The CT Midlines are indicated for Short-Term peripheral access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.
Device Story
CT Midline is a peripheral venous catheter designed for short-term vascular access. It facilitates intravenous therapy, blood sampling, and power injection of contrast media. The device is used in hospital settings by healthcare professionals. It consists of a catheter lumen (Tecothane), hub/suture wing (Pellethane), and purple-colored acetal clamps to signify power injection capability. The device is inserted into the peripheral venous system; clinicians monitor infusion rates printed on the catheter hub. Power injection allows for rapid delivery of contrast media during diagnostic imaging. The device benefits patients by providing a reliable, multi-purpose venous access point that supports both standard therapies and high-pressure contrast administration.
Clinical Evidence
Bench testing only. Performance verified via power injection flow rate testing and maximum static burst pressure testing in accordance with ISO 10555-1. Biocompatibility testing confirms materials meet ISO 10993 requirements.
Indicated for short-term peripheral venous access in patients requiring intravenous therapy, blood sampling, or power injection of contrast media. Contraindications not specified.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: This document is a 510(k) summary for a CT Midline device by MEDCOMP®, dated July 8, 2014. It includes the submitter's information, device name, common name, classification details, and predicate devices. The device is designed for peripheral vein catheterization and power injection of contrast media, available in 4Fx20cm single-lumen or 5Fx20cm double-lumen configurations.
The CT Midlines are indicated for Short-Term peripheral access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for medCoMP. The logo consists of a stylized geometric shape above the text "medCoMP". The geometric shape is made up of black lines and white space, forming an abstract design. The text "medCoMP" is in a serif font and is positioned below the geometric shape.
l 499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax: 215-256-1787
www.medcompnet.com
Indications for Use:
The CT Midlines are indicated for Short-Term peripheral access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.
# Comparison to Predicate Devices:
The CT Midline is substantially equivalent to the predicate devices in terms of intended use. materials, anatomical location, basic design, performance, labeling, manufacturing process and method of sterilization.
| Attribute | CT Midline<br>(Proposed) | Midline (Predicate)<br>K121094 | Pro-PICCCT<br>(Predicate) K091953 |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INDICATIONS<br>FOR USE: | The CT Midlines are<br>indicated for Short-<br>Term peripheral<br>access to the<br>peripheral venous<br>system for selected<br>intravenous therapies,<br>blood sampling, and<br>power injection of<br>contrast media. The<br>maximum<br>recommended<br>infusion rate varies<br>by catheter French<br>size and is printed on<br>the catheter. | The Midline<br>catheters are<br>indicated for short or<br>long term peripheral<br>access to the<br>peripheral venous<br>system for selected<br>intravenous therapies<br>and blood sampling.<br>(see<br>Contraindications)<br>For blood therapy it<br>is recommended that<br>a 4French or larger<br>catheter is used. | The PRO-PICCCT<br>catheter is indicated<br>for short term or<br>long term peripheral<br>access to the central<br>venous system for<br>intravenous therapy<br>and power injection<br>of contrast media<br>and allows for<br>central venous<br>pressure monitoring<br>when a 20gauge or<br>larger lumen is<br>used. For blood<br>sampling, infusion<br>or therapies use a<br>4F or larger<br>catheter. The<br>maximum<br>recommended<br>infusion rate varies<br>by catheter French<br>size and is printed<br>on the catheter. |
