ONE TOUCH BASIC/PROFILE/ONE TOUCH II TEST STRIPS

K031472 · Lifescan, Inc. · CGA · Aug 5, 2003 · Clinical Chemistry

Device Facts

Record IDK031472
Device NameONE TOUCH BASIC/PROFILE/ONE TOUCH II TEST STRIPS
ApplicantLifescan, Inc.
Product CodeCGA · Clinical Chemistry
Decision DateAug 5, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2

Intended Use

The ONE TOUCH® Basic/Profile/ONE TOUCH® II Test Strips are intended for use with ONE Touch ® Brand meters by healthcare professionals and laypersons with diabetes at home to provide a quantitative measurement of glucose in whole blood as an aid in managing their glycemic state.

Device Story

System consists of reagent-impregnated test strip and reflectance photometer. User inserts strip into meter; applies capillary blood drop. Glucose oxidase catalyzes reaction of glucose with oxygen to produce gluconic acid and hydrogen peroxide; hydrogen peroxide oxidizes dyes in presence of peroxidase to produce blue color. Meter measures light reflection intensity; converts to glucose concentration. User enters lot-specific calibration code. Used by healthcare professionals and patients at home for glycemic management. Modified version features altered dimensions for smaller blood sample volume.

Clinical Evidence

Bench testing only. Performance of the modified test strip was compared to the predicate device across the hematocrit range of 30% to 55% to demonstrate equivalence.

Technological Characteristics

In vitro diagnostic test strip; glucose oxidase and peroxidase enzyme chemistry; reflectance photometry measurement principle. Requires lot-specific calibration code entry. Compatible with ONE TOUCH® Basic, Profile, and II meters.

Indications for Use

Indicated for healthcare professionals and laypersons with diabetes to quantitatively measure glucose in whole blood to aid in glycemic management.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary | Submitter | LifeScan, Inc.<br>1000 Gibraltar Drive<br>Milpitas, CA 95035<br>Contact: Frank Peralta<br>Date Prepared: May 8, 2003 | |------------------|----------------------------------------------------------------------------------------------------------------------| | Device Name | ONE TOUCH® Basic/Profile/ONE TOUCH® II Test Strips<br>Common name: Glucose test strip | | Predicate Device | ONE TOUCH® Basic/Profile/ONE TOUCH® II Test Strips | ## Device Description The ONE TOUCH® Blood Glucose Monitoring System is an in vitro diagnostic product consisting of a test strip impregnated with reagents and a reflectance photometer for the determination of glucose in whole blood. Several different ONE Touch ® Brand reflectance photometers may be used (the ONE TOUCH ® Basic, ONE Touch ® II, or the ONE TOUCH ® Profile Meter). A quality control solution, ancillary devices to aid obtaining a capillary blood sample (lancets and lancing device) as well as data management computer software are also available. A ONE Touch ® Test Strip is inserted into the meter and a drop of blood applied to the test spot. Glucose oxidase present in the test strip catalyzes the reaction of qlucose in the sample with oxygen to yield gluconic acid and hydrogen peroxide. Hydrogen peroxide subsequently oxidizes dyes in the test strip in the presence of peroxidase, a second enzyme, to produce a blue colored form of the dyes. The intensity of the blue color produced is proportional to the amount of glucose in the sample. The meter measures the amount of light reflected by the blue product and converts it to a glucose concentration that is presented on the meter display. The user adjusts the meter response for each lot of test strips by entering a calibration code that is specific to that lot into the meter. No modifications to the intended use or to the fundamental scientific technology of the ONE Touch® Blood Glucose Monitoring System are being made. The description above applies to the currently marketed device and will apply equally to the modified device described in this submission. ## Intended Use The ONE TOUCH® Basic/Profile/ONE TOUCH® II Test Strips are intended for use with ONE Touch ® Brand meters by healthcare professionals and laypersons with diabetes at home to provide a quantitative measurement of glucose in whole blood as an aid in managing their glycemic state. {1}------------------------------------------------ ## Comparison to Predicate Device The dimensions of the ONE TOUCH® Test Strip have been altered to permit use of a smaller blood drop. The performance of the modified test strip has been shown to be equivalent to that of the predicate through the acceptable hematocrit range (30% to 55%). ## Conclusion The modified ONE TOUCH ®Test Strip is substantially equivalent to the predicate device. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is composed of three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the caduceus. Mr. Frank Peralta Senior Regulatory Submissions Specialist LifeScan, Inc. 1000 Gibraltar Drive Milpitas, California 95035-6312 AUG - 5 2003 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Re: k031472 > Trade/Device Name: One Touch® Basic/Profile/One Touch II Test Strip Regulation Number: 21 CFR § 862.1345 Regulation Name: System, Test, Blood Glucose, Over-the-Counter Regulatory Class: II Product Code: NBW. CGA Dated: May 8, 2003 Received: May 9, 2003 Dear Mr. Peralta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 - If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use Statement 510(k) Number: Ko31/72 ONE TOUCH® Basic/Profile/ ONE Touch® II Test Strips Device Name: Indications for Use: The ONE TOUCH® Basic/Profile/ONE TOUCH® II Test Strips are intended for use with ONE TOUCH ® Brand meters by healthcare professionals and laypersons with diabetes at home to provide a quantitative measurement of glucose in whole blood as an aid in managing their glycemic state. Alberto S Division Sign-Off For: Jean Cooper Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K031472 Concurrence of CDRH, Office of Device Evaluation Prescription Use (Per 21 CFR 801.109) OR Over-the-Counter Use _
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.