VITROS Chemistry Products GLU Slides

K163433 · Ortho Clinical Diagnostics · CGA · Sep 18, 2017 · Clinical Chemistry

Device Facts

Record IDK163433
Device NameVITROS Chemistry Products GLU Slides
ApplicantOrtho Clinical Diagnostics
Product CodeCGA · Clinical Chemistry
Decision DateSep 18, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

For in vitro diagnostic use only. VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS Systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia and of pancreatic islet cell carcinoma.

Device Story

VITROS Chemistry Products GLU Slides are multilayered analytical elements for quantitative glucose measurement. Device processes patient samples (serum, plasma, urine, CSF) deposited on slide; spreading layer distributes sample to underlying reagent layers. Glucose oxidase catalyzes glucose oxidation to hydrogen peroxide and gluconate; peroxidase catalyzes oxidative coupling with dye precursors to produce dye. Reflectance spectrophotometry on VITROS 4600 System measures dye intensity. Modification from predicate (k955286) reduces surface area, reagent amounts, and sample volume (from 10 μL to 6 μL). Used in clinical laboratories by trained personnel. Output is glucose concentration (mg/dL) used by clinicians for diagnosis and management of metabolic disorders.

Clinical Evidence

No clinical studies performed. Evidence consists of analytical bench testing including precision (n=88), linearity, limit of detection/quantitation, and interference studies (endogenous/exogenous substances) per CLSI guidelines (EP5-A3, EP6-A, EP9-A3, EP7-A2, EP17-A2). Method comparison against predicate showed high correlation (r ≥ 0.999) across serum, urine, and CSF matrices. Matrix comparison confirmed suitability of various plasma anticoagulants.

Technological Characteristics

Multilayered analytical element on polyester support. Reagents: glucose oxidase, peroxidase, 1,7-dihydroxynaphthalene, 4-aminoantipyrine hydrochloride. Principle: enzymatic endpoint colorimetric assay via reflectance spectrophotometry. Dimensions/form factor: slide-based. Connectivity: used on VITROS 4600 System. Sterilization: not applicable (in vitro diagnostic reagent).

