K162137 · VivaLNK, Inc. · FLL · Jan 19, 2017 · General Hospital
Device Facts
Record ID
K162137
Device Name
Fever Scout™ Continuous Monitoring thermometer
Applicant
VivaLNK, Inc.
Product Code
FLL · General Hospital
Decision Date
Jan 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The wireless Fever Scout™ Continuous Monitoring thermometer is a non-invasive and re-usable electronic device for home use. This product is intended for non-urgent ambulatory continuous armpit body temperature monitoring from ages 29 days and older.
Device Story
Fever Scout is a wearable, reusable electronic patch for continuous armpit temperature monitoring; intended for home use. Device consists of a flexible patch with a temperature sensor, control module, BLE communication module, and rechargeable lithium battery; includes a charging system with BLE relay function to extend range. Patch adheres to skin; transmits temperature data via BLE to a smartphone app. App provides GUI for temperature display, trending, and user-defined alerts. Healthcare providers or caregivers use the app to monitor patient temperature trends; facilitates non-urgent clinical decision-making regarding fever management. Benefits include continuous, non-invasive monitoring without manual spot-checks.
Clinical Evidence
Bench testing only. Performance demonstrated via product verification, software verification, usability validation, biocompatibility (ISO 10993-1), mechanical hazards, EMC (IEC 60601-1-2), and electrical safety (IEC 60601-1, IEC 60601-1-11). Accuracy verified against ASTM E1112-2000 (Reapproved 2011).
Technological Characteristics
Flexible PCB patch; temperature sensor; BLE communication; 3.0V DC rechargeable lithium battery. Standards: ASTM E1112, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62304, ISO 10993-1. Connectivity: BLE to smartphone app. Type B applied part.
Indications for Use
Indicated for non-urgent ambulatory continuous armpit body temperature monitoring in patients aged 29 days and older.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2017
Vivalnk Inc. Christine Kuo Director, RA/QA 4655 Old Ironsides Dr. #390 Santa Clara, California 95054
Re: K162137
Trade/Device Name: Fever Scout™ Continuous Monitoring thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: December 20, 2016 Received: December 23, 2016
Dear Christine Kuo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature that reads "Tina Kiang-S". The signature is written in a cursive style. There is a large, stylized loop to the left of the name. The background is white.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K162137
Device Name
Fever ScoutTM Continuous Monitoring thermometer
Indications for Use (Describe)
The wireless Fever Scout™ Continuous Monitoring thermometer is a non-invasive and re-usable electronic device for home use. This product is intended for non-urgent ambulatory continuous armpir body temperature monitioring from ages 29 days and older.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 – 510(k) Summary
# K162137
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92.
| Type of Submission: | Traditional 510(k); this is the first submission for this device |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter: | VivaLnk Inc.<br>4655 Old Ironsides Dr. #390<br>Santa Clara, CA 95054 USA<br>Phone: 408-868-2898 |
| Contact Person: | Christine Kuo |
| Date Prepared: | January 19, 2017 |
| Device Identification: | |
| Trade Name: | Fever Scout™ Continuous Monitoring thermometer |
| Device Common Name: | Armpit thermometer |
| Classification: | 21 CFR Sec. 880.2910, Clinical electronic thermometer |
| Product Code: | FLL |
| Classification Panel: | General Hospital |
| Class: | II |
| Indications for Use: | The wireless Fever Scout™ Continuous Monitoring thermometer<br>is a non-invasive and re-usable electronic device for home use.<br>This product is intended for non-urgent ambulatory continuous<br>armpit body temperature monitoring from ages 29 days and older. |
| Device Description: | Fever Scout Continuous Monitoring thermometer is an ambulatory<br>continuous body temperature monitor designed for armpit location,<br>and uses Bluetooth Low Energy (BLE) for wireless<br>communication with the Smartphone app for temperature display,<br>trending and alert. |
| | This product is composed of <ul><li>The Fever Scout VV-200 patch</li><li>The Smartphone app (supporting iOS)</li><li>AAA charging system with BLE relay function</li></ul> |
- . Eight adhesives
{4}------------------------------------------------
| | The patch hardware primarily includes the following function<br>modules: 1) temperature sensor, 2) control module, 3) Bluetooth<br>Low Energy (BLE) communication module and 4) battery. All are<br>integrated onto a flexible electronic printed circuit board. A MS<br>Lithium rechargeable battery is attached to the board to supply<br>3.0V DC internal power. |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The app includes 1) User account, profile and alert temperature<br>setup and 2) GUI for temperature display, trending and alerts. |
| | The charging system provides a cordless charging method to the<br>patch, a storage to the replaceable double-sided adhesives, as well<br>as an alternative BLE communication channel from the patch to<br>the charger and then to the iPhone, in order to extend<br>communication range. |
| | This product contains Type B applied part (per IEC 60601-1:<br>2012). All components, the patch, charger and adhesives are all<br>user accessible parts. However, the user is only expected to change<br>the charger AAA batteries, but not to open or modify any other<br>parts of the product. |
| Predicates: | TempTraq Model TT-100 by Blue Spark Technologies (K143267) |
| Trade Name: | TempTraq Model TT-100 |
| Device Common Name: | Armpit thermometer |
| Classification: | 21 CFR Sec. 880.2910, Clinical electronic thermometer |
| Product Code: | FLL |
| Classification Panel: | General Hospital |
| Class: | II |
#### Summary of Substantial Equivalence:
Fever Scout Continuous Monitoring thermometer is substantially equivalent to the predicate, TempTraq TT-100.
