K170761 · Lightmed USA, Inc. · IYO · Apr 13, 2018 · Radiology
Device Facts
Record ID
K170761
Device Name
LIGHTSonic BUBM
Applicant
Lightmed USA, Inc.
Product Code
IYO · Radiology
Decision Date
Apr 13, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
LIGHTSonic B | UBM is a non-invasion diagnostic ophthalmic ultrasound instrument specifically designed for measurements inside the ocular structures including orbital, anterior and posterior segment of eye.
Device Story
Portable ultrasound biometric ruler; used for ophthalmic applications. Inputs: ultrasonic echoes collected via B-scan or UBM probes (direct contact or immersion). Operation: microprocessor-controlled system converts echoes into digital pulses; generates 2D images and measurements (axial length, anterior chamber depth) based on time delays. Used in clinical settings by eye care professionals. Output: live 2D images and measurements displayed on LCD touch screen; supports DICOM connectivity for EMR integration, printing, and export (PDF, JPEG, AVI). Benefits: facilitates identification/measurement of ocular pathologies, particularly when view is obscured (e.g., cataracts).
Clinical Evidence
Bench testing only. No clinical data provided. Performance validated via biocompatibility testing (ISO 10993-1, -5, -10), electrical safety (AAMI/ANSI ES60601-1), EMC (IEC 60601-1-2), and software verification/validation per FDA guidance for moderate level of concern devices.
Indicated for non-invasive diagnostic ophthalmic ultrasound imaging and measurement of ocular structures, including orbital, anterior, and posterior segments of the eye.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
April 13, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure embracing a sphere. The FDA acronym is in a blue square, and the full name of the agency, "U.S. Food & Drug Administration," is in blue text to the right of the square.
Lightmed USA, Inc. % Ms. Angel Hsieh Regulatory Affairs Administrator 1130 Calle Cordillera SAN CLEMENTE CA 92673
Re: K170761
Trade/Device Name: LIGHTSonic B | UBM Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: March 1, 2018 Received: April 2, 2018
Dear Ms. Hsieh:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K170761
Device Name
LIGHTSonic B | UBM
Indications for Use (Describe)
LIGHTSonic B | UBM is a non-invasion diagnostic ophthalmic ultrasound instrument specifically designed for measurements inside the ocular structures including orbital, anterior and posterior segment of eye.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# Section 5 510(k) Summary
#### I. SUBMITTER
LIGHTMED USA, INC. 1130 Calle Cordillera, San Clemente, CA 92673, U.S.A. T: +1-949-218-9555 F: +1-949-218-9556 Contact Person: Katrina Hsu/Regulatory Affairs Administrator Date Prepared: June 30, 2017
### II. DEVICE
Trade name: LIGHTMED Ultrasound System Model name: LIGHTSonic B | UBM Classification name: System, Imaging, Pulsed Echo, Ultrasonic Classification Panel: Radiology Regulation number: 892.1560, 892.1570 Product code: IYO, ITX
#### III. PREDICATE DEVICE
Substantial equivalence to the following predicate device is as follows:
### IV. DEVICE DESCRIPTION
The LIGHTMED Ultrasound System is a portable ultrasound biometric ruler intended for use in ophthalmic applications. The system allows for the measurement of several key ocular features including axial length (AXL), anterior chamber depth. The device is used by coupling the probe / transducer to the eye either through direct contact or immersion methods. Available modes are biometric B-scan and UBM.
{4}------------------------------------------------
The B-Scan mode produces a live, two-dimensional image to facilitate the identification and measurement of ocular pathologies in the posterior-chamber of the eye, particularly when view of the chamber is obscured, such as is the case with cataracts.
The UBM mode can be used with a 35 MHz or a 50 MHz transducer. Because of the higher frequency of the transducer, it is expected that its greatest field of application will be in visualizing the anterior segment, because the focus area is about 11 mm from the transducer plane. The system will visualize other parts of the eye, but the resolution is not as high.
LIGHTMED Ultrasound System is a stand-alone system that runs on a Windows 10 platform and may be networked (by the user) for interface with electronic medical records systems, printing, and other purposes. The system consists of the LIGHTSonic B|UBM console, ultrasound probe(s) and transducer, and foot pedal.
