K171114 · Medos International SARL · MAI · Nov 27, 2017 · Orthopedic
Device Facts
Record ID
K171114
Device Name
HEALIX ADVANCE KNOTLESS Anchor
Applicant
Medos International SARL
Product Code
MAI · Orthopedic
Decision Date
Nov 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HEALIX ADVANCE KNOTLESS Anchors are indicated for use in the following procedures for reattachment of soft tissue to bone: Shoulder: Rotator Cuff, Biceps Tenodesis, Deltoid Repair Knee: Posterior Oblique, Medial Collateral Ligament (MCL), Lateral Ligament (LCL), Iliotibial (IT) Band Tenodesis Foot/Ankle*: Achilles Tendon Repair *HEALIX ADVANCE KNOTLESS PEEK Anchors only
Device Story
One-piece, cannulated, threaded suture anchor; secures soft tissue to bone. Provided pre-loaded on disposable inserter driver. Available in 4.75mm, 5.5mm, and 6.5mm sizes. Used by surgeons in orthopedic procedures. Implanted into bone to anchor sutures; facilitates tissue reattachment. Benefits include knotless fixation, reducing surgical complexity. Sterile, single-use device.
Clinical Evidence
Bench testing only. Performance testing included in-vitro anchor pull-out, time-zero anchor pull-out, and torque testing. Safety evaluations addressed biological, sterility, packaging, and shelf-life. Device confirmed non-pyrogenic per ANSI/AAMI ST-72:2011, USP, and EP standards using bacterial endotoxin testing.
Indicated for patients requiring soft tissue-to-bone reattachment in shoulder (rotator cuff, biceps tenodesis, deltoid repair), knee (posterior oblique, MCL, LCL, IT band tenodesis), and foot/ankle (Achilles tendon repair; PEEK anchors only).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
November 27, 2017
Medos International SARL % Ms. Julie Vafides Manager, Regulatory Affairs DePuy Mitek, a Johnson & Johnson Company 325 Paramount Drive Raynham, Massachusetts 02767
Re: K171114
Trade/Device Name: HEALIX ADVANCE KNOTLESS™ Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: October 19, 2017 Received: October 23, 2017
Dear Ms. Vafides:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K171114
Device Name
HEALIX ADVANCE™ KNOTLESS Anchor
Indications for Use (Describe)
The HEALIX ADVANCE KNOTLESS Anchors are indicated for use in the following procedures for reattachment of soft tissue to bone:
Shoulder: Rotator Cuff, Biceps Tenodesis, Deltoid Repair
Knee: Posterior Oblique, Medial Collateral Ligament (MCL), Lateral Ligament (LCL), Iliotibial (IT) Band Tenodesis
- Foot/Ankle*: Achilles Tendon Repair
*HEALIX ADVANCE KNOTLESS PEEK Anchors only
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY
## HEALIX ADVANCE™ KNOTLESS Anchor
| Submitter's<br>Name and<br>Address | Medos International SARL<br>Chemin-Blanc 38, Case Postale<br>Le Locle<br>CH 2400<br>Switzerland | | |
|-----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| | Date Prepared: April 13, 2017 | | |
| Contact Person | Julie Vafides<br>Manager, Regulatory Affairs<br>DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767, USA | | Telephone: 508-977-6645<br>e-mail: jvafides@its.jnj.com |
| Name of<br>Medical Device | Proprietary Name: | a) HEALIX ADVANCE™ KNOTLESS BR Anchor<br>b) HEALIX ADVANCE™ KNOTLESS PEEK Anchor | |
| | Classification Name: | a) Single/multiple component metallic bone fixation<br>appliances and accessories<br>b) Smooth or threaded metallic bone fixation fasteners | |
| | Common Name: | Suture Anchor | |
| Substantial<br>Equivalence | The HEALIX ADVANCE™ KNOTLESS Anchor is substantially equivalent to:<br>K130917, K131974 HEALIX ADVANCE™ KNOTLESS BR Anchor K130539, K131683 HEALIX ADVANCE™ KNOTLESS PEEK Anchor K101679, K082810, K071177, K061863, K051219 Arthrex PushLock Anchors | | |
| Device<br>Classification | > HEALIX ADVANCE™ KNOTLESS BR Anchor is classified as:<br>Single/multiple component metallic bone fixation appliances and accessories,<br>classified as Class II, product code MAI, regulated under 21 CFR 888.3030.<br>> HEALIX ADVANCE™ KNOTLESS PEEK Anchor is classified as:<br>Smooth or threaded metallic bone fixation fasteners, classified as Class II, product | | |
