CONCORD Plus Radiopaque Bone Cement

K162618 · Globus Medical, Inc. · NDN · Jan 23, 2017 · Orthopedic

Device Facts

Record IDK162618
Device NameCONCORD Plus Radiopaque Bone Cement
ApplicantGlobus Medical, Inc.
Product CodeNDN · Orthopedic
Decision DateJan 23, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3027
Device ClassClass 2
AttributesTherapeutic

Intended Use

CONCORD Plus™ Radiopaque Bone Cement is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

Device Story

CONCORD Plus™ is a high-viscosity, radiopaque, self-curing PMMA bone cement; provided as a two-component system (liquid and powder) mixed prior to use. Used in clinical settings by physicians during vertebroplasty or kyphoplasty procedures to stabilize vertebral fractures. The cement hardens in situ to provide structural support for the fractured vertebral body, aiming to alleviate pain and restore stability. Single-use device.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, chemical testing, and handling properties in accordance with the FDA Class II Special Controls Guidance Document for PMMA Bone Cement.

Technological Characteristics

High-viscosity, self-curing PMMA bone cement. Two-component system. Sterilization: powder (gamma radiation), liquid (filtration), outer packaging (ethylene oxide). Meets pyrogen limit ≤20 EU/device. Testing standards: ASTM D732-10, ASTM D5045-14, ASTM F2118-14, ASTM D638-10, ASTM F451-08, ASTM D2990-09, and ISO 5833:2002.

Indications for Use

Indicated for fixation of pathological vertebral body fractures via vertebroplasty or kyphoplasty in patients with painful vertebral compression fractures due to osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).

Regulatory Classification

Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 23, 2017 Globus Medical, Inc. Kelly J. Baker, Ph.D. Senior Vice President, Regulatory and Clinical Affairs 2560 General Armistead Avenue Audubon, Pennsylvania 19403 Re: K162618 Trade/Device Name: CONCORD Plus"11 Radiopaque Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN Dated: December 8, 2016 Received: December 9, 2016 Dear Dr. Baker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2 - Kelly J. Baker, Ph.D. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K162618 Device Name CONCORD Plus™ Radiopaque Bone Cement #### Indications for Use (Describe) CONCORD Plus™ Radiopaque Bone Cement is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma). | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary CONCORD Plus Radiopaque Bone Cement | Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800 | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Kelly J. Baker, Ph.D.<br>Senior Vice President, Regulatory and Clinical Affairs | | Date Prepared: | December 8, 2016 | | Device Name:<br>Classification: | CONCORD Plus™ Radiopaque Bone Cement<br>Per 21 CFR §888.3027 Polymethylmethacrylate<br>(PMMA) bone cement.<br>Product Code: NDN<br>Regulatory Class: II, Panel Code: Orthopedic 87 | | Predicate(s):<br>Primary<br>Additional | CONCORD™ Radiopaque Bone Cement (K042168)<br>VertaPlex® HV Radiopaque Bone Cement (K091606) | ## Purpose: The purpose of this submission is to request clearance of the CONCORD Plus™ Radiopaque Bone Cement. #### DEVICE DESCRIPTION: The CONCORD Plus™ Radiopaque Bone Cement is a high-viscosity, radiopaque, self-curing, PMMA bone cement. It is provided sterile, as a two component system, a liquid component and a powder component which are mixed together prior to use to form the cement. The bone cement is intended for single use only. #### INDICATION FOR USE: CONCORD Plus™ Radiopaque Bone Cement is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma). #### Performance Data: Performance data was conducted in accordance with the Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement, July 17, 2002. Performance data demonstrates substantial equivalence to the predicate. Performance testing, chemical testing and handling properties were conducted to characterize CONCORD Plus™ Radiopaque Bone Cement compared to the {4}------------------------------------------------ predicate in compliance with ASTM D732-10. ASTM D5045-14. ASTM F2118-14. ASTM D638-10, ASTM F451-08, ASTM D2990-09, and ISO 5833:2002. CONCORD Plus™ Radiopaque Bone Cement is provided sterile. The powder component is sterilized with gamma radiation to a Sterility Assurance Level (SAL) of 10°. The liquid component is sterilized with filtration methods to a SAL of 103. The outer packaging containing the liquid and powder components is sterilized by ethylene oxide gas to a SAL of 10-6. ## Technological Characteristics: The subject CONCORD Plus™ Radiopaque Bone Cement has the same fundamental scientific technology as the predicate CONCORD™ Radiopaque Bone Cement. The subject device has the same chemical composition, sterilization methods and meets the same pyrogen limit specification of ≤20 EU/device as the predicate. The subject CONCORD Plus™ is a high viscosity bone cement similar to the predicate VertaPlex HV Bone Cement. ## Basis of Substantial Equivalence: CONCORD Plus™ Radiopaque Bone Cement has been found to be substantially equivalent to the predicate device with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject bone cement to the predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%