K163522 · Maine Molecular Quality Controls, Inc. · PMN · Mar 14, 2017 · Microbiology
Device Facts
Record ID
K163522
Device Name
FilmArray Warrior Control Panel M290
Applicant
Maine Molecular Quality Controls, Inc.
Product Code
PMN · Microbiology
Decision Date
Mar 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3920
Device Class
Class 2
Indications for Use
FilmArray Warrior Control Panel M290 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bacillus anthracis (Chromosomal element, pXO1, pXO2), Coxiella burnetii, Francisella tularensis, Yersinia pestis, Ebola virus and Marburg virus on the FilmArray NGDS Warrior Panel performed on FilmArray 2.0 systems. FilmArray Warrior Control Panel M290 is composed of synthetic DNA specifically designed for and intended to be used solely with the FilmArray NGDS Warrior Panel assay. This product is not intended to replace manufacturer internal FilmArray NGDS Warrior Panel controls provided with the device.
Device Story
FilmArray Warrior Control Panel M290 provides external quality control for FilmArray NGDS Warrior Panel assays. Positive control contains synthetic DNA segments of biothreat agents (B. anthracis, C. burnetii, F. tularensis, Y. pestis, Ebola, Marburg) in buffer/stabilizers; negative control contains non-specific DNA. Used in clinical laboratories; processed by technicians like patient blood samples on FilmArray 2.0 instruments. Monitors qualitative amplification, detection, and identification steps. Results are compared against expected qualitative outcomes (Detected/Not Detected) to verify assay performance. Does not replace internal manufacturer controls. Benefits include ensuring test reliability for high-consequence pathogens.
Clinical Evidence
Bench testing only. Performance evaluated across 6 clinical sites using 3 lots of controls, 12 pouch lots, and multiple operators over 6 months. Total of 310 external controls tested (163 positive, 147 negative). Positive control percent correct was 98.8% (161/163); negative control percent correct was 100% (147/147). Internal studies (precision, reproducibility, lot-to-lot) confirmed consistent performance. Four invalid results occurred due to internal pouch control failures, which were excluded from final analysis.
Technological Characteristics
Ready-to-use liquid control panel. Positive control: synthetic DNA in buffers, stabilizers, preservatives. Negative control: non-specific DNA in buffers, stabilizers, preservatives. Designed for FilmArray 2.0 system. No specific material standards (e.g., ASTM) cited. Standalone control material; no connectivity or software components.
Indications for Use
Indicated for use as external positive and negative assayed quality control materials for laboratory personnel performing in vitro nucleic acid testing for biothreat pathogens (B. anthracis, C. burnetii, F. tularensis, Y. pestis, Ebola, Marburg) using the FilmArray NGDS Warrior Panel on FilmArray 2.0 systems. For prescription use only.
Regulatory Classification
Identification
An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.
Special Controls
An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K163522
B. Purpose for Submission:
To obtain a substantial equivalence determination for the FilmArray Warrior Control Panel M290.
C. Measurand:
Multi-analyte quality control materials
D. Type of Test:
FilmArray Warrior Control Panel M290 is intended for *in vitro* diagnostic use as external assayed quality control materials to monitor the qualitative amplification, detection and identification steps of the laboratory nucleic acid test, FilmArray NGDS Warrior Panel assay on the FilmArray 2.0 instrument, which detects biothreat pathogens: *Bacillus anthracis*, *Coxiella burnetii*, *Francisella tularensis*, *Yersinia pestis*, Ebola Virus and Marburg Virus.
E. Applicant:
Maine Molecular Quality Controls, Inc. (MMQCI)
F. Proprietary and Established Names:
FilmArray Warrior Control Panel M290
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3920, Assayed quality control material for clinical microbiology assays
2. Classification:
Class II (Special Controls)
3. Product code:
PMN
{1}
4. Panel:
83- Microbiology
H. Indication(s) for use:
1. Indications for use(s):
FilmArray Warrior Control Panel M290 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bacillus anthracis (Chromosomal element, pXO1, pXO2), Coxiella burnetii, Francisella tularensis, Yersinia pestis, Ebola virus and Marburg virus on the FilmArray NGDS Warrior Panel performed on FilmArray 2.0 systems. FilmArray Warrior Control Panel M290 is composed of synthetic DNA specifically designed for and intended to be used solely with the FilmArray NGDS Warrior Panel assay. This product is not intended to replace manufacturer internal FilmArray NGDS Warrior Panel controls provided with the device.