| Device<br>Description | The HEALIX ADVANCE™ KNOTLESS Anchor is a one piece implantable<br>cannulated, threaded anchor designed to secure soft tissue to bone. The anchor is<br>provided loaded on a disposable inserter driver device. The proposed anchor is offered<br>in three sizes, 4.75 mm, 5.5 mm and 6.5mm. The HEALIX ADVANCE™<br>KNOTLESS Anchors are molded from Biocryl Rapide® (BR) and<br>polyetheretherketone (PEEK) materials. The HEALIX ADVANCE™ KNOTLESS<br>Anchor is provided sterile and is for single use only. | | |
| Technological<br>Characteristics | The suture anchor design, anchor and suture materials, and finished good assembly<br>design remain the same when compared to the predicate HEALIX ADVANCE™<br>KNOTLESS Anchors (BR: K130917, K131974; PEEK: K130539, K131683). Minor<br>modifications have been made to the distal tip geometry of the BR Anchors.<br>Additional minor changes have been made to the disposable inserter to increase suture<br>management usability. | | |
| Indications for<br>Use | The HEALIX ADVANCE KNOTLESS Anchors are indicated for use in the following<br>procedures for reattachment of soft tissue to bone: | | |
| | Shoulder: | Rotator Cuff Repair, Biceps Tenodesis, Deltoid Repair | |
| | Knee: | Posterior Oblique, Medial Collateral Ligament (MCL), Lateral<br>Collateral Ligament (LCL), Iliotibial (IT) Band Tenodesis | |
| | Foot/Ankle*: | Achilles Tendon Repair | |
| | *HEALIX ADVANCE™ KNOTLESS PEEK Anchors only | | |
| Non-clinical<br>Testing | Non-clinical testing has been performed on the proposed device and / or its predicates.<br>Performance testing included in-vitro anchor pull out, time zero anchor pull out and<br>torque testing. Safety evaluations were conducted to address biological, sterility,<br>packaging and shelf-life. Note: The proposed device has been determined to be non-<br>pyrogenic per the requirements set forth in ANSI/AAMI ST-72:2011, United States<br>Pharmacopeia (USP), and European Pharmacopeia (EP) using the bacterial endotoxin<br>testing (BET) method. | | |
| Safety and<br>Performance | Results of safety and performance testing have demonstrated that the proposed devices<br>are suitable for their intended use. | | |
| | Based on similarities in the indications for use, technological characteristics, and<br>performance in comparison to the predicate devices, the proposed HEALIX<br>ADVANCE™ KNOTLESS Anchor has shown to be substantially equivalent to the<br>predicate device under the Federal Food, Drug and Cosmetic Act. | | |
| Proposed Device | | | |
| HEALIX<br>ADVANCE™<br>KNOTLESS Anchor | The HEALIX ADVANCE KNOTLESS Anchors are indicated for use in the following<br>procedures for reattachment of soft tissue to bone: | | |
| | Shoulder: | Rotator Cuff Repair, Biceps Tenodesis, Deltoid Repair | |
| | Knee: | Posterior Oblique, Medial Collateral Ligament (MCL), Lateral<br>Collateral Ligament (LCL), Iliotibial (IT) Band Tenodesis | |
| | Foot/Ankle*: | Achilles Tendon Repair | |
| | | * HEALIX ADVANCE™ KNOTLESS PEEK Anchors only | |
| Predicate Devices | | | |
| HEALIX<br>ADVANCE™<br>KNOTLESS Anchors<br>(K130917, K131974,<br>K130539, K131683) | Shoulder: | Rotator Cuff Repair, Biceps Tenodesis | |
| Arthrex PushLock™<br>(K101679, K082810,<br>K071177, K061863,<br>K051219) | Shoulder: | Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction | |
| | Foot/Ankle: | Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair,<br>Metatarsal Ligament Repair/Tendon Repair, Hallux Valgus Reconstruction, Mid-Foot<br>Reconstruction, Bunionectomy | |
| | Knee: | Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar<br>Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis | |
| | Hand/Wrist: | Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral<br>Ligament Reconstruction | |
| | Elbow: | Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral<br>Ligament Reconstruction, Lateral Epicondylitis Repair | |
| | Hip: | Capsular Repair, Acetabular Labral Repair | |
{5}------------------------------------------------
{6}------------------------------------------------
# INDICATIONS FOR USE
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.