K163204 · Kunshan Aoshida Electric Technology Co., Ltd. · ITI · May 5, 2017 · Physical Medicine
Device Facts
Record ID
K163204
Device Name
A08 Power wheelchair
Applicant
Kunshan Aoshida Electric Technology Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
May 5, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
The device is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
A08 power wheelchair is a battery-operated, motor-driven mobility device for indoor/outdoor use. It features a foldable welded-aluminum frame, two 250W DC 24V transaxle motors, and an electromagnetic braking system. The rider operates the device via a joystick-controlled electric motor controller (Yisheng Electric WS-1) mounted on an armrest to manage velocity and direction. Releasing the joystick automatically engages the electromagnetic brake to slow the device to a stop. It is powered by two 12V Li-ion batteries (20Ah) with an off-board charger. The device includes anti-tip wheels and a horn. It provides mobility for disabled or elderly users who are limited to a seated position. The healthcare provider or caregiver does not operate the device during use; the patient controls it directly. The device benefits the patient by providing independent mobility.
Clinical Evidence
Bench testing only. The device complied with ISO 7176 series (1-16, 21, 25) for powered wheelchairs, ISO 10993-1/5/10 for biocompatibility, ISO 14971 for risk management, and IEC 62133 for battery safety.
Technological Characteristics
Welded-aluminum frame; PU foam/nylon fabric seating; two 250W DC 24V transaxle motors; electromagnetic braking; Li-ion battery power (24V, 20Ah); joystick controller (Yisheng WS-1); non-pneumatic PU tires; 100kg weight capacity; max speed 7 km/hr; off-board charger (2A). Standards: ISO 7176 series, ISO 10993, ISO 14971, IEC 62133.
Indications for Use
Indicated for disabled or elderly persons limited to a seated position requiring mobility assistance.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is depicted in black and white, and the text is in a simple, sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2017
Kunshan Aoshida Electric Technology Co., Ltd % Mrs. Junnata Chang Official Correspondent IRC China 16f-2(16a), No. 462, Sec. 2, Chongde Rd., Beitun Dist. Taichung, 406 TW
Re: K163204
Trade/Device Name: A08 Power Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: March 30, 2017 Received: April 13, 2017
Dear Mrs. Junnata Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163204
Device Name A08 Power Wheelchair
Indications for Use (Describe)
The device is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Applicant
Name: Kunshan Aoshida Electric Technology Co., Ltd
Address: No. 65, Matang Road, Kunshan City, Jiangsu Province, 215333, China
Contact person: Shen Wenhua
Phone: +86-512-57816288
Fax: +86-512-57816388
Date prepared: Jan. 20, 2017
#### Device identification
- Trade name : A08 power wheelchair
- Common name: Powered wheelchair
- Classification name: Powered wheelchair
- Medical specialty (Panel): Physical Medicine Device
- Regulation number: 890.3860
- Product code: ITI
- Classification: Class II
#### Predicate devices
Trade name: PL001 power wheelchair
510(k) number: K113463
Manufacture: SUZHOU KD MEDICAL APPLIANCE CO. LTD.
Regulation number: 890.3860
Product Code: ITI
Classification: Class II
#### Intend use of device
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
#### Device description:
The A08 power wheelchair is an indoor/outdoor powered wheelchair that is battery operated. The design of this wheelchair is basically similar to other powered wheelchairs that are already on the market.
The A08 power wheelchair is with a 100 kg (220 lbs) weight capacity. It is basic
{4}------------------------------------------------
conventional rear wheels drive, rigid frame vehicle that are battery powered. It consists primarily of a foldable welded-aluminum frames, two sealed transaxle motors (250 W, DC 24V) drive system, electromagnetic braking system, electric motor controller (Yisheng Electric Co., Ltd., Model: WS-1) and two Li-ion batteries with an off-board battery charger (2 A). It is powered by two 12 volt Li-ion DC batteries with 20 km (12.5 miles) with 20 Ah which maximum speed upto 7 km/hr (4.4 mph). The controller with a joystick attaches to either armrest and allows the rider to control the movement and velocity of the powered wheelchair. When the joystick is released, the electromagnetic brake will be actuated and the power wheelchair is slow to stop.
#### Summary of non-clinical testing
The A08 power wheelchair complied with the requirements of the ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9. ISO 7176-10. ISO 7176-11. ISO 7176-13. ISO 7176-14. ISO 7176-15. ISO 7176-16, ISO 7176-21, ISO 7176-25, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 14971 and IEC 62133.
#### Statement of substantial equivalence
The A08 power wheelchair is substantially equivalent to the PL001 (K113463) manufactured by SUZHOU KD MEDICAL APPLIANCE CO. LTD.. They have same intended use of a motor driven, indoor and outdoor transportation vehicle to provide mobility to a disabled or elderly person limited to a seated position.