## V. INDICATIONS FOR USE
The LIGHTMED Ultrasound System is a non-invasion diagnostic ophthalmic ultrasound instrument specifically designed for measurements inside the ocular structures including orbital, anterior and posterior segment of eye.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A summary listing of the LIGHTMED Ultrasound System design characteristics that are identical as the predicate is provided below:
| product name | LIGHTMED Ultrasound System | Vupad System |
|--------------|----------------------------|--------------|
| model name | LIGHTSonic B UBM | Vupad |
{5}------------------------------------------------
| manufacturer | LIGHTMED USA, INC. | Sonomed Inc. |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K170761 | K140199 |
| intended use | * make measurements inside<br>the ocular structures including<br>- orbital structure<br>- anterior segment of eye<br>- posterior segment of eye | *visualize the interior of the eye<br>*make measurement inside the<br>eye such as<br>- axial length<br>- anterior chamber depth<br>- lens thickness |
| hard drive | 128 GB C-fast card | 128 GB SSD solid-state drive |
| operating<br>voltage | 100 - 240 VAC, 50/60 Hz<br>auto-switching medical-grade<br>power supply | 100 - 240 VAC, 50/60 Hz<br>auto-switching medical-grade<br>power supply |
| control<br>interface | Operator uses LCD<br>multi-touch control panel and<br>foot pedal switch to collect<br>exam data. | Operator uses LCD multi-touch<br>control panel and foot pedal<br>switch to collect exam data. |
| EMR<br>connectivity | DICOM | DICOM |
| general method<br>of operation | * Echoes converted to images<br>on a screen<br>* Measurement made by<br>time delays | * Echoes converted to<br>images on a screen<br>* Measurement made by<br>time delays |
| digital system | Echoes converted into digital<br>pulses, all operation carried out<br>digitally. | Echoes converted into digital<br>pulses, all operation carried out<br>digitally. |
| probe design | * B-scan: sealed probes<br>with focused transducer<br>* UBM: water path probe with<br>interchangeable focused<br>transducer | * B-scan: sealed probes<br>with focused transducer<br>* UBM: water path probe with<br>interchangeable focused<br>transducer |
| pulse repetition<br>frequency | B-scan & UBM: 2560 Hz | B-scan & UBM: 2560 Hz |
| export image | PDF ; JPEG; AVI | PDF ; JPEG; AVI |
| printer | Any Windows-compatible<br>printer (separate) | Any Windows-compatible printer<br>(separate) |
{6}------------------------------------------------
A summary listing of the LIGHTMED Ultrasound System design characteristics
| product name | LIGHTMED Ultrasound System | Vupad System |
|---------------|--------------------------------------------------|--------------------------------|
| model name | LIGHTSonic B UBM | Vupad |
| manufacturer | LIGHTMED USA, INC. | Sonomed Inc. |
| 510(k) number | K170761 | K140199 |
| unit | 12.75" w x 3" d x 8" h<br>(32.3 x 7.6 x 20.3 cm) | 13.3" w x 8.0" d x 2" h |
| device system | The LIGHTSonic B UBM is | Vupad is an innovative |
| | an advanced | ultrasonic system, composed of |
| | microprocessor-controlled | the following sub-systems: |
| | ultrasonic system, composed of | * LCD touch screen |
| | the following sub-systems: | * sealed A-probe |
| | * LCD control panel | * B-probes |
differ from the predicate is provided below:
{7}------------------------------------------------
| | * B-scan probes<br>* UBM probes<br>* footswitch | * UBM probes<br>* foot pedal |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| operating<br>system | Microsoft Windows 10 | Microsoft Windows 8 |
| annotation | image annotation with six tools:<br>orientation, angle, distance,<br>pointer, area, text | automatic annotation of images<br>and video clips |
| scan controls | fully adjustable time gain<br>control (TGC) | fully adjustable time-varied gain<br>(TVG), baseline, log gain, and<br>exponential gain (e-gain) |
| compliance<br>standard | * US Federal Performance<br>Standards 21 CFR 892.1560 for<br>Class II Ultrasonic products<br>- AAMI/ANSI<br>ES60601-1:2005/(R)2012 And<br>A1:2012, C1:2009/(R)2012<br>And A2:2010/(R)2012<br>- IEC 60601-1-2: 2007<br>- IEC 60601-1-6:2013<br>- IEC 60601-2-37:2007<br>- ISO 14971:2007<br>- IEC 62366:2014 | * US Federal Performance<br>Standards 21 CFR 892.1560 and<br>1570 for Class II Ultrasonic<br>products<br>- IEC 60601-1<br>- EN/IEC 60601-1-2 (2001)<br>- IEC 60601-2-37 |
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility testing
{8}------------------------------------------------
The patient contact components for B-scan and UBM probes along with LIGHTMED Ultrasound System conducted cytotoxicity, sensitization, and irritation testing in accordance with FDA guidance- Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." The patient contact materials for B-scan and UBM probes fall into surface device and the contact is mucosal membrane. Contact duration is A-limited. The results showed patient contact materials conform to ISO 10993-1, 10993-5 and ISO 10993-10.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on LIGHTMED Ultrasound System. It complies with the following applicable federal and international safety and performance standards.
- * AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text)
- * IEC 60601-1-2:2007
- * IEC 60601-1-6:2013
- * IEC 60601-2-37:2007
- * ISO 14971:2007
- * IEC 62366:2014
### Software verification and validation testing
Software verification and validation testing were performed, and documents are established as recommended by FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a Moderate Level of Concern, since failures or latent flaws in the software could directly result in non-serious injury to the patient, operator, and/or bystander.
{9}------------------------------------------------
## Measurement accuracy
LIGHTMED Ultrasound System has been validated and is safe and effective for the functions described in the indications for use.
# VIII. CONCLUSIONS
The LIGHTMED Ultrasound System is substantially equivalent to the predicate devices in technical characteristics, design features, operating principles, functional and performance characteristics, and for the intended uses in the stated medical specialties. The LIGHTMED Ultrasound System is designed to comply with applicable federal and international safety and performance standards. There are no new safety and effectiveness issues being raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.