| WHERE USED: | Hospital | Hospital | Hospital |
| STERILITY: | 100% Ethylene Oxide | 100% Ethylene Oxide | 100% Ethylene Oxide |
| BIOCOMPATIBILITY: | Materials are identical to<br>legally marketed | Legally Marketed<br>510(k) K121094 | Legally Marketed<br>510(k) K091953 |
| Attribute | CT Midline<br>(Proposed) | Midline (Predicate) | Pro-PICC CT<br>(Predicate) K091953 |
| | K121094 with the<br>exception of the Acetal<br>clamp which was cleared<br>in legally marketed<br>K091953.<br><br>Biocompatibility<br>summaries located in<br>Section 15 page 1. | K121094 | |
| MATERIALS<br>AND<br>ADDITIVES: | LUMEN:<br>Thermedics<br>Tecothane TT2095A<br>(30% Barium Sulfate) | LUMEN:<br>Thermedics<br>Tecothane TT2095A<br>(30% Barium<br>Sulfate) | |
| | PRINTING:<br>Markem, Black | PRINTING:<br>Markem, Black | |
| | HUB and SUTURE<br>WING: Dow<br>Pellethane 2363-80A | HUB and SUTURE<br>WING: Dow<br>Pellethane 2363-80A | |
| | LUERS: Isoplast<br>2510. Natural White<br>additives: Titanium<br>Dioxide. | LUERS: Isoplast<br>2510. Natural White<br>additives: Titanium<br>Dioxide. | CLAMPS: Halkey<br>Roberts Acetal,<br>Purple |
| | EXTENSIONS:<br>Dow Pellethane<br>2363-80A | EXTENSIONS:<br>Dow Pellethane<br>2363-80A | |
| | CLAMPS: Halkey<br>Roberts Acetal<br>Copolymer-<br>Purple - single lumen<br>Purple- double lumen | CLAMPS: Halkey<br>Roberts Acetal<br>Copolymer-<br>Natural - single<br>lumen<br>Natural- double<br>lumen | |
| | I.D. RING: ABS<br>Lustran 348 | I.D. RING: ABS<br>Lustran 348 | |
| | Drawings in Section<br>11. | Drawings in Section<br>11. | |
| Design<br>Specifications | LUMEN and<br>TAPER LENGTH<br>AVG.:<br>Single lumen 4F - | LUMEN and<br>TAPER LENGTH<br>AVG.:<br>Single lumen 4F - | N/A |
Medcomp CT Midline Summary 5-2
{2}------------------------------------------------
{3}------------------------------------------------
| Attribute | CT Midline<br>(Proposed) | Midline (Predicate)<br>K121094 | Pro-PICC CT<br>(Predicate) K091953 |
|-----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| | short taper 2.756 inch<br>(7cm) (Midline) | short taper 2.756 inch<br>(7cm) (Midline) | |
| | Double 5F – long<br>taper 3.44 to 5.44<br>inch. | Double 5F – long<br>taper 3.44 to 5.44<br>inch. | |
| | LUMEN I.D./O.D.<br>AVG.: Proximal to<br>taper (applies to both<br>short and long taper)<br>4F Single: I.D. .045<br>inch (1.14mm)<br>O.D. .082<br>inch (2.08mm)<br>5F Double*: I.D.<br>.039 inch (.99mm)<br>O.D. .092<br>inch (2.24mm)<br>* Equivalent<br>diameter of each<br>lumen based on D-<br>lumen cross-section<br>area. | LUMEN I.D./O.D.<br>AVG.: Proximal to<br>taper (applies to both<br>short and long taper)<br>4F Single: I.D. .045<br>inch (1.14mm)<br>O.D. .082<br>inch (2.08mm)<br>5F Double*: I.D.<br>.039 inch (.99mm)<br>O.D. .092<br>inch (2.24mm)<br>* Equivalent diameter<br>of each lumen based on<br>D-lumen cross-section<br>area. | |
| | TIP LUMEN<br>I.D./O.D. AVG.:<br>4F Single: I.D. .032<br>inch (.81mm)<br>O.D. .052<br>inch (1.32mm)<br>5F Double*: I.D.<br>.031 inch (.79mm)<br>O.D. .068<br>inch (1.73mm)<br>* Equivalent<br>diameter of each<br>lumen based on D-<br>lumen cross-section<br>area. | TIP LUMEN<br>I.D./O.D. AVG.:<br>4F Single: I.D. .032<br>inch (.81mm)<br>O.D. .052<br>inch (1.32mm)<br>5F Double*: I.D.<br>.031 inch (.79mm)<br>O.D. .068<br>inch (1.73mm)<br>* Equivalent diameter<br>of each lumen based on<br>D-lumen cross-section<br>area. | |
| | LUMEN LENGTH:<br>20cm (midline).<br>Open ended design.<br>DEPTH | LUMEN LENGTH:<br>20cm (midline).<br>Open ended design.<br>DEPTH | |
