The SyncVision System is an image acquisition and processing system. It is indicated for use as follows: · To provide quantitative information regarding the calculated dimensions of arterial segments. · To enhance visualization of the stent deployment region. · To be used in-procedure in the catheterization lab and off-line for post-procedural analysis. · To obtain a co-registration of an angiographic x-ray image and IVUS images. · To obtain a co-registration of an angiographic x-ray image and intravascular blood pressure values.
Device Story
SyncVision System is an image acquisition/processing workstation for interventional cardiology. Inputs include angiographic x-ray streams, IVUS image streams, FFR/iFR pressure values, and ECG signals. System transforms these inputs by automating co-registration of angiographic images with IVUS or pressure data, stabilizing/enhancing vessel images, and calculating arterial dimensions (diameter, length, % stenosis). Used in catheterization labs or control rooms by interventional cardiologists. Hardware includes workstation (PC with frame grabber, graphics/data acquisition cards), LCD monitor, keyboard, mouse, and procedure-room joystick. Output is displayed on a monitor side-by-side with native imaging, assisting clinicians in lesion evaluation, device positioning, and post-deployment assessment. Benefits include improved visualization of anatomical locations across modalities and automated workflow, reducing manual estimation requirements.
Clinical Evidence
No clinical data. Evidence consists of software verification and validation and simulated use validation testing.
Technological Characteristics
Workstation-based image processing system. Components: PC, high-resolution dual-channel frame grabber, graphics processing card, data acquisition card for ECG, LCD monitor, joystick. Connectivity: Wired integration with x-ray systems, IVUS systems, and FFR/iFR pressure measurement systems. Software-based processing for image enhancement, stabilization, and co-registration. No sterile or patient-contacting components.
Indications for Use
Indicated for use in the catheterization lab during coronary catheterizations to provide quantitative arterial measurements, enhance stent deployment visualization, and co-register angiographic x-ray images with IVUS images or intravascular blood pressure values.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping and connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 5, 2016
Volcano Corporation % Ms. Elaine Alan Regulatory Affairs Specialist 3721 Valley Centre Drive, Suite 500 SAN DIEGO CA 92130
Re: K161756
Trade/Device Name: SyncVision System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB Dated: June 27, 2016 Received: June 28, 2016
Dear Ms. Alan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161756
Device Name SyncVision System
### Indications for Use (Describe)
- The SyncVision System is an image acquisition and processing system. It is indicated for use as follows:
- · To provide quantitative information regarding the calculated dimensions of arterial segments.
- · To enhance visualization of the stent deployment region.
- · To be used in-procedure in the catheterization lab and off-line for post-procedural analysis.
- · To obtain a co-registration of an angiographic x-ray image and IVUS images.
- · To obtain a co-registration of an angiographic x-ray image and intravascular blood pressures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the only element in the image and it is centered.
# 510(k) Summary Revised
| SPONSOR: | Philips Volcano<br>3721 Valley Centre Dr., Suite 500<br>San Diego, CA 92130 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT/<br>SUBMITTER: | Elaine Alan<br>Sr. Regulatory Affairs Specialist<br>Philips Volcano<br>3721 Valley Center Dr.<br>San Diego, CA 92130<br>Tel: 858-764-1281<br>Email: ealan@volcanocorp.com |
| DATE OF<br>SUBMISSION: | June 24, 2016 |
| DEVICE: | Image-intensified fluoroscopic x-ray system |
| TRADE NAME: | SyncVision System |
| COMMON<br>NAME: | SyncVision System |
| PRODUCT CODE: | OWB |
| CLASSIFICATION: | 21 CFR 892.1650<br>Class II Device |
| PANEL: | Radiology |
| PREDICATE DEVICE: | Volcano SyncVision System, K151904 |
Image /page/3/Picture/3 description: The image shows the Philips logo. The logo is a blue shield shape with the word "PHILIPS" at the top. Inside the shield are four stars and three wavy lines.
