K130299 · Medacta International · KRO · Aug 15, 2013 · Orthopedic
Device Facts
Record ID
K130299
Device Name
GMK HINGE
Applicant
Medacta International
Product Code
KRO · Orthopedic
Decision Date
Aug 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3510
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GMK HINGE® knee prosthesis is designed for cemented use in total knee arthroplasty when the preoperative diagnosis of the joint determines that the bone and stability situation require the implantation of a constrained prosthesis. The GMK HINGE® knee system is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, . rheumatoid arthritis or polyarthritis associated with bone loss and/or severe joint instability - Considerable loss of function of the knee joint . - High-grade joint destruction requiring additional stabilization with stems and . reconstruction of bone defects with metal augmentation - Failure of a primary prosthesis (e.g. infection, loosening) . - Former revision arthroplasty . - Post traumatic loss of joint configuration - Avascular necrosis of femoral condyle Tibial augmentations are to be screwed to the tibial baseplate with both of the two provided fixing screws. When a GMK HINGE® implant is used it is mandatory to implant both the femoral and tibial components with an extension stem.
Device Story
GMK Hinge is a constrained, rotating total knee prosthesis for cemented implantation. System includes femoral component, tibial component, hinge post, hinge post extension, and UHMWPE bushing to prevent metal-on-metal contact. Tibial component features asymmetric design for bone coverage; tibial insert provides rotational control and varus/valgus constraint. Femoral and tibial augments allow restoration of joint line and filling of bone deficiencies; extension stems required for both components to ensure stable construct. Used by orthopedic surgeons in clinical settings for patients with poor bone quality or marked bone loss. Device restores joint function and stability through mechanical constraint and intramedullary fit and fill.
Clinical Evidence
Bench testing only. Mechanical performance evaluated against worst-case predicate devices per ISO 21534, ISO 7207-1, ASTM F1800, ASTM F2083-11, and ASTM F1223-08.
Technological Characteristics
Constrained, rotating knee prosthesis. Materials: CoCrMo (ISO 5832-4, ISO 5832-12), UHMWPE (ISO 5834-2 Type 1), Ti6Al4V (ISO 5832-3), M30NW (ISO 5832-9). Cemented fixation. Includes femoral/tibial components, hinge post, UHMWPE bushing, and optional augments. Requires extension stems.
Indications for Use
Indicated for patients requiring total knee arthroplasty with constrained prosthesis due to severe pain, joint disability, or destruction from arthritis (rheumatoid, traumatic, polyarthritis), avascular necrosis, post-traumatic loss of joint configuration, or failure of primary/revision arthroplasty associated with bone loss or severe instability.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Medacta International. The logo consists of two black triangles on the left, followed by the word "Medacta" in black lowercase letters. Below the word "Medacta" is a horizontal line, and below that is the word "International" in a smaller font, followed by a small cross symbol.
### 510(k) Summary
Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
AUG 1 5 2013
- Contact Person: Mr. Adam Gross Director of Regulatory and Quality Medacta USA 4725 Calle Quetzal, Unit B Camarillo, CA, 93012 Phone: (805)437-7085 Fax: (805)437-7553 Email: AGross@medacta.us.com
Date Prepared: February 5, 2013
DEVICE INFORMATION
Trade/Proprietary Name: GMK Hinge Common Name: Total Knee Prosthesis Classification Name: Prosthesis Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
> 21 CFR 888.3510 Class II Device Product Code: KRO
#### Section 5 - Page 2 of 4
{1}------------------------------------------------
| 510(k) | Product | 510(k) Holder | Clearance<br>Date |
|---------|-----------------------------|-------------------|-------------------|
| K003504 | RT-Plus | Plus Orthopedics | 5/11/2001 |
| K013385 | NexGen Rotating Hinge Knee | Zimmer | 1/9/2002 |
| K002552 | Modular Rotating Hinge Knee | Stryker/Howmedica | 11/13/2000 |
| K101815 | Enduro RHK | Aesculap | 12/20/2010 |
| K896048 | S-ROM/NOILES | Depuy | 03/14/1990 |
#### "Predicate Devices:
### Product Description
The GMK Hinge System is a hinged-knee system for total knee replacement. The GMK Hinge femoral component is designed to preserve bone stock. The hinge-post system is comprised of the hinge post and the hinge post extension and has been developed in order to reduce the impingement with the patellar tendon and provide mechanical resistance. There is a UHMWPE bushing between the femoral component and the hinge post in order to avoid contact between metal parts. The tibial component has an asymmetric design to provide coverage of the bone after tibia resection. The surface is mirror polished and the holes for the tibial augmentations are on the bottom. The UHMWPE tibial insert is fixed to the tibial component through a screw and is designed to provide high rotational control and a high varus/valgus constraint. The femoral and tibial augments allow the surgeon to selectively fill bone deficiencies and to aid in restoring the ioint line. The fixation between the augmentations and the femoral component or tibial trav is assured by a screw fixation. The GMK Hinge provides two solutions for the patella prosthesis: an inset patella and a resurfacing patella. The GMK Hinge System would be used most often in patients with poor bone quality or marked bone loss, and in these circumstances, extension stems are helpful to obtain a stable construct. An extension stem in combination with the right offset connector provides intramedullary femoral and tibial fit and fill.