Indications for Use

Indicated for quantitative measurement of glucose in serum, plasma, urine, and cerebrospinal fluid for the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 18. 2017 ORTHO CLINICAL DIAGNOSTICS DARLENE PHILLIPS MANAGER REGULATORY AFFAIRS 100 INDIGO CREEK DRIVE ROCHESTER NY 14626 Re: K163433 Trade/Device Name: VITROS Chemistry Products GLU Slides Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CGA Dated: August 15, 2017 Received: August 16, 2017 Dear Darlene Phillips: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, **Kellie B. Kelm -S** for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) k163433 Device Name VITROS Chemistry Products GLU Slides #### Indications for Use (Describe) For in vitro diagnostic use only. VITROS Chemistry Products GLU Slides are used on VITROS Systems to quantitatively measure glucose (GLU) concentration in serum, plasma, urine and cerebrospinal fluid. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. | Type of Use (Select one or both, as applicable) | |------------------------------------------------------------------------------------------------| | <span style="text-decoration: underline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 1. 510(k) Summary #### Submitter Information | Name | Ortho-Clinical Diagnostics, Inc. | |-----------------------------------------|-----------------------------------------------------| | Address | 100 Indigo Creek Drive<br>Rochester, New York 14626 | | Phone number | (585) 453-4253 | | Fax number | (585) 453-3368 | | Establishment<br>Registration<br>Number | 1319809 | | Name of<br>contact person | Darlene J Phillips | | Date prepared | September 18, 2017 | | Name of devices | | | Trade or proprietary names | VITROS Chemistry Products GLU Slides | | Common or usual name | VITROS GLU Slides assay | | Common or usual name | Glucose oxidase, glucose | | Classification name | Glucose test system | | Classification panel | Clinical Chemistry | | Regulation | 21 CFR 862.1345 | | Product Code | CGA | #### Legally marketed device(s) to which equivalence is claimed The VITROS® Chemistry Products GLU Slides (modified) are substantially equivalent to the VITROS® Chemistry Products GLU Slides (current). The FDA cleared the VITROS® Chemistry Products GLU Slides on May 17, 1996 (k955286). #### Reason for 510(k) submission The device modification is a proportional reduction in the amount of reagent used to manufacture each VITROS GLU Slide (or test). Each modified VITROS GLU Slide will have approximately 30% less ingredients compared to the current VITROS GLU Slide as a result of a smaller surface area. Since the reduction in ingredients per slide is due to a smaller surface area, the proportional concentration of ingredients of the modified VITROS GLU slide will be unchanged compared to the current VITROS GLU Slide. This device modification reduces the sample volume required per test from 10 µL per test to 6 µL per test. #### Device description The VITROS GLU Slide is a multilayered, analytical element coated on a polyester support. A drop of patient sample is deposited on the slide and is evenly distributed {4}------------------------------------------------ k163433 by the spreading layer to the underlying layers. The oxidation of sample glucose is catalyzed by glucose oxidase to form hydrogen peroxide and gluconate. This reaction is followed by an oxidative coupling catalyzed by peroxidase in the presence of dye precursors to produce a dye. The intensity of the dye is measured by reflected light. # Intended Use/ Indications for Use For in vitro diagnostic use only. VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS Systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia and of pancreatic islet cell carcinoma. Special Conditions for Use Statement: For prescription use only. # Comparison with Predicate Devices: | predicate device | | | |-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device<br>Characteristic | New Device<br>VITROS GLU Slide<br>(Modified) | Predicate Device<br>VITROS GLU Slide[k955286]<br>(Current) | | Intended Use | No Change. | For in vitro diagnostic use only.<br>VITROS® Chemistry Products GLU<br>Slides quantitatively measure<br>glucose (GLU) concentration in<br>serum, plasma, urine, and<br>cerebrospinal fluid. | | Sample volume | $6 \mu L$ | $10 \mu L$ | | Concentrations<br>of Slide<br>Reactive<br>Ingredients per<br>cm-squared | No Change | Glucose oxidase (Aspergillus sp.)