K170149 · Medacta International S.A. · JWH · Aug 18, 2017 · Orthopedic
Device Facts
Record ID
K170149
Device Name
3DMetal Tibial Cones
Applicant
Medacta International S.A.
Product Code
JWH · Orthopedic
Decision Date
Aug 18, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 3DMetal Tibial Cones are indicated for use with the GMK Revision and GMK Hinge knee systems, as well as the GMK tibial extension stems and offsets. Specific indications are as follows: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis. - Post traumatic loss of joint configuration. - Considerable loss of function of the knee joint. - High-grade joint destruction requiring additional stabilization and reconstruction of bone defects. - Primary implantation failure. - Former revision arthroplasty.
Device Story
3DMetal Tibial Cones are implantable titanium alloy fillers used in revision knee arthroplasty; designed to address severe bone deficiency by providing stabilization and reconstruction of bone defects. Devices are cemented to tibial trays of GMK Revision and GMK Hinge systems. Available in centered/eccentric versions, sizes XS-L, and heights 20-25mm. Manufactured via electron beam melting (EBM) to create porous structure; full-density layer prevents cement occlusion of pores. Used by orthopedic surgeons in clinical settings during revision knee surgery. Output is a physical structural support for the tibial tray; aids in restoring joint configuration and function. Benefits include improved stability and bone defect management in complex revision cases.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including dynamic fatigue testing (ASTM F1800-12), shear testing (ASTM F1044-05), shear fatigue testing (ASTM F1160-14), tensile testing (ASTM F1147-05), and stereological evaluation (ASTM F1854-09). Biocompatibility is supported by adherence to ISO 5832-3:1996 and ASTM F2924-12a.
Technological Characteristics
Titanium alloy (ISO 5832-3:1996, ASTM F2924-12a) manufactured via electron beam melting (EBM). Porous structure with full-density layer. Sizes XS, S, M, L; heights 20mm, 25mm. Cemented fixation. Sterile (gamma). Single-use.
Indications for Use
Indicated for patients with severe joint pain/disability due to arthritis (traumatic, rheumatoid, polyarthritis), post-traumatic joint loss, significant knee function loss, high-grade joint destruction requiring bone defect reconstruction, primary implantation failure, or prior revision arthroplasty.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image is a logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized image of three human profiles facing to the right, with flowing lines that suggest movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 18, 2017
Medacta International SA % Elizabeth Rose Manager, Regulatory Affairs Mapi Usa. Inc 2343 Alexandria Drive Suite 100 Lexington, Kentucky 40504
Re: K170149 Trade/Device Name: 3DMetal Tibial Cones Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: July 3, 2017 Received: July 3, 2017
Dear Elizabeth Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170149
Device Name 3DMetal Tibial Cones
### Indications for Use (Describe)
The 3DMetal Tibial Cones are indicated for use with the GMK Hinge knee systems, as well as the GMK tibial extension stems and offsets.
Specific indications are as follows:
- · Severely painful and/or disabled joint as a result of arthritis, theumatoid arthritis or polyarthritis.
- · Post traumatic loss of joint configuration.
- · Considerable loss of function of the knee joint.
- · High-grade joint destruction requiring additional stabilization and reconstruction of bone defects.
- · Primary implantation failure.
- · Former revision arthroplasty.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------|--|--|
| | | |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 3.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs Manager Date Prepared: January 13, 2017 Date Revised: July 3, 2017
#### II. Device
| Device Proprietary Name: | 3DMetal Tibial Cones |
|--------------------------|-----------------------------------------------------------------------------------------------|
| Common or Usual Name: | Total Knee Prosthesis |
| Classification Name: | Knee joint, patellofemorotibial polymer/metal/polymer semi<br>constrained cemented prosthesis |
| Regulation Number: | 21 CFR 888.3560 |
| Product Code: | JWH |
| Device Classification | 2 |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices: Primary Predicate:
- Trabecular Metal Tibial Cone Augments, K102896, Zimmer Trabecular Metal o Technology
- Regenerex Porous Titanium Sleeve Augments, K072336, Biomet Manufacturing Corp. ●
Reference Predicate
- GMK Total Knee System (also referred to as GMK Primary), K090988, Medacta ● International SA
- GMK Total Knee System-Revision (also referred to as GMK Revision), K102437, ● Medacta International SA
- GMK Hinge, K130299, Medacta International SA ●
- Delta TT Acetabular System, K141395, Limacorporate S.p.A. ●
{5}------------------------------------------------
#### IV. Device Description
The 3DMetal Tibial Cones are implantable devices to be used as fillers in cases of severe bone deficiency. The subject devices are to be cemented to the GMK Revision and GMK Hinge systems' tibial trays. The subject devices have a full-density layer that prevents cement occluding the pores of the external layer.
The purpose of this submission is to gain clearance for the tibial cones available in centred and eccentric versions. The tibial cones are available in sizes XS, S, M, and L with heights of 20 mm and 25 mm.
The subject devices are intended to be used with the cleared indications for use and the tibial trays of Medacta's GMK Revision (K102437) and GMK Hinge (K130299) systems.
The 3DMetal Tibial Cones are manufactured with titanium alloy substrate, which is identical to predicate devices Regenerex Porous Titanium Sleeve Augments (K072336). However the subject devices' material is processed using electron beam melting (EBM) similar to the material of referenced device Delta TT Acetabular System (K141395).
