The VuPad ultrasound system is a multi-purpose computer-based ultrasonic diagnostic system for ophthalmic application, intended to both visualize the interior of the eye by means of ultrasound and to make measurements inside the eye, including the measurement of axial length for determination of IOL Power.
Device Story
VuPad is a portable, computer-based ophthalmic ultrasound system; operates on Windows 8 platform. Inputs: ultrasound signals from probes/transducers (10 MHz A-scan; 12.5, 20, 35, 50 MHz B-scan). Operation: probe coupled to eye via direct contact or immersion; echoes converted to digital pulses/images. Features: biometric A-scan (axial length, anterior chamber depth, lens thickness), UBM (high-frequency B-scan), IOL power calculation formulas, and 50-frame video clip storage. Used in clinical settings by eye care professionals. Output: live A-scan traces, 2D B-scan images, and measurement data displayed on integrated LCD touch screen. Assists clinicians in diagnosing ocular pathologies (e.g., when view is obscured by cataracts) and surgical planning. Benefits: provides diagnostic imaging and biometric data in a single, portable, ergonomic unit.
Clinical Evidence
Bench testing only. Performance data includes A-scan measurement accuracy (±0.1 mm clinical, ±0.02 mm theoretical) and acoustic output (ISPTA.3 < 50 mW/cm2, MI < 0.23). No clinical study data presented.
Technological Characteristics
Portable ultrasound system with integrated LCD touch screen. Transducers: 10 MHz (A-scan); 12.5, 20, 35, 50 MHz (B-scan). Connectivity: networked for EMR/printing. Software: Windows 8 platform. Standards: IEC 60601-1, IEC 60601-2-37, EN/IEC 60601-1-2.
Indications for Use
Indicated for ophthalmic patients requiring visualization of the interior of the eye and ocular measurements, including axial length for IOL power determination.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K092637 — PACSCAN PLUS, MODELS 300P AND 300AP · Sonomed, Inc. · Sep 24, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K140199
Page 1 of 7
# 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92
## Date: March 12, 2014
1. Company and Correspondent making the submission: Sonomed Inc. Name: Address: 1979 Marcus Ave Lake Success, NY, 11798 U.S.A. Telephone: 516-354-0900 Fax: 516-354 Website: www.sonomed.com Contact: Mr. Charles C. O'Neal, Quality Manager E-mail: coneal@escalonmed.com
#### 2. Device:
| Trade/proprietary name: VuPad | |
|--------------------------------|------------------------------------------|
| 510(k) Number: | K140199 |
| Common Name: | Diagnostic ultrasound system |
| Classification Name: | System, imaging, pulsed echo, ultrasonic |
#### 3. Predicate Devices:
| Manufacturer: | Sonomed, Inc. |
|----------------|---------------------------------------|
| Device: | E-Z Scan 5500+ A-Scan / B-Scan System |
| 510(k) Number: | K040668 |
Manufacturer: Sonomed, Inc. Device: VuMax System 510(k) Number: K060626
- Classification Names & Citations: 4. Classification: Class 2 Classification Code: 21CFR 892.1560, 1570. IYO, ITX, system, imaging, pulsed echo, ultrasonic, .
## 5. Description:
The VuPad is a portable ultrasound biometric ruler intended for use in ophthalmic applications. The system allows for the measurement of several key ocular features including axial length (AXL), anterior chamber depth, and lens thickness while also aiding in the calculation of associated IOL power for implanted lenses.
The device is used by coupling the probe / transducer to the eye either through direct contact or immersion methods. Available modes are biometric A-scan, and UBM (high frequency B-scan).
The A-scan mode of the system features a live A-scan trace with storage for up to five scans. There are five (auto/ manual) examination modes that use three different tissue velocities to calculate individual intraocular distances within the eye (ACD, Lens, and Vitreous). Other features include: post examination review of scans & measurement; five IOL formulas (six refractive and three post refractive) for lens power calculations; an immersion scanning capability for zero corneal compression of the eye while scanning and storage for up to five different user profiles.
> Sonomed. Inc 1979 Marcus Avenue Lake Success, NY 11042, USA Tel 800-227-1285 / 516-354-0900 Fax 516-354-5902
{1}------------------------------------------------
The B-Scan mode produces a live, two-dimensional image to facilitate the identification and measurement of ocular pathologies in the posterior-chamber of the eye, particularly when view of the chamber is obscured, such as is the case with cataracts.
