Retrospective pediatric clinical cases (US and OUS)
Retrospective analysis of clinical cases was conducted to support the safety and short-term efficacy of the SurgiCase Orthopaedics system for the expanded pediatric indication.
Retrospective analysis; Pediatric clinical cases; Safety and efficacy
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis of US and OUS pediatric clinical cases; Retrospective analysis
Pediatric patients
Not applicable for this study
Safety and short-term efficacy
Indications for Use
The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to assist in pre-operative planning and/ or in guiding the marking of bone and/or guide surgical instruments in non-acute, non-joint replacing osteotomies · For adult patients; in upper extremity orthopedic surgical procedures and orthopedic surgical procedures around the knee. · For pediatic patients 7 years of age and older; in orthopedic surgical procedures involving the radius and ulna. SurgiCase Guides are intended for single use only.
Device Story
System comprises SurgiCase Connect software and SurgiCase Guides (patient-specific templates). Surgeon uses Connect software to visualize, measure, and edit medical data for pre-surgical simulation. Guides are designed/manufactured based on patient anatomy from pre-surgical plan. During surgery, guides assist in marking bone and directing cutting/drilling instruments. Stability Model algorithm identifies optimal base plate position on patient anatomy to ensure guide stability. Used in clinical settings for non-acute, non-joint replacing osteotomies. Benefits include improved surgical accuracy and planning. Pediatric use requires strict 3-week useful life limit between scan and surgery to account for bone growth.
Clinical Evidence
Retrospective analysis of US and OUS pediatric clinical cases supported safety and short-term efficacy. Non-clinical testing included 3D-printed pediatric bone model fit tests to evaluate guide stability against simulated growth. Bench testing confirmed biocompatibility, sterilization, and functional elements (drill sleeves, cutting slots) remain identical to the predicate.
Technological Characteristics
Patient-specific templates (SurgiCase Guides) and planning software (SurgiCase Connect). Guides feature drill sleeves, cutting slots, and fixation holes. Stability Model algorithm used for base plate positioning. Anatomy-independent design. Single-use. Software-driven design process.
Indications for Use
Indicated for adult patients undergoing upper extremity or knee-related orthopedic osteotomies, and pediatric patients (≥7 years) undergoing radius and ulna osteotomies.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 18, 2017
Materialise N.V. Oliver Clemens Regulatory Officer Technologielaan 15 Leuven, BE 3001 Belgium
Re: K163156
Trade/Device Name: Surgicase Orthopaedics, Surgicase Connect, Surgicase Guides Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: PBF Dated: November 4, 2016 Received: November 10, 2016
Dear Mr. Clemens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K163156
Device Name SurgiCase Orthopaedics system
The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to assist in pre-operative planning and/ or in guiding the marking of bone and/or guide surgical instruments in non-acute, non-joint replacing osteotomies
· For adult patients; in upper extremity orthopedic surgical procedures and orthopedic surgical procedures around the knee.
· For pediatic patients 7 years of age and older; in orthopedic surgical procedures involving the radius and ulna.
SurgiCase Guides are intended for single use only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
# 510(k) Summary
# Submitter information
| Company name | Materialise N.V. |
|-----------------------------------|-----------------------------------|
| Establishment registration number | 3003998208 |
| Street Address | Technologielaan 15 |
| City | Leuven |
| Postal code | 3001 |
| Country | Belgium |
| Phone number | +32 16 39 62 80 |
| Fax number | +32 16 39 66 06 |
| Contact name | Oliver Clemens |
| Contact title | Regulatory officer |
| Contact e-mail address | regulatory.affairs@materialise.be |
### Submission date
The date of the Traditional 510(k) submission is November 4th, 2016.
### Submission information
| Trade Name | SurgiCase Orthopaedics system, SurgiCase Connect, SurgiCase<br>Guides |
|---------------------|-----------------------------------------------------------------------|
| Common Name | Orthopaedic surgical planning and instruments guide |
| Classification Name | Orthopaedic surgical planning and instruments guide |
| Product code | PBF |
### Predicate devices
| Trade or proprietary or model name | SurgiCase Orthopaedics system |
|------------------------------------|-------------------------------|
| 510(k) number | K132290 |
| Decision date | 04/10/2014 |
| Product code | PBF |
| Manufacturer | Materialise N.V. |
# Device Information
#### Description and functioning of the device
The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to transfer a pre-surgical plan to surgeries involving osteotomies in upper extremity orthopedic surgical procedures and orthopedic surgical procedures around the knee.
