K170770 · Medical Components, Inc. (dba MedComp) · PND · Jul 25, 2017 · General Hospital
Device Facts
Record ID
K170770
Device Name
CT Midline
Applicant
Medical Components, Inc. (dba MedComp)
Product Code
PND · General Hospital
Decision Date
Jul 25, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CT Midlines are indicated for Short-Term peripheral access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media.
Device Story
CT Midline is a peripheral venous catheter designed for short-term access, blood sampling, and power injection of contrast media. The device features an open-ended lumen made of radiopaque polyurethane (with barium sulfate), a soft hub with suture wings, extension tubes with clamps, and female luer connectors. It is available in 4F (single-lumen) and 5F (double-lumen) configurations, both 20cm in length. The lumen design includes a reverse taper near the hub to improve kink resistance and provide a mechanical obstruction to venotomy bleeding. Depth and numerical markings are provided for placement guidance. The device is used in clinical settings for peripheral vein catheterization. It is supplied sterile and includes necessary insertion accessories. The device is identical to the predicate (K141151) in design, materials, and performance; this submission solely removes a previously cited contraindication from the instructions for use to align with current Journal of Infusion Nursing standards.
Clinical Evidence
No clinical data provided. Bench testing and biocompatibility were previously established for the predicate device. Biocompatibility testing for the device was performed in accordance with ISO 10993-1 and ISO 10993-12:2012.
Indicated for short-term peripheral venous access for intravenous therapy, blood sampling, and power injection of contrast media in patients requiring such access.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 25, 2017
Medical Components, Inc. (dba Medcomp) Courtney Nix Regulatory Affairs Manager. North America and Europe 1499 Delp Drive Harleysville, Pennsylvania 19438
Re: K170770
Trade/Device Name: CT Midline Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: PND Dated: July 5, 2017 Received: July 6, 2017
Dear Courtney Nix:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Tara A. Ryan -S
for Lori Wiggins Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) N/A - Unknown
K170770
Device Name CT Midline
Indications for Use (Describe)
The CT Midlines are indicated for Short-Term peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"> ☒ </span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
| Section 6 | | | 510(k) SUMMARY | Traditional 510K | |
|-----------|--------------------------------------------------------|--|-------------------------------------------------------------------------------------------------------------|------------------|-----|
| A. | Submitter Information: | | | | |
| | Submitter: | | Medcomp®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Tel: (215) 256-4201, x 2285<br>Fax: (215) 256-9191 | | |
| | Registration Number: | | 2518902 | | |
| | Contact: | | Courtney Nix<br>Cnix@Medcompnet.com<br>Regulatory Affairs Manager: North America and EU | | |
| | Date Prepared: | | 03/13/2017 | | |
| в. | Proposed or Subject Device Information: | | | | |
| | Trade Name: | | CT Midline | | |
| | Device: | | Midline Catheter | | |
| | Product Code: | | PND | | |
| | Regulation Description | | Intravascular Catheter | | |
| | C.F.R. Section: | | 880.5200 | | |
| | Class: | | ll | | |
| | Regulation Medical<br>Specialty and Review<br>Panel: | | General Hospital | | |
| B. | Predicate Device Information: | | | | |
| | 510(k) Number: | | K141151 | | |
| | 510(k) Holder: | | Medcomp® | | |
| | Trade Name: | | CT Midline | | |
| | Device: | | Catheter, Intravascular, Therapeutic, Long-Term<br>Greater Than 30 Days | | |
| | Product Code: | | LJS | | |
| | Medcomp®: CT Midline<br>Section 6: 510(k) Site Summary | | | | 6-1 |
K170770
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a stylized geometric shape in pink, resembling a house or an arrow pointing upwards. Below the shape, the word "med" is written in pink, lowercase letters, followed by "COMP" in gray, uppercase letters. The overall design is clean and modern.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Requlation Description:
Percutaneous, implanted, long-term intravascular catheter
880.5970 C.F.R Section:
Class:
D.
E.
11
Requlation Medical Specialty and Review Panel: General Hospital
Purpose for Submission:
The purpose of this submission is to remove a contraindication that was once cited in the Journal of Infusion Nursing but has since been revised.
Device Description:
The CT Midline Catheter is designed for peripheral vein catheterization and power injection of contrast media. The lumen is an open-ended design comprised of a soft radiopaque polyurethane material with barium sulfate for radiopacity. The lumen is connected to the extensions via a soft pliable hub with suture wing for secure placement. Clamps are provided on the extension tubes to prevent air/fluid contamination. Female luer connectors provide the connection for intravenous administration.
The CT Midline Catheter is available in a 4Fx20cm single-lumen, or 5Fx20cm double-lumen configuration. The outside diameter of the lumen has a reverse taper increasing gradually near the hub to aid in kink resistance and to provide a mechanical obstruction to bleeding from the venotomy. The lumen has depth marks every centimeter and numerical marks every fifth centimeter. The CT Midline is packaged sterile with the necessary accessories to facilitate insertion.