<br>$0.77$ U; peroxidase (horseradish root)<br>$3.6$ U; 1,7- dihydroxynaphthalene<br>(dye precursor) $67 \mu g$ and 4-<br>aminoantipyrine hydrochloride (dye<br>precursor) $0.11$ mg. | | Amount of Slide<br>Reactive<br>Ingredients per<br>Slide (test) | Glucose oxidase<br>(Aspergillus sp.) $0.77$ U;<br>peroxidase (horseradish root)<br>$3.6$ U; 1,7-<br>dihydroxynaphthalene (dye<br>precursor) $67 \mu g$ and 4-<br>aminoantipyrine<br>hydrochloride (dye<br>precursor) $0.11$ mg. | Glucose oxidase (Aspergillus sp.)<br>$0.99$ U; peroxidase (horseradish root)<br>$4.7$ U; 1,7- dihydroxynaphthalene<br>(dye precursor) $86 \mu g$ and 4-<br>aminoantipyrine hydrochloride (dye<br>precursor) $0.14$ mg. | ### Summary of the technological characteristics of the device compared to the nredicate device {5}------------------------------------------------ | Device<br>Characteristic | New Device<br>VITROS GLU Slide<br>(Modified) | Predicate Device<br>VITROS GLU Slide[k955286]<br>(Current) | |------------------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device<br>Description | No Change | The VITROS GLU assay is<br>performed using the VITROS<br>Chemistry Products GLU Slides and<br>the VITROS Chemistry Products<br>Calibrator Kit 1 on the VITROS<br>Systems. The VITROS GLU Slide is<br>a multi-layered, analytical element<br>coated on a polyester support. The<br>method is based on an enzymatic<br>detection. All reactions necessary for<br>a single quantitative measurement of<br>glucose take place within the multi-<br>layered analytical element of a<br>VITROS GLU Slide. A drop of<br>sample fluid is metered onto the slide<br>and a reaction occurs which<br>ultimately results in the oxidation of<br>dye precursors by peroxidase. The<br>concentration of glucose in the<br>sample is determined by measuring<br>the intensity of the dye using<br>reflectance spectrophotometry. | | Basic Principle | No Change | Enzymatic Endpoint test using<br>reflectance spectrophotometry. | | Sample type | No Change | Serum, plasma, urine, and<br>cerebrospinal fluid | | Assay Range | No Change | Serum: 20.0-625.0 mg/dL<br>Urine and CSF: 20.0-650.0 mg/dL | | Incubation time<br>and temperature | No Change | 5 minutes at 37°C | # Summary of the technological characteristics of the device compared to the predicate device {6}------------------------------------------------ ### Non-Clinical Testing Analytical Performance: ### Method comparison Predicate vs Feature To demonstrate substantial equivalence with the predicate, human serum, urine and CSF samples were tested with the modified VITROS Chemistry Products GLU Slides (feature) and the results obtained were compared to those obtained using the current VITROS Chemistry Products GLU Slides (predicate). Method Comparison testing followed CLSI Protocol EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples. The correlation between the predicate and the modified VITROS GLU Slides on the VITROS 4600 Chemistry System is summarized below. | | No. Obs. | Slope | y-intercept | Correlation | |-------|----------|-------|-------------|-------------| | Serum | 113 | 1.01 | -0.80 | 1.000 | | Urine | 105 | 1.03 | -0.81 | 0.999 | | CSF | 165 | 1.04 | -2.74 | 1.000 | Units (mg/dL) #### Precision Long term precision testing followed CLSI Protocol EP05-A3, Evaluation of Precision Performance of Quantitative Methods; Approved Guideline—Third Edition, using the modified VITROS Chemistry Products GLU Slides on the VITROS Systems. The test included a minimum of 80 observations (2 replicates per run, 2 runs per day over 20 days) for serum, urine and CSF specimens. The long term precision of the modified VITROS GLU Slides on the VITROS 4600 Chemistry System is summarized below. #### Precision | Serum | Mean<br>Conc. | Within<br>day SD* | Within lab<br>SD** | Within Lab<br>CV% | No.<br>Obs. | No. Days | |--------------|---------------|-------------------|--------------------|-------------------|-------------|----------| | Serum Pool 1 | 29.4 | 0.24 | 0.39 | 1.3 | 88 | 22 | | PV I | 90.3 | 0.63 | 1.47 | 1.6 | 88 | 22 | | PV II | 287.2 | 1.56 | 3.35 | 1.2 | 88 | 22 | | Serum Pool 2 | 571.2 | 3.28 | 4.99 | 0.9 | 88 | 22 | Units (mg/dL) | Urine | Mean<br>Conc. | Within<br>day SD* | Within lab<br>SD** | Within Lab<br>CV% | No.<br>Obs. | No.<br>Days | |--------------|---------------|-------------------|--------------------|-------------------|-------------|-------------| | Control 1 | 31.6 | 0.34 | 0.47 | 1.5 | 88 | 22 | | Control 2 | 301.4 | 2.84 | 4.48 | 1.5 | 88 | 22 | | Urine Pool 1 | 574.5 | 4.05 | 5.46 | 1.0 | 88 | 22 | Units (mg/dL) {7}------------------------------------------------ | CSF | Mean<br>Conc. | Within<br>day SD* | Within lab<br>SD** | Within Lab<br>CV% | No.<br>Obs. | No.