The VuPad is a stand-alone system that runs on a Windows 8 platform and may be networked (by the user) for interface with electronic medical records systems, printing, and other purposes. The system consists of the VuPad console, ultrasound probe(s) and transducer(s), and foot pedal.
#### ි. Indications for Use:
The VuPad ultrasound system is a multi-purpose computer-based ultrasonic diagnostic system for ophthalmic application, intended to both visualize the interior of the eye by means of ultrasound and to make measurements inside the eye, including the measurement of axial length for determination of IOL Power.
#### 7. Comparison with predicate device:
Sonomed. Inc. believes that the technologies incorporated into the VuPad are substantially equivalent to those of the E-Z Scan 5500+ A-scan / B-scan system and the VuMax System.
A summary listing of design characteristics that are shared between the VuPad and the established predicate devices is provided on the following pages.
A summary listing of design characteristics that differ between the VuPad and the established predicate devices is also provided on the following pages.
{2}------------------------------------------------
A summary listing of VuPad design characteristics that are the same as established predicate devices is provided below:
| Parameter. | VuPad<br>(K140199 ) | EZScan AB5500+<br>(K040668) | VuMax<br>(K060626) |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarities | | | |
| Intended Use | The VuPad ultrasound<br>system is a multi-<br>purpose computer-<br>based ultrasonic<br>diagnostic system for<br>ophthalmic application,<br>intended to both<br>visualize the interior of<br>the eye by means of<br>ultrasound and to make<br>measurements inside<br>the eye, including the<br>measurement of axial<br>length for determination<br>of IOL Power. | The EZ Scan ultrasound<br>system is a multi-purpose<br>computer based ultrasonic<br>diagnostic system for<br>ophthalmic applications,<br>intended to both visualize<br>the interior of the eye by<br>means of ultrasound and to<br>make measurements inside<br>the eye, including the<br>measurement of axial<br>length for determination of<br>IOL Power. | The VuMAX<br>ultrasound syst5em<br>is a multi-purpose<br>computer -based<br>ultrasonic diagnostic<br>system for<br>ophthalmic<br>applications,<br>intended to both<br>visualize the interior<br>of the eye by means<br>of ultrasound and to<br>make measurements<br>inside the eye. |
| Ultrasound Modes | Ophthalmic A and B<br>Scans | Ophthalmic A and B Scans | Ophthalmic A and B<br>Scans |
| Technology | Visualization by<br>Ultrasound | Visualization by Ultrasound | Visualization by<br>Ultrasound |
| General Method of<br>Operation | Echoes converted to<br>images on a screen.<br>Measurement made by<br>time delays | Echoes converted to<br>images on a screen.<br>Measurement made by<br>time delays | Echoes converted to<br>images on a screen.<br>Measurement made<br>by time delays |
| Digital System | Echoes converted into<br>digital pulses, all<br>operation carried out<br>digitally | Echoes converted into<br>digital pulses, all operation<br>carried out digitally | Echoes converted<br>into digital pulses, all<br>operation carried out<br>digitally |
| Ability To Make<br>Measurements | Can make<br>measurements using A-<br>scan technology | Can make measurements<br>using A-scan technology | Can make<br>measurements using<br>A-scan technology |
| Eye-transducer Interface | Sealed probes with<br>scanning transducer<br>behind ultrasound<br>transparent membrane<br>(10, 12.5, 20 MHz); and<br>Stand-off or "nose-<br>scone" separates<br>exposed transducer<br>from patient (35, 50<br>MHz) | Sealed probes with<br>scanning transducer behind<br>ultrasound transparent<br>membrane (10 MHz) | Stand-off or "nose-<br>scone" separates<br>exposed transducer<br>from patient (35, 50<br>MHz) |