For pediatric patients 7 years of age and older, it is intended to be used in osteotomies involving the radius and ulna.
The SurgiCase Orthopaedics system is composed of two components: SurgiCase Connect (software) and SurgiCase Guides (hardware).
SurgiCase Connect is a medical device for Materialise and a surgeon for pre-surgical simulation of surgical treatment options. This includes transferring, visualizing, measuring, and editing medical data.
{4}------------------------------------------------
The SurgiCase Guides are patient specific templates that are designed and manufactured based on a pre-surgical software plan for a specific patient. In surgery these guides are used to assist a surgeon in guiding the marking of bone and/or guiding surgical instruments to cut and drill according to the pre-surgical plan.
All guides are individually designed and manufactured for each patient using a design and manufacturing process with strict procedures and work instructions. Part of this process is a scientific Stability Model which measures the sensitivity of a guide to movement during surgery. The use of this Stability Model provides a way to find the most stable position of the base plate on the individual patient's anatomy for accurate guiding of surgical instruments. The Stability Model is anatomy independent, thus it can be applied to any bony structure in upper and lower extremity surgical procedures.
### Intended use
The SurgiCase Orthopaedics system is intended to be used as a surgical instrument to assist in pre-operative planning and/or in guiding the marking of bone and/or guide surgical instruments in non-acute, non-joint replacing osteotomies
- For adult patients; in upper extremity orthopedic surgical procedures and orthopedic surgical procedures around the knee.
- . For pediatric patients 7 years of age and older; in orthopedic surgical procedures involving the radius and ulna.
SurgiCase Guides are intended for single use only.
### Summary of technological characteristics
The proposed device has the same design, functionality, fundamental technology, materials, and software components as the predicate device (K13290). The only difference between the proposed device and the predicate device (K132290) is the intended patient population of the proposed device which has been expanded to the radius and ulna in pediatric patients 7 years of age and older. Since pediatric patients may not be fully grown, the guide might not adequately fit on the patient when too much time has passed between scan and surgery. Performance testing was conducted to address this concern.
### Performance data
#### Non-clinical testing
In terms of biocompatibility, sterilization dimensional stability, cleaning, packaging and shipment, the SurgiCase Guides are identical to the predicate device. No new tests on these topics are needed.
The functional elements on the guides, i.e. drill sleeves, cutting slots, fixation holes, remain identical to the predicate device when used on a pediatric patient. The only part of the guide that requires extra testing is the base plate: the guide might not fit on the patient when too much time has passed between scan and surgery as the anatomy of the pediatric patient is likely to be growing. This changing anatomy is not a new risk compared to the predicate device, but the anatomical changes for pediatric patients are more profound. Several risk reduction measures are in place to account for a changing anatomy. One of the risk reduction measures is the useful life. The useful life period depends on the age of the patient, and the surgical area. The useful life is determined based on a literature study, covering both growth in diameter of the bone, and fit tests. The goal of the fit tests was to determine the maximal allowed growth, which still ensures a good fit. The maximal allowed growth cannot be investigated by a sawbones or a cadaver lab, since those bones cannot grow. The only alternative is to simulate growth for pediatric bones coming from pediatric clinical cases. As such, a fit test is performed in which guides were placed on the grown, 3D-printed, pediatric bone models and evaluated.
Based on this maximal growth and the literature study the is calculated as the period that can pass between scan date and surgery date without posing any risks with respect to the guide fit.
{5}------------------------------------------------
Due to the lack of literature on bone growth rates in pediatric patients, many extrapolations were made to the worst-case growth rate calculations and thereby the useful life calculations. Because of these extrapolations, multiple safety factors were incorporated into the useful life calculations, and in addition the shortest calculated useful life amongst all age/gender categories, i.e., 3 weeks, is to be used in all indicated pediatric patients. With a useful life of 3 weeks, indicated pediatric patients are not expected to have bone growth that would affect the performance of the proposed device. The proposed device is expected to perform substantially equivalent in indicated pediatric patients to the predicate device (K132290).
Like the predicate device, the additional use of a scientific Stability Model provides a way to find the most stable position of the base plate on the individual patient's anatomy for accurate guiding of surgical instruments. This Stability Model is anatomy independent and thus is applicable for pediatric bones as well.
#### Clinical testing
Retrospective analysis of US and OUS pediatric clinical cases helped to further support the safety and short-term efficacy of the proposed device.
#### Summary
All non clinical testing and the retrospective analysis of clinical cases indicate that the subject device is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.