The All-in-One Health Monitor PC-303 is a device designed for spot-checking measuring of the patient's physiological parameters, such as Non-Invasive Blood Pressure (NIBP), Oxygen saturation (SpO2), Pulse Rate (PR) and Body Temperature (TEMP): Additionally, the device is available to communicate with the compatible Blood Glucose Monitoring System and ECG monitor to make the measurement. This device is applicable for Adult and Pediatic (age>3 years old) use in clinical institutions and has no conditions or factors of contraindication.
Device Story
Mobile modular patient monitor; measures SpO2/PR via photoelectric sensors, NIBP via oscillometric method, and body temperature via infrared probe. Optional external connectivity via USB to compatible blood glucose meters and ECG monitors. Used in clinical institutions by healthcare providers. Device receives digital signals from internal modules and external linked devices; processes data for display on 4.3" segmented LCD and storage of up to 999 records. Provides real-time parameter monitoring; assists clinicians in spot-checking patient vitals; aids in clinical decision-making through immediate physiological data acquisition.
Clinical Evidence
Bench testing only. Performance testing validated device operation within specified parameters, including wireless/USB data transfer and compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-30, and ISO 80601-2-61. Biocompatibility testing (ISO 10993-5, ISO 10993-10) confirmed no cytotoxicity, irritation, or sensitization. NIBP accuracy verified against standard mean deviation requirements.
Technological Characteristics
Modular patient monitor; 4.3" segmented LCD; powered by AC or 3.7VDC rechargeable lithium battery. Sensing: photoelectric (SpO2), oscillometric (NIBP), infrared (TEMP). Connectivity: USB interface for external glucose/ECG modules. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 80601-2-61 (SpO2), IEC 60601-2-30 (NIBP), ISO 10993 (biocompatibility).
Indications for Use
Indicated for spot-checking physiological parameters (NIBP, SpO2, PR, TEMP) in adult and pediatric patients (age ≥3 years) in clinical institutions. Optional connectivity for blood glucose and ECG monitoring.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2017
Shenzhen Creative Industry Co., Ltd. % Mr. Charles Mack Principal Engineer IRC, LLC 7808 Rush Creek Drive Pasco. Washington 99301
Re: K170047
Trade/Device Name: All-in-One Health Monitor, PC-303 Regulation Number: 21 CFR 870.2300 Regulation Name: Physiological Patient Monitor (Without Arrhythmia Detection Or Alarms) Regulatory Class: Class II Product Code: MWI, DQA, DXN, FLL, NBW, DSH Dated: July 22, 2017 Received: August 1, 2017
Dear Mr. Charles Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Wilhelm
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170047
Device Name All-in-One Health Monitor, PC-303
#### Indications for Use (Describe)
The All-in-One Health Monitor PC-303 is a device designed for spot-checking measuring of the patient's physiological parameters, such as Non-Invasive Blood Pressure (NIBP), Oxygen saturation (SpO2), Pulse Rate (PR) and Body Temperature (TEMP):
Additionally, the device is available to communicate with the compatible Blood Glucose Monitoring System and ECG monitor to make the measurement.
This device is applicable for Adult and Pediatic (age>3 years old) use in clinical institutions and has no conditions or factors of contraindication.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Date: April14, 2017
#### 1. Company and Correspondent making the submission:
Name - Shenzhen Creative Industry Co., Ltd. Address - 2/F, Block 3, Nanyou Tian'an Industry Town Shenzhen, China 518054 Tel: +86-755-26431671 Fax: +86-755-26435433 Contact - Mrs. Jia Wang General Manager US Agent and Correspondent Mr. Charles Mack PRINCIPAL ENGINEER IRC USA 7808 Rush Creek Drive, Pasco, Wa. 99301 USA Tel: 931-6254938 Email: charliemack@irc-us.com
#### 2. Device :
| Trade/proprietary name: | All-in-One Health Monitor, PC-303 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Patient Monitor |
| Classification Name: | Monitor, Physiological, Patient (without arrhythmia<br>detection or alarms) |
| Classification ProCode: | MWI |
| Device Class: | II |
| Regulation Number: | 870.2300 |
| Other ProCodes: | DQA, Oximeter<br>DXN, System Mearurement, Blood Pressure, non-invasive<br>FLL, Thermometer, electronic, clinical<br>NBW, System, test, blood glucose OTC<br>DSH, Recorder, magnetic tape for ECG |
{4}------------------------------------------------
### 3. Predicate Devices :
| Manufacturer | PredicateDevice | 510(k)<br>Number | SubmittedDevice |
|----------------------------------------------|----------------------------|------------------|--------------------------------------|
| Shenzhen<br>Creative<br>Industry Co.,<br>Ltd | UP-7000 Patient<br>Monitor | K123711 | All-in-One Health<br>Monitor, PC-303 |
### 4. Classifications Names & Citations :
Primary: MWI, monitor,physiological,patient(without arrhythmia detection or alarms)
Secondary: DQA, SpO2 Pulse Rate; DXN, NIBP; FLL, Temperature; NBW, Blood Glucose; DSH, ECG
#### 5. Device Description:
#### Device Description:
The All-in-One Health Monitor, PC-303 is mobile modular designed patient monitor. It monitors the patient's blood oxygen saturation (SpO2) and pulse rate (PR) non-invasively by the photoelectric method. It can also measure non-invasive blood pressure (NIBP, the pressures of systolic, diastolic and mean) by the oscillating method and body temperature (TEMP) by the infrared radiation energy technology.
