K163382 · Oxitone Medical , Ltd. · DQA · May 9, 2017 · Cardiovascular
Device Facts
Record ID
K163382
Device Name
Oxitone 1000
Applicant
Oxitone Medical , Ltd.
Product Code
DQA · Cardiovascular
Decision Date
May 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The Oxitone Model 1000 Pulse Oximeter is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate. It is intended for spot-checking of adult patients with wrist between 13-22 cm circumferences in hospitals, clinics, long-term care, and home use.
Device Story
Oxitone 1000 is a wrist-worn pulse oximeter for spot-checking SpO2 and pulse rate in adults. Device uses reflectance photoplethysmography (PPG); bi-color LEDs (640nm red, 940nm infrared) and photodetector are positioned on the wrist near the ulnar bone. Light is transmitted through tissue, partially absorbed by oxyhemoglobin, and sensed by the photodetector. Electronics compute SpO2 and pulse rate from these signals. Device features an LCD display for digital values; no pulse amplitude display. Powered by internal rechargeable Li-Ion battery. Used by clinicians or patients in hospitals, clinics, long-term care, or home environments. Output provides immediate physiological data for spot-check assessment. Benefits include non-invasive, portable monitoring of oxygenation and heart rate.
Clinical Evidence
Clinical accuracy testing performed on 10 patients in accordance with ISO 80601-2-61. Results demonstrated compliance with standard requirements for SpO2 and pulse rate accuracy.
Technological Characteristics
Wrist-worn reflectance pulse oximeter. Materials: biocompatible (ISO 10993-1). Sensing: Red (640nm) and IR (940nm) LEDs; photodetector. Power: Rechargeable Li-Ion battery. Connectivity: Standalone. Standards: AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-61, IEC 62471, IEC 62133. Form factor: Watch-like device with elastic concave surface for ulnar bulge alignment.
Indications for Use
Indicated for spot-checking functional oxygen saturation (%SpO2) and pulse rate in adult patients (wrist circumference 13-22 cm) in clinical and home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Nonin PalmSAT with Purelight Forehead 8000R sensor (K050056)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 9, 2017
Oxitone Medical Ltd. % Paul Dryden Consultant Promedic, LLC 24301 Woodsage Dr. Bonita Springs, Florida 34134
Re: K163382
Trade/Device Name: Oxitone 1000 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: March 28, 2017 Received: March 29, 2017
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang" written in a clear, legible font. Below the name, there is a small "-s" written. The signature is complex and stylized, featuring looping lines and curves. The background of the image has a faint watermark.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K163382
Device Name
Oxitone 1000
Indications for Use (Describe)
The Oxitone Model 1000 Pulse Oximeter is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate.
It is intended for spot-checking of adult patients with wrist between 13-22 cm circumferences in hospitals, clinics, long-term care, and home use.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
|Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14
SC Publishing Services (301) 443-6740
{3}------------------------------------------------
#### 510(k) Summary May 8, 2017 Page 1 of 6
| Official Contact: | Leon Eisen, PhD Founder and CEO<br>Oxitone Medical Ltd.<br>17 Atir Yeda Street<br>Kfar Saba, Israel 4464312<br>Tel: +972 (9) 834 6731 |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | Oxitone 1000 |
| Common/Usual Name: | Oximeter |
| Classification Name: | Oximeter<br>DQA, Class II, CFR 870.2700 |
| Predicate Device:<br>Reference Device: | Nonin Model 9590 - K112843<br>Nonin 8000R Forehead sensor - K050056 |
#### Device Description:
The Oxitone 1000 pulse oximeter is an instrument for photoelectrically determining the oxygenation of blood in a part of the body, in this case the wrist.
This Oxitone 1000 contains the electronics, interface and sensor. The Oxitone pulse oximeter is mounted on a watch-like device and measures oxygen saturation at the wrist location. The Oxitone pulse oximeter exploits a reflectance sensor that includes the photodetector located at the wearer's Ulnar Bone near the Ulnar Bulge and a bi-color light source (LED's) that are located around the top of the wearer's Ulnar "Bulge", to align the light source properly to the Ulnar Bulge the LEDs surrounded by elastic concave shape surface. The Red and infrared light being transmitted through the tissue is partially absorbed by the blood oxyhemoglobin and is sensed by the photodetector. The blood oxygen saturation is measured using the well-established technology where the red and infrared light is absorbed in different amounts depending on the oxygenation of the blood that enables to calculate the blood functional pulse oximetry.
