Single Use Circular Stapler, Single Use Hemorrhodial Circular Stapler, Single Patient Use Linear Stapler and Reload, Single Patient Use Linear Cutter and Reload, Single Patient Use Endo Cutter and Reload, Single Patient Use Transverse Cutter and Reload
K162710 · Tgs Medical Co., Ltd. · GDW · Feb 9, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162710
Device Name
Single Use Circular Stapler, Single Use Hemorrhodial Circular Stapler, Single Patient Use Linear Stapler and Reload, Single Patient Use Linear Cutter and Reload, Single Patient Use Endo Cutter and Reload, Single Patient Use Transverse Cutter and Reload
Applicant
Tgs Medical Co., Ltd.
Product Code
GDW · General, Plastic Surgery
Decision Date
Feb 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Single Use Circular Stapler has application throughout the alimentary tract for end to side and side to side anastomoses. The Single Patient Use Endo Cutter and Reload is intended for transection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, thoracic, and pediativ surgical procedures. It can be used with tissue buttressing materials. The Single Patient Use Linear Cutter and Reload are intended for use in multiple open or minimally invasive surgical procedures for the transection of tissues. Linear cutters are also intended for creation of anastomoses in these procedures. The Single Patient Use Linear Stapler and Reload has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection of tissue and creation of anastomosis, including occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas. The Single Use Hemorrhoidal Circular Stapler has application throughout the anal canal to perform surgical treatment of hemorrhoidal disease. The Single Patient Use Transverse Cutter and Reload is intended for transection and or creation of anastomoses. The instrument has application in multiple open or minimally invasive general (gastrointestinal and skeletal muscle stapling), gynecologic, urologic, and thoracic surgical procedures.
Device Story
Manual surgical staplers; used in open or minimally invasive procedures; perform tissue transection, resection, and anastomosis. Devices include circular, linear, and transverse cutters/staplers. Operation: manual activation (squeezing handles) places titanium staples in staggered rows; integrated circular or linear knives simultaneously cut tissue between staple lines. Used by surgeons in OR settings. Benefits: facilitates tissue closure and anastomosis; reduces procedure time. Some models allow reloading for multiple firings per procedure. Safety features include reset knobs or safety releases to prevent mis-firing.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing, including biocompatibility (cytotoxicity, irritation, sensitization), endotoxin testing, and mechanical performance verification (seal strength).
Technological Characteristics
Manual surgical staplers; materials: unalloyed titanium, polycarbonate, stainless steel. Energy source: manual. Form factor: various lengths (30-100mm) and diameters (21-36mm). Sterilization: irradiation. No software or electronic components.
Indications for Use
Indicated for patients undergoing open or minimally invasive surgical procedures requiring tissue transection, resection, or anastomosis in general, gynecologic, urologic, thoracic, and pediatric surgery, including hemorrhoidal disease treatment and left atrial appendage occlusion.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
PROXIMATE® PPH Hemorrhoidal Circular Stapler and Accessories (K030411)
PROXIMATE® Curved and Straight Intraluminal Staplers (K983536)
ENDOPATH and PROXIMATE Linear Cutters and Staplers (K020779)
Echelon Endoscopic Linear Cutter Reload, Black (K112056)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, layered on top of each other. The profiles are connected by a flowing, ribbon-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2017
TGS Medical Co., Ltd. % Ms. Diana Hong Mid-Link Consulting Co.,Ltd P.O. Box 120-119 Shanghai, 200120 China
Re: K162710
Trade/Device Name: Single Use Circular Stapler, Single Use Hemorrhodial Circular Stapler, Single Patient Use Linear Stapler And Reload, Single Patient Use Linear Cutter And Reload, Single Patient Use Endo Cutter And Reload, Single Patient Use Transverse Cutter And Reload
Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: January 6, 2017 Received: January 10, 2017
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K162710
K16271U
### Device Name
Single Use Circular Stapler, Single Use Hemorrhodial Circular Stapler, Single Patient Use Linear Stapler and Reload, Single Patient Use Linear Cutter and Reload, Single Patient Use Endo Cutter and Reload, Single Patient Use Transverse Cutter and Reload
### Indications for Use (Describe)
The Single Use Circular Stapler has application throughout the alimentary tract for end to side and side to side anastomoses.
The Single Patient Use Endo Cutter and Reload is intended for transection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, thoracic, and pediativ surgical procedures. It can be used with tissue buttressing materials.
The Single Patient Use Linear Cutter and Reload are intended for use in multiple open or minimally invasive surgical procedures for the transection of tissues. Linear cutters are also intended for creation of anastomoses in these procedures.