Additionally, the device is also capable to link and communicate with the compatible Blood Glucose Meter and ECG monitor to make the measurement.
This device system measure the patient's physical parameters with variety modules. It takes digital signals from NIBP, SpO2, Temperature module through COM ports, and receives ECG data, blood glucose data from compatible legally marketed ECG Monitor or BLOOD GLUCOSE MONITORING Device. The software to process the data, then display parameters on the screen and storage some useful information.
This device is applicable for use in clinical institutions and has no conditions or factors of contraindication.
#### Device Features:
This monitoring system may be used to monitor up to 10 of a patient's physical parameters such as Pulse Oxygen Saturation (SpO2), Pulse Rate; Non-invasive blood pressure (NIBP); Body temperature (TEMP); ECG, heart rate and BLOOD GLUCOSE:
- Large numeric display, segmented LCD panel, real-time clock display. Blood pressure measurements can be activated or canceled by one shortcut button.
- Oximetry technique displays SpO2 & pulse rate measurements by smart sensors.
{5}------------------------------------------------
- Smart infrared temperature probe presents measurements of body temperature.
- o Blood pressure, oxygen saturation, pulse rate and temperature can be measured simultaneously.
- Blood Glucose meter option can be connected to the device.
- The Easy ECG Monitor option can be connected to the device.
- Data storage with recall, up to 999 groups of records can be stored and recognized by patient ID.
- . Power management with power saving mode, auto power off and low battery indicator:
| | Function | Product<br>Code | Module 510K Status |
|--------------------------------------|--------------------|-----------------|-----------------------------------------------------|
| Built-in Function: | | | |
| Built-in | SpO2<br>Pulse Rate | DQA | Creative SpO2 sensor only (K063641) |
| Built-in | NIBP | DXN | Creative B01-WK126 NIBP Module |
| Built-in | TEMP | FLL | Valeo VT-601E IR Ear thermometer (K111463) |
| External Linked Function (Optional): | | | |
| External Linked Function | Blood<br>Glucose | NBW | G-777G BLOOD GLUCOSE MONITORING SYSTEM<br>(K113077) |
| External Linked Function | ECG | DSH | PC-80 Easy ECG Monitor (K073152) |
Below is a brief tabular display of the monitor functions and operations.
### 6. Indication for use:
The All-in-One Health Monitor, PC-303 is a device designed for spot-checking measuring of the patient's physiological parameters, such as Non-Invasive Blood Pressure (NIBP), Oxygen saturation (SpO2), Pulse Rate (PR) and Body Temperature (TEMP);
Additionally, the device is available to communicate with the compatible Blood Glucose Monitoring System and ECG monitor to make the measurement.
This device is applicable for Adult and Pediatric (age≥3 years old) use in clinical institutions and has no conditions or factors of contraindication.
### 7. Comparison with predicate device:
Shenzhen Creative Industry Co., Ltd. believes that the All-in-One Health Monitor, PC-303 is substantially equivalent to the (K123711) Patient Monitor (Shenzhen Creative Industry Co., Ltd.).