#### G. Indications for Use:
The CT Midlines are indicated for Short-Term peripheral access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.
#### H. Intended Use:
The CT Midlines are indicated for Short-Term peripheral access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media.
- -Comparison to Predicate Devices:
Medcomp®: CT Midline Section 6: 510(k) Site Summary
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the MedComp logo. The logo features a stylized, geometric shape in pink, resembling a simplified house or building outline. Below the shape, the word "med" is written in pink, followed by "COMP" in gray, indicating the company name.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
The proposed CT Midline is substantially equivalent to the predicate device. CT Midline (K141151), in all aspects, including but not limited to, indications for use, intended use, anatomical location, basic design, dimensions, lengths performance, biocompatibility, materials, manufacturing process and method of sterilization. All technology, including features, materials, and principles of operation are identical to what was cleared in K141151, CT Midline. The only difference between the predicate and the proposed device is a revision to the instructions for use (labeling). This revision is to remove a contraindication. The revision to the labeling does not change the intended use of the device, both revisions are in are in alight with the Journal of Infusion Nursing.
The difference or changes between the proposed CT Midline and the predicate, CT Midline (K141151), is that the revision to the instructions for use, below is a comparison matrix demonstrating substantial equivalence.
Table 5.1: 510(K) Summary: Design Comparison Matrix
| Device | Proposed Device:<br>CT Midline | Predicate Device:<br>CT Midline (K141151) | Substantially<br>Equivalent<br>Comparison |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Dimensions<br>or Lengths | Lumen: 4F Single and 5F<br>Double<br>Lumen Length: 20cm | Lumen: 4F Single and 5F<br>Double<br>Lumen Length: 20cm | Identical, no<br>change |
| Indications<br>for Use | The CT Midlines are<br>indicated for Short-Term<br>peripheral access to the<br>peripheral venous system<br>for selected intravenous<br>therapies, blood sampling,<br>and power injection of<br>contrast media. The<br>maximum recommended<br>infusion rate varies by<br>catheter French size and<br>is printed on the catheter. | The CT Midlines are<br>indicated for Short-Term<br>peripheral access to the<br>peripheral venous system<br>for selected intravenous<br>therapies, blood sampling,<br>and power injection of<br>contrast media. The<br>maximum recommended<br>infusion rate varies by<br>catheter French size and<br>is printed on the catheter. | Identical, no<br>change |
| Sterilization<br>Method | ETO | ETO | Identical, no<br>change |
| Materials<br>and<br>Processes | LUMEN: Tecothane<br>HUB and SUTURE WING:<br>Pellethane<br>LUERS: Isoplast<br>EXTENSIONS: Pellethane<br>CLAMPS: Acetal<br>Purple: single lumen<br>Purple: double lumen<br>I.D. RING: Lustran | LUMEN: Tecothane<br>HUB and SUTURE WING:<br>Pellethane<br>LUERS: Isoplast<br>EXTENSIONS: Pellethane<br>CLAMPS: Acetal<br>Purple: single lumen<br>Purple: double lumen<br>I.D. RING: Lustran | Identical, no<br>change |
| Performance<br>Testing | Power Injection:<br>5cc/sec (4F) | Power Injection:<br>5cc/sec (4F) | Identical, no<br>change |
Medcomp®: CT Midline Section 6: 510(k) Site Summary 6-3
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a pink geometric shape that resembles a stylized house or building. Below the shape, the word "med" is written in pink, while "COMP" is written in gray. The logo appears to be for a medical company.
-
K.
L
- 1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
7cc/sec (5F)
7cc/sec (5F)
## Bench / Performance Data / Non-Clinical Testing:
No performance testing was required for the revision to the instructions in conjunction with the substantial equivalence claims, effectively demonstrate the proposed device, CT Midline, is equivalent to the predicate device, CT Midline (K141151).
## Biocompatibility:
Biocompatibility was performed on the CT Midline in accordance with ISO 10993-1. According to ISO 10993-1 the final, finished 5F Double lumen CT midline was tested as an implant device with blood contact and prolonged contact duration (greater than 24 hours less than 30 days). Biocompatibility was performed on the finished and sterilized device in accordance with ISO 10993-12:2012. Biological Evaluation of Medical Devices - Part 12: Sample Preparation and Reference Materials, and 21CFR. Part 58 - Good Laboratory Practices for Nonclinical Laboratory Studies by a contracted laboratory that is ISO 17025 certified.
- Summary of Substantial Equivalence:
The proposed CT Midline is substantially equivalent to the predicate device, CT Midline (K141151), in all aspects, including but not limited to, indications for use, intended use, anatomical location, basic design, dimensions, lengths performance, biocompatibility, materials, manufacturing process and method of sterilization. The only difference between the predicate and the proposed device is a revision to the instructions for use (labeling). This revision is to remove a contraindication.
The proposed device, CT Midline, meets the performance criteria of design verification as specified by ISO standards, quidance documents and internal test protocols. The proposed device is substantially equivalent to the indicated legally marketed predicate as defined in paragraph above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.