K151761 · Phillips Medical Systems · DSJ · Sep 29, 2015 · Cardiovascular
Device Facts
Record ID
K151761
Device Name
SureSigns VS3, SureSigns VS4
Applicant
Phillips Medical Systems
Product Code
DSJ · Cardiovascular
Decision Date
Sep 29, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1100
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility. The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
Device Story
SureSigns VS3/VS4 are multi-parameter patient monitors used by healthcare professionals in clinical environments and transport. Devices collect physiological data (NBP, SpO2, temperature, CO2, RRa, SpHb) via sensors; process signals to provide real-time monitoring, recording, and alarming. Modifications include connectivity to IntelliVue GuardianSoftware (IGS) for documentation, charting, and decision support; IGS provides early warning scores (MEWS), single-parameter scores, and alert messages to the monitor. Additional features include RFID patient ID entry, expanded record panes, and temporal temperature measurement (VS3). Output is displayed on the monitor for clinician review to support clinical decision-making and patient management. Benefits include enhanced workflow integration and standardized early warning scoring.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, performance tests, and safety testing based on hazard analysis. Results confirmed that modified devices meet all safety, reliability, and performance requirements, supporting substantial equivalence.
Technological Characteristics
Multi-parameter monitors; connectivity via RS232 (thermometer) and network (IGS). Software version A.06 compliant with IEC 62304. Temporal temperature measurement uses Exergen TemporalScanner (ISO 80601-2-56). Features RFID for patient ID. Connectivity to IGS for decision support.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring monitoring of NBP, SpO2, and temperature (VS3/VS4), plus CO2, RRa, and SpHb (VS4 only) in healthcare environments and during transport.
Regulatory Classification
Identification
A blood pressure alarm is a device that accepts the signal from a blood pressure transducer amplifier, processes the signal, and emits an alarm when the blood pressure falls outside a pre-set upper or lower limit.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 29, 2015
Phillips Medical Systems Greg Li Regulatory Affairs Specialist 3000 Minuteman Rd. Andover, Massachusetts 01810
Re: K151761
Trade/Device Name: SureSigns VS3, SureSigns VS4 Regulation Number: 21 CFR 870.1100 Regulation Name: Blood Pressure Alarm Regulatory Class: Class II Product Code: DSJ, DSK, FLL, CCK, DQA, DSA, DXN Dated: August 21, 2015 Received: August 28, 2015
Dear Greg Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510 (k) Number (if known):
### Device Name:
SureSigns VS3 (reference numbers: 863071, 863072, 863073, 863074) SureSigns VS4 (reference number: 863283)
### Indications for Use for SureSigns VS3 Vital Signs Monitor
The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
| | Patient Types | | |
|-------------|---------------|-----------|----------|
| Parameter | Adult | Pediatric | Neonatal |
| NBP | √ | √ | √ |
| SpO2 | √ | √ | √ |
| Temperature | √ | √ | √ |
### Indications for Use for SureSigns VS4 Vital Signs Monitor
The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
| | Patient Types | | |
|-------------|---------------|-----------|----------|
| Parameter | Adult | Pediatric | Neonatal |
| NBP | √ | √ | √ |
| SpO2 | √ | √ | √ |
| Temperature | √ | √ | √ |
| CO2 | √ | √ | √ |
| RRa | √ | √ | |
| SpHb | √ | √ | |
Prescription Use: YES NO (Part 21 CFFR 801 Subpart D) (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
Page ____ of _________________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Philips SureSigns VS3 and VS4 Special 510K
010 Indications for Use Statement
{3}------------------------------------------------
# 011_510(k) Summary
The following section contains the 510(k) Summary as defined by the Medical Devices Act of 1990 and 21 CFR 807-92(c).
{4}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).
- 1. The submitter and contact person of this pre-market notification is:
Greg Li Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 United States
Tel: 978-659-4227 Fax: 978-659-7323 Email: Greg.li@philips.com
- 2. Date this summary is prepared: This summary is prepared on 10 June, 2015.
- The names of the subject devices are as following: 3. SureSigns VS3 SureSigns VS4
- 4. The trade names of the devices are SureSigns VS3 and SureSigns VS4.
