K163213 · Ge Medical Systems, LLC · JAK · Dec 16, 2016 · Radiology
Device Facts
Record ID
K163213
Device Name
Revolution CT
Applicant
Ge Medical Systems, LLC
Product Code
JAK · Radiology
Decision Date
Dec 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different angles and the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, liver, kidney, pancreas, etc. The system may acquire data using Axial, Cine, Helical, Cardiac, and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further, the images can be post processed to produce additional imaging planes or analysis results. The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications. The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy. If the spectral imaging option is included on the system can acquire CT images using different kV levels of the same anatomical region of a patient in a single rotation from a single source. The differences in the energy dependence of the attenuation coefficient of the different materials provide information of body materials. This approach enables images to be generated at energies selected from the visualize and analyze information about anatomical and pathological structures. GSI provides information of the chemical composition of renal calculation and graphical display of the spectrum of effective atomic number. GSI Kidney stone characterization provides addin the characterization of uric acid versus nonuric acid stones. It is intended to be used as an adjunct to current standard methods for evaluating stone etiology and composition.
Device Story
Revolution CT is a 256-detector row CT scanner; includes gantry, patient table, console, and cabinet. Incorporates Gemstone Spectral Imaging (GSI) option; utilizes fast kVp switching x-ray source and ultra-fast Gemstone detector to acquire dual-energy data. Used in clinical environments by radiologists/technicians for head, body, cardiac, and vascular imaging. Transforms x-ray transmission data into cross-sectional images via computer reconstruction. GSI enables material decomposition (water, iodine, uric acid, fat, hydroxyapatite), monochromatic images, and virtual unenhanced (VUE) images. Includes Metal Artifact Reduction (MAR) algorithms. Output provides diagnostic information for disease/trauma assessment and therapy planning. Benefits include improved material differentiation, artifact reduction, and optimized contrast-to-noise ratio.
Clinical Evidence
Clinical assessment involved 5 subjects at two sites (US/Canada). Images evaluated by 6 board-certified radiologists using a 5-point Likert scale for diagnostic acceptance, image quality, MAR performance, and VUE iodine suppression. Results demonstrated acceptable diagnostic imaging performance.
Technological Characteristics
Multi-slice (256 detector row) CT scanner. Features fast kVp switching x-ray source and Gemstone detector. Connectivity: networked. Software: GSI projection-based material decomposition, MAR, and VUE algorithms. Standards: IEC 60601-1 Ed. 3, NEMA XR-25, XR-26, XR-28, XR-29, 21 CFR Subchapter J.
Indications for Use
Indicated for head, whole body, cardiac, and vascular CT imaging in patients of all ages. Used for diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy. Spectral imaging option aids in material characterization, including renal calculi (uric acid vs. non-uric acid).
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2016
GE Medical Systems, L.L.C. % Ms. Helen Peng Regulatory Affairs Manager 3000 N. Grandview Blvd. WAUKESHA WI 53188
Re: K163213 Trade/Device Name: Revolution CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: November 15, 2016 Received: November 16, 2016
Dear Ms. Peng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
## K163213
Device Name Revolution CT
#### Indications for Use (Describe)
The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different ans the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, liver, kidney, pancreas, etc. The system may acquire data using Axial, Cine, Helical, Cardiac, and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further, the images can be post processed to produce additional imaging planes or analysis results
The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
If the spectral imaging option is included on the system can acquire CT images using different kV levels of the same anatomical region of a patient in a single rotation from a single source. The differences in the energy dependence of the attenuation coefficient of the different materials provide information of body materials. This approach enables images to be generated at energies selected from the visualize and analyze information about anatomical and pathological structures.
GSI provides information of the chemical composition of renal calculation and graphical display of the spectrum of effective atomic number. GSI Kidney stone characterization orovides addin the characterization of uric acid versus nonuric acid stones. It is intended to be used as an adjunct to current standard methods for evaluating stone etiology and composition.