® VOLCANO
Philips Volcano, 3721 Valley Centre Drive, Suite 500, San Diego, CA 92130 USA
www.volcanocorp.com, Tel 800 228 4728, Fax 916 638 8112
Philips Volcano
{4}------------------------------------------------
#### DEVICE DECRISPTION:
The current standard procedure in the catheterization laboratory for utilizing x-ray, IVUS image streams, and FFR/iFR pressure measurements calls for referencing the native x-ray, IVUS imaging streams, and FFR/iFR intravascular blood pressure values of the patient's vasculature separately on different displays or monitor. This configuration requires the physician, usually an interventional cardiologist, to estimate an identical location on the patient's vasculature location on both the x-ray and IVUS image streams/FFR/iFR intravascular blood pressure values visually on adjacent windows on the same display. The SyncVision System is displayed on a monitor that is situated next to the native x-ray and IVUS image streams/FFR/iFR intravascular blood pressure values in either the catheterization lab or in the control room. The System provides a means to view the angiographic and IVUS image streams or FFR/IFR intravascular blood pressure values on a single display and automates the manual registration processes performed by the physician.
The System is comprised of a workstation, LCD monitor, keyboard, mouse and an isolation transformer installed in the control room of the catheterization lab, as well as a user input device (joystick) installed in the procedure room. The workstation is comprised of a PC equipped with an internal high-resolution dual channel frame grabber, a high-performance graphic processing card and a data acquisition card for ECG signal acquisition. The joystick allows the user to control all of the functions of the System from the procedure room. Cables connect the workstation to data sources which are an input device; the joystick previously mentioned, the x-ray imaging system, the IVUS imaging/FFR/iFR intravascular pressure measurement system, and the ECG monitor. Cables also connect the workstation to an existing output monitor (not supplied with the System) situated in the procedure room and provide the user in the procedure room with the visual output of the System side-by-side to the existing displays.
The SyncVision System is an image acquisition and processing workstation situated in the coronary catheterization lab and intended to be used during coronary catheterizations. The SyncVision System captures angiographic and intravascular ultrasound (IVUS) image streams and FFR/iFR displayed value streams and performs the following display functions for assisting the interventional cardiologist:
- During Lesion Evaluation; angiogram and IVUS image selection, quantitative coronary measurements (lesion diameters, length, % stenosis), vessel region enhancement and vessel region stabilization are performed instantly and online.
- . During Device Positioning, Deployment and Post-deployment; an on-line image stream derived from the native angiographic image stream that is enhanced and stabilized, is displayed side-by-side to the native angiographic and IVUS image streams.
- . Import and display of image data from endoluminal modalities, leading to a joint display of images acquired by x-Ray and endoluminal imaging and corresponding to same selected luminal locations or segments (also known as co-registration of such modalities).
The Co-Registration is an automation of a manual workflow process used by interventional cardiologists today. The manual procedure in the catheterization lab for utilizing x-Ray, IVUS
{5}------------------------------------------------
image streams and FFR/iFR displayed value streams calls for referencing the native x-ray and IVUS image streams/FFR/iFR displayed value streams of the patient's vasculature separately on different displays. This common configuration requires the cardiologist to estimate an identical location on the patient's vasculature on the two native image streams (x-ray and IVUS/FFR/iFR) on two separate displays. The SyncVision System automates this process by providing an option to display an identical anatomical location on both the x-Ray and IVUS image or FFR/iFR displayed value streams automatically on adjacent windows on the same display.
# INTENDED USE:
The SyncVision System is an image acquisition and processing system. It is indicated for use as follows:
- To provide quantitative information regarding the calculated dimensions of arterial segments.
- To enhance visualization of the stent deployment region.
- To be used in-procedure in the catheterization lab and off-line for post-procedural analysis.
- To obtain a co-registration of an angiographic x-ray image and IVUS images.
- To obtain a co-registration of an angiographic x-ray image and intravascular blood pressure values.
