K170307 · Sun Nuclear Corporation · IYE · Oct 25, 2017 · Radiology
Device Facts
Record ID
K170307
Device Name
SunCHECK
Applicant
Sun Nuclear Corporation
Product Code
IYE · Radiology
Decision Date
Oct 25, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
SunCHECK is a software platform intended to collect, detect, compare, calculate, analyze, display, and store radiotherapy quality assurance and dosimetry data.
Device Story
SunCHECK is a server-based web application platform for radiation therapy departments; integrates patient QA, machine QA, and data management workflows. Inputs include DICOM-RT beam intensity values, 3D volumetric imaging, and beam output data from EPID, kV, and Cone Beam CT imaging systems. Device performs independent dose calculations, compares results to Treatment Planning System (TPS) dose distributions, and monitors consistency between fraction deliveries. Operates on a centralized server accessible via networked PCs; provides a single GUI for data display and storage. Enables radiation therapy professionals to verify treatment delivery accuracy and manage departmental QA efforts. Benefits include centralized oversight of QA workflows and improved verification of radiation dose delivery.
Clinical Evidence
Bench testing only. No clinical data provided. Testing demonstrated that the device performs within design specifications and equivalently to predicate devices.
Technological Characteristics
Server-based web application; utilizes DICOM-RT data, EPID, kV, and Cone Beam CT imaging inputs. Hardware requirements: Dual Intel Xeon 2.4 GHz processors, 64 GB RAM, Dual NVidia Quadro M4000 GPUs, Windows Server 2012 R3. Connectivity via 1 GB/s Ethernet. Software modules include PerFRACTION (EPID-based secondary calculation), DoseCHECK (3D volumetric dose computation), and SNC Machine QA. Data storage and display functionality included.
Indications for Use
Indicated for use by radiation therapy professionals to perform quality assurance and dosimetry data management for radiotherapy treatment delivery devices and associated imaging systems.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
Sun Nuclear Corporation % Ms. Rene' Hardee Regulatory Affairs Specialist III 3275 Suntree Blvd. MELBOURNE FL 32940
October 25, 2017
### Re: K170307
Trade/Device Name: SunCHECK Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: October 13, 2017 Received: October 16, 2017
Dear Ms. Hardee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
### Page 2 - Ms. Rene' Hardee
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170307
Device Name SunCHECK
Indications for Use (Describe)
SunCHECK is a software platform intended to collect, detect, compare, calculate, analyze, display, and store radiotherapy quality assurance and dosimetry data.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Sun Nuclear Corporation. The logo consists of a blue square with a yellow circle inside, with horizontal lines across the circle. To the right of the square is the text "SUN NUCLEAR" in blue, with the word "corporation" underneath in a smaller font, also in blue. The logo is clean and professional, suggesting a company in the technology or energy sector.
# 510(k) Summary
Provided in accordance with 21 CFR 807.92 (c)
#### 1 General Provisions Date Prepared:
September 27, 2017
Submitted by:
Sun Nuclear Corporation 3275 Suntree Blvd. Melbourne, FL 32940 Ph: 321-259-6862 Fax: 321-259-7979 Web: www.sunnuclear.com
Contact Person:
Rene' Hardee ReneHardee@sunnuclear.com
Ph: 321-259-6862 extension 2304
Classification Name:
Accelerator, Linear, Medical
- Common Name:
Secondary check QA software
Proprietary Names:
Model 1299028 SunCHECK
Establishment Registration Number:
1038814
Classification:
Regulation Number: 21 CFR 892.5050 Name: Accelerator, Linear, Medical
Product code: IYE
Class II
{4}------------------------------------------------
Predicate Devices:
Primary Predicate:
| 1. | Model Name: | PerFRACTION |
|----|---------------|-----------------------------|
| 2. | Common Name: | Secondary check QA software |
| 3. | 510(k) # | K141800 |
| 4. | Manufacturer: | Sun Nuclear Corporation |
| 5. | Submitted: | July 03, 2014 |
To our knowledge, this predicate has not been subject to a design-related recall.
Reference Device:
| 2. | Model Name: | DoseCHECK |
|----|---------------|-----------------------------|
| | Common Name: | Secondary check QA software |
| | 510(k) # | K161946 |
| | Manufacturer: | Sun Nuclear Corporation |
| | Submitted: | July 15, 2016 |
To our knowledge, this predicate has not been subject to a design-related recall.
#### 2 Description and Use:
SunCHECK is a server-based Web application which is accessible from any networked PC. It is intended to provide radiation therapy professionals with a platform that integrates patient QA, machine QA and data management workflows. This platform consists of a single GUI and database that is intended to provide a centralized view of a radiation therapy department's QA efforts.
#### 3 Intended Use Statement:
SunCHECK is a software platform intended to collect, detect, compare, calculate, analyze, display, and store radiotherapy quality assurance and dosimetry data.
{5}------------------------------------------------
#### 4 Technological Characteristics
The primary technological characteristics of the Model 1299028 SunCHECK can be broken up into 4 modules of functionality:
| Primary Software Technological Characteristics |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Utilization of the beam output from imaging systems including electronic portal<br>imaging device (EPID), kV and Cone Beam CT that are part of the treatment<br>delivery device to perform standardized QA of the treatment delivery device and<br>associated imaging systems. |
| 2. The usage of three dimensional volumetric imaging information and beam intensity<br>values in DICOM-RT format to compute a dose volume (also in DICOM-RT format)<br>and a comparison of this independent calculation to the TPS dose distribution. |
| 3. Utilization of the beam output from an electronic portal imaging device (EPID) that is<br>part of the treatment delivery device to perform a consistency check between<br>fraction delivery. |
- 4. Data storage and display.