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESSBELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMBnumber."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white swirls around the letters. The logo is simple and recognizable.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
| Date: | November 15th, 2016 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems, LLC (GE Healthcare<br>3000 N. Grandview Blvd.<br>Waukesha, WI 53188 |
| Primary Contact: | Helen Peng<br>Senior Regulatory Affairs Manager<br>GE Medical Systems, LLC<br>Tel: 262-424-8222<br>e-mail: hong.peng@ge.com |
| Secondary Contact: | John Jaeckle<br>Chief Regulatory Affairs Strategist<br>GE Medical Systems, LLC<br>Tel: 262-424-9547<br>e-mail: john.jaeckle@ge.com |
| | Laura Mixon |
Regulatory Affairs Leader GE Medical Systems, LLC Tel: 262-548-4588 e-mail: laura.mixon@ge.com
### PRODUCT IDENTIFICATION
Device Trade Name:
Revolution CT
Regulation Name: Regulation: Classification : Product Code: Manufacturer / Design Location:
Computed Tomography X-ray System 21CFR 892.1750 Class II JAK GE Medical Systems, LLC (GE Healthcare) 3000 N. Grandview Blvd. Waukesha, WI 53188
Manufacturing Location(s):
GE Medical Systems, LLC (GE Healthcare) 3000 N. Grandview Blvd. Waukesha, WI 53188
{5}------------------------------------------------
# GE Healthcare
510(k) Premarket Notification Submission- Revolution CT
Image /page/5/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle is surrounded by a series of curved lines, resembling water droplets or stylized flourishes, also in blue. The logo is simple, clean, and recognizable.
### Marketed Devices:
Revolution CT with GSI option is a CT device with modifications made to the predicate device Revolution CT (K133705) to enable Gemstone Spectral Imaging (GSI) option. It is of comparable type and substantially equivalent to its predicate device and GE's other currently marketed Computed Tomography X-ray Systems that comply with the same standards. This 510(k) submission for the Revolution CT system includes the same intended use and substantially equivalent indications for use as its predicate device, the GE Revolution CT (K133705) and reference device GE Discovery CT750 HD (K120833). The existing indications for use of Revolution CT are expanded to include the GSI capability, which is verbatim copy from that of the reference device Discovery CT 750HD. The system is labeled as the Revolution CT.
### Predicate Device:
| Device Name: | Revolution CT |
|------------------|-----------------------------------------------------------|
| 510(k) number: | K133705 cleared 4/11/2014 |
| Manufacturer: | GE Medical Systems, LLC (same as that of proposed device) |
| Regulation Name: | Computed Tomography X-ray System |
| Regulation: | 21CFR 892.1750 |
| Classification: | Class II |
| Product Code: | JAK |
### Reference Device:
| Device Name: | Discovery CT750 HD |
|------------------|-----------------------------------------------------------|
| 510(K) number: | K081105 cleared 5/9/2008, K120833 cleared 6/12/2012 |
| Manufacturer: | GE Medical Systems, LLC (same as that of proposed device) |
| Regulation Name: | Computed Tomography X-ray System |
| Regulation: | 21CFR 892.1750 |
| Classification: | Class II |
| Product Code: | JAK |
## Device Description: Revolution CT with GSI Option
The Revolution CT is a multi-slice (256 detector row) CT scanner consisting of a gantry, patient table, scanner desktop (operator console), system cabinet, power distribution unit (PDU), and interconnecting cables. The system includes image acquisition hardware, image acquisition and reconstruction software, and associated accessories.
GE has modified the cleared Revolution CT (K133705) within our design controls to include the Gemstone™ Spectral Imaging (GSI) Option. GSI is the state-of-the-art single source, projection-based, spectral CT application. It is GE's unique dual energy design and implementation which offers clear advantage over traditional dual source Dual Energy implementation. This feature has been previously cleared on Discovery CT750 HD (K081105, K120833) and it is fundamentally the same technology on Revolution CT. Revolution CT however offers a few technology improvements to enable Volume GSI with up to 80mm GSI zcollimation, 245mm/s GSI volumetric scan speed, dose neutrality and more improved workflow to support GSI in routine scanning.