The design and technological characteristics of this device is basically similar to the predicate device. They are 4-wheeled power wheelchair with basic conventional rear wheel drive which with free-wheel mode. The motor of both devices is equipped with differential mechanism. The anti-tip wheels are equipped behind the rigid frame of both devices. The power supply of both devices is Li-ion batteries to provide power that can be recharged by an off-board charger that can be plugged into an AC outlet when the devices are not in use. The horn is also equipped on both devices have the same user interface, the patient uses the joystick to engage and disengage the power wheelchair motion in both the forward and reverse directions. When the joystick is released, the electromagnetic brake will be actuated and the power wheelchair is slow to stop.
Both seats of the devices are equipped with removable flip-up armrest and there is no headrest on backrest. Seat cushions and backrests of both devices used the same materials, which is PU foam covered by nylon fabric cloth. The seat, seat cushion, backrest, joystick of the device will contact to the patient. The materials of the patient contacting parts used for the A08 power wheelchair and predicate device are exactly
{5}------------------------------------------------
same and no additional processing different from predicate device, so no new safety and efficacy concerning.
The main difference for the two devices is the rated power of the motor, maximum curb height and the maximum speed. While there are minor differences between the devices including max load capacity, dimension, wheels size, braking distance and pre-charging distance, do not alter the intended use function and use of the device. Moreover, the non-clinical tests and the predicate comparisons demonstrate that these differences in their technological characteristics do not raise any questions as to the safety and effectiveness.
#### Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Kunshan Aoshida Electric Technology Co., Ltd concludes that, A08 power wheelchair is substantially equivalent to predicate devices as described herein.
{6}------------------------------------------------
| | PL001 (K113463) | A08 | Comparison to the<br>predicate device |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------------------|
| Indications for use | They are motor driven, indoor and outdoor transportation vehicle<br>with the intended use to provide mobility to a disabled or elderly<br>person limited to a seated position. | | Same indications for use |
| Maximum load capacity | 114 kg (251 lbs) | 100 kg (220 lbs) | Smaller load capacity |
| Overall length | 900 mm (35.4") | 890 mm (35") | Smaller overall length |
| Overall width | 600 mm (23.6") | 603 mm (23.7") | Larger overall width |
| Stowage length | 106 cm(41.7") | 324 mm(12.8") | Larger stowage length |
| Stowage width | 510 mm(20.1") | 603 mm(23.7") | Larger stowage width |
| Stowage height | 450 mm(17.7") | 670 mm(26.4") | Larger stowage height |
| Rising | 150 mm(5.9") | 40 mm(1.57") | Smaller rising |
| Total mass: | with battery : 26 kg (57.2 lbs)<br>without battery : 23 kg(50.6 lbs) | with battery : 28 kg(61.7 lbs)<br>without battery : 24 kg(53 lbs) | Heavier mass |
| Mass of heaviest part | 23 kg (50.6 lbs) body frame<br>with motor | 24 kg(53 lbs) body frame with<br>motor | Heavier mass of heaviest part |
| Pivot width | 250 cm(98.4") | 150.5 cm(59.3") | Smaller pivot width |
| Reversing width | 160 cm(63") | 233 cm(91.7") | Larger reversing width |
| Turning diameter | 160 cm(63") | | Same turning diameter |
| Ground clearance | 60 mm(2.4") | 63.5 mm(2.5") | Larger ground clearance |
| Required width of angled corridor | 120 cm(47.2") | 100 cm(39.4") | Smaller required width of<br>angled corridor |
| Required doorway entry depth | 106 cm(41.7") | 160 cm(63") | Larger required doorway entry<br>depth |
| Required corridor width for side<br>opening | 190 cm(74.8") | 120 cm(47.2") | Smaller required corridor<br>width for side opening |
| (Continuous)<br>The substantial equivalence comparison of the A08 and PL001 (K113463) | | | |
| | PL001 (K113463) | A08 | Comparison to the<br>predicate device |
| Seat plane angle | 5° | 2° | Smaller seat plane angle |
| Effective seat depth | 350 mm(13.8") | 430 mm (17") | Larger effective seat depth |
| Seat width | 350 mm(13.8") | 450 mm (17.7") | Larger seat width |
| Effective seat width | 307 mm ~ 430 mm(12" ~ 17") | 450 mm(17.7") | Larger effective seat width |