2. Special conditions for use statement(s):
For in vitro diagnostic use only
For prescription use only
3. Special instrument requirements:
FilmArray Warrior Control Panel M290 was evaluated on FilmArray 2.0 instrument.
I. Device Description:
FilmArray Warrior Control Panel M290 is a quality control panel consisting of two Controls: FilmArray Warrior Positive Control and FilmArray Warrior Negative Control. The Positive Control contains non-infectious surrogate control material; a solution of synthetic DNA in buffers, stabilizers and preservatives. The DNA carries segments of all biothreat agents detected by the FilmArray NGDS Warrior Panel assay (see Table below) on FilmArray 2.0 systems. The Negative Control is comprised of non-specific DNA in buffers, stabilizers and preservatives. Each liquid control of FilmArray Warrior Control Panel M290 is processed separately according to FilmArray NGDS Warrior Panel assay manufacturer's Instructions for Use for Blood Samples.
The FilmArray Warrior Control Panel M290 is prepared nucleic acid concentrations of 5X-10X LoD for each of the organisms detected by the FilmArray NGDS Warrior Panel assay.
2
{2}
| Biothreat pathogens detected by FilmArray NGDS Warrior Panel |
| --- |
| Bacillus anthracis: Chromosomal element, pXO1, pXO2 |
| Coxiella burnetii |
| Francisella tularensis |
| Yersinia pestis |
| Ebola Virus |
| Marburg Virus |
## J. Substantial Equivalence Information:
1. Predicate device name(s):
FilmArray RP EZ Control Panel M265
2. Predicate 510(k) number(s):
K161573
3. Comparison with predicate:
{3}
| Similarities | | |
| --- | --- | --- |
| Item | FilmArray Warrior Control Panel M290 (K163522) | FilmArray RP EZ Control Panel M265 (K161573) |
| Intended Use | FilmArray Warrior Control Panel M290 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bacillus anthracis (Chromosomal element, pXO1, pXO2), Coxiella burnetii, Francisella tularensis, Yersinia pestis, Ebola virus and Marburg virus on the FilmArray NGDS Warrior Panel performed on FilmArray 2.0 systems. FilmArray Warrior Control Panel M290 is composed of synthetic DNA specifically designed for and intended to be used solely with the FilmArray NGDS Warrior Panel assay. This product is not intended to replace manufacturer internal FilmArray NGDS Warrior Panel controls provided with the device. | FilmArray RP EZ Control Panel M265 is intended for use as external positive and negative, surrogate assayed quality control materials to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/ Enterovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype H1-2009, Influenza B, Parainfluenza Virus, Respiratory Syncytial Virus, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae on the FilmArray RP EZ assay performed on the FilmArray systems. The control panel also contains a negative control. This product is not intended to replace manufacturer controls provided with the device. |
| Physical format | Ready-to-Use Liquid | Same |
| Directions for Use | Process like patient Sample | Same |
| Composition | Synthetic DNA | Synthetic RNA transcripts |
| Number of targets monitored in one assay | Multiple | Same |
| Differences | | |
| --- | --- | --- |
| Item | FilmArray RP EZ Control Panel M265 (K163522) | FilmArray RP EZ Control Panel M265 (K161573) |
| Assay Steps Monitored | Amplification, detection, identification | Same plus reverse transcription |
{4}
K. Standard/Guidance Document Referenced (if applicable):
None were referenced
L. Test Principle:
Not applicable
M. Performance Characteristics (if/when applicable):
1. External Site Testing:
Performance characteristics of the FilmArray Warrior Control Panel M290 on the FilmArray instrument 2.0 were evaluated at five clinical laboratory sites. Testing was conducted by multiple operators for over a six month period, consisting of three lots of FilmArray Warrior Control Panel M290 and nine FilmArray Warrior Panel pouch lots. A total of 310 external controls were tested (163 positive and 147 negative). The results are shown in the table below:
| Summary of External Test Results | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| External Site | Total Tests | Percent Total Correct Results | Correct Positive Control Result | Incorrect Positive Control Result | Percent Correct Positive Control | Correct Negative Control Result | Incorrect Negative control Result | Percent Correct Negative Control |
| A | 41 | 100% | 20 | 0 | 100% | 21 | 0 | 100% |
| B | 104 | 99.1% | 58 | 1 | 98.3% | 45 | 0 | 100% |
| C | 51 | 100% | 26 | 0 | 100% | 25 | 0 | 100% |
| D | 60 | 100% | 31 | 0 | 100% | 29 | 0 | 100% |
| E | 54 | 98.1% | 26 | 1 | 96.3% | 27 | 0 | 100% |
| All Sites | 310 | 99.4% | 161 | 2 | 98.8% | 147 | 0 | 100% |
2. Analytical performance:
a. Precision:
An internal precision study for the FilmArray Warrior Control Panel M290 was conducted over multiple days by testing three FilmArray Warrior Control Panel M290 lots with four FilmArray Warrior Panel assay lots performed by three operators. The results are shown in the table below:
{5}
6
| FilmArray Warrior Control Panel Summary of Internal Precision Test Results | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Control | Control Lot # | No. of Tests | Invalid | Correct Results | Incorrect Results | Percent Correct^{1} |
| FilmArray Warrior Positive Control | C18MAR16 | 20 | 0 | 20 | 0 | 100% |
| FilmArray Warrior Negative Control | A02NOV15 | 20 | 0 | 20 | 0 | 100% |
| FilmArray Warrior Positive Control | C01JUL16 | 20 | 0 | 20 | 0 | 100% |
| FilmArray Warrior Negative Control | B01JUL16 | 21 | 1 | 20 | 0 | 100% |
| FilmArray Warrior Positive Control | B15JUL16 | 22 | 2 | 20 | 0 | 100% |
| FilmArray Warrior Negative Control | A15JUL16 | 21 | 1 | 20 | 0 | 100% |
| | TOTAL | 124 | 4 | 120 | 0 | 100% |
¹The 4 Invalid samples were re-tested according to BioFire instructions and were not included in the Percent Correct analysis.
There appears to be no significant differences in mean Cp value when testing different control lots on different days. Precision studies are acceptable.
b. Linearity/Assay Reportable Range:
Not applicable
c. Lot-to-Lot Testing: A subset of testing in the internal precision study was used to demonstrate Lot-to-lot reproducibility by testing 3 lots of FilmArray Warrior Control Panel M290 using the same pouch lot on the same instrument at MMQCI. The results are shown in the table below:
| FilmArray Warrior Control Panel Summary of Lot-to-Lot Testing | | | | |
| --- | --- | --- | --- | --- |
| Control | Control Lot # | Number of Tests | Pouch Lot | Correct Results |
| FilmArray Warrior Positive Control | C18MAR16 | 7 | 160314 | 7/7 |
| FilmArray Warrior Negative Control | A02NOV15 | 7 | 160314 | 7/7 |
| FilmArray Warrior Positive Control | C01JUL16 | 7 | 160314 | 7/7 |
| FilmArray Warrior Negative Control | B01JUL16 | 8 (1 Invalid) | 160314 | 7/7 |
| FilmArray Warrior Positive Control | B15JUL16 | 7 | 160314 | 7/7 |
| FilmArray Warrior Negative Control | A15JUL16 | 7 | 160314 | 7/7 |
d. Within-run Testing:
Within-run reproducibility was demonstrated in a separate study conducted by one operator testing one lot each of FilmArray Warrior Positive Control and FilmArray Warrior Negative Control with one lot of NGDS Warrior Panel pouches on the FilmArray 2.0, each within one day at MMQCI.
{6}
| FilmArray Warrior Control Panel Summary of Within-run Reproducibility Results | | | | |
| --- | --- | --- | --- | --- |
| Control | Control Lot# | Number of Tests | Pouch Lot | Correct Results |
| FilmArray Warrior Positive Control | C01JUL16 | 6 | 151216 | 6/6 |
| FilmArray Warrior Negative Control | B01JUL16 | 6 | 151216 | 6/6 |
Within-run reproducibility studies for the FilmArray Warrior Control Panel M290 are acceptable.
e. Linearity/assay reportable range:
Not applicable
f. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
Not applicable
Stability:
Open Vial Stability: Not applicable because FilmArray Warrior Control Panel M290 is packaged for single use.
Closed Vial Real-time Stability: An accelerated stability study was performed to establish the shelf life stability claims for the FilmArray Warrior Control Panel M290. Based on this study, the FilmArray Warrior Control Panel M290 is expected to be stable until the expiration date (12 months) when stored frozen $(-20^{\circ}\mathrm{C}$ or colder) and unopened. This product is for single use.