# COMPARISON OF CHARACTERISTICS:
This submission is for software and labeling changes, and expansion of the Indications for Use. The devices are identical in terms of design, materials, specifications, principles of operation, and fundamental scientific technology. There are no hardware or component changes. The software changes add new features, administrative, and infrastructure changes. A summary of the major changes and compatibility is provided in Table 1, Comparison Matrix of Changes, below with the changes in bold text.
| Technical<br>Attributes | Proposed SyncVision System<br>SW 4.0/HW 2.1 | Predicate Device<br>SyncVision System<br>SW3.2/HW2.1<br>K132558 | Comment |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The SyncVision System is an image<br>acquisition and processing system. It<br>is indicated for use as follows:<br>• To provide quantitative<br>information regarding the calculated<br>dimensions of arterial segments.<br>• To enhance visualization of the<br>stent deployment region.<br>• To be used in-procedure in the<br>catheterization lab and off-line for<br>post-procedural analysis.<br>• To obtain a co-registration of an<br>angiographic x-ray image and IVUS<br>images.<br>• To obtain a co-registration of an<br>angiographic x-ray image and | The SyncVision System is an<br>image acquisition and processing<br>system. It is indicated for use as<br>follows:<br>• To provide quantitative<br>information regarding the<br>calculated dimensions of arterial<br>segments.<br>• To enhance visualization of the<br>stent deployment region.<br>• To be used in-procedure in the<br>catheterization lab and off-line<br>for post-procedural analysis.<br>• To obtain a co-registration of<br>an angiographic x-ray image and<br>IVUS images. | Expansion of<br>the Indications<br>for Use to<br>include the<br>FFR/iFR Co-<br>Registration<br>functionality |
# Table 1: Comparison Matrix of Changes
{6}------------------------------------------------
| Technical<br>Attributes | Proposed SyncVision System<br>SW 4.0/HW 2.1 | Predicate Device<br>SyncVision System<br>SW3.2/HW2.1<br>K132558 | Comment |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| | intravascular blood pressure<br>values | | |
| Compatibility<br>with the<br>catheter-<br>laboratory<br>fluoroscopy<br>Systems | - Siemens Axiom Artis & Siemens<br>Artis Zee<br>- Philips Allura Xper FD10/20 &<br>Philips Allura Clarity<br>- GE Innova<br>- Toshiba Infinix<br>- Shimadzu Trinias/Voyager | - Siemens Axiom Artis & Siemens<br>Artis Zee<br>- Philips Allura Xper FD10/20 &<br>Philips Allura Clarity<br>- GE Innova<br>- Toshiba Infinix | Increased<br>through x-ray<br>qualification<br>process |
| Compatible<br>with the<br>following IVUS<br>system &<br>catheters | - Volcano s5/CORE System with SW<br>version 3.2.2 and above, and FM<br>software version 2.4.1<br>- Eagle Eye Platinum IVUS catheter<br>- Verrata pressure guide wire | - Volcano s5 / Core (mobile or<br>integrated configurations) with<br>SW version 3.2.2 and above<br>- Eagle Eye Platinum IVUS<br>catheter | Added<br>compatibility<br>due to the<br>new FFR/iFR<br>features |
| Image Source | Angiography/fluoroscopy/<br>Intravascular Ultrasound (IVUS)/<br>FFR/iFR pressure wires | Angiography/fluoroscopy/<br>Intravascular Ultrasound (IVUS) | Added new<br>FFR/iFR<br>pressure wire<br>as a source |
| Image Frame<br>Selection for<br>Measurement<br>S | Angiogram - The system<br>automatically presents an image<br>frame, and in most cases this will be<br>a frame in which the vessel is filled<br>with contrast media<br>IVUS – User selectable<br>FFR/iFR - user selectable | Angiogram - The system<br>automatically presents an image<br>frame, and in most cases this will<br>be a frame in which the vessel is<br>filled with contrast media<br>IVUS – User selectable | Added FFR/iFR<br>Co-<br>Registration<br>functionality |
# PERFORMANCE DATA:
Applicable testing was performed as required by the Quality System to evaluate the modifications to the SyncVision System software. The following tests were conducted:
- