K170395 · Meditech Spine, LLC · MAX · Jun 2, 2017 · Orthopedic
Device Facts
Record ID
K170395
Device Name
Talos® Lumbar (HA) PEEK IBF Devices
Applicant
Meditech Spine, LLC
Product Code
MAX · Orthopedic
Decision Date
Jun 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Talos® Lumbar (HA) PEEK IBF Devices are intervertebral body devices intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Talos® Lumbar (HA) PEEK IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft. The Talos® Lumbar (HA) IBF PEEK Device is to be used in patients who have had six months of non- operative treatment. Talos® Lumbar (HA) PEEK IBF Devices are to be implanted via a direct posterior, transforaminal, lateral, or anterior approach in the lumbosacral spine. The Talos®-A (HA), Talos®-L (HA), Talos®-P (HA), Talos®-P WB (HA), Talos®-T (HA) and Talos®-TL (HA) are intended to be used with supplemental fixation. Additionally, the use of Hyperlordotic Talos®-A (HA) devices (lordotic angle greater than 20°) are intended to be used exclusively with anterior supplemental fixation.
Device Story
Intervertebral body fusion device; implanted in lumbar spine via posterior, transforaminal, lateral, or anterior approach; used with autograft/allograft bone graft; requires supplemental fixation. Device provides structural support to disc space to facilitate fusion. Surgeon implants device to treat discogenic back pain. Tantalum or titanium markers allow radiological positioning confirmation. HA-impregnated PEEK material promotes integration.
Clinical Evidence
Bench testing only. Performed static/dynamic compression and compression shear testing per ASTM F2077 and subsidence testing per ASTM F2267. Bacterial endotoxin testing performed per ANSI/AAMI ST-72:2011 (limit <20EU/device).
Technological Characteristics
Hydroxyapatite impregnated polyetheretherketone (HA PEEK) polymer; tantalum or titanium radiographic markers; ridged teeth for pullout resistance; various sizes and lordotic angles. Testing standards: ASTM F2077, ASTM F2267, ANSI/AAMI ST-72:2011.
Indications for Use
Indicated for skeletally mature patients with lumbar Degenerative Disc Disease (DDD) and up to Grade 1 spondylolisthesis at one or two contiguous levels (L2-S1) who have failed six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Talos® Intervertebral Body Fusion Devices (K150788)
Talos® Intervertebral Body Fusion Devices (K090707)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. The profiles are all connected, and they form a single, continuous line. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profiles.
June 2, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Meditech Spine, LLC Bruce Dunaway Chief Design Engineer 1447 Peachtree Street NE Suite 440 Atlanta, Georgia 30309
Re: K170395
Trade/Device Name: Talos® Lumbar (HA) PEEK IBF Devices Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: March 4, 2017 Received: March 8, 2017
Dear Mr. Dunaway:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely,
## Vincent J. Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170395
Device Name Talos® Lumbar (HA) PEEK IBF Devices
#### Indications for Use (Describe)
The Talos® Lumbar (HA) PEEK IBF device is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Talos® Lumbar (HA) PEEK IBF devices are intended to be used with autograft and/ or allograft comprised of cancellous and/or corticocancellous bone graft.
The Talos® Lumbar (HA) PEEK IBF Device is to be used in patients who have had six months of non- operative treatment.
Talos® Lumbar (HA) PEEK IBF devices are to be implanted via a direct posterior, transforaminal, laterior or anteriolateral approach in the lumbosacral spine. The Talos®-P WB (HA), Talos®-TL (HA), Talos®-T (HA), Talos®-L (HA), and Talos®-A (HA) are intended to be used with supplemental fixation.