The Oxitone 1000 has the ability to determine both the percent of saturated hemoglobin and the pulse rate. It performs these functions on adult patient populations. It is designed for spot checking both the SpO2 and Pulse rate information. To perform this, the Oxitone 1000 displays digital values for both the SpO2 and Pulse.
Pulse amplitude is not displayed. The Oxitone 1000 is powered by an integral rechargeable Lithium Ion battery pack. The wavelength of red LED is 640nm and Infrared LED is 940 nm with maximum optical output power of less than 1 mW.
The device incorporates a battery state indicator and provides a visual indication of low battery.
#### Indications for Use:
The Oxitone Model 1000 Pulse Oximeter is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate.
It is intended for spot-checking of adult patients with wrist between 13-22 cm circumferences in hospitals, clinics, long-term care, and home use.
#### Contraindications:
None
{4}------------------------------------------------
#### Substantial Equivalence Discussion
The Oxitone 1000 Pulse Oximeter is similar to other small worn SpO2 monitors. There are pulse oximeters that use sensors that measure SpO2 at the extremities - fingers, toes, ear lobes, etc. They also have sensors which are placed on the forehead using reflectance technology for measuring SpO2
We have selected 2 devices for comparative purposes for demonstrating substantial equivalence. Table 1 outlines the rationale.
| Feature | Primary Predicate | Reference | Subject device |
|-----------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| | Nonin<br>Model 9590<br>(K112843) | Nonin PalmSAT with<br>Purelight<br>Forehead 8000R sensor<br>(K050056) | Oxitone 1000 |
| Classification | DQA<br>CFR 870.2700<br>Pulse oximeter | DQA<br>CFR 870.2700<br>Pulse oximeter | DQA<br>CFR 870.2700<br>Pulse oximeter |
| Intended Use | Measure functional oxygen<br>saturation of arterial<br>hemoglobin (%SpO2) and<br>pulse rate | Measure functional oxygen<br>saturation of arterial<br>hemoglobin (%SpO2) and<br>pulse rate | Measure functional oxygen<br>saturation of arterial<br>hemoglobin (%SpO2) and<br>pulse rate |
| | For spot-checking | For spot-checking | For spot-checking |
| Patients | Adult<br>Pediatrics | Adult<br>Pediatrics | Adult |
| Sensor location | Digits | Digits<br>Forehead | Wrist |
## Table 1 – Predicate and Reference Devices
The rationale for the selected devices is that the predicate has the indications for use and the reference shows that SpO2 measurements can be measure in other anatomical locations other than digits.
We demonstrate device pereformance by testing the subject device according to ISO 80601-2-61 and the FDA's Guidance for Pulse Oximeters (March 4, 2013).
## Table of Comparison and Differences
The Oxitone 1000 Pulse Oximeter is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate.
It is intended for spot-checking of adult patients with wrist between 13-22 cm circumferences in hospitals, clinics, long-term care and home use.
The monitor features an easy-to-read display that presents patient data, status information, and alarms when appropriate.
The Oxitone 1000 permits spot checking of functional oxygen saturation (SpO2) and Pulse Rate (PR) in clinical and non-clinical settings.
{5}------------------------------------------------
#### 510(k) Summary May 8, 2017 Page 3 of 6
The software monitors the signals, computes the PR and SpO2 values and displays the PR and SpO2 values.
