Talos Intervertebral Body Fusion Devices, Talos -C Cervical Intervertebral Body Fusion Devices, Talos -C(HA) Cervical Intervertebral Body Fusion Devices
Applicant
Meditech Spine, LLC
Product Code
ODP · Orthopedic
Decision Date
Oct 14, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K150788 · Oct 14, 2015
Talos Intervertebral Body Fusion Devices, Talos -C Cervical Intervertebral Body Fusion Devices, Talos -C(HA) Cervical Intervertebral Body Fusion Devices
Meditech Spine, LLC
Published retrospective clinical literature
The sponsor provided published retrospective clinical data to demonstrate that the use of allograft in interbody fusion procedures for degenerative disc disease does not pose new risks, supporting the expansion of the indications for use to include allograft.
Retrospective clinical data; Allograft; Intervertebral body fusion; Degenerative disc disease
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published retrospective clinical data for lumbar and cervical intervertebral body fusion devices; Retrospective clinical study (literature)
Patients with degenerative disc disease treated with intervertebral body fusion devices
Not applicable for this study
Safety and clinical outcomes of allograft use in interbody fusion
Indications for Use
Talos® IBF Device is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Talos® IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft. The Talos® IBF Device is to be used in patients who have had six months of non-operative treatment. Talos® IBF devices are to be implanted via a direct posterior, transforaminal, lateral, or anterior approach in the lumbosacral spine. The Talos®-L, Talos®-L, Talos®-P and Talos®-T are intended to be used with supplemental fixation. Talos®-C Cervical Intervertebral Body Fusion Device is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C Cervical IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft. Non-operative treatment prior to treatment with Talos®-C Cervical Intervertebral Body Fusion Devices is six (6) weeks. Talos®-C Cervical IBF Devices are to be implanted via an open anterior approach. Talos®-C Cervical IBF Devices are also to be used with supplemental fixation. Talos®-C (HA) Cervical Intervertebral Body Fusion Devices are intervertebral body devices intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C (HA) Cervical IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft. Non-operative treatment prior to treatment with Talos®-C (HA) Cervical Intervertebral Body Fusion Devices is six (6) weeks. Talos®-C (HA) Cervical IBF Devices are to be implanted via an open anterior approach. Talos®-C (HA) Cervical IBF Devices are also to be used with supplemental fixation.
Device Story
Talos® Intervertebral Body Fusion (IBF) devices are spinal implants designed to facilitate fusion in lumbar or cervical spine segments. Devices are manufactured from PEEK or HA-impregnated PEEK, featuring ridged teeth for stability and metallic markers (tantalum or titanium) for radiographic visualization. Implants are inserted by surgeons via various approaches (posterior, transforaminal, lateral, or anterior for lumbar; anterior for cervical) to replace degenerated discs. Devices are used in conjunction with autograft or allograft bone and require supplemental fixation. The device provides structural support to the intervertebral space, promoting bone fusion and alleviating symptoms of nerve root or spinal cord compression. Benefits include stabilization of the spinal segment and reduction of discogenic pain.
Clinical Evidence
No new clinical or mechanical testing was performed. Substantial equivalence is supported by published retrospective clinical data for similar lumbar and cervical intervertebral body fusion devices, demonstrating that the use of allograft in interbody fusion procedures poses no new risks.
Technological Characteristics
Materials: PEEK or HA-impregnated PEEK; titanium or tantalum markers. Design: Open configuration with ridged teeth for pullout resistance; rectangular or curved lateral walls; flat and lordotic angles. Sterilization: Not specified. Connectivity: None. Software: None.
Indications for Use
Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine (L2-S1, up to Grade 1 spondylolisthesis) or cervical spine (C2-T1, one level). Lumbar DDD defined as discogenic back pain; cervical DDD defined as intractable radiculopathy/myelopathy with nerve root/spinal cord compression. Requires 6 months (lumbar) or 6 weeks (cervical) of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Orthofix CONSTRUX Mini PEEK Spacer System, CONSTRUX Mini PEEK Ti Spacer System, Cervical Stand Alone Device (K142152)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 14, 2015
Meditech Spine, LLC Mr. Jason Owens Director of Engineering 1447 Peachtree Street Northeast, Suite 440 Atlanta, Georgia 30309
Re: K150788
Trade/Device Name: Talos® Intervertebral Body Fusion Devices, Talos®-C Cervical Intervertebral Body Fusion Devices, Talos®-C(HA) Cervical Intervertebral Body Fusion Devices Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP, MAX Dated: July 15, 2015 Received: July 20, 2015
Dear Mr. Owens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K150788
#### Device Name
Talos® Intervertebral Body Fusion Devices Talos®-C Cervical Intervertebral Body Fusion Devices Talos®-C (HA) Cervical Intervertebral Body Fusion Devices
#### Indications for Use (Describe)
Talos® Intervertebral Body Fusion Devices:
The Talos® IBF Device is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Talos® IBF Devices are intended to be used with autograft comprised of cancellous and/or corticocancellous bone graft.
The Talos® IBF Device is to be used in patients who have had six months of non-operative treatment. Talos® IBF devices are to be implanted via a direct posterior, transforaminal, lateral, or anterior approach in the lumbosacral spine. The Talos®-A, Talos®-L, Talos®-T are intended to be used with supplemental fixation.
Talos®-C Cervical Intervertebral Body Fusion Devices:
The Talos®-C Cervical Intervertebral Body Fusion Device is an intervertebral body device intended for use in sketally mature patients with Degenerative Disc (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radior myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C Cervical IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft.
Non-operative treatment prior to treatment with Talos®-C Cervical Intervertebral Body Fusion Devices is six (6) weeks. Talos®-C Cervical IBF Devices are to be implanted via an open anterior approach. Talos®-C Cervical IBF Devices are also to be used with supplemental fixation.
Talos®-C (HA) Cervical Intervertebral Body Fusion Devices:
Talos®-C (HA) Cervical Intervertebral Body Fusion Devices are intervertebral body devices intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radior myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C (HA) Cervical IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft.
Non-operative treatment prior to treatment with Talos@-C (HA) Cervical Intervertebral Body Fusion Devices is six (6) weeks.
Talos®-C (HA) Cervical IBF Devices are to be implanted via an open anterior approach. Talos®-C (HA) Cervical IBF Devices are also to be used with supplemental fixation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# meditech spine, IIa
1447 Peachtree St NE Suite 440, Atlanta GA 30309 | p (678) 974-5287 | f (404) 759-2104 | support@meditechspine.com
### 510(k) Summary
As required by section 807.92(c)
Meditech Spine, LLC is requesting clearance for the Indication for Use change to existing Talos® products. Summary prepared on October 9, 2015.
| A. Sponsor/Manufacturer: | Meditech Spine, LLC |
|--------------------------|---------------------------------|
| | Registration Number: 3009405289 |
| | Jason Owens, Operations Manager |
| | 1447 Peachtree St NE Suite 440 |
| | Atlanta, GA 30309 |
| | 678-974-5287 Phone |
| | 404-759-2104 Fax |
| | |
- Talos® Intervertebral Body Fusion Devices B. Trade Name: Talos®-C Cervical Intervertebral Body Fusion Devices Talos®-C (HA) Cervical Intervertebral Body Fusion Devices Common Name: Spinal Implant Classification Name: Intervertebral body fusion device (21 CFR 888.3080 Class II, Product Code ODP) Intervertebral body fusion device (21 CFR 888.3080 Class II, Product Code MAX)
- ﻥ Predicate Devices: K 142152: Orthofix CONSTRUX Mini PEEK Spacer System, CONSTRUX Mini PEEK Ti Spacer System, Cervical Stand Alone Device (Primary Predicate - Cervical) K 143163: Stryker AVS® Spacers and Cages (1st Additional -Lumbar) K 090707: Meditech Advisors, LLC Talos® Intervertebral Body Fusion Devices K 122850: Meditech Advisors, LLC Talos®-C Cervical Intervertebral Body Fusion Devices K 142345: Meditech Spine, LLC Talos®-C (HA) Cervical Intervertebral Body Fusion Devices
### D. Device Description:
The Talos® IBF Device is made of polymer, polyetheretherketone (PEEK). The Talos® IBF Device is available in four configurations: Talos®-P, Talos®-L, and Talos®-A. The devices are open devices with ridged teeth on superior and inferior ends to resist implant pullout. The Talos® –P and Talos® –L are rectangular devices and the Talos® – A have curved lateral walls and rounded edges. The implants are available in a range of sizes, as well as flat and lordotic angled implants to accommodate variations in patient's anatomy. In addition, tantalum markers at the opposite ends are offered which allow the Talos® IBF radiological confirmation for proper positioning.
{5}------------------------------------------------
## meditech spine, IIc
1447 Peachtree St NE Suite 440, Atlanta GA 30309 | p (678) 974-5287 | f (404) 759-2104 | support@meditechspine.com
The Talos®-C Cervical Intervertebral Body Devices (Talos®-C Cervical IBF Devices) are made of the polymer, polyetheretherketone (PEEK). The devices are open devices with ridged teeth on superior and inferior ends to resist implant pullout. The Talos®-C Cervical IBF Devices are rectangular devices and have curved lateral walls and rounded edges. The implants are available in a range of sizes as well as flat and lordotic angled implants to accommodate variations in patient's anatomy. In addition, titanium markers at the opposite ends are offered which allows the Talos®-C Cervical IBF Device radiological confirmation for proper positioning.
The Talos® -C (HA) Cervical Intervertebral Body Devices (Talos®-C (HA) Cervical IBF Devices) are made of the polymer, hydroxyapatite impregnated polyetheretherketone (HA PEEK). The devices are open devices with ridged teeth on superior and inferior ends to resist implant pullout. The Talos®-C (HA) Cervical IBF Devices are rectangular devices and have curved lateral walls and rounded edges. The implants are available in a range of sizes as well as flat and lordotic angled implants to accommodate variations in patient's anatomy. In addition, titanium markers at the opposite ends are offered which allows the Talos®-C (HA) Cervical IBF Device radiological confirmation for proper positioning.