### Device Comparison
| Characteristics | Proposed<br>Device<br>Fever Scout | Primary<br>Predicate<br>TempTraq<br>TT-100 | Reference<br>Predicate<br>iThermonitor<br>WT-701 | Comparison | |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) Number | NA | K143267 | K132761 | NA | |
| Product Code | FLL | | | Identical | |
| Regulation # | 21CFR880.2910 | | | | Identical |
| Class | Clinical electronic thermometers | | | | Identical |
| Indications for Use | The wireless<br>Fever Scout<br>Continuous<br>Monitoring<br>thermometer<br>is a non-<br>invasive and<br>re-usable<br>electronic<br>device for<br>home use.<br>This product<br>is intended for<br>non-urgent<br>ambulatory<br>continuous<br>armpit body<br>temperature<br>monitoring<br>from ages 29<br>days and<br>older. | The Wireless<br>thermometer,<br>model TT-100,<br>is<br>a battery-<br>operated<br>electronic<br>device<br>with intended<br>use<br>of measuring<br>human body<br>temperature<br>precisely. This<br>device is single<br>use<br>and intended<br>for armpit<br>temperature<br>measurement for<br>persons of all<br>age. | The Wireless<br>Thermometer<br>is a battery<br>operated<br>electronic<br>device with<br>intended use<br>of measuring<br>human armpit<br>temperature<br>continuously<br>via wireless<br>signal<br>transmission<br>of the<br>measuring<br>results. This<br>system is<br>reusable and<br>intended for<br>armpit<br>monitoring for<br>persons over<br>two years old. | Equivalent.<br><br>All three devices<br>are for human<br>armpit<br>temperature<br>measurements.<br>Difference are<br>demonstrated to<br>not raise<br>questions of<br>safety and<br>effectiveness per<br>requirements set<br>forth in the<br>standards listed<br>in Section 5-<br>510(k)<br>Summary | |
| Display Use Specification | iOS device<br>display | iOS and<br>Android device<br>display | iOS device<br>display | Equivalent<br><br>iOS is what all<br>three devices<br>support. Not<br>supporting<br>Android will not<br>impact safety or<br>efficacy<br>concerns, but<br>limit market<br>share (number<br>of users who can<br>use this device). | |
| Working Voltage | 3.0 V DC | | | Identical | |
| Battery | MS Lithium<br>Rechargeable<br>Battery 3.0V | Two (2)<br>Blue<br>Spark 1.5 V<br>batteries<br>(103-<br>UT1) | The button<br>battery 3.0 V,<br>210mAh | Equivalent per<br>UL 1642 5th<br>Edition | |
| Reuse | Yes | No | Yes | Equivalent | |
| | | | | is demonstrated<br>to not raise<br>questions of<br>safety and<br>efficacy through<br>functional<br>verification and<br>cleaning testing. | |
| Measurement Range | 35 ~ 42°C | 30 ~ 42.4°C | 25 ~ 45° C | Equivalent<br>All three devices<br>cover the needed<br>range per<br>ASTM_1112<br>Temperature<br>range. | |
| Accuracy | ±0.1°C<br>From 37 ~<br>39°C, ±0.2°C<br>from 35 ~<br>37°C and 39 ~<br>42°C | ±0.1°C<br>between 30<br>~ 42.4°C | ±0.05° C (35 -<br>38.5° C);<br>±0.1° C (25 -<br>34.99° C);<br>±0.1° C<br>(38.51 -45° C) | Equivalent<br>All three devices<br>meet the<br>maximum error<br>temperature<br>ranges per<br>ASTM_1112. | |
| Signal Transmission | Wireless 2.4G<br>Bluetooth<br>BLE | | | Identical | |
| Receiver | Wireless 2.4G<br>Bluetooth<br>BLE enabled<br>smart devices<br>running Apple<br>operating<br>system<br>iPhone 5S+ or<br>newer. | Wireless<br>2.4G<br>Bluetooth<br>BLE<br>enabled<br>smart<br>devices<br>running<br>Apple<br>operating<br>system iOS<br>7.1 through<br>8.1 or<br>Android<br>operating<br>system 4.3<br>through<br>4.4.4 | Wireless 2.4G<br>Bluetooth<br>BLE enabled<br>smart devices<br>running Apple<br>operating<br>system iPhone<br>4S, iPhone 5, iPad<br>(3rd<br>generation),<br>iPad (4th<br>generation),<br>iPad mini,<br>iPod (5th<br>generation) | Equivalent<br>All three devices<br>support 2.4G<br>BLE enabled<br>devices. The<br>difference in<br>supported<br>operating<br>systems doesn't<br>impact safety<br>and efficacy. | |