{6}------------------------------------------------
| Table of Comparison to Predicate Device | | |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Specifications<br>Characteristics | | |
| Name and model | Subject Device<br>All-in-One Health Monitor, PC-303 | Predicate Device<br>UP-7000 Patient Monitor |
| 510K Applicant | Shenzhen Creative Industry Co., LTD. | Shenzhen Creative Industry Co., LTD. |
| 510(K) Number | N/A | K123711 |
| Physical dimension(mm)<br>/weight(kg) | 165(L) X 96(W) X 68(H)/ 0.44kg | 300 (L) x150 (W) x290 (H)/ 5.2kg |
| Display | 4.3" Segment LCD | 12.1" Color TFT |
| Type, Degree of<br>protection against<br>electric shock | Class II with internal electric power supply.<br>SpO2/NIBP/TEMP: Type BF applied part. | Class I with internal electric power supply.<br>Where the integrity of the external protective<br>earth(ground) in the installation or its conductors is in<br>doubt, the equipment shall be operated from its internal<br>electric power supply(batteries)<br>CO2: BF (defibrillation proof)<br>SpO2/NIBP /TEMP/ECG/RESP: CF (defibrillation proof) |
| Power supply | Battery or AC | Battery or AC |
| Power requirement | (100-240) VAC, 50/60Hz, 15VA;<br>Rechargeable lithium battery, 3.7VDC | (100-240) VAC 50/60Hz 100VA;<br>Rechargeable sealed lead-acid battery, 12V 2.3AH |
| Battery charging<br>indicator | Available | Available |
| Low battery indicator | Available | Available |
| Battery charge time, typ. | Maximum 8 hours for charging | Maximum 13-15 hours for charging |
| Flammable anesthetics | N/A | N/A |
| Operating condition | Temperature: 5°C ~ 40°C<br>Humidity: 30% ~ 80%, noncondensing<br>Atmospheric pressure: 70.0kPa~106.0kPa | Temperature: 5°C ~ 40°C<br>Humidity: 30% ~ 80%, noncondensing<br>Atmospheric pressure: 70.0kPa~106.0kPa |
| Storage condition | Temperature: -20°C~60°C<br>Humidity: 10%~95%, noncondensing<br>Atmosphere: 53kPa~106kPa | Temperature: -20°C~60°C<br>Humidity: 10%~95%, noncondensing<br>Atmosphere: 53kPa~106kPa |
| EMC | The equipment meets the requirements of IEC 60601-1-2 | The equipment meets the requirements of IEC 60601-1-2 |
| Optional printer | N/A | Available |
| Characteristics | Subject Device | Predicate Device |
| Name and model | All-in-One Health Monitor, PC-303 | UP-7000 Patient Monitor |
| Manufacturer | Shenzhen Creative Industry Co., LTD. | Shenzhen Creative Industry Co., Ltd. |
| 510(K) Number: | N/A | K123711 |
| SpO2 module | Identical | Creative SpO2 module with the same principle as PC-<br>60(K063641) Creative SpO2 sensor only |
| Patient type | Identical | Adult, pediatric patients |
| SpO2 measurement<br>accuracy | Adult and Pediatric:<br>$\pm$ 3% (during 70%-100%)<br>Undefined (during 0-70%) | Adult and Pediatric:<br>$\pm$ 3% (during 70%-100%)<br>Undefined (during 0-70%) |
| Alarm range (%) | No Alarm | High Limit: 1%-100%<br>Lower Limit. 0%-99% |
| Pulse rate measurement<br>range | 30 bpm-240 bpm | 30 bpm-240 bpm |
| Pulse rate accuracy | $\pm$ 2bpm or $\pm$ 2% (whichever is greater) | $\pm$ 2bpm or $\pm$ 2% (whichever is greater) |
| Alarm range-Pulse rate<br>(bpm) | No Alarm | High Limit: 1bpm-300bpm<br>Lower Limit. 0bpm-299bpm |
{7}------------------------------------------------
# Build-in Function:
#### SpO₂ / Pulse Rate
#### NIBP
| Characteristics | Subject Device | Predicate Device |
|-------------------------------------------|-------------------------------------------------------------------------|----------------------------------------------------------|