<br>Days | |--------------|---------------|-------------------|--------------------|-------------------|-------------|-------------| | CSF Pool 1 | 37.2 | 0.29 | 0.56 | 1.5 | 88 | 22 | | Liquid PV I | 49.2 | 0.45 | 0.64 | 1.3 | 88 | 22 | | Liquid PV II | 88.9 | 0.58 | 1.11 | 1.2 | 88 | 22 | | CSF Pool 2 | 283.0 | 1.69 | 3.55 | 1.3 | 88 | 22 | | CSF Pool 3 | 609.3 | 2.58 | 4.03 | 0.7 | 88 | 22 | Units (mg/dL) * Within Day precision was determined using two runs per day with two replications. ** Within Lab precision was determined using a single lot of slides and calibrating weekly. # Limit of Blank, Detection and Quantitation (LoB/LoD/LoQ) The Limit of Blank (LOB), Limit of Detection (LOD), and Limit of Quantitation (LOO) of the VITROS GLU assay were determined according to CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. The following limits of the modified VITROS GLU Slides assay are summarized below: #### Limit of Blank, Detection and Quantitation Calculated | Fluid Type | LOB (mg/dL) | LOD (mg/dL) | LOQ (mg/dL) | |------------|-------------|-------------|-------------| | Serum | 1 | 2 | 9 | | Urine | 7 | 8 | 18 | | CSF | 4 | 5 | 20 | The Limit of Quantitation is defined as the lowest amount of analyte in the sample that can be quantitatively determined with stated acceptable precision and trueness, under stated experimental conditions. A properly operating VITROS System should exhibit a LoQ value of 20 mg/dL with serum, plasma, urine and CSF specimens. # Linearity Linearity was determined following CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures: a Statistical Approach: Approved Guideline. A high GLU pool was intermixed with a low pool to generate 14 concentration levels each tested in six replicate determinations for each body fluid type, serum, urine and CSF. Linear results were compared to 2nd and 3rd order polynomial fits against a pre-specified allowable error. Analysis by linear regression indicated that the assay is linear across the measuring range of each fluid type. {8}------------------------------------------------ #### Matrix Comparison The performance of plasma specimens were compared to serum using a paired serum/plasma blood draw. The results indicated that serum (stored removed from the clot), EDTA, both sodium and lithium heparin, and sodium fluoride/potassium oxalate plasma are suitable for use with the VITROS Chemistry Products GLU Slides assay. #### Measuring Range and Expected Values The modification does not impact the measuring range and expected values of the modified VITROS GLU Slides assays. Current claims will apply as summarized below. Each laboratory should confirm the validity of these intervals for the population it serves. #### Measuring (Reportable or Dynamic) Range | Serum | 20.0 – 625.0 mg/dL | |-------|--------------------| | Urine | 20.0 – 650.0 mg/dL | | CSF | 20.0 – 650.0 mg/dL | #### Expected Values | Serum | | |----------------|----------------| | Fasting Adults | 74 - 106 mg/dL | | Urine | | | Random | < 30 mg/dL | | 24-hour | < 500 mg/day* | | CSF | 40 - 70 mg/dL | *Glucose concentration (md/dL) x 24-hour volume (dL) = mg/day {9}------------------------------------------------ #### Interfering Substances Specimens Not Recommended Urine with the following preservatives: - . Acetic acid - . Hydrochloric acid The VITROS Chemistry Products GLU Slide assay was screened for interfering substances following CLSI Protocol EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. The presence of N-Acetylcysteine, cholesterol, glutathione, dextran and total protein in serum are known to cause bias in assay results. High levels of ascorbic acid in urine will cause a large negative bias, potentially resulting in values less than the measuring range. The degree of bias is proportional to the ascorbic acid concentration. The presence of hemoglobin in CSF is known to cause a positive bias in assay results. Substances exhibiting bias less than the stated acceptance criteria have been cited as substances that do not interfere. It is possible that other interfering substances may be encountered. These results are representative; however results may differ somewhat due to test-to-test variation. The degree of interference at concentrations other than those listed may not be predictable. #### Conclusion: The conclusions drawn from the nonclinical tests (discussed above) demonstrate the modified VITROS Chemistry Products GLU Slides assay is as safe, effective, and perform as well as the predicate device, current VITROS Chemistry Products GLU Slides assay. The information submitted in the premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...