These technological characteristics are believed to be substantially equivalent to the predicate devices as seen in the below table.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Sun Nuclear Corporation. The logo consists of a blue square with a yellow sun inside on the left, and the words "SUN NUCLEAR" in blue on the right. Below the words "SUN NUCLEAR" is the word "corporation" in a smaller blue font.
### Your Most Valuable QA & Dosimetry Tools
| Areas of<br>Comparison | Subject Device:<br>Sun Nuclear Model 1299028<br>SunCHECK | Primary Predicate Device:<br>Sun Nuclear PerFRACTION<br>(K141800) | Reference Predicate Device:<br>Sun Nuclear DoseCHECK<br>(K161946) | Similarities and Differences |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology | The primary technological<br>characteristics of the Model<br>1299028 SunCHECK are:<br>• PerFRACTION Technology<br>• DoseCHECK Technology<br>• SNC Machine Technology<br>• Data Storage<br>• Data Display<br>Application Server:<br>• Dual Intel Xeon Processor 2.4<br>GHz (6 Cores each with<br>hyperthreading)<br>• 64 GB (8x8 GB) 2133 MHz DDR4<br>RDIMM<br>• Dual NVidia Quadro M4000 8GB<br>• OS: 200 GB minimum usable<br>free space<br>• Recommendation for Data<br>Drive: 5.76 TB minimum RAID 5e<br>• Minimum 1 GB/s Ethernet with<br>teaming considered<br>• Microsoft Windows Server 2012<br>R3<br>Client Machine:<br>• Google Chrome browser<br>(recommended) or Internet<br>Explorer 11.0 (depending upon the<br>computer's operating system)<br>• Pentium 4 Dual Core<br>• CPU Speed: 1.6 GHz<br>• Total RAM: 2 GB<br>• Display resolution: 1280 x 1024<br>• Color depth: 32-bit<br>• Minimum Windows 7 OS | The primary technological<br>characteristics of PerFRACTION<br>are:<br>• The usage of EPID data to<br>perform an independent<br>secondary calculation.<br>• Comparison of the calculation to<br>the TPS calculated data.<br>Application Server:<br>• Dual Intel Xeon Processor 2.4 GHz<br>(6 Cores each with hyperthreading)<br>• 64 GB (8x8 GB) 2133 MHz DDR4<br>RDIMM<br>• Dual NVidia Quadro M4000 8GB<br>• OS: 200 GB minimum usable<br>free space<br>• Recommendation for Data<br>Drive: 5.76 TB minimum RAID 5e<br>• Minimum 1 GB/s Ethernet with<br>teaming considered<br>• Microsoft Windows Server 2012<br>R3<br>Client Machine:<br>• Google Chrome browser<br>(recommended) or Internet<br>Explorer 11.0 (depending upon the<br>computer's operating system)<br>• Pentium 4 Dual Core<br>• CPU Speed: 1.6 GHz<br>• Total RAM: 2 GB<br>• Display resolution: 1280 x 1024<br>• Color depth: 32-bit<br>• Minimum Windows 7 OS | The primary technological<br>characteristics of DoseCHECK are:<br>• the usage of three dimensional<br>volumetric imaging information and<br>beam intensity values in DICOM-RT<br>format to compute a dose volume<br>(also in DICOM-RT format)<br>• to support electron point dose<br>calculations<br>Application Server:<br>• Dual Intel Xeon Processor 2.4 GHz (6<br>Cores each with hyperthreading)<br>• 64 GB (8x8 GB) 2133 MHz DDR4<br>RDIMM<br>• Dual NVidia Quadro M4000 8GB<br>• OS: 200 GB minimum usable<br>free space<br>• Recommendation for Data<br>Drive: 5.76 TB minimum RAID 5e<br>• Minimum 1 GB/s Ethernet with<br>teaming considered<br>• Microsoft Windows Server 2012 R3<br>Client Machine:<br>• Google Chrome browser<br>(recommended) or Internet Explorer<br>11.0 (depending upon the computer's<br>operating system).<br>• Pentium 4 Dual Core<br>• CPU Speed: 1.6 GHz<br>• Total RAM: 2 GB<br>• Display resolution: 1280 x 1024<br>• Color depth: 32-bit<br>• Minimum Windows 7 OS | Similarities<br>The SunCHECK Technology<br>encompasses both the<br>PerFRACTION and DoseCHECK<br>Technologies.<br>Differences<br>SunCHECK includes the<br>additional technology of Class I<br>listed SNC Machine and data<br>storage and display. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Sun Nuclear Corporation. The logo consists of a blue square with a yellow sun inside on the left, and the words "SUN NUCLEAR" in blue on the right. Below the words "SUN NUCLEAR" is the word "corporation" in a smaller blue font.
#### 5 Performance Data and Comparison with Predicate
Model 1299028 SunCHECK has been tested using appropriate bench testing methods. Test results of the modified device have demonstrated that the device performs within its design specifications and equivalently to the predicate devices.
#### 6 Summary
Model 1299028 SunCHECK is believed to be substantially equivalent to the predicate devices due to the similarities in function, technology, and performance. The intended use, performance testing, safety and effectiveness reviews demonstrate that Model 1299028 SunCHECK is as safe, as effective, and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.