K170839 · Given Imaging , Ltd. · NEZ · May 26, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K170839
Device Name
RAPID Web
Applicant
Given Imaging , Ltd.
Product Code
NEZ · Gastroenterology, Urology
Decision Date
May 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1300
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The PillCam Capsule Endoscopy System with a PillCam SB 2 and SB 3 capsules is intended for visualization of the small bowel mucosa. PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy. PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy. The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas. The PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age. The PillCam Capsule Endoscopy System with PillCam ESO 3 capsules is intended for the visualization of esophageal mucosa in adults and children from 18 years of age. The PillCam UGI capsule endoscopy system is intended for visualization of the upper gastrointestinal tract (esophagus, stomach, duodenum). It may be used for visualization of blood in the upper gastrointestinal tract (esophagus, stomach, duodenum) in patients who are hemodynamically stable and at least 18 years of age. The PillCam COLON 2 capsule endoscopy system is intended to provide visualization of the colon. It may be used for detection of colon polyps in patients after an incomplete optical colonoscopy with adequate preparation, and a complete evaluation of the colon was not technically possible. In addition, it is intended for detection of colon polyps in patients with evidence of gastrointestinal bleeding of lower GI origin. This applies only to patients with major risks for colonoscopy or moderate sedation, but who could tolerate colonoscopy and moderate sedation in the event a clinically significant colon abnormality was identified on capsule endoscopy.
Device Story
RAPID Web is a client-server software application for reviewing and analyzing capsule endoscopy videos. It functions as a web-based interface for studies previously captured by PillCam systems and stored in a central repository. Physicians access the system via secure internet or intranet browsers on workstations, laptops, or tablets (Apple iPads > 9.7"). The software allows clinicians to view, manage, and report on capsule endoscopy findings. By enabling remote access to study data, it facilitates efficient clinical review and decision-making regarding GI abnormalities, potentially improving diagnostic workflows and patient management. It does not perform initial capsule initialization or video creation, which are handled by the cleared RAPID 8.0 workstation software.
Clinical Evidence
Comparative performance study conducted with trained professionals identifying pathologies in identical video segments using both RAPID Web and RAPID 8.0 across various platforms (workstations, laptops, iPads). Analysis of matched and unmatched pathologies demonstrated that RAPID Web met pre-specified sensitivity requirements. Software validation performed per FDA guidance.
Technological Characteristics
Client-server software architecture; web-based deployment. Compatible with Windows (Chrome, Firefox, IE) and MacOS (Safari) environments. Supports viewing on workstations, laptops, and iPads (minimum 9.7" screen). Requires specific display standards (e.g., 20" monitor, 1600x900 resolution for workstations; 107 PPI for laptops). Connectivity via private intranet or secure internet. Does not include hardware; relies on existing PillCam capsule imaging hardware.
Indications for Use
Indicated for visualization of GI mucosa (small bowel, esophagus, upper GI, or colon) in adults and children (age 2+ for SB, 18+ for ESO/UGI/COLON). Used for monitoring Crohn's, obscure bleeding, iron deficiency anemia, or polyp detection post-incomplete colonoscopy. Contraindicated for patients unable to tolerate colonoscopy/sedation if required for follow-up.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 26, 2017
Given Imaging Ltd. Hilla Debby Director Clinical & Regulatory 2 Hacarmel St. New Industrial Park POB 258 Yoqneam, 20692 Israel
Re: K170839 Trade/Device Name: RAPID Web Regulation Number: 21 CFR§ 876.1300 Regulation Name: Ingestible Telemetric Gastrointestinal Capsule Imaging System Regulatory Class: II Product Code: NEZ, NSI, PGD Dated: March 16, 2017 Received: March 21, 2017
Dear Hilla Debby:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
RAPID Web
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known)
K170839
Device Name
### RAPID Web
Indications for Use (Describe)
With PillCam SB 2 and SB 3 Capsules
The PillCam Capsule Endoscopy System with a PillCam SB 2 and SB 3 capsules is intended for visualization of the small bowel mucosa.
- o PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
- PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules ● may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
- PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.
The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas.
The PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age.
### With PillCam ESO 3 Capsule
The PillCam Capsule Endoscopy System with PillCam ESO 3 capsules is intended for the visualization of esophageal mucosa in adults and children from 18 years of age..
### With PillCam UGI Capsule
The PillCam UGI capsule endoscopy system is intended for visualization of the upper gastrointestinal tract (esophagus, stomach, duodenum). It may be used for visualization of blood in the upper gastrointestinal tract (esophagus, stomach, duodenum) in patients who are hemodynamically stable and at least 18 years of age.