The Single Patient Use Linear Stapler and Reload has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection of tissue and creation of anastomosis, including occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.
The Single Use Hemorrhoidal Circular Stapler has application throughout the anal to perform surgical treatment of hemorrhoidal disease.
The Single Patient Use Transverse Cutter and Reload is intended for transection and or creation of anastomoses. The instrument has application in multiple open or minimally invasive general (gastrointestinal and skeletal muscle stapling), gynecologic, urologic, and thoracic surgical procedures.
Type of Use (*Select one or both, as applicable*)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
# 510(k) Summary
- 1. Sponsor Identification
## TGS Medical Co., Ltd.
Building 17, No.8 Jinfeng Road, High-tech Zone, Suzhou, Jiangsu Province, China
Establishment Registration Number: Not yet registered
Contact Person: Hua Cao Position: General Manager Tel: +86-512-86867776 Fax: +86- 512-86867070 Email: caohua@tgsmedical.com
- 2. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### Identification of Proposed Device 3.
Trade Name: Single Use Circular Stapler
Single Use Hemorrhoidal Circular Stapler Single Patient Use Linear Stapler and Reload Single Patient Use Linear Cutter and Reload Single Patient Use Endo Cutter and Reload Single Patient Use Transverse Cutter and Reload
Common Name: Stapler and Reload
Regulatory Information Classification Name: Staple, Implantable Classification: II Product Code: GDW Regulation Number: 21CFR 878.4750 Review Panel: General& Plastic Surgery
Classification Name: Stapler, Surgical; Classification: I Exempt Regulation Number: 21CFR 878.4800 Review Panel: General & Plastic Surgery
Intended Use Statement:
The Single Use Circular Stapler has application throughout the alimentary tract for end to side and side to side anastomoses.
The Single Patient Use Endo Cutter and Reload is intended for transection, and/or creation of The instrument has application in multiple open or minimally invasive general, anastomoses. gynecologic, urologic, thoracic, and pediatric surgical procedures. It can be used with tissue buttressing materials.
The Single Patient Use Linear Cutter and Reload are intended for use in multiple open or surgical procedures for the transection of tissues. Linear cutters are also intended for creation of anastomoses in these procedures.
The Single Patient Use Linear Stapler and Reload has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection of tissue and creation of
{5}------------------------------------------------
anastomosis, including occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.
The Single Use Hemorrhoidal Circular Stapler has application throughout the anal canal to perform surgical treatment of hemorrhoidal disease.
The Single Patient Use Transverse Cutter and Reload is intended for transection, resection and/or creation of anastomoses. The instrument has application in multiple open or general (gastrointestinal and skeletal muscle stapling), gynecologic, urologic, and thoracic surgical procedures.
Device Description
The Single Patient Use Transverse Cutter and Reload places four staggered curved row of titanium staples on the tissue upon activation, and cut the tissue between staple lines. The device is available in 40mm length with reload in 3.8mm and 4.5mm staple size. The device may be reloaded and fired up to 5 times in a single procedure.
Single Use Circular stapler places a double staggered, circular row of titanium staples upon activation, which was achieved by squeezing the handles firmly as far as they could go. Immediately after formation of the staples, the excess tissue will be resect by the circular knife, and then a circular anastomosis is created. The stapler are available in 21mm, 29mm, 32mm and 34mm six specifications. The staple sizes are 4.8mm and 5.5mm.
Single Patient Use Endo Cutter and Reload places two, triple-staggered rows of titanium staples and simultaneously divides the tissue from central line. The devices are available in 30mm, 45mm and 60mm three lengths. Reloads are available in six staple sizes to accommodate various tissue thicknesses: 2.0mm, 2.5mm, 3.5mm, 3.8mm, 4.1mm, 4.2mm. The device may be reloaded and fired up to 5 times in a single procedure.
Single Patient Use Linear Stapler and Reload places a double staggered row of titanium staples and is available in 30mm, 45mm, 60mm, 75mm and 90mm staple for use in various applications. Three staple sizes (2.5mm, 3.8mm and 4.5mm) are available to accommodate various tissue thicknesses. It may be reloaded and fired up to 4 times for a total 5 firing in a single procedure.
Single Patient Use Linear Cutter and Reload place two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. The devices are available in 55mm, 60mm, 75mm, 80mm and 100mm lengths. Reloads are available in four staple sizes to accommodate various tissue thicknesses: 2.5mm, 4.2mm and 4.5mm. It may be reloaded and fired up to 4 times for a total 5 firings in a single procedure.