Table 2 outlines the features of the Oxitone 1000 and compares it to the predicate and reference devices to establish substantial equivalence.
| Table 2 Device Comparison | | | | | |
|---------------------------|--|--|--|--|--|
|---------------------------|--|--|--|--|--|
| CHARACTERISTICS | Oxitone 1000 | Predicate<br>Nonin<br>Model 9590 (K112843) | Reference<br>Nonin PalmSAT with<br>Forehead 8000R sensor<br>(K050056) |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Indications for Use | The Oxitone Model 1000<br>Pulse Oximeter is a small,<br>wrist-worn device indicated<br>for use in measuring and<br>displaying functional<br>oxygen saturation of arterial<br>hemoglobin (% SpO2) and<br>pulse rate.<br>It is intended for spot-<br>checking of adult patients<br>with wrist between 13-22<br>cm circumferences in<br>hospitals, clinics, long-term<br>care, and home use. | The Nonin Onyx Vantage 9590<br>Finger Pulse Oximeter is a small,<br>lightweight, portable device indicated<br>for use in measuring and displaying<br>functional oxygen saturation of<br>arterial hemoglobin (%Sp02) and<br>pulse rate of patients who are well or<br>poorly perfused. It is intended for<br>spot-checking of adult and pediatric<br>patients on digits, including the<br>thumb and toes, that are between 0.3<br>-1.0 inch (0.8 - 2.5 cm) thick. The<br>device's intended use environments<br>include hospitals, clinics, long-term<br>care facilities, skilled nursing<br>facilities, emergency medical<br>services, and home healthcare<br>services. | Indicated for<br>non-invasive spot-checking<br>and/or continuous<br>monitoring |
| Type of use | Spot checking | Spot-checking | Spot checking<br>Continuous use |
| Motion | Non-motion | Not specified | Non-motion<br>Motion |
| Patient Population | Adults | Adult<br>Pediatric | Adult<br>Pediatric |
| Perfusion | Well | Well<br>Poorly | Well<br>Poorly |
| Environment of Use | Hospitals, clinics, long-term<br>care and home use | Hospitals, clinics, long-term care<br>facilities, skilled nursing facilities,<br>emergency medical services, and<br>home healthcare services | Hospitals, long-term<br>care and home use |
| Technology | Reflectance | Transmissive | Transmissive<br>Reflectance |
{6}------------------------------------------------
#### 510(k) Summary May 8, 2017 Page 4 of 6
| CHARACTERISTICS | Oxitone 1000 | Predicate<br>Nonin<br>Model 9590 (K112843) | Reference<br>Nonin PalmSAT with<br>Forehead 8000R sensor<br>(K050056) |
|---------------------------------------|------------------------------------------------------------------------------------------|-------------------------------------------------------|-----------------------------------------------------------------------|
| System components / Configurations | | | |
| Batteries | Rechargeable Li-On | 2 x 1.5 AAA batteries | Batteries<br>Sensor has no power |
| SpO2 Display Range | 0% to 100% SpO2 | 0% to 100% SpO2 | The referenced sensor can<br>be placed on the forehead |
| Pulse rate declared<br>accuracy range | 30-250 BPM | 20-250 BPM | |
| Accuracy | | | Determined by monitor |
| SpO2 Accuracy Range | 70% to 100% $\pm$ 3% | 70% to 100% $\pm$ 3% | |
| Pulse rate | 30-250 $\pm$ 3 | 20-250 $\pm$ 2 | |
| Display | | | |
| LCD | Multi-pixel 3 digits | Multi-pixel 3 digits | |
| Pulse strength indicator | No pulse strength indicator.<br>LCD, readings or dashes<br>give pulse quality indication | Tricolor LED gives indication of<br>quality | |
| Application site | Wrist | Digits | Digits<br>Forehead |
| Data output | Front panel easy-to-read<br>display (LCD) | Front panel and USB | Monitor dependent |
| Operation mode | Spot checking | Spot checking | Spot checking<br>Continuous |
| LED wavelengths<br>(multiple) | 640-940 nm | 660 and 910 nm | 660-910 nm |
| Radiant Power | Red 1.05 mW IR 0.95 mW | Red 0.8 mW IR 1.2 mW | Determined by monitor |
| Type of protection | Internally powered | Internally powered | Class II |
| Degree of protection -<br>sensor | Type BF - applied part | Type BF - applied part | Type BF - applied part |
| Functional and safety<br>testing | ES 60601-1, IEC 60601-1-<br>11, ISO 80601-2-61 | IEC 60601-1, IEC 60601-1-11,, ISO<br>9919 | IEC ISO 9919 |
| Biocompatibility | Surface contact<br>Skin<br>Limited duration (<24 hrs) | Surface contact<br>Skin<br>Limited duration (<24 hrs) | Surface contact<br>Skin<br>Limited duration (<24 hrs) |
## Substantial Equivalence Discussion and Rationale
In Table 2 we have compared the Oxitone 1000 to the predicate and reference devices for equivalence of:
Indications - The Oxitone Model 1000 Pulse Oximeter is a small, wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate. It is intended for spot-checking of adult patients with wrist between 13-22 cm circumferences in hospitals, clinics, long-term care, and home use.