P
Medcomp CT Midline Summary 5-4
{4}------------------------------------------------
| | | CT Midline 15<br>(Proposed) | Midline (Predicate);<br>K121094 .; I | |
|------------------------|----------------------------|--------------------------------------------|--------------------------------------|--|
| | | MARKING: | MARKING: | |
| | | Number every Scim | Number every Scm | |
| | | and depth mark every | and depth mark every | |
| | | cm. | cm. | |
| | | HUB: With suture | HUB: With suture | |
| | | wing, all models. | wing, all models. | |
| medCoMP | | Contains French size<br>on hub. | Contains French size<br>on hub. | |
| 1499 Delp Drive | | | | |
| Harleysville, PA 19438 | | LUER: Easy Grip<br>design. | LUER: Easy Grip™<br>design. | |
| | | | | |
| Tel: 215-256-4201 | | EXTENSION | EXTENSION | |
| Fax: 215-256-1787 | | AVG.: Clear with | AVG .: Clear with | |
| | | clamp. | clamp. | |
| www.medcompnet.com | | 4F Single - 19 gauge | 4F Single - 19 gauge | |
| | | 5F Double - 18 | 5F Double - 18 | |
| | | gauge | gauge | |
| | | ALL SINGLE, AND | ALL SINGLE, AND | |
| | | DOUBLE | DOUBLE | |
| | | EXTENSIONS: | EXTENSIONS: | |
| | | I.D .: . 070 inches | I.D . 070 inches | |
| | | O.D .: . 106 inches | O.D .: 106 inches | |
| | | I.D. RING WITHIN | I.D. RING WITHIN | |
| | | CLAMP: On CT | CLAMP: On | |
| | | Midline contains | Midline contains | |
| | | product name and | extension gauge and | |
| | | Max rate of 7cc/sec. | French size with | |
| | | | lumen length. | |
| | | Drawings in Section | Drawings in Section | |
| | | ] l . | l 2. | |
| | MECHANICAL/<br>PERFORMANCE | AIR LEAKAGE: | | |
| | TESTING: | 4F | | |
| | | All versions passed in accordance with ISO | | |
| | | 10555-1. Annex D. Reference testing | | |
| | | summaries and protocols in Section 18. | | |
| | | SF | | |
| | | All versions passed in accordance with ISO | | |
| | | 10555-1. Annex D. Reference testing | | |
| | | summaries and protocols in Section 18. | | |
| | | LIQUID LEAKAGE: | | |
ﺴﺴ
{5}------------------------------------------------
| Attribute | CT Midline<br>(Proposed) | Midline (Predicate)<br>K121094 | Pro-PICC CT<br>(Predicate) K09195- |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | All versions passed in accordance with ISO 10555-<br>1, Annex C. Reference testing summaries and<br>protocols in Section 18. | | |
| Power Injection<br>Capability<br>Testing | POWER INJECTION<br>FLOW RATE: | | POWER INJECTION<br>FLOW RATE: |
| | 4F:<br>V-9131:<br>Power Injection Flow<br>Rate / Injection<br>Pressure (Avg.)<br>Flow Rate (cc/sec)- 4.9<br>Machine Pressure (psi)-<br>213<br>HYDAC Pressure (psi)-<br>168 | | 4F:<br>4Fx50cm Single<br>Power Injection Flow<br>Rate / Injection<br>Pressure (Avg.)<br>Flow Rate (cc/sec)-<br>3.9<br>Machine Pressure<br>(psi)- 183 |
| | 5F:<br>V-9077:<br>Power Injection Flow<br>Rate / Injection<br>Pressure (Avg.)<br>Flow Rate (cc/sec)- 6.9<br>Catheter Pressure (psi)-<br>244<br>HYDAC Pressure (psi)-<br>181 | | 5F:<br>5Fx55cm Double<br>Power Injection<br>Flow Rate / Injection<br>Pressure (Avg.)<br>Flow Rate (cc/sec)-<br>4.9<br>Machine Pressure<br>(psi)- 212 |
| | MAX STATIC BURST: | | MAX STATIC<br>BURST: |
| | 4F:<br>V-9043:<br>The average maximum<br>burst pressure was<br>302±5psi. The range of<br>burst pressures was 292-<br>312 psi. | | 4F:<br>The average maximum<br>burst pressure was<br>288±4psi. The range<br>of burst pressures was<br>279-293 psi. All<br>samples burst along<br>the lumen. |
| | 5F:<br>V-9073:<br>The average maximum<br>burst pressure was<br>249±8psi. The range of<br>burst pressures was 241-<br>262 psi. | | 5F:<br>The average maximum<br>burst pressure was<br>269±4psi. The range<br>of burst pressures was<br>262-278 psi. All<br>samples failed by<br>rupture of the lumen. |