{6}------------------------------------------------
Image /page/6/Picture/15 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are also some white swirls around the letters. The logo is simple and recognizable.
510(k) Premarket Notification Submission- Revolution CT
The GSI Option on Revolution CT is also commercially referenced as GSI Xtream or Volumetric GSI.
The Revolution CT with GSI is designed to be a head and whole body powerful Volume High Definition CT scanner that is designed to provide best-in-class technologies for whole organ coverage, high image quality and responsible dose performance. Revolution CT with GSI option incorporates the same basic fundamental operating principles as the existing marketed products Revolution CT and Discovery CT750 HD.
### Intended Use
The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.
### Indications for Use:
The system is intended to produce cross-sectional images of the body by computer reconstruction of x-rav transmission data from the same axial plane taken at different angles. The system has the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, liver, kidney, pancreas, etc. The system may acquire data using Axial, Cine, Helical, Cardiac, and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of transaxial and reformatted planes. Further, the images can be post processed to produce additional imaging planes or analysis results
The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
If the spectral imaging option is included on the system, the system can acquire CT images using different kV levels of the same anatomical region of a patient in a single rotation from a single source. The differences in the energy dependence of the attenuation coefficient of the different materials provide information about the chemical composition of body materials. This approach enables images to be generated at energies selected from the available spectrum to visualize and analyze information about anatomical and pathological structures.
GSI provides information of the chemical composition of renal calculi by calculation and graphical display of the spectrum of effective atomic number. GSI Kidney stone characterization provides additional information to aid in the characterization of uric acid versus non-uric acid stones. It is intended to be used as an adjunct to current standard methods for evaluating stone etiology and composition.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by four teardrop-shaped elements, also in blue, positioned symmetrically around the circle.
## Technology of GSI Feature:
Gemstone Spectral Imaging is GE's unique implementation of the dual energy scan modes that utilizes fast kVp switching x-ray source (switching between two different energy levels of X-rays from view to view during a single rotation) and ultra-fast response Gemstone detector to acquire almost perfectly registered dual energy CT data. The data is then processed through projection domain material decomposition algorithms to generate material density maps (MD), monochromatic imaqes (MC) and virtual unenhanced images (VUE). This data can be utilized to identify material specific differences in attenuation in terms of Water, Iodine, Uric Acid, Fat and Hydroxyapatite (HAP) basis-pair images, allowing monochromatic and material representations. Metal Artifact Reduction (MAR) algorithms can also be applied to all GSI images to reduce artifacts due to the presence of metal.
GSI can provide:
- excellent temporal and spatial registration to avoid mis-registration artifacts due to ● motion in dual energy CT
- . Advanced material differentiation, classification and quantification
- Optimization of contrast-to-noise ratio (CNR)
- Reduction in artifacts due to beam hardening and metal.
### Determination of Substantial Equivalence:
The main change in the proposed device for this submission is the addition of the GSI capability. The table below summarizes the substantive feature/technological differences between the predicate device, the reference device and the proposed device:
| Features | Discovery CT750 HD<br>(Reference Device,<br>K081105, K128033) | Revolution CT<br>(Predicate Device,<br>K133705) | Revolution CT with GSI<br>Option<br>(Proposed Device) |
|----------|---------------------------------------------------------------|-------------------------------------------------|-------------------------------------------------------|
|----------|---------------------------------------------------------------|-------------------------------------------------|-------------------------------------------------------|
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. The letters are stylized with curved lines, giving them a classic and recognizable appearance.