Discussion - The indications are similar to the predicate Nonin 9590 except the Nonin includes poor perfusion and pediatrics. The reference Nonin PalmSAT monitor, K050056, includes neonates but the use of the reference is for the Forehead sensor to demonstrate that other sensor location sites have been cleared.
Patient Population – Though the predicate Nonin 9590 is indicated for pediatrics. The Oxitone 1000 and predicate are both indicated for adults.
{7}------------------------------------------------
## 510(k) Summary May 8, 2017 Page 5 of 6
Environment of Use – The environment of use for the subject device is narrower than the predicate and reference, but can still be considered substantially equivalent.
Prescriptive – The Oxitone 1000 and Nonin 9590 predicate and reference Nonin PalmSAT with forehead sensor are prescriptive.
Design and Technology – The Oxitone 1000 and Nonin 9590 have equivalent technological and design features. They both calculate SpO2 and pulse by use of the same technology - the ratio of red and infrared signals of light propagated through the tissue between light sources and detector. Both devices use Photoplethysmography (PPG) signals of red and infrared light through tissue to calculate SpO2 and pulse rate
The Oxitone 1000 and Nonin 9590 differ in sensor placement, the Oxitone 1000 being placed on the wrist and the Nonin 9590 on a finger. This placement is similar to the reference device Nonin Purelight Forehead 8000R in that they are both based on reflected light (red and IR).
Oxitone's reflectance optical sensor so that the optical elements (LEDs and detector) are placed at the same side of the wrist thereby having the same optical path features are similar to the reflectance used for the forehead sensors.
Performance Specifications - The Oxitone 1000 and Nonin 9590 predicate have equivalent specifications. The reference sensor does not have listed specifications, its performance is based upon the monitor to which it is attached.
Compliance with standards The Oxitone 1000 complies with AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO-80601-2-61, IEC 62471, and IEC 62133. The predicate complies with IEC 60601-1-2 and ISO 9919. The Oxitone 1000 complies with all applicable currently recognized standards.
## Performance Testing
## Non-clinical
Bench - We have performed bench tests and found that the Oxitone 1000 met all requirements specifications and standards requirements and was found to be equivalent in comparison to the predicate. Testing includes the following:
- . Verification Testing
- AAMI ES 60601-1 ●
- IEC 60601-1-2,
- IEC 60601-1-11
- . IEC 80601-2-61
The results demonstrate that the devices perform as intended are substantially equivalent to the performance of the predicate and in accordance with applicable standards.
Biocompatibility - The patient contacting materials of the Oxitone 1000 have been tested in accordance with ISO 10993-1 and FDA Guidance. The tests included Cytotoxicity, Sensitization, and Irritation, which is similar to the predicate and reference devices.
Clinical Testing - Testing to insure clinical accuracy of the device in accordance with ISO 80601-2-61 was performed. This testing was performed on 10 patients with results showing compliance to the standard.
{8}------------------------------------------------
## 510(k) Summary May 8, 2017 Page 6 of 6
## Differences –
The identified difference:
- Narrower environment of use .
- . Narrower patient population - adults only
- Narrower indications non-motion only .
- Detection on the wrist vs. digits of forehead .
Have been evaluated and tested and confirm that these differences do not raise different questions of safety or effectiveness when compared to the predicate and reference devices for the proposed indications for use.
#### Substantial Equivalence Conclusion
The Oxitone 1000 is substantially equivalent to the predicate in: indications for use, patient population, environment of use, technology characteristics, materials, specifications / performance and compliance with international standards. The differences do not raise different questions of safety or effectiveness when compared to the predicate and reference devices for the proposed indications for use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.