{6}------------------------------------------------
Single Use Hemorrhoidal Circular Stapler is a set of instruments that place a double staggered, circular row of titanium staples. Immediately after the formation of staples, the circular knife blade resects the excess of compressed mucosa. The stapler are available in 30mm, 34mm and 36mm four specifications. The staple size is 4.5mm. It cannot be reloaded.
- 4. Identification of Predicate Devices
Predicate Device 1 510(k) Number: K030411 Product Name: PROXIMATE® PPH Hemorrhoidal Circular Stapler and Accessories
Predicate Device 2 510(k) Number: K983536 Product Name: PROXIMATE® Curved and Straight Intraluminal Staplers
Predicate Device 3 510(k) Number: K020779 Product Name: ENDOPATH and PROXIMATE Linear Cutters and Staplers
Predicate Device 4 510(k) Number: K111825 Product Name: DST Series™ Staplers
Predicate Device 5 510(k) Number: K061156 Product Name: ENDOPATH Linear Cutters and Staplers
Predicate Device 6 510(k) Number: K062869 Product Name: CONTOUR™ Curved Cutter Stapler and Reloads
- న. Identification of Reference Devices
Reference Device 1 510(k) Number: K080839 Product Name: Echelon™ Gray Cartridge-45mm
Reference Device 2 510(k) Number: K112056 Product Name: Echelon Endoscopic Linear Cutter Reload, Black
{7}------------------------------------------------
### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
- > ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- > USP 38-NF 33 <85> Bacterial Endotoxins Tests
- > ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;
- > ISO 11137-2: 2013 Sterilization of health care products -Radiation- Part 2: Establishing the sterilization dose
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{8}------------------------------------------------
### 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device | Predicate Device 6 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Product Code | GDW | K062869 |
| Regulation Number | 21 CFR 878.4750 | Same |
| Intended Use | The Single Patient Use Transverse Cutter and Reload is intended for transection, resection and/or creation of anastomoses. The instrument has application in multiple open or general (gastrointestinal and skeletal muscle stapling), gynecologic, urologic, and thoracic surgical procedures. | Same |
| Cutting Mechanism | Curved Knife | Same |
| Operation Principle | Manual | Same |
| Safety Mechanism | Reset Knob is used for preventing from mis-firing | Same |
| Closed staple height | 1.5mm, 2.0mm | Same |
| Closed staple form | Image: staple form | Image: staple form |
| Patient-contact material | Unalloyed Titanium<br>Polycarbonate<br>Stainless Steel | Unknown |
| Sterilization | Irradiation Sterilization | EO Sterilization |
| Endotoxin Limit | 20 EU | Same |
| Labeling | Conforms with 21 CFR 801 | Same |
# Table 1 Comparison of Technology Characteristics for Single Patient Use Transverse Cutter and Reload
{9}------------------------------------------------
| Item | Proposed Device | Predicate Device 2 | Item | Proposed Device | Predicate | Reference | Reference | Item | Proposed Device | Predicate Device 4 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|---------------------|---------------------|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| | | K983536 | | | Device 5<br>K061156 | Device 1<br>K080839 | Device 2<br>K112056 | Product Code | GDW | K111825 |
| Product Code | GDW | Same | Product Code | GDW | Same | Same | Same | Regulation Number | 21 CFR 878.4750 | Same |
| Regulation Number | 21 CFR 878.4750 | Same | Regulation<br>Number | 21 CFR 878.4750 | Same | Same | Same | Intended Use | The Single Patient Use Linear Stapler<br>and Reload has applications in<br>abdominal, gynecological, pediatric<br>and thoracic surgical procedures for<br>resection or transection of tissue and<br>creation of anastomosis, including<br>occlusion of the left atrial appendage<br>in open procedures. They may be<br>used for transection and resection of<br>pancreas. | Same |
| Intended Use | The Single Use Circular Stapler has<br>application throughout the alimentary<br>tract for end to end, end to side and<br>side to side anastomoses. | Same | Intended Use | The Single Patient Use Endo Cutter<br>and Reload is intended for<br>transection, resection, and/or<br>creation of anastomoses. The<br>instrument has application in<br>multiple open or minimally invasive<br>general, gynecologic, urologic,<br>thoracic, and pediatric surgical<br>procedures. | Same | Same | Same | Cutting Mechanism | N.A. | Same |