#### V. Indications for Use
The 3DMetal Tibial Cones are indicated for use with the GMK Revision and GMK Hinge knee systems, as well as the GMK tibial extension stems and offsets.
Specific indications are as follows:
- o Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis.
- Post traumatic loss of joint configuration.
- Considerable loss of function of the knee joint. ●
- o High-grade joint destruction requiring additional stabilization and reconstruction of bone defects.
- o Primary implantation failure.
- o Former revision arthroplasty.
#### VI. Comparison of Technological Characteristics
The 3DMetal Tibial Cones and the predicate devices share the following characteristics:
- Indications for Use ●
- Materials
- Packaging
- Design ●
- Use with tibial trays ●
- Sterile
- Sizes ●
{6}------------------------------------------------
The 3DMetal Tibial Cones are technologically different from the predicate devices as follows:
- Heights available
- o Manufacturing Process
Biocompatibility testing conducted on the predicate and reference devices for the same material, supports the biological safety of the 3DMetal Tibial Cones. Additional biocompatibility testing was deemed unnecessary because the materials and manufacturing process are identical to the predicate devices described below and meet ISO 5832-3:1996 Implants For Surgery – Metallic Materials - Part 3: Wrought Titanium 6-Aluminum 4-Vanadium Alloy and ASTM F2924-12a Standard Specification For Additive Manufacturing Titanium-6 Vanadium With Powder Bed Fusion.
A comparison of the subject and predicate devices is provided in the table below.
| Parameters | 3DMetal Tibial<br>Cones<br>(Subject Device) | Trabecular Metal<br>Tibial Cone<br>Augments<br>K102896<br>(Predicate Device) | Regenerex Porous<br>Titanium Sleeve<br>Augments<br>K072336<br>(Predicate Device) |
|-----------------------------------------|------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Design/Types | Cemented to cone of<br>tibial tray; Centred<br>and Eccentric | Cemented to cone of<br>tibial tray; Full and<br>Stepped | Cemented to cone of<br>tibial tray |
| Material | Titanium Alloy | Tantalum | Titanium Alloy |
| Manufacturing<br>Process | Electron Beam<br>Melting | Additive<br>Manufacturing | Chemical Vapor<br>Deposition |
| Sizes | Sizes XS, S, M, L | Sizes XS, S, M, L | Sizes XS, S, M, L |
| Heights | 20 mm, 25 mm | 15 mm, 30 mm | 25mm, 40mm |
| Compatibility<br>with implant<br>system | GMK Revision and<br>GMK Hinge | NexGen Complete<br>Knee Solution,<br>Legacy LCCK and<br>Rotating Hinge Knee<br>Systems | Vanguard SSK<br>Revision Knee<br>System |
| Device usage | Single Use | Single Use | Single Use |
| Shelf Life | 5 years | 5 years | 5 years |
| Biocompatibility | Implant with<br>permanent >30 day<br>(Equivalency<br>determined) | Implant with<br>permanent >30 day | Implant with<br>permanent >30 day |
| Sterilization | Gamma | Gamma | Gamma |
| Packaging | Individual packaging | Individual packaging | Individual packaging |
Technological comparison
{7}------------------------------------------------
## Discussion
As seen above, the differences between the subject and predicate devices are that the subject devices have different heights than the predicate devices. Although the subject devices and predicate devices are made of titanium alloy, the manufacturing process method is different because the subject devices use electron beam melting to process the titanium. This technological difference does not raise new questions of safety or effectiveness and a comparison evaluation shows there are no new risks associated with the subject devices design.
#### VII. Performance Data
Based on the risk analysis and pre-submission submitted to FDA to review testing protocols, testing was conducted to written protocols with acceptance criteria that were based on standards and FDA guidance documents. The following mechanical tests are being provided in support of a substantial equivalence determination:
Non-Clinical Studies
- . Dynamic Fatigue Test: ASTM F1800-12 Standard Test Method For Cyclic Fatigue Testing Of Metal Tibial Tray Components Of Total Knee Joint Replacements
- Characterization Test
- o Shear Testing: ASTM F1044-05 (Reapproved 2011) Standard Test Method For Shear Testing of Calcium Phosphate Coatings And Metallic Coatings
- Shear Fatigue Testing: ASTM F1160-14 Standard Test Method For Shear and о Bending Fatigue Testing of Calcium Phosphate And Metallic Medical And Composite Calcium Phosphate/Metallic Coatings
- Tensile Testing: ASTM F1147-05 (Reapproved 2011) Standard Test Method For o Tension Testing of Calcium Phosphate And Metal Coatings
- Stereological Evaluation: ASTM F1854-09 Standard Test Method For O Stereological Evaluation Of Porous Coatings On Medical Implants
- Pyrogenicity ●
- o Medacta uses both the Bacterial Endotoxin Test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>) and the Pyrogen Test according to USP chapter <151> for pyrogenicity determination.
- Medacta has no intentions of labeling the subject devices as non-pyrogenic or o pyrogen free.
Clinical Studies
- . No clinical studies were conducted.
{8}------------------------------------------------
## VIII. Conclusion
The information provided above supports that the 3DMetal Tibial Cones are as safe and effective as the predicate devices. The subject devices are manufactured using the Electron Beam Melting (EBM) process in with titanium alloy, as compared to the predicate devices, does not raise any new questions of safety and effectiveness. Therefore, it is concluded that the 3DMetal Tibial Cones are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.