| Valid Transmission | Up to 40<br>meters (with<br>relay) | Up to 40<br>feet | Up to 5 meters | Equivalent<br>The device is<br>demonstrated to | |
| Operating Condition | 10~40°C<br>15-85%<br>humidity | 16~40°C<br>15 -95%<br>humidity | 5~40 °C<br>15-85% humidity | intended.<br>Equivalent<br><br>The operating<br>range is<br>compliant with<br>ASTM_1112. | |
| Anatomical Application | Armpit<br>peel-and-stick contact thermometer sensor | | | Identical/<br>Equivalent | |
| Temperature Measurement<br>Interval | Continuous<br>transmitter<br>measures body<br>temperature<br>every 15<br>seconds | Continuous<br>transmitter<br>measures<br>body<br>temperature<br>every 10<br>seconds | Continuous<br>transmitter<br>measures body<br>temperature every<br>4 seconds | Equivalent<br><br>The maximum<br>difference in<br>temperature<br>measurements is<br>11 seconds,<br>which does not<br>raise questions<br>of safety and<br>efficacy. | |
| Safety and<br>Performance | Biocompatibility | Conformed to ISO 10993-1, ISO 10993-5, and ISO 10993-10 | | Identical/<br>Equivalent<br><br>Test against the<br>same standards<br>for safety and<br>performance | |
| | Electrical Safety | Conformed to IEC 60601-1 | | Identical/<br>Equivalent<br><br>Test against the<br>same standard<br>for safety and<br>performance | |
| | Electromagnetic<br>Compatibility | Conformed to IEC 60601-1-2 and FCC | | Identical/<br>Equivalent<br><br>Test against the<br>same standard<br>for safety and<br>performance | |
| | Performance | Conformed to ASTM E1112 | | Identical/<br>Equivalent<br><br>Test against the<br>same standard<br>for safety and<br>performance | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| Technological Characteristics: | Fever Scout Continuous<br>Monitoring thermometer is technologically substantially equivalent<br>to the predicate device. Both devices use equivalent temperature<br>technology to sense, process and transmit temperature data<br>continuously from the patch affixed to the skin under the armpit to<br>a mobile application. |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Data: | The substantially equivalent performance of Fever Scout<br>Continuous Monitoring thermometer has been demonstrated by<br>product verification, software verification, user/usability validation,<br>biocompatibility test, mechanical hazards test, EMC test and IEC<br>60601-1 safety compliance test. |
| | Performance data demonstrates conformance with 21 CFR Part<br>1020 or compliance with voluntary standards includes but not<br>limited to the following: |
| | IEC 60601-1:2005/(R) 2012: Medical Electrical Equipment -- Part 1:<br>General Requirements For Basic Safety And Essential Performance |
| | IEC 60601-1-2:2014: Medical Electrical Equipment - Part 1-2: General<br>Requirements For Basic Safety And Essential Performance - Collateral<br>Standard: Electromagnetic Compatibility – Requirements And Tests |
| | IEC 60601-1-11: 2015 Medical electrical equipment - Part 1-11:<br>General requirements for basic safety and essential performance -<br>Collateral standard: Requirements for medical electrical equipment and<br>medical electrical systems used in the home healthcare environment |
| | IEC 62304:2006 Medical Device Software – Software lifecycle Processes |
| | ISO 10993-1: 2009 Biological Evaluation of Medical Devices - Part 1:<br>Evaluation and Testing Within a Risk Management Process |
| | ASTM E1112-2000 (Reapproved 2011), standard specification for<br>electronic thermometer for intermittent determination of patient<br>temperature. |
| Conclusion: | VivaLnk concludes that the device is substantially equivalent to<br>the currently legally marketed predicate devices. Fever Scout<br>Continuous Monitoring thermometer has equivalent indications for<br>use and technological characteristics, and does not introduce<br>different questions of safety and effectiveness. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.