| Cutting Mechanism | Circular Knife | Same | Cutting<br>Mechanism | Linear Knife | Same | Same | Same | Operation Principle | Manual | Same |
| Operation Principle | Manual | Same | Operation<br>Principle | Manual | Same | Same | Same | Safety Mechanism | Safety release for preventing from<br>mis-firing | Same |
| Safety Mechanism | Insurance is used for preventing from<br>mis-firing | Same | Safety<br>Mechanism | Green button for preventing from<br>mis-firing | Same | Same | Same | Closed staple height | 1.0mm, 1.5mm, 2.0mm | Same |
| Closed staple height | 1.0~2.5mm | Same | Closed staple<br>height | 0.75, 1.0, 1.5, 1.8, 2.0, 2.3mm | 1.0, 1.5,<br>1.8, 2.0mm | 0.75mm | 2.3mm | Closed staple form | Image: staple form | Image: staple form |
| Closed staple form | Image: staple form | Image: staple form | Closed staple<br>form | Image: staple form | Image: staple form | Image: staple form | Image: staple form | Patient-contact<br>material | Unalloyed Titanium<br>Polycarbonate<br>Stainless Steel | Unknown |
| Patient-contact<br>material | Unalloyed Titanium<br>Polycarbonate<br>Stainless Steel | Unknown | Patient-contact<br>material | Unalloyed Titanium<br>Acrylonitrile-Butadiene-Styrene<br>Stainless Steel | Unknown | Unknown | Unknown | Sterilization | Irradiation Sterilization | EO Sterilization |
| Sterilization | Irradiation Sterilization | EO Sterilization | Sterilization | Irradiation Sterilization | EO<br>Sterilization | EO<br>Sterilization | EO<br>Sterilization | Endotoxin Limit | 20 EU | Same |
| Endotoxin Limit | 20 EU | Same | Endotoxin<br>Limit | 20 EU | Same | Same | Same | Labeling | Conforms with 21 CFR 801 | Same |
| Labeling | Conforms with 21 CFR 801 | Same | Labeling | Conforms with 21 CFR 801 | Same | Same | Same | | | |
Table 2 Comparison of Technology Characteristics for Single Use Circular Stapler
{10}------------------------------------------------
Table 3 Comparison of Technology Characteristics for Single Patient Use Endo Cutter and Reload
{11}------------------------------------------------
# Table 4 Comparison of Technology Characteristics for Single Patient Use Linear Stapler and Reload
{12}------------------------------------------------
| Item | Proposed Device | Predicate Device 3 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Product Code | GDW | K020779 |
| Regulation Number | 21 CFR 878.4750 | Same |
| Intended Use | The Single Patient Use Linear Cutter<br>and Reload are intended for use in<br>multiple open or surgical procedures<br>for the transection and resection of<br>tissues. Linear cutters are also<br>intended for creation of anastomoses<br>in these procedures. | Same |
| Cutting Mechanism | Linear Knife | Same |
| Operation Principle | Manual | Same |
| Safety Mechanism | Safety release for preventing from<br>mis-firing | Same |
| Closed staple height | 1.0mm, 1.5mm, 1.8mm, 2.0mm | Same |
| Closed staple form | Image: staple form | Image: staple form |
| Patient-contact<br>material | Unalloyed Titanium<br>Polycarbonate<br>Stainless Steel | Unknown |
| Sterilization | Irradiation Sterilization | EO Sterilization |
| Endotoxin Limit | 20 EU | Same |
| Labeling | Conforms with 21 CFR 801 | Same |
# Table 5 Comparison of Technology Characteristics for Single Patient Use Linear Cutter and Reload
{13}------------------------------------------------
| Item | Proposed Device | Predicate Device 1 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Product Code | GDW | K030411 |
| Regulation Number | 21 CFR 878.4750 | Same |
| Intended Use | The Single Use Hemorrhoidal Circular Stapler has application throughout the anal canal to perform surgical treatment of hemorrhoidal disease. | Same |
| Cutting Mechanism | Circular Knife | Same |
| Operation Principle | Manual | Same |
| Safety Mechanism | Safety release for preventing from mis-firing | Same |
| Closed staple height | 0.8~2.0mm | Same |
| Closed staple form | Image: staple form | Image: staple form |
| Patient-contact material | Unalloyed Titanium<br>Polycarbonate<br>Stainless Steel | Unknown |
| Sterilization | Irradiation Sterilization | EO Sterilization |
| Endotoxin Limit | 20 EU | Same |
| Labeling | Conforms with 21 CFR 801 | Same |
Table 6 Comparison of Technology Characteristics for Single Use Hemorrhoidal Circular Stapler
#### Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.