# GE Healthcare
510(k) Premarket Notification Submission- Revolution CT
| Features | Discovery CT750 HD<br>(Reference Device,<br>K081105, K128033) | Revolution CT<br>(Predicate Device,<br>K133705) | Revolution CT with GSI<br>Option<br>(Proposed Device) |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Scan Modes | Scout<br>Axial<br>Helical<br>Cine<br>Cardiac<br>Gated<br>High Definition<br>Interventional (Step+Shoot)<br>Interventional(SmartView)<br>GSI<br>GSI Cardiac | Scout<br>Axial<br>Helical<br>Cine<br>Cardiac<br>Gated<br>High Definition | Scout<br>Axial<br>Helical<br>Cine<br>Cardiac<br>Gated<br>High Definition<br>Interventional(Step+Shoot)<br>GSI |
| GSI | Axial up to 40 mm<br>Helical up to 40 mm<br>Cardiac up to 40mm<br>High Res | No | Axial up to 40 mm<br>Helical up to 80mm |
| GSI MAR<br>(Metal Artifact<br>Reduction) | Yes | No | Yes |
| GSI VUE<br>(Virtual<br>UnEhnahced) | Yes | No | Yes |
| SmartStep | Yes | No | Yes |
| High Pitch<br>Helical<br>(hyper Drive) | No | No | Yes |
The Revolution CT with GSI option has completed testing and is in compliance with IEC 60601-1 Ed. 3 and its associated collateral and particular standards, 21CFR Subchapter J, and NEMA XR-25, XR-26, XR-28 and XR-29. The proposed device has successfully completed all testing per our quality system as well as comparison testing to the predicate device. It was designed and is manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- · Required Reviews
- · Design Reviews
- · Testing on unit level (Module verification)
- · Integration testing (System verification)
- · Performance testing (Verification)
- · Safety testing (Verification)
- · Simulated use testing (Validation)
GE believes the Revolution CT system is of comparable type and substantially equivalent to our currently marketed system Revolution CT (K133705) and Discovery CT750 HD (K120833).
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are small white flourishes around the edge of the circle. The logo is simple and recognizable.
The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.
### Summary of Additional Testing
In addition to the verification and validation testing successfully completed as required by GE Healthcare's quality system, additional engineering (non-Clinical testing) and clinical performance testing was performed to provide the requisite data to substantiate performance claims, the revised indications, and ultimately substantial equivalence.
### Non-Clinical Testing
The performance evaluation testing used a variety of phantoms representing head and body, and clinical datasets to provide technical substantiation of the Gemstone Spectral Imaging (GSI) performance on the Revolution CT. Various mathematical and statistical analyses were performed to demonstrate that each performance item was successfully verified and substantiated for data acquisition and optimized image generation.
### Clinical Testing
Sample clinical data was collected from 51 subjects at two sites (one in the US and one in Canada) with the approval of appropriate ethics committee and in accordance with applicable US and Health Canada regulations as well as GE Healthcare's quality system's procedures for such evaluations.
The intent of the study was to obtain a sample set of clinical images across different patient populations, clinical scenarios, and scanning protocols/techniques and to evaluate these images for image quality related to diagnostic use, reduction of metal artifacts using the MAR algorithm, and suppression of iodine in contrast enhanced acquisitions using VUE algorithm. Patients were selected for potential recruitment to meet these needs. Any patient who met these criteria stated in the Protocol and who voluntarily signed the Informed Consent Form was recruited.
This sample image data was representative of a wide range of anatomical coverage and patient indications, and was representative of how the Revolution CT with GSI may be used in a clinical environment. The images were evaluated by 6 board certified and qualified radiologists at different institutions in the United States of America for clinical acceptance and image quality using a 5 point Likert scale. Each data set was read by three different radiologists depending on their area of expertise.
The results of this clinical assessment demonstrate the acceptable diagnostic imaging performance of the GE Healthcare Revolution CT with GSI option.
### Substantial Equivalence Conclusion:
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by decorative swirls, giving it a classic and recognizable appearance.
# GE Healthcare
510(k) Premarket Notification Submission- Revolution CT
Based on the conformance to standards, development under our quality system, and the engineering and clinical testing provided, GE Medical Systems believes that the Revolution CT with GSI option is as safe and effective, and performs in a substantially equivalent manner to the predicate device Revolution CT (K133705).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.