{3}------------------------------------------------
### With PillCam COLON Capsule
The PillCam COLON 2 capsule endoscopy system is intended to provide visualization of the colon. It may be used for detection of colon polyps in patients after an incomplete optical colonoscopy with adequate preparation, and a complete evaluation of the colon was not technically possible. In addition, it is intended for detection of colon polyps in patients with evidence of gastrointestinal bleeding of lower GI origin. This applies only to patients with major risks for colonoscopy or moderate sedation, but who could tolerate colonoscopy and moderate sedation in the event a clinically significant colon abnormality was identified on capsule endoscopy.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D) CFR 801 Subpart C)
_ Over-The-Counter Use (21
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
#### I. SUBMITTER
| Submitter Name and Address: | Given Imaging Ltd.(GI Solutions,<br>Medtronic) |
|----------------------------------|------------------------------------------------|
| 2 Hacarmel Street | |
| New Industrial Park | |
| PO Box 258 | |
| Yokneam 20692 | |
| Israel | |
| Contact Person: | Hilla Debby |
| Director, Clinical & Regulatory | |
| Email: Hilla.Debby@medtronic.com | |
| Phone Number: | 972 (4) 909-7774 |
| Fax Number: | 972 (73) 2501533 |
| Date Prepared: | March 16, 2017 |
#### II. DEVICE
Device Trade Name: RAPID Web
Common or Usual Name, Classification Name, Regulatory Class, and Product Code:
21 CFR 876.1300 Ingestible telemetric gastrointestinal capsule imaging system 21 CFR 876.1330 Colon capsule endoscopy system NEZ - System, imaging, gastrointestinal, wireless, capsule, class II
{5}------------------------------------------------
NSI - System, imaging, esophageal, wireless, capsule, class II
PGD - colon capsule imaging system, class II
#### III. PREDICATE DEVICES
RAPID 8.0 Software (K123864)
Reference Devices: UGI Capsule (K140284) and COLON 2 Capsule (K153466)
#### DEVICE DESCRIPTION IV.
The RAPID Web is a client server based version of the RAPID video review and report creation components of the cleared RAPID 8 software. The RAPID Web is designed to allow users to review and analyze videos created using the cleared RAPID software, and to create reports over a private intranet network or over a secure internet connection. The PillCam capsule studies are created at workflow-relevant locations and stored at a central repository. The RAPID Web server is installed on a central server in an intranet or internet environment with access to the study repository. After the capsule endoscopy procedure was conducted using the cleared RAPID software, Given Workstation/User PC, PillCam capsule, and DR 3 PillCam recorder, end users will use an internet browser with authorization login to access, review, and manage the studies in the repository using workstation, user PC, a laptop or an Apple™ (screen size > 9.7").
#### INTENDED USE / INDICATIONS FOR USE V.
### With PillCam SB 2 and SB 3Capsules
The PillCam Capsule Endoscopy System with a PillCam SB 2 and SB 3 capsules is intended for visualization of the small bowel mucosa.
{6}------------------------------------------------
- · PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
- · PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
- PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsule may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.
The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas.
The PillCam Capsule Endoscopy System with PillCam SB 2 and SB 3 capsules may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age.
### With PillCam ESO 3 Capsule
The PillCam Capsule Endoscopy System with PillCam ESO 3 capsules is intended for the visualization of esophageal mucosa in adults and children from 18 years of age.
## With PillCam UGI Capsule
The PillCam UGI capsule endoscopy system is intended for visualization of the upper gastrointestinal tract (esophagus, stomach, duodenum). It may be used for visualization of blood in the upper gastrointestinal tract (esophagus, stomach, duodenum) in patients who are hemodynamically stable and at least 18 years of age.
{7}------------------------------------------------
# With PillCam COLON Capsule
The PillCam COLON 2 capsule endoscopy system is intended to provide visualization of the colon. It may be used for detection of colon polyps in patients after an incomplete optical colonoscopy with adequate preparation, and a complete evaluation of the colon was not technically possible. In addition, it is intended for detection of colon polyps in patients with evidence of gastrointestinal bleeding of lower GI origin. This applies only to patients with major risks for colonoscopy or moderate sedation, but who could tolerate colonoscopy and moderate sedation in the event a clinically significant colon abnormality was identified on capsule endoscopy.
#### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS VI.
