VariAx Distal Radius Plating System, VariAx 2 System
Applicant
Stryker GmbH
Product Code
HRS · Orthopedic
Decision Date
Feb 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VariAx Distal Radius Plate System is intended for internal fixation of small bone fractures, primarily including distal radius fractures. The VariAx 2 System is intended for internal fixation.
Device Story
System consists of metallic bone plates and screws for internal fracture fixation. Used by surgeons in clinical settings to stabilize small bone fractures, primarily distal radius. Implants are manufactured from titanium and titanium alloy. Device provides mechanical support to bone fragments to facilitate healing. No changes to device design or materials; submission introduces new 'Xpress' packaging configuration combining previously cleared plates, screws, and class I exempt instrumentation. Implants provided sterile or non-sterile.
Clinical Evidence
No clinical data. Bench testing limited to LAL testing to confirm endotoxin limits (20EU/device).
Technological Characteristics
Materials: Commercially Pure Titanium Grade 2 (ASTM F67) and Titanium Alloy (ASTM F136). Components: Bone plates and screws. Sterilization: Provided sterile and non-sterile. No changes to design or materials from predicates.
Indications for Use
Indicated for internal fixation of small bone fractures (distal radius, radius, ulna, clavicle, humerus, foot, ankle, distal tibia, fibula). Includes compression, intra-articular, extra-articular, displaced, single, segmental, and comminuted fractures; osteotomies; non-unions; mal-unions; replantation; joint fusions; arthrodesis; and reconstruction. Applicable to normal or osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 21, 2017
Stryker GmbH Irene Bacalocostantis, PhD Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430
Re: K162841 Trade/Device Name: VariAx Distal Radius Plate System, VariAx 2 System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 17, 2017 Received: January 18, 2017
Dear Dr. Bacalocostantis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known) K162481
Device Name
### VariAx Distal Radius Plating System
Indications for Use (Describe)
The VariAx Distal Radius Plating System including the XXL Volar Distal Radius Plates is intended for internal fixation of small bone fractures, primarily including distal radius fractures.
Indications include:
- · compression fractures
- · intra-articular and extra-articular fractures
- · displaced fractures
Following additional indications apply only for the XXL Volar Distal Radius Plates: Osteotomies, non-unions, and malunions.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162841
Device Name VariAx 2 System
Indications for Use (Describe)
The Stryker VariAx 2 System screws, when used in conjunction with VariAx Plating Systems; or used independently in a lag screw technique, are indicated for:
· Internal fracture fixation;
· Osteotomies;
· Revision procedures such as non-unions or mal-unions;
In addition, the following indications are specific to the devices listed below:
· T8 2.4mm Screws & T8 2.0mm Locking Peg: For use in small bones, primarily including the Distal Radius, in the treatment of:
o Compression fractures:
o Intra-articular and extra-articular fractures;
o Displaced fractures:
o Reconstruction procedures:
· T8 2.7mm Screws: For use in small bones, including the Distal Radius as well as the fore, mid- and hind Foot and Ankle, in the treatment of:
o Intra-articular and extra-articular fractures of the Distal Radius.
o Displaced and compression fractures of the Distal Radius:
o Replantation, joint fusions or arthrodesis and corrective osteotomies in the Foot & Ankle;
o Reconstruction procedures in the Foot & Ankle and Distal Radius:
· T10 3.5mm and T10 2.7mm Screws: For use in the Radius, Ulna, Clavicle, Humerus, Foot and Ankle, Distal Tibia and Fibula, in the treatment of:
o Intra-articular and extra-articular fractures of the Distal Humerus and Proximal Ulna;
- o Single, segmental and comminuted fractures;
o Replantation, joint fusions or arthrodesis and corrective osteotomies in the Foot & Ankle;
o Normal bone density or osteopenic bone.