| IOL Power Calculation | Various formulas<br>available | Various formulas available | Not available |
| Method of generating A-<br>Scans | Separate A-Scan<br>transducer, A-scan<br>measuring system | Separate A-Scan<br>transducer, A-scan<br>measuring system | Line traced on B-<br>Scan, A-scan shown,<br>caliper for<br>measurement |
| Focus feature | Improves resolution by<br>reducing transducer<br>ringing by software | Not available. | Improves resolution<br>by reducing<br>transducer ringing by<br>software |
| A-Scan Probe Design | Closed Fixed Single-<br>Element with Internal<br>Fixation Light | Closed Fixed Single-<br>Element with Internal<br>Fixation Light | N/A |
| Parameter | VuPad<br>(K140199) | EZScan AB5500+<br>(K040668) | VuMax<br>(K060626) |
| Similarities (cont.) | | | |
| A-Scan Transducer<br>Frequency | 10 MHz | 10 MHz | N/A |
| A-Scan Available Probe<br>Configurations | Solid Tip<br>(for immersion scan)<br>Soft-Touch<br>(for direct contact scan) | Solid Tip<br>(for immersion scan)<br>Soft-Touch<br>(for direct contact scan) | N/A |
| A-Scan Measurements | Anterior Chamber<br>Depth<br>Lens Thickness<br>Axial Length<br>Manual Measurements | Anterior Chamber Depth<br>Lens Thickness<br>Axial Length<br>Manual Measurements | N/A |
| A-Scan Measurement<br>Accuracy | ±0.1 mm (clinical)<br>±0.02 mm (theoretical) | ±0.1 mm (clinical)<br>±0.023 mm (theoretical) | N/A |
| A-Scan Measurement<br>Range | 18 - 40 mm | 18 - 40 mm | N/A |
| A-Scan Measurement<br>Modes | Automatic - Cataract<br>Automatic - Dense<br>Cataract<br>Automatic - Aphakic<br>Automatic -<br>Pseduophakic<br>Manual | Automatic - Cataract<br>Automatic - Dense<br>Cataract<br>Automatic - Aphakic<br>Automatic - Pseduophakic<br>Manual | N/A |
| A-Scan IOL Formulas | Manual<br>Holladay<br>Regression-II<br>Theoretic-T<br>Binkhorst<br>Hoffer-Q<br>Haigis<br>Latkany Mypoic Post-<br>Refractive<br>Latkany Hyperoptic<br>Post-Refractive<br>Aramberri Double-K<br>Post-Refractive | Holladay<br>Regression-II<br>Theoretic-T<br>Binkhorst<br>Hoffer-Q<br>Haigis<br>Latkany Mypoic Post-<br>Refractive<br>Latkany Hyperoptic Post-<br>Refractive<br>Aramberri Double-K Post-<br>Refractive | N/A |
| A-Scan Tissue Velocity<br>Constants | Anterior Chamber<br>Lens<br>Vitreous | Anterior Chamber<br>Lens<br>Vitreous | N/A |
| A-Scan User Constants | Personalized A-<br>Constants<br>Surgeon Factors | Personalized A-Constants<br>Surgeon Factors | N/A |
| A-Scan Acoustic Output<br>Global Maximum | ISPTA.3 < 50 mW/cm2<br>MI < 0.23 | ISPTA.3 < 50 mW/cm2<br>MI < 0.23 | N/A |
| B-Scan Probe Design | Sealed Pivoting Single-<br>Element, and Open<br>Pivoting Single-<br>Element | Sealed Pivoting Single-<br>Element Only | Open Pivoting<br>Single-Element Only |
| B-Scan Lines per Scan | 256 | 128 or 256 | 256 |
| Selectable A-Scan Vector<br>while in B-Scan mode | Yes | Yes | Yes |
| Parameter | VuPad (K140199) | EZScan AB5500+ (K040668) | VuMax (K060626) |
| Similarities (cont.) | | | |
| B-Scan Scan Display Controls | Fully adjustable time-varied gain (TVG),<br>baseline, log gain, and<br>exponential gain | Fully adjustable time-varied gain (TVG), log gain, and<br>exponential gain | Fully adjustable time-varied gain (TVG),<br>baseline, log gain, and<br>exponential gain |
| B-Scan Video Clips | Capture and store 50-frame video clips | N/A | Capture and store<br>50-frame video clips |
Sonomed, Inc
1979 Marcus Avenue Lake Success, NY 11042, USA Tel 800-227-1285 / 516-354-0900 Fax 516-354-5902
{3}------------------------------------------------
.
·
،
{4}------------------------------------------------
.
and which and the
.
