Mako Partial Knee Application
K170584 · Mako Surgical Corp. · OLO · May 11, 2017 · Neurology
Device Facts
| Record ID | K170584 |
| Device Name | Mako Partial Knee Application |
| Applicant | Mako Surgical Corp. |
| Product Code | OLO · Neurology |
| Decision Date | May 11, 2017 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 882.4560 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Partial Knee Application (PKA), for use with the Mako System, is intended to assist the surgeon in providing software defined spatial boundaries for orientation to anatomical structures during orthopedic procedures. The Partial Knee Application (PKA), for use with the Mako System, is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include unicondylar knee replacement and/or patellofemoral knee replacement.
Device Story
Stereotactic instrument system for orthopedic surgery; includes robotic arm, optical detector, computer, and planning laptop. Uses patient CT data for pre-surgical planning and intraoperative navigation. Robotic arm acts as intelligent tool guide, providing passive constraint via software-defined spatial boundaries during implant preparation. Used in clinical settings by surgeons. Modification introduces 'eRequest LifeCycle' portal for preoperative case management and file transfer. System assists in implant placement and anatomical orientation; benefits patient through precise, software-guided surgical execution.
Clinical Evidence
Bench testing only. Validation included product specification verification and simulated-use testing of the eRequest LifeCycle workflow to ensure integration with the Mako System. All tests passed.
Technological Characteristics
Stereotactic instrument; robotic arm; optical detector; preoperative planning laptop. Uses CT imaging for navigation. Software-defined spatial boundaries for tool guidance. Connectivity via eRequest LifeCycle portal for case management.
Indications for Use
Indicated for surgical knee procedures, including unicondylar knee replacement and patellofemoral knee replacement, where stereotactic surgery is appropriate and rigid anatomical bony structures can be identified relative to a CT-based model.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
- Mako Partial Knee Application (K112507)
Related Devices
- K170581 — Mako Total Knee Application · Mako Surgical Corp. · May 11, 2017
- K170593 — Mako Total Hip Application · Mako Surgical Corp. · Apr 18, 2017
- K260222 — Mako Total Knee Application · Mako Surgical Corp. · Feb 25, 2026
- K171651 — EXCELSIUS GPS · Globus Medical, Inc. · Aug 16, 2017
- K052851 — VOYAGER LINUX WITH TACTILE GUIDANCE SYSTEM (TGS) · Mako Surgical Corp. · Nov 18, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Mako Surgical Corp. Shikha Khandelwal Staff Regulatory Affairs Specialist 2555 Davie Rd Fort Lauderdale, Florida 33317
May 11, 2017
Re: K170584
Trade/Device Name: Mako Partial Knee Application Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: February 27, 2017 Received: February 28, 2017
Dear Shikha Khandelwal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K170584
Device Name
Partial Knee Application
Indications for Use (Describe)
The Partial Knee Application (PKA), for use with the Mako System, is intended to assist the surgeon in providing software defined spatial boundaries for orientation to anatomical structures during orthopedic procedures.
The Partial Knee Application (PKA), for use with the Mako System, is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include unicondylar knee replacement and/or patellofemoral knee replacement.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold, black letters. Below "MAKO" is the text "SURGICAL CORP" in smaller, black letters. Above the word "MAKO" is a blue, curved shape that resembles a wave or an arc.
55 Davie Road · Ft. Lauderdale, FL 3 2044 · Fax 954.927.0 ww.makosuraical.com
510(K) SUMMARY
| Sponsor: | MAKO Surgical Corp.<br>2555 Davie Road<br>Fort Lauderdale, FL 33317 |
|----------|---------------------------------------------------------------------|
|----------|---------------------------------------------------------------------|
Contact Person:
Shikha Khandelwal, PhD Shikha.khandelwal@stryker.com Phone: 201-831-6921 Fax: 201-831-3921
Date Prepared: February 24, 2017
Proprietary Name: Mako Partial Knee Application
Common Name: Partial Knee Application (PKA)
Regulation Name: Stereotaxic Instrument
Regulation Number: 21 CFR 882.4560
Device Classification: Class II
Product Code: OLO
### Substantial Equivalence Claimed To:
The Mako Partial Knee Application is substantially equivalent to Mako Surgical's Partial Knee Application cleared via K112507.
Device Modification: An alternate online portal for case management and file transfer known as eRequest LifeCycle is being implemented for use during the preoperative planning phase of the Mako Partial Knee Application.
### Description:
The Mako System with the Partial Knee Application is a stereotactic instrument that includes a robotic arm, an integrated cutting system, an optical detector, a computer, dedicated instrumentation, operating software, a planning laptop, and tools and accessories.
{4}------------------------------------------------
The system's architecture is designed to support total and partial knee procedures and total hip procedures. With application specific hardware and software, the system provides stereotactic guidance during orthopedic surgical procedures by using patient CT data to assist a surgeon with pre-surgical planning, implant placement and interpretive/intraoperative navigation of the patient's anatomy.
The Mako robotic arm, once configured for a specific application, can serve as surgeon's "intelligent" tool holder or tool guide by passively constraining the preparation of an anatomical site for an orthopedic implant with software-defined spatial boundaries.
Summary of Technological Characteristics Compared to Predicate Devices: The technological characteristics of the Mako Partial Knee Application compared to the predicate device are listed below:
| Technological<br>Characteristics | Mako Partial Knee Application | Mako Partial Knee Application<br>(K112507) |
|----------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Major<br>Components | Guidance Module, robotic arm,<br>camera stand, cutting system,<br>preoperative planning laptop. | Guidance Module, robotic arm,<br>camera stand, cutting system,<br>preoperative planning laptop. |
| Tools/accessories | Various reusable and disposable<br>instruments | Various reusable and disposable<br>instruments |
| Image Use | CT | CT |
### Intended Use/Indications for Use:
The Partial Knee Application (PKA), for use with the Mako System, is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
The Partial Knee Application (PKA), for use with the Mako System, is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include unicondylar knee replacement and/or patellofemoral knee replacement.
### Performance Data:
| Validation / Verification<br>Method | Purpose | Validation / Verification<br>Results |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Product Specification<br>Verification | Verify that PKA fields and<br>values implemented into<br>the eRequest application<br>must match the THA<br>Product Specifications. | Pass |
| eRequest - Full System | Verify the integration of | Pass |
{5}------------------------------------------------
| Run Through for PKA<br>Application | the eRequest Lifecycle<br>into the Mako System<br>provides adequate<br>functionality to<br>successfully complete the<br>pre-operative planning<br>workflow | |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| eRequest LifeCycle PKA<br>Validation | Validate in a simulated-<br>use environment, with<br>appropriate user, that the<br>implementation of<br>eRequest LifeCycle into<br>the Mako System provides<br>adequate functionality to<br>successfully complete the<br>pre-operative workflow<br>and satisfies the customer<br>requirements. | Pass |
### Conclusions of Performance Testing:
Performance testing has demonstrated that the characteristics of the Mako Partial Knee Application are equivalent to the predicate device, and that the device is as safe and as effective as the predicate device and does not raise different questions of safety and effectiveness, and therefore, supports a determination of Substantial Equivalence.
Image /page/5/Picture/3 description: The image shows a light blue, abstract shape against a white background. The shape is curved and tapers at both ends, resembling a stylized wing or a swooping arc. The color is a soft, pastel blue, and the overall impression is minimalist and clean.