K162248 · Asahi Kasei Medical Co., Ltd. · KDI · May 3, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K162248
Device Name
Asahi ViE-U Series Dialyzer
Applicant
Asahi Kasei Medical Co., Ltd.
Product Code
KDI · Gastroenterology, Urology
Decision Date
May 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VIE-U is intended for use in hemodialysis for the treatment of patients who have acute or chronic renal failure. ViE-U is intended for single use only.
Device Story
High flux hollow fiber hemodialyzer; utilizes vitamin E-coated hollow fiber membrane. Operates by facilitating solute clearance and ultrafiltration of blood during hemodialysis. Used in clinical settings by healthcare professionals. Input is patient blood; output is dialyzed blood. Vitamin E coating intended to provide biocompatibility benefits. Device is single-use and sterile.
Clinical Evidence
Prospective, open-label, non-randomized, single-armed, controlled study (n=17 enrolled, 12 completed). Compared ViE-21U to standard-of-care control dialyzer over 12 weeks. Endpoints: KuF, removal rates for urea, creatinine, albumin, and beta 2-microglobulin; biocompatibility (WBC, platelets, complement activation); safety. Results showed equivalence to control. One patient experienced possibly related adverse events (pruritus, rash) resolved with medication.
Technological Characteristics
High flux hollow fiber hemodialyzer. Materials (hollow fiber, housing, headers, potting, stoppers) identical to APS predicate. Vitamin E coating. Sterilization per ISO 11137-2. Biocompatibility per ISO 10993-1. Performance testing per ISO 8637.
Indications for Use
Indicated for patients with acute or chronic renal failure requiring hemodialysis.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an emblem featuring three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 3, 2017
Asahi Kasei Medical Co., Ltd. % Megan Shackelford Principal, Consulting Services Boston Biomedical Associates, LLC 100 Crowley Drive, Suite 216 Marlborough, MA 01752
Re: K162248
> Trade/Device Name: Asahi ViE-U Series Dialyzer Regulation Number: 21 CFR§ 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: II Product Code: KDI Dated: April 3, 2017 Received: April 4, 2017
Dear Megan Shackelford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162248
Device Name Asahi ViE-U Series Dialyzer
Indications for Use (Describe)
VIE-U is intended for use in hemodialysis for the treatment of patients who have acute or chronic renal failure. ViE-U is intended for single use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) summary
# AsahiKASEI
## ASAHI KASEI MEDICAL CO., LTD.
(As Required By 21 CFR 807.92)
Date Prepared: April 3, 2017
#### I. SUBMITTER
| Submitter Name and Address | Contact |
|-------------------------------|--------------------------------|
| Asahi Kasei Medical Co., Ltd. | Megan M. Shackelford |
| 1-105, Kanda Jinbocho, | Principal, Consulting Services |
| Chiyoda-ku, Tokyo 101-8101 | Boston Biomedical Associates |
| Japan | 100 Crowley Drive Suite 216 |
| | Marlborough, MA 01752 |
#### II. DEVICE
| Trade Name: | Asahi ViE-U Series Dialyzer |
|----------------|--------------------------------------------------------|
| Common Name: | Hemodialyzer, High Permeability<br>Hemodialysis System |
| Classification | 21 CFR 876.5860, Class II |
| Product Code | KDI |
| Panel | Gastroenterology and Urology |
#### III. PREDICATE DEVICES
| Asahi APS Series Dialyzers: |
|--------------------------------------------------------------|
| K001250, August 16, 2000, and K041726, July 23, 2004 |
| CLIRANS E-Series Hollow Fiber Dialyzers (Terumo Corporation) |
| K003425, February 1, 2001, and K013550, November 20, 2001 |
#### DEVICE DESCRIPTION IV
The ViE-U is a high flux hollow fiber hemodialyzer. The device is sold sterile and is intended for single use only.
K162248 – 510(K) Summary
{4}------------------------------------------------
## AsahiKASEI
All the materials including the hollow fiber membrane, housing, headers, potting material and stoppers are identical to those previously cleared in the predicate device (APS). The additional coating of vitamin E on the hollow fiber membrane uses an equivalent material and technology to the predicate device (CLIRANS).
#### V. INDICATIONS FOR USE
ViE-U is intended for use in hemodialysis for the treatment of patients who have acute or chronic renal failure. ViE-U is intended for single use only.
### COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE VI PREDICATE DEVICE
The proposed device (ViE-U) is substantially equivalent to the predicate device (APS) in terms of materials, principles of operation and device design. The proposed device was tested and compared with the predicate device (APS) for performance specifications. Additionally, the vitamin E coating of the proposed device uses an equivalent material and technology to that of the predicate device (CLIRANS). Considering these factors together, the proposed device was shown to be substantially equivalent to the predicate devices currently cleared through the 510(k) process.
#### PERFORMANCE DATA VII
Asahi has performed testing in conformance with the special controls for High Permeability Hemodialysis System, 21 CFR 876.5860, and the FDA document "Guidance for the Content of Premarket Notifications for Conventional and High Permeability Hemodialyzers'" including:
- . Biocompatibility testing for external communicating devices, circulating blood, prolonged contact per ISO 10993-1;
- In vitro performance testing per ISO 8637 for
- o Clearance (urea, creatinine, phosphate, and vitamin B12 at blood flow rates from 200 up to 500 mL/min.)
- o Pressure drop
- o Kuf
- o Priming volume (Blood compartment volume)
- o Maximum TMP
- o Structural integrity
{5}------------------------------------------------
- Sterilization validation per ISO 11137-2; .
- Pyrogenicity testing; .
- Expiration date testing: .
- Packaging and transportation testing per ASTM D4169-14; .
- Clinical performance including in vivo KuF and removal rates for urea, albumin, and ● beta 2-microglobulin.
The clinical study was a prospective, open-label, non-randomized, single-armed, controlled study at a single site in Canada per ISO 14155. Seventeen patients were enrolled in the study and received treatment sessions in three study phases. Each patient began the study by receiving treatments with their current standard of care (control dialyzer) for 6 initial sessions (two weeks), followed by treatments with the ViE-21U for 36 sessions (12 weeks), and concluded the study with 6 final treatments with the control dialyzer (two weeks). The clinical performance of the ultrafiltration coefficient (KuF) and the removal rates for urea, creatinine, albumin and beta 2-microglobulin were evaluated for both the control and the ViE-21U dialyzers. In addition to performance data, biocompatibility and safety data were collected during the study, including: measurements of white blood cells, platelets and complement activation; type and number of adverse events; and type and number of device malfunctions. The clinical performance and biocompatibility were calculated based on the 12 patients who completed the 36 sessions with the ViE-21U dialyzer, while the safety evaluation was performed on all 17 enrolled patients. The clinical performance and biocompatibility evaluation showed equivalence of the ViE-21U and control dialyzers. Two adverse events determined to be possibly related (possible delayed hypersensitivity reaction) to the use of the ViE-21U occurred in one patient: pruritus and full body rash. These events were resolved with medication (diphenhydramine and prednisone) without sequelae, and the patient was transitioned back to the original hemodialyzer. There were no serious adverse events determined to be related to the ViE-21U. The results of this clinical study show that the ViE-U was used effectively as a hemodialyzer in patients who have renal failure and was shown to be safe as compared to the control dialyzer.
# VIII CONCLUSION
All design verification tests and in vivo validation of clinical study demonstrated that ViE-U is substantially equivalent in intended use, design, principle of operation, technology, materials, specifications, and performance to APS cleared under K001250 and K041726 and CLIRANS cleared under K003425 and K013550.
AsahiKASEI
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.