| Name and model | All-in-One Health Monitor, PC-303 | UP-7000 Patient Monitor |
| Manufacturer | Shenzhen Creative Industry Co., LTD. | Shenzhen Creative Industry Co., Ltd. |
| 510(K) Number: | N/A | K123711 |
| NIBP module | Creative B01-WK126 NIBP Module | SunTech Advantage 2.0 NIBP Module |
| Method | Oscillometric method | Oscillometric method |
| Patient type | Adult and Pediatric patients | Adult and Pediatric patients |
| Unit of measure | mmHg & kPa | mmHg & kPa |
| Pressure measurement<br>range - Systolic | 60 mmHg - 255mmHg | Adult: 40 mmHg - 260mmHg<br>Pediatric: 40 mmHg - 160mmHg |
| Pressure measurement<br>range - Diastolic | 30 mmHg - 195mmHg | Adult: 20 mmHg - 200mmHg<br>Pediatric: 20 mmHg - 120mmHg |
| BP accuracy | Mean deviation values: ±5 mmHg; | Mean deviation values: ±5 mmHg; |
| | Standard deviation <= 8 mmHg. | Standard deviation <= 8 mmHg. |
| Cuff pressure range | 0 to 300mmHg | 0 to 300mmHg |
| Over pressure protector | Cuff pressure exceeds 300mmHg (Adult & Pediatric<br>modes) at any time. | Internal operating software ensures that: -- |
| | | Maximum cuff inflation time is limited to 75<br>Seconds |
{8}------------------------------------------------
| | | | | Duration of blood pressure reading is limited to<br>• 130 seconds (Adult mode), 90 seconds (Pediatric<br>mode)<br>Additional redundant safety circuitry oversees normal<br>operation and will override to abort a reading if:<br>• Cuff pressure exceeds 300 mmHg (Adult &<br>Pediatric modes) at any time.<br>• The cuff has been inflated for 180 seconds (Adult<br>& Pediatric modes). | |
|----------------|-----------|----------------|----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Alarm<br>setup | Systolic | High<br>Limit | No Alarm | Adult: 31 mmHg -280 mmHg<br>Pediatric: 30 mmHg -200 mmHg | |
| | | Lower<br>Limit | No Alarm | Adult: 29 mmHg -279 mmHg<br>Pediatric: 29 mmHg -199 mmHg | |
| | Diastolic | High<br>Limit | No Alarm | Adult: 11 mmHg -232 mmHg<br>Pediatric: 11 mmHg -150 mmHg | |
| | | Lower<br>Limit | No Alarm | Adult: 10 mmHg -231 mmHg<br>Pediatric: 10 mmHg -149 mmHg | |
| Alarm method | | | No Alarm | Sound light alarm and record the alarm status for review | |
{9}------------------------------------------------
#### TEMP
| Characteristics | Subject Device | Predicate Device |
|-----------------------------------|------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Name and model | All-in-One Health Monitor, PC-303 | UP-7000 Patient Monitor |
| Manufacturer | Shenzhen Creative Industry Co., Ltd. | Shenzhen Creative Industry Co., Ltd. |
| 510(K) Number: | N/A | K123711 |
| TEMP Module | Valeo VT-601E IR Ear thermometer cleared under K111463 | TEMP module cleared under K982140 |
| Fundamental scientific technology | Infrared technology | Thermo-resistor type |
| Patient type | Adult, Pediatric | Adult, Pediatric |
| Unit of measure | °C or °F | °C or °F |
| Measurement site | Ear | Body surface |
| Temperature measurement range | 32.0°C to 43.0°C (90°F to 109.5°F) | 32.0°C to 43.9°C (90°F to 109.9°F) |
| Temperature measurement accuracy | ±0.2°C (36.0°C to 39.0°C),<br>±0.3°C (the rest);<br>±0.4°F (96.8°F to 102.2°F),<br>±0.5°F (the rest) | ±0.1°C (35.0°C to 39.0°C)<br>±0.2°F (95.0°F to 102.2°F) |
{10}------------------------------------------------
External Linked Functions (Optional):
The subject device is also capable to link and communicate with the legally marketed compatible Blood Glucose Meter and EGG monitor to take measurements.
