The purpose and function of the Mediana ARGUS VCM vital signs monitor is to monitor noninvasive blood pressure (systolic, diastolic, and mean arterial pressures), functional arterial oxygen saturation, pulse rate for adult, pediatric and neonate patients and temperature for adult and pediatric patients in all hospital areas and hospital-type facilities. It may be used during hospital transport and in mobile, land-based environments, such as ambulances, within the specification of the environmental characteristics. Note: Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters. Note: Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
Device Story
Compact, lightweight vital signs monitor (2.7 kg) for hospital and mobile environments; operated by medically trained personnel. Inputs: NIBP (via cuff), SpO2 (via pulse oximetry sensor), pulse rate (derived from NIBP or SpO2), and temperature (via thermometry probe). Device integrates third-party modules: Welch Allyn POEM for NIBP, Nellcor MP506 for SpO2, and Welch Allyn SureTemp for temperature. Data processed and displayed on 7-segment and LED displays. Provides real-time patient monitoring; assists clinicians in assessing patient status during care or transport. Benefits include continuous, non-invasive tracking of vital parameters to support clinical decision-making.
Clinical Evidence
Bench testing only. System verification and validation performed per plan #PA300-09000. Risk analysis and mitigation documented per design control procedures. Quality system certified to 21CFR820, ISO 9001, and ISO 13485.
Technological Characteristics
Dimensions: 180x278x130 mm; Weight: 2.7 kg. Power: AC mains (100-240VAC) or internal battery. Modules: Welch Allyn POEM (NIBP), Nellcor MP506 (SpO2), Welch Allyn SureTemp (thermometry). Display: 7-segment and LED. Connectivity: Standalone monitor.
Indications for Use
Indicated for monitoring non-invasive blood pressure, functional arterial oxygen saturation, and pulse rate in adult, pediatric, and neonate patients, and temperature in adult and pediatric patients. Used in hospital settings, hospital-type facilities, and during land-based transport by medically trained personnel.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Welch Allyn Protocol Inc. Vital Signs Monitors, Model 53000 Series (K031740)
Welch Allyn Protocol Inc. Vital signs Monitors, Model Propaq LT 802 Series (K033378)
Nellcor Puritan Bennett Pulse Oximeter, Model N-550 (K021090)
Welch Allyn Inc. Vital Signs Monitor (VSM) (K024005)
Submission Summary (Full Text)
{1}------------------------------------------------
# JUL 2 8 2005
# 510(k) Summary of Safety and Effectiveness
#### Submitter
Mediana Co.,Ltd. 231-6 Nonhyun-Dong, Kangnam-Ku, Seoul, Korea 135-010 Tel) +822-542-3375 Fax) +822-542-1621 Company Contact: Gina Y.J.LEE, Specialist of Regulatory Affairs Date Summary Prepared: March 2nd, 2005
#### Device Name
| Trade Name: | ARGUS VCM series, Vital Signs Monitor |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Vital signs measurement devices |
| Classification Name: | NIBP measurement system (21CFR870.1130), also contains<br>non-invasive pulse oximetry, SpO2 (21CFR870.2700) and<br>clinical electronic thermometer (21CFR870.2910) |
| Classification: | Class II |
#### Predicate Devices (Legally Marketed Devices)
The predicate devices for Vital signs monitors, Model ARGUS VCM series are:
- Welch Allyn Protocol Inc. Vital Signs Monitors, Model 53000 Series cleared by FDA through 510(k) No. K031740, and
- Welch Allyn Protocol Inc. Vital signs Monitors, Model Propaq LT 802 Series cleared by FDA through 510(k) No. K033378, and
- Nellcor Puritan Bennett (division of Tyco Healthcare Inc.) Pulse Oximeter, Model N-550 cleared by FDA through 510(k) No. K021090, and
- Welch Allyn Inc. Vital Signs Monitors, Welch Allyn Vital Signs Monitor (VSM), cleared by FDA through 510(k) No. K024005.
#### Device Description
The ARGUS VCM vital signs monitor is to monitor non-invasive blood pressure, pulse rate, non-invasive functional oxygen saturation of arterial hemoglobin (SpO2) for adult, pediatric and neonate patients and body temperature for adult and pediatric patients in general hospital and alternate care facilities by medically trained personnel. This monitor is available for sale only upon the order of a physician or licensed health care professional.
The Mediana ARGUS VCM vital signs monitor is a lightweight and compact device (180×278×130 mm and 2.7 kg) powered by AC mains (100-240VAC, 50-60Hz) and also powered by internal battery. The monitor provides patient data and monitoring status on 7segment and LED displays.