- 5. The common usual name for both the VS3 and the VS4 is multi-parameter patient monitor
- 6. The Classification names are as follows:
| Device Panel | Classification | ProCode | Description | Applicable subject devices |
|--------------------------------------|----------------|---------|----------------------------------------------------|----------------------------|
| Cardiovascular | 870.1100, II | DSJ | Alarm, Blood Pressure | VS3, VS4 |
| | 870.1110, II | DSK | Computer, Blood Pressure | VS3, VS4 |
| | 870.1130, II | DXN | System, Measurement, Blood Pressure, Non-Invasive | VS3, VS4 |
| | 870.2700, II | DQA | Oximeter | VS3, VS4 |
| | 870.2900, II | DSA | Cable, transducer and electrode, patient connector | VS3, VS4 |
| General Hospital and Personal Use | 880.2910, II | FLL | Thermometer, Electronic, Clinical | VS3, VS4 |
| Anesthesiology & Respiratory Therapy | 868.1400, II | CCK | Analyzer, Gas | VS4 |
{5}------------------------------------------------
- 7. The modified devices are substantially equivalent to previously cleared devices from Philips:
SureSigns VS3 Vital Signs Monitor cleared under K133961 SureSigns VS4 Vital Signs Monitor cleared under K133961
- 8. The modifications are as follows:
- a) The subject VS3 and VS4 will add connectivity to IntelliVue GuardianSoftware (IGS). This IGS connectivity includes new QuickCheck work flows, QuickAlert prompts, and configurable Modified Early Warning Scores (MEWS).
IGS is a documentation, charting, and decision support software that can be configured to collect vital signs data from the VS3. VS4 monitors. The monitors validate the patient ID from IGS, send records to IGS, and receive early warning scores, single-parameter scores, and alert messages from IGS. The IGS connectivity to VS3 and VS4 is a change of software only.
The IntelliVue GuardianSoftware (IGS) is a software medical device owned by Philips Medizine Systeme Boeblingen GmbH, in Germany. The IGS was FDA cleared through 510(k) K111905 for Rev. A and 510(k) K122223 for Rev. B. The IGS has been revised to Rev. C which will allow IGS and VS3, VS4 to be connected. Philips Medizine Systeme Boeblingen GmbH will submit its own 510(k) for the Rev. C of IGS.
The VS3 and VS4 software is verified by Philips Medical Systems following out QMS process. The modified software for VS3 and VS4 is version A.06 and it complies with IEC 62304: 2006- Medical Device Software life-cycle processes.
- b) Add Temporal Temperature measurement to VS3. This is achieved by using the OEM Exergen TemporalScanner Thermometer. This Exergen TemporalScanner Thermometer is FDA cleared under K011291. The OEM thermometer has a RS232 cable added to connect the thermometer to the VS3 monitor. This same Temporal Temperature measurement was implemented on VS4 which was 510(k) cleared in June 2014 through k133961. Both VS3 and VS4 use the same solution for this measurement. and both complies with ISO 80601-2-56: 2009 + CORR. 2 (2007). Medical electrical equipment - Part 2-56: Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement.
- c) Other enhancements to VS3 and VS4 are described as following:
- . Additional Title Bar Indicators
- . Expanded Patients Records Pane
- Display Patient Name or Primary ID
{6}------------------------------------------------
- Expanded QuickCapture Entries ●
- . Default CO2 Setting Alignment (VS4 only)
- Temperature High and Low Alarms Color Change .
- Increased SpO2 Alarm Delay Defaults
- . Option to Prevent Inadvertently Clearing NBP Programs
- RFID patient ID entry.
- 9. The subject device VS3 has the same Intended Use and Indications for Use as the legally marketed predicate device, currently marketed VS3:
### Indications for Use for SureSigns VS3 Vital Signs Monitor
The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
| | Patient Types | | |
|-------------|---------------|-----------|----------|
| Parameter | Adult | Pediatric | Neonatal |
| NBP | | | |
| SpO2 | | | |
| Temperature | | | |
The subject device VS4 has the same Intended Use and Indications for Use as the legally marketed predicate device, currently marketed VS4:
### Indications for Use for SureSigns VS4 Vital Signs Monitor
The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
| | Patient Types | | |
|-------------|---------------|-----------|----------|
| Parameter | Adult | Pediatric | Neonatal |
| NBP | √ | √ | √ |
| SpO2 | √ | √ | √ |
| Temperature | √ | √ | √ |
| CO2 | √ | √ | √ |
| RRa | √ | √ | |
| SpHb | √ | √ | |
{7}------------------------------------------------
- 10. The subject devices VS3 and VS4 have the same fundamental technological characteristics as the leqally marketed predicate devices VS3 and VS4. There are no specification change to VS4. The added Temporal Temperature measurement to VS3 will have the same specification with that from the predicate VS4. The subject devices use the same algorithms for the measurements as the predicate devices.
- 11. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject device. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the subject device and test results showed substantial equivalence. The results demonstrate that the Philips SureSigns VS3 Vital Signs monitor and the Philips SureSigns VS4 Vital Signs monitor meet all safety and reliability requirements, performance claims and support a determination of substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.