MOJAVE Expandable Interbody System

K163364 · K2m · MAX · Mar 15, 2017 · Orthopedic

Device Facts

Record IDK163364
Device NameMOJAVE Expandable Interbody System
ApplicantK2m
Product CodeMAX · Orthopedic
Decision DateMar 15, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MOJAVE Expandable Interbody System implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy. Additionally, the MOJAVE lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. MOJAVE lumbar implants are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine.

Device Story

MOJAVE Expandable Interbody System consists of expandable titanium implants; designed for lumbar spine fusion. Implants feature titanium endplates to facilitate bone ingrowth; adjustable intraoperatively to match patient vertebral anatomy. Used by surgeons in clinical settings as an adjunct to fusion; requires supplemental spinal fixation systems cleared for lumbosacral use. Benefits include anatomical fit and stabilization of vertebral segments in patients with DDD or degenerative scoliosis.

Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F2077 (static/dynamic compression, torsion, compression shear) and ASTM F2267 (subsidence, expulsion). Bacterial endotoxin testing (LAL) conducted per ANSI/AAMI/ST72:2011. No clinical data provided.

Technological Characteristics

Expandable interbody fusion implants. Materials: medical grade titanium alloy (ASTM F1472, ASTM F136, ASTM F3001) and cobalt chromium alloy (ASTM F1537). Features titanium endplates for bone ingrowth. Mechanical testing per ASTM F2077 and ASTM F2267.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1) or degenerative scoliosis. DDD defined as discogenic back pain with radiographic degeneration; may include up to Grade I spondylolisthesis or retrolisthesis. Requires six months of prior nonoperative therapy. Used with autograft/allogenic bone graft and supplemental spinal fixation.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a bird-like figure, possibly representing an eagle or other national symbol, with three overlapping profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 15, 2017 K2M Nancy Giezen Manager Regulatory Affairs 600 Hope Parkway SE Leesburg, Virginia 20175 Re: K163364 Trade/Device Name: MOJAVE Expandable Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: February 9, 2017 Received: February 10, 2017 Dear Ms. Giezen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K163364 #### Device Name MOJAVE Expandable Interbody System #### Indications for Use (Describe) The MOJAVE Expandable Interbody System implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of carcellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy. Additionally, the MOJAVE lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. MOJAVE lumbar are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary MOJAVE Expandable Interbody System K2M, Inc. #### Submitter K2M, Inc. 600 Hope Pkwy SE Leesburg, VA 20175 Contact Person: Nancy Giezen Telephone: (571) 919-2000 Date Prepared: 3/13/2017 #### Classification | Trade Name: | MOJAVE Expandable Interbody System | |-------------------|----------------------------------------------| | Common Name: | Intervertebral Fusion Device with Bone Graft | | Regulatory Class: | Class II | Classification Name(s): Intervertebral Fusion Device with Bone Graft, lumbar (21 CFR 888.3080, Product Code: MAX) #### Predicate Device(s) Primary Predicate: Globus Caliber (K102293) Additional Predicates: K2M Aleutian (K082698, K133614) K2M Cascadia (K150481, K160125) ## Device Description The MOJAVE Expandable Interbody System is comprised of expandable titanium implants designed to allow for intraoperative adjustment to aid the surgeon in matching implant fit to the vertebral anatomy in the lumbar spine. The implants have titanium endplates designed to allow for bone ingrowth and engagement with the vertebral body end plates. The implants are manufactured from medical grade titanium alloy (ASTM F1472, ASTM F136, ASTM F3001) and cobalt chromium alloy (ASTM F1537) and are available in a variety of heights to accommodate anatomical variations. #### Indications for Use The MOJAVE Expandable Interbody System implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have six months of {4}------------------------------------------------ nonoperative therapy. Additionally, the MOJA VE lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. MOJAVE lumbar implants are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine. #### Technological Comparison to Predicate(s) The MOJAVE Expandable Interbody System was compared to predicate systems and were found to be comparable to these systems in design, function, intended use, materials, and size. #### Non-clinical Performance Evaluation The worst MOJAVE Expandable Interbody System implants were tested in static compression, torsion, and compression shear; dynamic compression shear (per ASTM F2077); subsidence (per ASTM F2267) and expulsion and performed similarly to or better than predicate devices. In addition, bacterial endotoxin testing (BET), also known as limulus amebocyte lysate (LAL) testing, was conducted in accordance with ANSI/AAMI/ST72:2011 #### Conclusion There are no significant differences between the MOJAVE spacers and other devices currently being marketed which would adversely affect the use of the product. Therefore the MOJAVE Expandable Interbody System implants are substantially equivalent to predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...