·
Sonomed, Inc 1979 Marcus Avenue Lake Success, NY 11042, USA Tel 800-227-1285 / 516-354-0900 Fax 516-354-5902
:
{5}------------------------------------------------
.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
A summary listing of VuPad design characteristics that differ from established predicate devices is provided below:
| Parameter | VuPad | EZScan AB5500+<br>K040668 | VuMax |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Differences | | | |
| Hardware<br>Configuration and<br>Components | System consists of<br>unit with integrated<br>LCD touch screen, A-<br>probe, sealed B-<br>probe, open<br>transducer water path<br>B-probe, calibration<br>cylinder, probe holder,<br>and foot pedal | System consists of unit<br>with integrated LCD<br>touch screen, A-probe,<br>sealed B-probe,<br>calibration cylinder,<br>probe holder, and foot<br>pedal | Systems consists of tower PC with<br>integrated ultrasound electronics,<br>PC monitor, open transducer water<br>path B-probe, keyboard, mouse,<br>and foot pedal |
| Control Interface | Operator uses tablet<br>PC with LCD touch<br>screen and foot pedal<br>switch to collect exam<br>data. | Operator uses LCD<br>touch screen and foot<br>pedal switch to collect<br>exam data. | Operator uses desktop computer<br>control interface to collect exam<br>data. Hardware includes monitor,<br>keyboard, mouse, and foot pedal<br>switch. |
| System Dimensions | 13.3" x 8.0" x 2.0" | 9.4" x 8.9" x 2.8" | Dimensions vary for multiple<br>components |
| Display Screen | Integrated LCD Panel<br>(10.1" diagonal wide-<br>screen, 1920 x 1080<br>pixel resolution) | Integrated LCD Panel<br>(5.25" x 3.4", 640x480) | PC Monitor<br>(17" Diagonal, 1280 x 1024 pixel<br>resolution) |
| Data Storage<br>Location | Storage within<br>software database<br>with ability to recall<br>patient exam records | Internal storage of a<br>single patient exam<br>(previous exam data<br>overwritten when new<br>exam initiated)<br>Capability to export exam<br>record to PC via serial<br>connector for long-term<br>storage | Storage within software database<br>with ability to recall patient exam<br>records |
| Printer | Any Windows-<br>compatible printer<br>(separate) | Sony UP-897MD video<br>printer (separate) | Any Windows-compatible printer<br>(separate) |
| A-Scan<br>Measurement<br>Accuracy | $\pm$ 0.1 mm (clinical)<br>$\pm$ 0.02 mm theoretical) | $\pm$ 0.1 mm (clinical)<br>$\pm$ 0.023 mm (theoretical) | N/A |
| B-Scan Transducer<br>Frequencies | 12.5 MHz, 20 MHz, 35<br>MHz, 50MHz | 10MHz | 35 MHz, 50MHz |
| B-Scan Transducer<br>drive and receiver | Circuitry suitable for<br>12.5-50MHz | Circuitry suitable for<br>10MHz | Circuitry suitable for 10-50 MHz |
| B-Scan Axial<br>Accuracy<br>(Theoretical) | 12.5 MHz: 0.2034 mm<br>35 or 50 MHz:<br>0.0146mm | 10 MHz: 0.2088 mm | 35 or 50 MHz: 0.0146mm |
{6}------------------------------------------------
K140199
Page 7 of 7
## 8. Conclusions:
The goal in designing the VuPad system was to combine the primary features of two previously marketed predicate devices into a single self-contained package along with enhancements in ergonomics and utility that embody the current state of the industry.
The core technologies incorporated into the VuPad are primarily based on two predicate devices that are currently legally marketed by Sonomed. Inc: the E-Z Scan 5500+ A-Scan / B-Scan system (K040668) and the VuMax System (K060626). The similarities in intended usage, method of application, and system capability between the VuPad and predicate devices are evidenced in the comparison tables provided herein.
Both the E-Z Scan 5500+ A/B system and the VuPad system provide Operators with a combination of B-scan and A-scan ultrasound. The A-scan technology incorporated into the VuPad system is fundamentally identical to that currently in use by the E-Z Scan 5500+ A/B system.
Similarly, the Ultrasound Biomicroscopy (UBM) capabilities of the VuPad System are fundamentally identical to that of the VuMAX. The VuPad utilizes the exact same models of 35Mhz and 50Mhz transducers as the VuMAX system to collect live B-scan images of the anterior segment of the eye. The intended usage and method of application for these transducers are the same for both systems.
The VuPad also provides users with the additional option of using 12.5 MHz or 20 MHz Bscan transducers for enhanced B-Scan imaging of the posterior segment of the eve. The intended use and application of these transducers are exactly the same as the 10 MHz Bscan transducer provided for use with the previously cleared E-Z Scan AB5500+. The dual transducer frequencies made available with the VuPad system (12.5 MHz and 20 MHz) provide users with superior image resolution and greater exam flexibility while maintaining a comparable scanning depth when compared to the use of a single 10 MHz transducer.
