K163649 · Philips Medical Systems · DSJ · Mar 31, 2017 · Cardiovascular
Device Facts
Record ID
K163649
Device Name
SureSigns VS3; SureSigns VS4
Applicant
Philips Medical Systems
Product Code
DSJ · Cardiovascular
Decision Date
Mar 31, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1100
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility. The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
Device Story
SureSigns VS3/VS4 are multi-parameter patient monitors used by healthcare professionals in clinical settings for real-time monitoring, recording, and alarming of physiological data. Inputs include NBP, SpO2, Temperature, CO2, RRa, and SpHb. The devices process these signals to provide continuous patient status updates. The primary modification is the integration of 'QuickAlert' functionality, which calculates and displays configurable Modified Early Warning Scores (MEWS) on-device. This output assists clinicians in identifying patient deterioration, facilitating timely clinical decision-making. The monitors are intended for use in healthcare facilities and during intra-facility transport.
Clinical Evidence
Bench testing only. Verification and validation activities, including system-level performance tests and risk assessment-based testing, were conducted. Results demonstrated that the modified VS3 and VS4 monitors met all performance claims and passed all specified verification tests.
Technological Characteristics
Multi-parameter patient monitors. Features include NBP, SpO2, Temperature, CO2, RRa, and SpHb measurement modules. Connectivity includes on-device display and alarming. Software includes the QuickAlert feature for calculating Modified Early Warning Scores (MEWS). No changes to measurement specifications or fundamental sensing principles compared to predicate devices.
Indications for Use
Indicated for health care professionals to monitor, record, and alarm physiological parameters in healthcare environments and during intra-facility transport. VS3 supports NBP, SpO2, and Temperature for adult, pediatric, and neonatal patients. VS4 supports NBP, SpO2, Temperature, CO2, RRa, and SpHb for adult, pediatric, and neonatal patients (with specific exclusions for RRa/SpHb in neonates). Contraindicated for transport outside the healthcare facility.
Regulatory Classification
Identification
A blood pressure alarm is a device that accepts the signal from a blood pressure transducer amplifier, processes the signal, and emits an alarm when the blood pressure falls outside a pre-set upper or lower limit.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 19, 2017
Philips Medical Systems Mr. Greg Li Regulatory Affairs Specialist 3000 Minuteman Rd. Andover, Massachusetts 01810
Re: K163649
Trade/Device Name: SureSigns VS3 Vital Signs Monitor (reference numbers: 863071, 863072, 863073, 863074) SureSigns VS4 Vital Signs Monitor (reference number: 863283)
Regulation Number: 21 CFR 870.1100 Regulation Name: Blood Pressure Alarm Regulatory Class: Class II Product Codes: DSJ, DSK, DXN, DOA, DSA, FLL and CCK (for VS4 only) Dated: December 16, 2016 Received: December 23, 2016
Dear Mr. Greg Li:
This letter corrects our substantially equivalent letter of March 31, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mude Jellm
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163649
#### Device Name
SureSigns VS3 Vital Signs Monitor (reference numbers: 863071, 863073, 863074) SureSigns VS4 Vital Signs Monitor (reference number: 863283)
#### Indications for Use (Describe)
Indications for Use for SureSigns VS3 Vital Signs Monitor
The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
Parameter (patient types): NBP (Adult, Pediatric, Neonatal); SpO2 (Adult, Pediatric, Neonatal); Temperature (Adult, Pediatric, Neonatal)
Contraindications: Not for transport outside the healthcare facility
Indications for Use for SureSigns VS4 Vital Signs Monitor
The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
Parameter (patient types): NBP (Adult, Pediatric, Neonatal); SpO2 (Adult, Pediatric, Neonatal); Temperature (Adult, Pediatric. Neonatal); CO2 (Adult, Pediatric. Neonatal); RRa (Adult, Pediatric); SpHb (Adult, Pediatric)
Contraindications: Not for transport outside the healthcare facility
Type of Use (Select one or both, as applicable)
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510K Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).