| Seat surface height at front edge | 440 mm~516 mm(17.3"~20.3") | 470 mm(18.5") | Larger seat surface height at<br>front edge |
| Back support angle | 12° | 20° | Larger back support angle |
| Back support height | 395 mm(15.6") | 470 mm(18.5") | Smaller back support height |
| Back support width | 350 mm(13.8") | 440 mm (17.3") | Larger back support width |
| Foot support to seat distance | 345 ~ 421 mm(13.6" ~ 16.6") | 400 mm(15.7") | Predicate device is adjustable |
| Foot support clearance | 70 mm(2.8") | 55 mm(2.2") | Smaller foot support clearance |
| Arm support height | 155 mm(6.1") | 200 mm(7.9") | Larger arm support height |
| Front of arm support to back<br>support | 250 mm(9.8") | 350 mm(13.8") | Larger front of arm support to<br>back support |
| Arm support length | 260 mm(10.2") | 335 mm(13.2") | Larger arm support length |
| Arm support width | 50 mm(2") | | Same arm support width |
| Arm support angle | 0° | 12° | Larger arm support angle |
| Distance between arm supports | 420 mm(16.5") | 480 mm(18.9") | Larger distance |
| Horizontal location of the wheel<br>axle | 219 mm(8.6") | 120 mm(4.7") | Smaller horizontal location of<br>the wheel axle |
| Vertical location of the wheel axle | 284 mm(11.2") | 320 mm(12.6") | Larger vertical location of the<br>wheel axle |
| Caster wheel diameter | 200 mm(8") | 176 mm(7") | Smaller diameter |
| (Continuous) | The substantial equivalence comparison of the A08 and PL001 (K113463) | | |
| | PL001 (K113463) | A08 | Comparison to the<br>predicate device |
| Material | | | |
| Seat<br>Seat cushion<br>Backrest | PU foam covered by nylon fabric cloth | | Same material |
| Differential mechanism | Differential rate: 22 : 1 | | Same differential rate |
| Free-wheel mode | Yes | | Both devices have free- wheel<br>mode |
| Lights | | | |
| Head/Tail lights | No | | Both devices do not have lights |
| Signal light | No | | |
| Warning light | No | | |
| Horn | Yes | | Both devices have horn |
| Anti-tip wheels | Yes | | Both devices have anti-tip<br>wheels |
| Armrest | Removable flip-up armrest | | Both devices have adjustable<br>armrest |
| Headrest | No | | Both devices do not have<br>headrest |
| Motor | | | |
| Rated power | 150 W | 250 W | Larger rated power |
| Input voltage | DC 24V | | Same input voltage |
| Amount | 2 Pcs | | Same amount |
| Brake system | Intelligent regenerative electromagnetic brake | | Same brake system |
| (Continuous) | The substantial equivalence comparison of the A08 and PL001 (K113463) | | |
| | PL001 (K113463) | A08 | Comparison to the predicate device |
| Controller | Yisheng Electric Co. Ltd , WS-1, 40A | | Same controller |
| Rear wheel drive | Sealed transaxle direct drive | | Same drive mode |
| Braking distance | Forward:1.5 m (59") at max speed | Forward: 1 m (39.4") at max speed | Smaller braking distance |
| Per-charge distance | Up to 20 km (12.5 miles) | | Same pre-charging distance |
| Maximum speed | Up to 6 km/h (3.75 mph) , variable | Up to 7 km/hr (4.4 mph), variable | Larger max. speed |
| Maximum curb height | 30 mm (1.2") | 40 mm(1.57") | Larger curb height |
| Battery | | | |
| Type | Li-ion, Rechargeable | | Same type |
| Rated power | 20 Ah | | Same rated power |
| Output voltage | 24 VDC | | Same output voltage |
| Amount | DC 12V x 2 Pcs | | Same amount |
| Battery level indicator | Yes | | Both devices have battery level indicator |
| Charger | | | |
| Type | off-board, Automatic Type | | Same type |
| Main plug | Standard US-plug, 2 pins | | Same type |
| Input power requirement | 110-220 V / 50-60 Hz | | Same input power requirement |
| Output voltage / current | DC 24V / 2A | | Same output voltage / current |
| (Continuous) | The substantial equivalence comparison of the A08 and PL001 (K113463) | | |
| | PL001 (K113463) | A08 | Comparison to the predicate device |
| Front wheels | | | |
| Diameter | 8" × 2" | 7" × 2" | Smaller diameter |
| type | non-pneumatic tires | non-pneumatic tires | Same type |
| Material | Polyurethane (PU) | Polyurethane (PU) | Same materials |
| Amount | 2 | 2 | Same amount |
| Rear wheels | | | |
| Diameter | 8" × 2.4" | 10" × 2.35" | Larger diameter |
| type | non-pneumatic tires | non-pneumatic tires | Same type |
| Material | Polyurethane (PU) | Polyurethane (PU) | Same materials |
| Amount | 2 | 2 | Same amount |
The substantial equivalence comparison of the A08 and PL001 (K113463)
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.