Blood Glucose (K113077)
| Name and model: | G-777G BLOOD GLUCOSE MONITORING SYSTEM | |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Manufacturer | BIOLAND TECHNOLOGY LTD. | |
| 510(K) Number: | K113077 | |
| Physical dimension/ weight | 81(L) X 62(W) X 19(H)/ 60g | |
| Test sites: | Fingertip, forearm | |
| Power requirement | CR2032 3V lithium coin battery x1 | |
| Patient type | Adult, pediatric patients | |
| Glucose Operating Range | (1.1-33.3) mmol/L or (20-600) mg/dL | |
| Glucose Sample Types | Whole Blood | |
| Glucose Sample volume | 0.7µl | |
| Glucose Hematocrit Compensation | N/A | |
| Glucose Assay Time | <10s | |
| Calibration | N/A | |
| Data Upload | USB | |
| Compatible Test Strips | N/A | |
| Operating condition | Temperature: 10℃ - 40℃<br>Humidity: 20%-80%RH | |
| Connection Method with the subject<br>device | Connection Method:<br>Using the optional link cable for the Blood Glucose Meter, connect the Blood Glucose Meter to the<br>connector on the right side of the All-in-One Health Monitor marked "GLU" ("if"). When the device shows<br>', this indicates that the blood glucose meter is linked successfully.<br>5-<br>Principle: The PC-303 All-in-One Health Monitor receives data from Blood Glucose Meter via the USB data<br>interfaces for data communication, and the PC-303 looks over measurement result stored in Glucose<br>Meter in real time. | |
| | | |
{11}------------------------------------------------
ECG (K073152)
| Name and model: | PC-80 Easy ECG Monitor |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Shenzhen Creative Industry Co., Ltd. |
| 510(K) Number: | K073152 |
| Physical dimension/ weight | 115(L) X 58(W) X 18(H)/100g |
| Type, Degree of protection against<br>electric shock | Internal power source, Type B |
| Power requirement | Two 1.5Vdc Alkaline battery (AAA) |
| Patient type | Adult, pediatric patients |
| Number of electrodes employed | 3 embedded metal electrodes or using 3 adhesive ECG electrodes by connection to the lead wire |
| Heart rate measuring range | 30bpm-240bpm |
| Resolution | 1bpm |
| Heart rate measuring precision | ±2bpm or ±2%, whichever is greater |
| Sweep speed | 20mm/s±10% |
| Signal bandwidth | 0.5Hz-40Hz |
| Internal noise level | ≤30uVp-p |
| CMRR | ≥60dB |
| Display sensitivity | 4.8mm/mV±10% |
| Operating condition | Temperature: 5°C-40°C<br>Humidity: ≤80%, noncondensing<br>Atmospheric pressure: 86.0kPa-106.0kPa |
| Storage condition | Temperature: -20°C-55°C<br>Humidity: ≤95%, noncondensing<br>Atmosphere: 50kPa-106kPa |
| EMC | The equipment meets the requirements of IEC 60601-1-2 |
| Connection method with the subject<br>device | Connection Method:<br>1. Connect the ECG accessory to the connector on the upper side of device marked "ECG".<br>2. Choose one of the methods to take the ECG measurement.<br>3. When the ECG accessory is connected to All-in-One Health Monitor successfully, press the "Start"<br>button on the ECG accessory to activate the ECG measurement.<br>4. When "ECG" appears on the display screen of the All-in-One Health Monitor, it means the ECG<br>accessory has begun to take the ECG measurement.<br>5. 30 Seconds later, the result will display on the screen of the terminal device, and the measurement will<br>terminate.<br>Principle: The PC-303 All-in-One Health Monitor receives data from Easy ECG Monitor via the USB data<br>interfaces for data communication, and the PC-303 looks over measurement result stored in Easy ECG<br>Monitor in real time. |
{12}------------------------------------------------
# 8. Safety and Performance Data:
| Serial<br>Number/Version | Standard and Description | Scope |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| IEC 60601-1:<br>1990+A1+A2+A13 | Medical devices Part1: General requirements for safety<br>and Amendment 1, Amendment 2 and Amendment 13 | General<br>requirements:<br>Electrical Safety |
| IEC 60601-1-2:<br>2007 | Medical electrical equipment-<br>part 1-2: General requirements for basic safety and<br>essential performance – Collateral standards:<br>Electromagnetic compatibility –Requirements and tests | General<br>requirements:<br>EMC |
| IEC 60601-2-30:<br>2000 | Medical electrical equipment - Part 2-30: Particular<br>requirements for the basic safety and essential<br>performance of automated non-invasive<br>sphygmomanometers | General<br>requirements:<br>Software |
| IEC80601-2-<br>30_NIBP | Medical electrical equipment - Part 2-30: Particular<br>requirements for the basic safety and essential<br>performance of automated non-invasive<br>sphygmomanometers | General<br>requirements:<br>Usability |
| ISO80601-2-<br>61_SpO2 | Medical electrical equipment - Part 2-61: Particular<br>requirements for basic safety and essential<br>performance of pulse oximeter equipment | General<br>requirements:<br>Alarm system |
| IEC/TR60878:<br>2003 | Graphical symbols for electrical equipment in medical<br>practice | General<br>requirements:<br>Graphical symbols |
| ISO 10993-1:2009 | Biological evaluation of medical devices--Part 1:<br>Evaluation and testing | Biological<br>evaluation |
| ISO 10993-5:2009 | Biological Evaluation of Medical Devices - Part 5 Tests<br>for In Vitro Cytotoxicity. | General<br>requirements:<br>Biological<br>Estimation |
| ISO 10993-<br>10:2010 | Biological Evaluation of Medical Devices – Part 10:<br>Tests for Irritation and Delayed-Type Hypersensitivity | General<br>requirements:<br>Biological<br>Estimation |
| ISO 13485: 2003 | Medical devices-Quality management systems-<br>Requirements for regulatory purposes | Quality system |
All of the test results noted in the table above were satisfactory.
