrace is slightly different is that the recording session time is preprogrammed with a default of 60 minutes for a prescribed exercise session. The mobile App logs out and stops data recording when the session ends. This feature reduces the risk posed by off label use of the device in critical care scenarios. The 60 minutes recording session was shown to be safe and effective during the clinical study.
The technological differences are mainly the electrode design including the materials used. The electrode performance has been tested as per the FDA's electrode guidance document. It was also validated during a clinical study and found comparable to that of the predicate device. All the materials have passed biocompatibility and cleaning validation tests.
Therefore, Physiotrace Smart is as safe and effective as the predicate device and can therefore be found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.