#### Indications for Use
The GMK HINGE® knee prosthesis is designed for cemented use in total knee arthroplasty when the preoperative diagnosis of the joint determines that the bone and stability situation require the implantation of a constrained prosthesis.
The GMK HINGE® knee system is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, . rheumatoid arthritis or polyarthritis associated with bone loss and/or severe joint instability
- Considerable loss of function of the knee joint .
- High-grade joint destruction requiring additional stabilization with stems and . reconstruction of bone defects with metal augmentation
- Failure of a primary prosthesis (e.g. infection, loosening) .
- Former revision arthroplasty .
{2}------------------------------------------------
- · Post traumatic loss of joint configuration
- Avascular necrosis of femoral condyle
Tibial augmentations are to be screwed to the tibial baseplate with both of the two provided fixing screws.
When a GMK HINGE® implant is used it is mandatory to implant both the femoral and tibial components with an extension stem.
### Comparison to Predicate Devices
The indications for use, design features, and materials of the subject device are substantially equivalent to those of the predicate devices.
The design features of both the GMK Hinge and the predicates are a constrained, rotating knee prosthesis that have similar sizes and are intended to be used for cemented use only.
The GMK Hinge and the predicates use the same materials: CoCrMo (ISO 5832-4) UHMWPE (ISO 5834-2 Type 1) Ti6Al4V (ISO 5832-3) CoCrMo (ISO 5832-12) M30NW (ISO 5832-9)
The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the GMK Hinge are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
## Performance Testing
The GMK Hinge was compared to the worst case of the predicate devices. GMK Hinge was tested according to the following standards:
- . ISO 21534
- ISO 7207-1 .
- ASTM F1800 .
- ASTM F2083-11 .
- ASTM F1223-08 .
The comparison to predicate devices and the mechanical testing performed demonstrate that the GMK Hinge does not introduce any issues in regards to safety and effectiveness.
#### Conclusion:
Based on the above information, the GMK Hinge can be considered as substantially equivalent to its predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure embracing a globe, symbolizing the department's mission to protect and promote the health and well-being of the nation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 15, 2013
Medacta International SA % Medacta United States of America Mr. Adam Gross Director of Regulatory and Quality 4725 Calle Quetzal, Unit B Camarillo, California 93012
Re: K130299
Trade/Device Name: GMK Hinge Regulation Number: 21 CFR 888.3510 Regulation Name: Knee joint femorotibial metal/polymer constrained cemented prosthesis Regulatory Class: Class II Product Code: KRO Dated: June 20, 2013 Received: June 21, 2013
Dear Mr. Gross:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Adam Gross
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin Keith
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
K130299 510(k) Number (if known):
Device Name: GMK Hinge
Indications for Use:
The GMK HINGE® knee prosthesis is designed for cemented use in total knee arthroplasty when the preoperative diagnosis of the joint determines that the bone and stability situation require the implantation of a constrained prosthesis.
The GMK HINGE® knee system is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, . rheumatoid arthritis or polyarthritis associated with bone loss and/or severe joint instability
- Considerable loss of function of the knee joint .
- High-arade joint destruction requiring additional stabilization with stems and . reconstruction of bone defects with metal augmentation
- Failure of a primary prosthesis (e.g. infection, loosening) .
- Former revision arthroplasty .
- Post traumatic loss of joint configuration .
- Avascular necrosis of femoral condyle .
Tibial augmentations are to be screwed to the tibial baseplate with both of the two provided fixing screws.
When a GMK HINGE® implant is used it is mandatory to implant both the femoral and tibial components with an extension stem.
Prescription Use X (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
Division of Orthopedic Devices
GMK Hinge 510(k) February 5, 2013
Section 4 - Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.