Type of Use (Select one or both, as applicable)
> Prescription Use {Part 21 CFR 801 Subpart D}
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
| Proprietary Name: | VariAx Distal Radius Plating System<br>VariAx 2 System |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Bone Plates<br>Bone Screws |
| Regulation Number &<br>Description: | 21 CFR 888.3030; Single/multiple component<br>metallic bone fixation appliances and accessories |
| | 21 CFR 888.3040; Smooth or threaded metallic bone<br>fixation fastener |
| Product Code: | HRS (Plate, Fixation, Bone)<br>HWC (Screw, Fixation, Bone) |
| Device Class: | Class II |
| Sponsor: | Stryker Trauma GmbH<br>Bohnackerweg 1<br>2545 Selzach / Switzerland |
| Contact Person: | Irene Bacalocostantis<br>Sr. Regulatory Affairs Specialist<br>325 Corporate Drive<br>07430 Mahwah / New Jersey<br>irene.bacalocostantis@stryker.com<br>Phone: +1 201 831 6797 |
| Additional Contact Person: Tina Mornak | Sr. Regulatory Affairs Specialist<br>tina.mornak@stryker.com<br>Phone: +1 201 831 6365 |
| Date Prepared: | December 13, 2016 |
| Primary Predicate:<br>Reference Predicate: | VariAx Distal Radius Plating System, K141430<br>VariAx 2 System (screws), K140769 |
| Description | |
This Traditional 510(k) submission is intended to provide information on a new packaging configuration of the VariAx Distal Radius Plates. The components of the VariAx Distal
{6}------------------------------------------------
Radius System remain identical to those in the most recently cleared in K141430. No new components have been added to the system.
Briefly, the VariAx Distal Radius Plating System consists of several different plates and screws manufactured from Commercially Pure Titanium Grade 2 (ASTM F67) and Titanium Alloy (ASTM F136), respectively. The VariAx 2 System (screws) (cleared under K140769) include a series of Titanium Alloy (ASTM F136) screws intended to be used across the entire VariAx line. The VariAx 2 T8 2.0 mm peg, T8 2.4 mm and T8 2.7 mm screws, and washer are specifically indicated for use with VariAx Distal Radius Plating System. All implants are provided sterile and non-sterile.
The new packaging configuration, called the VariAx 2 Xpress Distal Radius Kit, includes already cleared sterile packed VariAx Distal Radius Plates and VariAx 2 T8 2.7 mm screws along with single use class I exempt instrumentation.
# Intended Use
The VariAx Distal Radius Plate System is intended for internal fixation of small bone fractures, primarily including distal radius fractures. The VariAx 2 System is intended for internal fixation.
## Indications for Use
## VariAx Distal Radius Plating System
The VariAx Distal Radius Plating System including the XXL Volar Distal Radius Plates is intended for internal fixation of small bone fractures, primarily including distal radius fractures.
Indications include:
- compression fractures
- intra-articular and extra-articular fractures
- displaced fractures
Following additional indications apply only for the XXL Volar Distal Radius Plates: Osteotomies, non-unions, and mal-unions.
## VariAx 2 System
The Stryker VariAx 2 System screws, when used in coniunction with VariAx Plating Systems; or used independently in a lag screw technique, are indicated for:
- . Internal fracture fixation;
- Osteotomies;
- Revision procedures such as non-unions or mal-unions;
{7}------------------------------------------------
In addition, the following indications are specific to the devices listed below:
- T8 2.4mm Screws & T8 2.0mmn Locking Peg: For use in small bones, primarily ● including the Distal Radius, in the treatment of:
- Compression fractures; o
- Intra-articular and extra-articular fractures; O
- Displaced fractures; o
- Reconstruction procedures; o
- . T8 2.7mm Screws: For use in small bones, including the Distal Radius as well as the fore, mid- and hind Foot and Ankle, in the treatment of:
- Intra-articular and extra-articular fractures of the Distal Radius, o
- Displaced and compression fractures of the Distal Radius; o
- Replantation, joint fusions or arthrodesis and corrective osteotomies in the O Foot & Ankle;
- Reconstruction procedures in the Foot & Ankle and Distal Radius; o
- TI0 3.5mmn and T10 2.7mm Screws: For use in the Radius, Ulna, Clavicle, Humerus, Foot and Ankle, Distal Tibia and Fibula, in the treatment of:
- Intra-articular and extra-articular fractures of the Distal Humerus and O Proximal Ulna;
- Single, segmental and comminuted fractures; o
- Replantation, joint fusions or arthrodesis and corrective osteotomies in the O Foot & Ankle;
- Normal bone density or osteopenic bone. o
# Summary of Technologies
There have not been any modifications made to the VariAx Distal Radius Plating System or the VariAx 2 System since of the VariAx Distal Radius Plating System (K141430) and of the VariAx 2 System (K140769).
# Non-Clinical Test
LAL testing has been performed to establish that the subject device meets the specified 20EU/device limit.
No additional non-clinical laboratory testing was required for this submission as there are no new components being added to the system.
## Clinical Testing
Clinical testing was not required for this submission because there are no new components being added to the system.
{8}------------------------------------------------
Conclusion
This submission is only intended to introduce a new packaging configuration where the VariAx Distal Radius Plating System and VariAx 2 T8 2.7mm screws are packaged together. The VariAx Distal Radius System and the VariAx 2 System remain identical to the predicate system cleared under K141430 and K140769, respectively.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.