- Intended Use: نن
#### Talos® Intervertebral Body Fusion Devices:
The Talos® IBF Device is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Talos® IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft.
The Talos® IBF Device is to be used in patients who have had six months of non-operative treatment.
Talos® IBF devices are to be implanted via a direct posterior, transforaminal, lateral, or anterior approach in the lumbosacral spine. The Talos®-L, Talos®-L, Talos®-P and Talos®-T are intended to be used with supplemental fixation.
#### Talos®-C Cervical Intervertebral Body Fusion Devices:
The Talos®-C Cervical Intervertebral Body Fusion Device is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C Cervical IBF Devices
{6}------------------------------------------------
## meditech spine, IIc
1447 Peachtree St NE Suite 440, Atlanta GA 30309 | p (678) 974-5287 | f (404) 759-2104 | support@meditechspine.com
are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft.
Non-operative treatment prior to treatment with Talos®-C Cervical Intervertebral Body Fusion Devices is six (6) weeks.
Talos®-C Cervical IBF Devices are to be implanted via an open anterior approach. Talos®-C Cervical IBF Devices are also to be used with supplemental fixation.
### Talos®-C (HA) Cervical Intervertebral Body Fusion Devices:
Talos®-C (HA) Cervical Intervertebral Body Fusion Devices are intervertebral body devices intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level from C2-T1. Cervical degenerative disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on the posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by history and radiographic studies. Talos®-C (HA) Cervical IBF Devices are intended to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft.
Non-operative treatment prior to treatment with Talos®-C (HA) Cervical Intervertebral Body Fusion Devices is six (6) weeks.
Talos®-C (HA) Cervical IBF Devices are to be implanted via an open anterior approach. Talos®-C (HA) Cervical IBF Devices are also to be used with supplemental fixation.
#### F. Technological Characteristics:
The technological characteristics of the Talos® Intervertebral Body Fusion Devices, Talos®-C Cervical Intervertebral Body Fusion Devices, and Talos®-C (HA) Cervical Intervertebral Body Fusion Devices are identical to the predicate devices in terms of intended use and design. The indications for use are similar. There is a marker material change in the Talos®-C Cervical Intervertebral Body Fusion Devices and in the Talos®-C (HA) Cervical Intervertebral Body Fusion Devices to titanium.
The addition of the allograft comprised of cancellous and/or corticocancellous bone graft to the Talos® Intervertebral Body Fusion Devices, Talos® -C Cervical Intervertebral Body Fusion Devices, and Talos®-C (HA) Cervical Intervertebral Body Fusion Devices does not increase risk. The same risks occur in the subject Talos® Intervertebral Body Fusion Devices, Talos®-C Cervical Intervertebral Body Fusion Devices, and Talos®-C (HA) Cervical Intervertebral Body Fusion Devices as in the predicate devices. Titanium is biocompatible material and commonly used. The cervical predicate is an intervertebral body fusion device that uses titanium markers.
- G. Non-clinical Testing:
{7}------------------------------------------------
There have been no design changes made to the Talos® Intervertebral Body Fusion Devices, Talos®-C Cervical Intervertebral Body Fusion Devices, and Talos®-C (HA) Cervical Intervertebral Body Fusion Devices. The purpose of this 510(k) submission is to obtain clearance to use allograft bone comprised of cancellous and/or corticocancellous bone in addition to autologous bone and to change the material of the markers in Talos®-C Cervical Intervertebral Body Fusion Devices and Talos®-C (HA) Cervical Intervertebral Body Fusion Devices. No mechanical testing was performed to determine substantial equivalence.
- H. Clinical Testing:
Published retrospective clinical data for lumbar and cervical intervertebral body fusion devices similar to the Talos® Intervertebral Body Fusion Devices, Talos®-C Cervical Intervertebral Body Fusion Devices, and Talos®-C (HA) Cervical Intervertebral Body Fusion Devices is provided in support of this application. The published clinical outcomes demonstrated that the use of allograft in interbody fusion procedures to treat patients with degenerative disc disease as defined above poses no new risks to patients. No design changes were made to the existing devices, nor were any new components added to the system. Therefore, no additional testing was required or performed.
- l. Conclusion:
The Talos® Intervertebral Body Fusion Devices, Talos®-C Cervical Intervertebral Body Fusion Devices, and Talos® C (HA) Cervical Intervertebral Body Fusion Devices have identical intended use, technological characteristics, design, and principles of operation as their predicate devices; as well as similar indications for use. The proposed devices will have Titanium markers instead of tantalum, which is the same as the predicate device.
There have been no changes to the Talos® Intervertebral Body Fusion Devices, Talos®-C Cervical Intervertebral Body Fusion Devices, and Talos®-C (HA) Cervical Intervertebral Body Fusion Devices since the last 510(k) clearance. The purpose of this 510(k) submission is to obtain clearance to use either autograft and/or allograft bone comprised of cancellous and/or corticocancellous bone graft as a substitute/addition to autologous bone and to change the markers from tantalum to titanium in the cervical implants.