{13}------------------------------------------------
| Serial Number | Standard and Description | Scope |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| ECG K073152 | | |
| IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) | General requirements:<br>Electrical Safety |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests | General requirements:<br>EMC |
| AAMI EC 38 | Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems | Basic safety and essential performance of ambulatory electrocardiographic (ECG) systems |
| Temp-K111463 | | |
| ASTM<br>E1965:1998(2009) | Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature | General Thermometer Performance |
| IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) | General requirements:<br>Electrical Safety |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests | General requirements:<br>EMC |
{14}------------------------------------------------
| Serial Number | Standard and Description | Scope |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| SPO2 K063641 | | |
| IEC60601-1:1988<br>A1:1991,+<br>A2:1995 | General requirements for Safety and Essential Performance | General<br>requirements:<br>Electrical Safety |
| IEC 60601-1-4 | Medical Devices Part 1-4: General Requirements for Safety -<br>Programmable Medical Electrical Equipment | General<br>requirements<br>Software |
| IEC 60601-1-<br>1:2000 | Medical electrical equipment -- Part 1-1: General requirements for<br>safety - Collateral standard: Safety requirements for medical electrical<br>systems | General<br>requirements:<br>Electrical Safety |
| IEC60601-1-<br>2:2007 | Medical Electrical Equipment-Part 1-2:General Requirements for Safety<br>- 2.Collateral Standard-Electromagnetic compatibility - Requirements<br>and tests | General<br>requirements:<br>EMC |
{15}------------------------------------------------
| Serial Number | Standard and Description | Scope |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| IEC 61010-1: 2010 | General requirements for Safety and Essential Performance | General<br>requirements:<br>Electrical Safety |
| IEC 61010-2-101 | Safety Requirements for Electrical Equipment for Measurement, Control<br>and Laboratory Use - Part 2-101: Particular Requirements for In Vitro<br>Diagnostic (IVD) Medical Equipment | IVD General Safety<br>Requirements |
| IEC 61326-1:<br>2012 | Electrical equipment for measurement, control and laboratory use - EMC<br>requirements - Part 1: General requirements | General<br>requirements:<br>EMC |
| IEC 61326-2-6 | Electrical equipment for measurement, control and laboratory use. EMC<br>requirements. Particular requirements. In vitro diagnostic (IVD) medical<br>equipment | General<br>requirements:<br>EMC |
| ISO 15197 | In vitro diagnostic test systems -- Requirements for blood-glucose<br>monitoring systems for self-testing in managing diabetes mellitus | IVD Glucose<br>performance and<br>Safety |
| CLSI GP42-A6:<br>2008 | Procedures and devices for the collection of diagnostic capillary blood<br>specimens | General technique<br>for capillary blood<br>specimens |
| IEEE ISO 11073-<br>10417 | Health informatics -- Personal health device communication Part 10417:<br>Device Specialization -- Glucose Meter | Communication<br>between personal<br>telehealth glucose<br>meter devices and<br>compute engines |
| ISO 10993-1:<br>2009 | Biological evaluation of medical devices -- Part 1: Evaluation and testing<br>within a risk management process | Biological<br>Evaluation |
{16}------------------------------------------------
### 9. Testing Summary:
IEC 60601-1: 1990+A1+A2+A13: Testing was performed to validate the safety of this medica…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.