Real-Time Stability Program: Real-time stability studies are ongoing to support product claims and to monitor potential assay modifications for which the FilmArray Warrior Control Panel M290 is indicated for use. Real-time stability study protocols and acceptance criteria were reviewed and found to be acceptable.
Shipping Stability: MMQCI ships the FilmArray Warrior Control Panel M290 on dry ice with overnight delivery, ensuring that the control material remains frozen upon receipt. The frozen control material is then to be stored at $-20^{\circ}\mathrm{C}$ , as indicated in the FilmArray Warrior Control Panel M290 Package insert. A shipping study was performed to confirm the shipping process and to investigate the outcome of a possible shipping delay and subsequent arrival with no dry ice. The study evaluated two (2) lots of the FilmArray Warrior Positive Control placed in dry ice and was stored for five (5) days at ambient temperature then tested using the FilmArray Warrior Panel Assay. To simulate a shipping delay, two (2) lots of the FilmArray
{7}
Warrior Positive Control were stored for up to 5 days at ambient temperature and tested using the FilmArray Warrior Panel Assay.
The study demonstrated that the FilmArray Warrior Control Panel M290 is stable for two (2) days on dry ice in MMQCI's standard shipping box. FilmArray Warrior Positive Control is stable after five (5) days at ambient temperatures of approximately $19 - 21^{\circ}\mathrm{C}$ . The FilmArray Warrior Positive Control should be stored frozen $(-20^{\circ}\mathrm{C}$ or colder) as indicated in the package insert.
# Expected Values:
FilmArray Warrior Positive Control is a qualitative control and the expected results are listed in the tables below.
FilmArray Warrior Control Panel M290 Result Summary
| FilmArray Warrior Positive Control | |
| --- | --- |
| Bacillus anthracis | Detected |
| pXO1 | Detected |
| pXO2 | Detected |
| Coxiella burnetii | Detected |
| Francisella tularensis | Detected |
| Yersinia pestis | Detected |
| Ebola virus | Detected |
| Marburg virus | Detected |
| FilmArray Warrior Negative Control | |
| Bacillus anthracis | Not Detected |
| pXO1 | Not Detected |
| pXO2 | Not Detected |
| Coxiella burnetii | Not Detected |
| Francisella tularensis | Not Detected |
| Yersinia pestis | Not Detected |
| Ebola virus | Not Detected |
| Marburg virus | Not Detected |
# Matrix Effects:
A study was performed to evaluate potential matrix effects of the FilmArray Warrior Control Panel M290. To confirm that clinical matrix (blood) does not significantly affect the results of the FilmArray NGDS Warrior Panel assay, inactivated Francisella tularensis was spiked into the FilmArray Warrior Negative Control, which contains the identical matrix found in the FilmArray Warrior Positive Control, and whole blood to simulate a clinical sample. Both sample types were tested in triplicate by the FilmArray NGDS Warrior Panel assay.
{8}
Results demonstrated that the FilmArray Warrior Control Panel M290 matrix has no effect on the assay. Study results are shown in the table below:
| Biothreat Pathogen Analyte | Spiked Francisella tularensis Crossing Point (Cp) Data | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Whole Blood | | | | Mean Cp | MMQCI Matrix | | | | Mean Cp |
| FTT2 | Positive | 22.4 | 21.2 | 22.0 | 21.9 | Positive | 21.5 | 21.1 | 22.2 | 21.6 |
| FTT3 | Positive | 27.0 | 27.0 | 27.0 | 27.0 | Positive | 23.8 | 27.0 | 25.0 | 25.3 |
| PCR2 | Positive | 18.3 | 18.2 | 18.5 | 18.3 | Positive | 18.3 | 18.2 | 18.1 | 18.2 |
| YeastRNA | Positive | 17.7 | 20.9 | 18.7 | 19.1 | Positive | 18.9 | 20.8 | 19.0 | 19.6 |
g. Detection limit:
Not applicable
h. Analytical Reactivity (Inclusivity):
Not applicable
i. Cross Reactivity:
Not applicable
j. Interference:
Not applicable
k. Assay cut-off:
Not applicable.
3. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
4. Clinical Studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
{9}
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
5. Clinical cut-off:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and satisfies the requirements of 21 CFR Parts 801 and 809 and the specials controls for this device type.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.