Additionally, the use of Hyperlordotic Talos®-A (HA) devices (lordotic angle greater than 20°) are intended to be used exclusively with anterior supplemental fixation.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
|--------------------------------------------------------------------------------------|-----------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
# meditech spi
1447 Peachtree St NE Suite 440, Atlanta GA 30309 | p (678) 974-5287 | f (404) 759-2104 | support@meditechspine.com
## 510(k) Summarv
As required by section 807.92(c)
Meditech Spine, LLC is requesting marketing clearance for the Talos® Lumbar (HA) PEEK IBF.
| A. Sponsor/Manufacturer: | Meditech Spine, LLC<br>Registration Number: 3009405289<br>Bruce Dunaway, Chief Design Engineer<br>1447 Peachtree St NE Suite 440<br>Atlanta, GA 30309<br>678-974-5287 Phone<br>404-759-2104 Fax |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B. Trade Name: | Talos® Lumbar (HA) PEEK IBF Devices |
| Common Name: | Intervertebral Body Fusion Device |
| Classification Name: | Intervertebral body fusion device (21 CFR 888.3080 Class II,<br>Product Code MAX) |
| C. Predicate Devices: | K 150788: Meditech Spine, LLC Talos® Intervertebral Body Fusion<br>Devices (Primary)<br>K090707: Meditech Advisors, LLC Talos® Intervertebral Body<br>Fusion Devices (Additional) |
### D. Device Description:
The Talos® Lumbar (HA) PEEK IBF Devices are made of the polymer, hydroxyapatite impregnated polyetheretherketone (HA PEEK). The Talos® Lumbar (HA) PEEK IBF Device is available in six configurations: Talos®-P (HA), Talos®-P WB (HA), Talos®-T (HA), Talos®-TL (HA), Talos®-L (HA), and Talos®-A (HA). The devices are open devices with ridged teeth on superior and inferior ends to resist implant pullout. The Talos®-P (HA), Talos®-P WB (HA), Talos® –TL (HA) and Talos® –L are rectangular devices. The Talos® –T (HA) and Talos® – A (HA) have curved lateral walls and rounded edges. The implants are available in a range of sizes, as well as flat and lordotic angled implants to accommodate variations in patient anatomy. In addition, tantalum or titanium markers at the opposite ends are offered which allow radiological confirmation for proper IBF positioning.
- E. Intended Use:
Talos® Lumbar (HA) PEEK IBF Devices are intervertebral body devices intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1.
{4}------------------------------------------------
# meditech spine, IIc
1447 Peachtree St NE Suite 440, Atlanta GA 30309 | p (678) 974-5287 | f (404) 759-2104 | support@meditechspine.com
Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Talos® Lumbar (HA) PEEK IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft.
The Talos® Lumbar (HA) IBF PEEK Device is to be used in patients who have had six months of non- operative treatment.
Talos® Lumbar (HA) PEEK IBF Devices are to be implanted via a direct posterior, transforaminal, lateral, or anterior approach in the lumbosacral spine. The Talos®-A (HA), Talos®-L (HA), Talos®-P (HA), Talos®-P WB (HA), Talos®-T (HA) and Talos®-TL (HA) are intended to be used with supplemental fixation.
Additionally, the use of Hyperlordotic Talos®-A (HA) devices (lordotic angle greater than 20°) are intended to be used exclusively with anterior supplemental fixation.
- F. Technological Characteristics:
The technological characteristics of the Talos® Lumbar (HA) PEEK IBF devices are identical to the predicate devices in terms of indications for use and intended use. Size offerings and choices of lordotic angles have been expanded for the Talos® Lumbar (HA) PEEK IBF Devices. The significant difference between the Talos® Lumbar (HA) PEEK IBF Devices and the predicate is in material composition: hydroxyapatite impregnated polyetheretherketone (HA PEEK) vs. polyetheretherketone (PEEK)
- G. Non-clinical Testing:
Testing per ASTM F2077 (Static Compression, Static Compression Shear, Dynamic Compression, Dynamic Compression Shear) and ASTM F2267 (Subsidence) was performed on the Talos® Lumbar (HA) PEEK IBF devices to establish equivalency to the predicate device. Confirmation was achieved that the Talos® Lumbar (HA) PEEK IBF devices are substantially equivalent in safety and efficacy to the predicate device. In addition, bacterial endotoxin testing (BET) has been performed. BET as specified in ANSI/AAMI ST-72:2011 confirm an endotoxin limit less than 20EU per device.
- H. Conclusion:
The Talos® Lumbar (HA) PEEK IBF devices have identical indications for use, intended use, technological characteristics, design, and principles of operation as their predicate devices; as well as similar size and geometric offerings. The differentiating characteristic is the device material, which has been proven to perform equivalently to the predicate device material. The Talos® Lumbar (HA) PEEK IBF devices are substantially equivalent in safety and efficacy to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.