Comparison to Predicate Devices (cont.):
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a geometric design above the word "medCOMP." The geometric design consists of a stylized shape, possibly representing a stylized letter or symbol, with a combination of black and white areas creating a sense of depth. The word "medCOMP" is written in a serif font, with the letters in lowercase except for the "COMP" part, which is in uppercase. There is a period at the end of the word.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4207
Fax: 215-256-1787
www.medcompnet.com
The proposed device is substantially equivalent to the predicate devices because the data demonstrates the proposed device matches the power injection indication and performance of the predicate K091953 while being otherwise identical to the K121094 predicate device.
## Performance Standards:
Performance standards have not been established by the FDA under section 514 of the Federal Food, Drug and Cosmetic Act.
# Nonclinical Performance Tests:
The following tests were performed to establish the device's equivalence to the relevant predicate device:
- o Power Injection Flow Rate
- � Max Static Burst
These tests highlight the relevant difference between the proposed device and predicate K121094 by testing for the safety and effectiveness of the proposed device with regard to the expanded indication for use of power injection. As the predicate K121094 was not indicated for power injection, the K091953 performance data is discussed in this submission to establish that the proposed device is substantially equivalent in terms of power injection performance.
# Biocompatibility:
Results for all biocompatibility testing demonstrate the materials used meet the requirements of ISO 10993.
## Technological Characteristics:
The principles of operation are the same as the predicate devices, with the exception that the proposed device is indicated for power injection of contrast media. Fundamentally, the proposed device is physically identical to the predicate K 121094 aside from the clamp, which is now colored purple to indicate power injection. The purpose of this submission is to establish that the proposed device can be properly indicated for power injection as supported by the provided data. There are no new questions raised regarding the safety or effectiveness of the device.
### Summary of Substantial Equivalence:
The proposed device meets the performance criteria of design verification as specified by ISOstandards, guidance documents and test protocols. The proposed device has the same intended use, operation and function as the predicates. There are no differences that raise new issues of safety and effectiveness. The proposed device is substantially equivalent to the logally marketed predicate devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# July 8, 2014
Medical Components, Inc. Timothy Holwick International Principal Regulatory Associate 1499 Delo Drive Harleysville, PA 19438
Re: K141151
Trade/Device Name: CT Midline Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long-term intravascular catheter Regulatory Class: II Product Code: LJS Dated: May 5, 2014 Reccived: May 6, 2014
Dear Mr. Holwick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Holwick
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K141151
Device Name
Medcomp CT Midline
Indications for Use (Describe)
The CT Midlines are indicated for Short-Term peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
.
Digitally signed by Richard C. Chapman -S Date: 2014.07.08 11:11:27 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.