- 1. The submitter and contact person of this pre-market notification is:
Greg Li Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 United States
Tel: 978-659-4227 Fax: 978-659-7323 Email: Greg.li@philips.com
- 2. Date this summary is prepared: This summary is prepared on 16 December, 2016.
- The names of the subject devices are as following: 3. SureSigns VS3 Vital Signs Monitor SureSigns VS4 Vital Signs Monitor
- 4. The trade names of the devices are SureSigns VS3 and SureSigns VS4.
- 5. The common usual name for both the VS3 and the VS4 is multi-parameter patient monitor
| Device Panel | Classification | ProCode | Description | Applicable<br>subject devices |
|-----------------------------------------|----------------|---------|-------------------------------------------------------|-------------------------------|
| Cardiovascular | 870.1100, II | DSJ | Alarm, Blood Pressure | VS3, VS4 |
| | 870.1110, II | DSK | Computer, Blood Pressure | VS3, VS4 |
| | 870.1130, II | DXN | System, Measurement, Blood<br>Pressure, Non-Invasive | VS3, VS4 |
| | 870.2700, II | DQA | Oximeter | VS3, VS4 |
| | 870.2900, II | DSA | Cable, transducer and<br>electrode, patient connector | VS3, VS4 |
| General Hospital and<br>Personal Use | 880.2910, II | FLL | Thermometer, Electronic,<br>Clinical | VS3, VS4 |
| Anesthesiology &<br>Respiratory Therapy | 868.1400, II | CCK | Analyzer, Gas | VS4 |
- 6. The Classification names are as follows:
{4}------------------------------------------------
- 7. The modified devices are substantially equivalent to previously cleared devices from Philips:
SureSigns VS3 Vital Signs Monitor cleared under k151761 SureSigns VS4 Vital Signs Monitor cleared under k151761
- 8. The modifications are as following:
The subiect VS3 and VS4 will be modified with the current QuickAlert to include On-device multi-parameter configurable Modified Early Warning Scores (MEWS).
- 9. The subject device VS3 has identical Intended Use and Indications for Use as the legally marketed predicate device, currently marketed VS3:
### Indications for Use for SureSigns VS3 Vital Signs Monitor
The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
| | Patient Types | | |
|-------------|---------------|-----------|----------|
| Parameter | Adult | Pediatric | Neonatal |
| NBP | √ | √ | √ |
| SpO2 | √ | √ | √ |
| Temperature | √ | √ | √ |
Contraindications: Not for transport outside the healthcare facility
The subject device VS4 has identical Intended Use and Indications for Use as the legally marketed predicate device, currently marketed VS4:
{5}------------------------------------------------
## Indications for Use for SureSigns VS4 Vital Signs Monitor
The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
| | Patient Types | | |
|-------------|---------------|-----------|----------|
| Parameter | Adult | Pediatric | Neonatal |
| NBP | ✓ | ✓ | ✓ |
| SpO2 | ✓ | ✓ | ✓ |
| Temperature | ✓ | ✓ | ✓ |
| CO2 | ✓ | ✓ | ✓ |
| RRa | ✓ | ✓ | |
| SpHb | ✓ | ✓ | |
Contraindications: Not for transport outside the healthcare facility
- 10. The subject devices VS3 and VS4 have the same fundamental technological characteristics as the legally marketed predicate devices VS3 and VS4. There is no addition, deletion or change to the current measurements or alarming, no specification change to current SureSigns VS3 or SureSigns VS4 monitor.
- 11. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject device. Testing involved system level tests, performance tests and tests according risk assessment. Pass/Fail criteria were based on the specifications cleared for the subject device and test results showed substantial equivalence. The modified VS3 and VS4 monitors passed all specified verification tests. The results demonstrate that the Philips SureSigns VS3 Vital Signs monitor and the Philips SureSigns VS4 Vital Signs monitor meet all performance claims and support a determination of substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.