The RAPID Web and RAPID 8.0 have the same intended use and indications, and similar technological characteristics and principles of operation. The only technological differences between the RAPID Web and its predicates are: relocation of software from local hard drive to web-based server, increased range of viewing platforms, removal of features intended for use on mainframes and PCs, removal of viewing modes and features not critical to diagnosis, and the addition of several user friendly features. These differences do not present any new issues of safety or effectiveness because they are not critical to diagnosis and primarily affect convenience. Thus, the RAPID Web is substantially equivalent to the RAPID 8.0.
| Intended Use | RAPID Web<br>(subject) | RAPID 8.0<br>(K123864) |
|---------------------------------------------------------------------------------|-----------------------------------------|--------------------------------------------|
| Viewing of capsule<br>endoscopy studies | Viewing of capsule<br>endoscopy studies | Viewing of<br>capsule endoscopy<br>studies |
| Capsule endoscopy procedure<br>(Check-in, initialization, video<br>creation) | Not supported | Supported |
| Capsule endoscopy post-procedure<br>steps (Video review and report<br>creation) | Supported | Supported |
| Installation | Installed on a server | Installed on a<br>work station |
{8}------------------------------------------------
| Hardware | Workstation, PC,<br>laptops, Apple iPads ><br>9.7" screen | Workstation, PC |
|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|-----------------|
| Main video reading features (Open,<br>play, viewing modes, FICE & Blue | Supported | Supported |
| Workstation related features<br>(keyboard, CD, etc.) | Not supported | Supported |
| Optional viewing aids (mosaic view,<br>image enhancement control options,<br>Lewis score, atlas, compare<br>thumbnails etc.) | Not supported | Supported |
| Additional features (dynamic player<br>control, findings preview) | Supported | Not Supported |
### Minimum Requirements
Minimum Display Requirements for Workstations Monitors
| | Minimum<br>Requirement |
|-------------------|------------------------|
| Diagonal | 20" |
| Response time | 8 ms |
| Advanced contrast | 15,000:1 |
| Resolution max | 1600x900 |
# Minimum Display Requirements for Laptops
| | Minimum Requirement |
|-----------------|--------------------------|
| Diagonal | 12.5" |
| Static contrast | 300:1 |
| Brightness | 200 nits |
| PPI | 107 PPI |
| Resolution max | 1366 x 768<br>1280 x 800 |
| Refresh rate | 60 Hz |
{9}------------------------------------------------
Minimum Display Requirements for iPads
| | Minimum<br>Requirement |
|----------------|------------------------|
| Diagonal | 9.7" |
| Resolution max | 264 PPI |
| Diagonal | 2048 x 1536 |
Compatible operating systems and browsers:
- 1. For Windows 7 SP1, and Windows 8 or win 8.1:
- 1.1. Chrome v38.
- 1.2. Firefox v33.
- 1.3. Internet Explorer 11.0.4.
Hardware specifications: All Windows related tests were performed on Intel x86-based computer hardware that meets the operating system minimum requirements.
- 2. For MacOS v10.10 "Yosemite":
- 2.1. Safari 8.0
- 3. For MacOS v10.8 "Mountain Lion" and v10.9 "Mavericks": 3.1. Safari 7.1
MAC Hardware specifications:
- -MacBook Pro 13", 2.4GHz dual-core Intel Core i5 processor, 4GB RAM
- Macbook air 13", 1.8HGZ dual-core Intel Core i7, 4GB RAM -
- 4. For iPad with iOS 8.1:
- 4.1. Safari 8.0
iPad Hardware specifications:
- iPad 4: Dual-core A6X with quad-core graphics and 16GB storage
- iPad air: A7 chip with 64-bit architecture and 16GB storage -
- 5. Color depth: 24 bit
{10}------------------------------------------------
#### VII. PERFORMANCE DATA
A comparative performance study wherein trained professionals identified pathologies in the same video and video segments that we displayed using RAPID Web and RAPID 8.0 was performed. RAPID 8.0 performance was assessed on multiple platforms, including workstations, laptops, and iPads. An analysis assessing the matched and un-matched pathologies showed that RAPID 8.0 met the pre-specified sensitivity.
In addition, the software was validated and documented per FDA's software guidance document.
#### VIII. CONCLUSIONS
The RAPID Web is as safe and effective as the RAPID 8.0. The RAPID Web has the same intended uses and indications, and similar technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended diagnostic use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the RAPID Web and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the RAPID Web is as safe and effective as the RAPID 8.0. Thus, the RAPID Web is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.