Olecranon Plating System

K162898 · Miami Device Solutions, LLC · HRS · Mar 17, 2017 · Orthopedic

Device Facts

Record IDK162898
Device NameOlecranon Plating System
ApplicantMiami Device Solutions, LLC
Product CodeHRS · Orthopedic
Decision DateMar 17, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The MDS Olecranon Plating System is intended for internal fixation of fractures of the Olecranon.

Device Story

Internal fixation system for Olecranon fractures; consists of pre-contoured, side-specific plates (2-hole and 4-hole), 2.7mm diameter screws, and locking caps. Used by orthopedic surgeons in clinical settings to stabilize bone fractures. Implants are single-use; system includes reusable and single-use instruments. Provides mechanical stabilization of bone fragments to facilitate healing.

Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical laboratory performance testing and engineering analysis comparing subject device to predicates.

Technological Characteristics

Materials: Titanium alloy and stainless steel. Components: Pre-contoured plates, 2.7mm screws, locking caps. Testing standards: ASTM F382-99 (bone plates), ASTM F543-07 (bone screws). Mechanical fixation principle.

Indications for Use

Indicated for internal fixation of fractures of the Olecranon.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 17, 2017 Miami Device Solutions, LLC Ms. Michelle Montesino Regulatory Affairs Associate 7620 NW 25th Street, Unit 3 Miami. Florida 33122 Re: K162898 Trade/Device Name: Olecranon Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: March 7, 2017 Received: March 9, 2017 Dear Ms. Montesino: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements {1}------------------------------------------------ as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K162898 Device Name Olecranon Plating System Indications for Use (Describe) The MDS Olecranon Plating System is intended for internal fixation of fractures of the Olecranon. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY | Submitter Name: | Miami Device Solutions, LLC | |-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter Address: | 7620 NW 25 Street, Unit 3;<br>Miami, FL 33122 | | Contact Person: | Michelle Montesino<br>Phone: (786) 422-1400 Ext. 106<br>Fax: (786) 422-1401 | | | Anderson Giraldo<br>Phone: (786) 422-1400 Ext. 115<br>Fax: (786) 422-1401 | | Date of Submission: | 3/7/2017 | | Manufacturer Name: | Miami Device Solutions, LLC | | Manufacturer Address: | 7620 NW 25th Street, Unit 3;<br>Miami, FL 33122 | | Registration Number: | 3009222247 | | Contact Name: | Markku Biedermann | | Title: | President | | Device Trade Name: | Olecranon Plating System | | Device Common Name: | Primary: Plate, Fixation, Bone<br>Secondary: Screw, Fixation, Bone | | Classification Names: | Primary: Plate, Fixation, Bone<br>Secondary: Screw, Fixation, Bone | | Classification Code: | Primary: HRS – Class II<br>Secondary: HWC – Class II | | Classification Panel: | Orthopedic | | Regulation Number: | Primary: 21 CFR section 888.3030 – Single/multiple component<br>metallic bone fixation appliances and accessories | | | Secondary: 21 CFR section 888.3040 – Smooth or threaded<br>metallic bone fixation fastener | | Predicate Device: | Primary: Acumed Small Fragment Base Set, K143394<br>Additional: Miami Device Solutions Distal Radius Plating<br>System, K161292<br>Additional: Synthes Variable Angle LCP Elbow System,<br>K120070 | {4}------------------------------------------------ #### Reference Devices: Paragon 28, Inc.; The Monster Screw System: Instrument Reprocessing Instructions for Reusable Instruments, K151418 # Device Description: The Olecranon Plating System is an internal fixation system to be used for the treatment of Olecranon fractures. The system consists of plates, screws, and locking caps, as well as single use and reusable instruments. The Olecranon plates are pre-contoured and side-specific, available in two different sizes; 2 hole and 4 hole. The Olecranon Plating System Screws are 2.7mm in diameter and available in various lengths. The MDS implants are single-use ONLY. Materials: Titanium alloy and stainless steel #### Intended Use: The MDS Olecranon Plating System is intended for internal fixation of fractures of the Olecranon. ### Summary of Technologies: The Olecranon Plating System has the same intended use, similar performance characteristics, and is similar in design and material to the predicate devices listed above with the exception of the Miami Device Solutions Distal Radius Plating System (K161292) which is also used for fixation of bone fragments but it is indicated for the distal radius. The Miami Device Solutions Distal Radius Plating System (K161292) is considered a predicate because it utilizes the same screws and locking technology as the subject device. # Performance Data: #### Non-Clinical Performance and Conclusions: The results of non-clinical (laboratory/performance) testing as well as engineering analysis for subject devices demonstrate that the device is as safe and as effective as the predicate. Substantial equivalence is demonstrated in the performance testing section of the submission by comparing subject and predicated designs, as well as testing according to ASTMF382-99, Standard Test Method for Metallic Medical Bone Plates and ASTM F543-07, Standard Specification and Test Method for Metallic Bone Screws. Comparison of the design, intended use, and testing demonstrate that the Olecranon Plating System performs as well as the predicate devices and should thereby be considered substantially equivalent. #### Clinical Performance and Conclusions: Clinical testing was not performed for this device. #### Conclusion The Miami Device Solutions Olecranon Plating System is substantially equivalent to the predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%