#### Intended Use
The purpose and function of the Mediana ARGUS VCM vital signs monitor is to monitor noninvasive blood pressure (systolic, diastolic, and mean arterial pressures), functional arterial oxygen saturation, pulse rate for adult, pediatric and neonate patients and temperature for adult and pediatric patients in all hospital areas and hospital-type facilities. It may be used during hospital transport and in mobile, land-based environments, such as ambulances, within the specification of the environmental characteristics.
{2}------------------------------------------------
- Note: Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters.
- Note: Intra-hospital transport includes transport of a patient within the hospital or hospitaltype facility.
### Summary of Technical Characteristics of the Device Compared to the Predicate Devices (Legally Marketed Devices)
The Mediana vital signs monitors, Model ARGUS VCM Series are substantially equivalent to the Welch Allyn Protocol vital signs monitors, Model 53000 Series and Model Propaq LT 802 Series, Nellcor Puritan Bennett (division of Tyco Healthcare Inc.) Pulse Oximeter Model N-550, and Welch Allyn Inc. Vital Signs Monitor, Welch Allyn Vital signs Monitor (VSM).
- The Non-Invasive Blood Pressure (NIBP) measurement specifications and performance are equivalent to the Welch Allyn Protocol vital signs monitors, Model 53000 Series and Model Propaq LT 802 Series. Welch Allyn NIBP module POEM is the same NIBP module used in the Mediana vital signs monitors, Model ARGUS VCM Series.
- The Pulse rate specifications and performance derived from either Non-Invasive Blood Pressure (NIBP) or Pulse Oximetry (SpO2) are equivalent to the Welch Allyn Protocol vital signs monitors, Model 53000 Series, and the Nellcor Puritan Bennett Pulse Oximeter Model N-550.
- The Pulse Oximetry (SpO2) specifications and performance are equivalent to the Nellcor Puritan Bennett Pulse Oximeter Model N-550. Nellcor Oximetry module MP506 is the same SpO2 module used in the Mediana vital signs monitors, Model ARGUS VCM Series.
- The Temperature specifications and performance are equivalent to the Welch Allyn Protocol vital signs monitors, Model 53000 Series and the Vital Signs Monitor (VSM). Welch Allyn thermometry module SureTemp® is the same temperature module used in the Mediana vital signs monitors, Model ARGUS VCM Series.
#### Summary of Performance Testing
The Mediana vital signs monitors, Model ARGUS VCM Series substantially have been tested in accordance with the system V & V plan (#PA300-09000) and summary (#PA300-09013) included with the submission using production equivalent units prior to release to market.
A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Mediana design control procedure. Mediana's quality system confirms to 21CFR820, ISO 9001, ISO13485 and CMDCAS ISO 13485 certified by DNV (Det Norske Veritas) and NSAI (National Standards Authority of Irelands).
#### Conclusions
As stated above, the Mediana vital signs monitors, Model ARGUS VCM Series are safe and effective and comply with the appropriate medical device standards and are substantially equivalent to the earlier identified predicate devices.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, U.S.A." arranged around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mediana Co., LTD. c/o Mr.Charlie Mack International Regulatory Consultants 231-6 Nonhyun-Dong Kangnam-Ku, Seoul, Kerea 135-110
#### Re: K051375
Trade Name: Argus VCM series, vital signs monitors Regulation Number: 21 CFR 870.1130, 21 CFR 870.2700, 21 CFR 880.2910 Regulation Name: System, Measurement, Blood Pressure, Non-Invasive Regulatory Class: Class II (two) Product Code: DXN Dated: May 24, 2005 Received: May 26, 2005
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
JUL 2 8 2005
{4}------------------------------------------------
Page 2 - Mr.Charlie Mack
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0295. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bfimmermon for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K051375
Device Name: Arqus VCM Series, Vital Signs Monitor
Indications For Use: The purpose and function of the Mediana ARGUS VCM vital signs monitor is to monitor non-invasive blood pressure (systolic, diastolic, and mean arterial pressures), functional arterial oxygen saturation, pulse rate for adult, pediatric and neonate patients and temperature for adult and pediatric patients in all hospital areas and hospital-type facilities. It may be used during hospital transport and in mobile, land-based environments, such as ambulances, within the specification of the environmental characteristics.
- Note: Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters.
- Note: Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Hammonds
ivision Sign-Off) Division of Cardlovascular Devices KOSIS IS 510(k) Number__
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.