The differences in Hardware Configuration, Control Interface, System Dimensions, Display Screen. Data Storage Location, and Printing Capability evident in the VuPad system do not render the device substantially different from the predicate devices because they do not establish a new intended usage, nor do they significantly alter the core A-scan and B-scan scan technologies employed by the system. These variances in form factor and data presentation have been evaluated by Sonomed's Risk Management Team and shall be addressed and fully detailed within the Operator's Instruction Manual that shall accompany the system.
In accordance with the Federal Food, Drug and Cosmetic Act 21 CFR Part 807 and based on the information provided in this premarket notification Sonomed, Inc. has concluded that the VuPad is safe and effective and substantially equivalent to predicate devices as described herein.
- 9. Safety, EMC and Performance Data:
Electrical, mechanical, environments safety and performance testing according to standard IEC 60601-1, IEC 60601-2-37, and EN/IEC 60601-1-2(2001) are currently pending.
- 10. Sonomed Inc. will update and include in this summary any other information deemed reasonably necessary by the FDA.
{7}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the eagle.
Food and Drug Administration
Public Health Service
10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 1, 2014
Sonomed. Inc. % Mr. Charles O'Neal Quality Manager 1979 Marcus Avenue, Suite 105C LAKE SUCCESS NY 11042
Re: K140199
Trade/Device Name: VuPad Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: March 12, 2014 Received: March 20, 2014
Dear Mr. O'Neal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
Page 2-Mr. O'Neal
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Orug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Stalement below
#### 510(k) Number (if known) K140109
Device Name
VirPad
### Indications for Use (Describe)
The VePad ultrasound system is a multi-purpose computer-based ultrasonic system for ophitaline application, intended to both visualize the interior of the eye by neans of ultrasound and to make neasurements inside the eve. including the measurement of axial length for determination of IOL Power.
System: VuPad Transducer_10 Mhz
| Clinical Application | | | Mode of Operation | | | | | |
|---------------------------|----------------------------|---|-------------------|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | A-Mode (NEW) |
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Smh.7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden lime for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the dala needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@tda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
{10}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved. OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known) K140199
K140179
Device Name
VuPad
#### Indicalions for Use (Describe)
The Vullad ultrasound system is a multi-purpose computer-based ultrasonic system for ophitalian. intended to both visualize the interior of the eye by means of ultrasound and to make measurements inside the eye. including the measurement of axial length for determination of IOL. Power.
System: VuPad Transducer 12,5 Mhz
| Clinical Application | Specific<br>(Tracks 1 & 3) | Mode of Operation | | | | | | |
|---------------------------|----------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | ✓ | | | | | | |
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subparl D)
Over . The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Sm.h.7).
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources. gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimale or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Acl (PRA) Staff PRAStaff@lda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to. a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
{11}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) ייצולון א
Device Name Val¹ad
Indications for Use (Describe)
The Vullad ultrasound system is a multi-purpose computer-based ultrasmic system for ophication. intended to both visualize the interior of the eye by means of ultrasound and to make measurements inside the eye, including the measurement of axial length for determination of IOL Power.
System: VuPad Transducer_20 Mhz
| Clinical Application | | Mode of Operation | | | | | | |
|------------------------|-------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| General (Track 1 Only) | Specific (Tracks 1 & 3) | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | ✓ | | | | | | |
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
ﺎ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
•DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.'
The burden lime for this collection of information is estimated to average 79 hours per response, including the lime to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection.of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, lo:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{12}------------------------------------------------
DEPARTMENT OF HEAL.TH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known) K140199
Devico Name VuPad
Indications for Use (Describe)
The VuPad ultrisound system is a multi-purpose computer-based ultrasonic system for ophitalinic application. intended to both visualize the interior of the eye by newns of ultrassund and to make measurements inside the eye. including the measurement of axial length for determination of IOI. Power.
System: VuPad Transducer_ 35 Mhz
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|----------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | ✓ | | | | | | |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing dala sources, gather and mainlain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection. including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStall@ida.hhs.gov
"An agency may not conduct or sponsor. and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{13}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statumont below.
510(k) Number (if known) ド140199
Device Name VuPad
Indications for Use (Describe)
The Vullad ultrasound system is a multi-purpose computer-based ultrasonic system for ophitalinic application, intended to both visualize the interior of the eye by means of ultrasound and to make measurements inside the eye, including the measurement of axial length for determination of IOL Power.
System: VuPad Transducer_50 Mhz
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|----------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
Type of Use (Select one or both, as applicable)
الط Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
'DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.'
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, galher and maintain the data needed and complete line to review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.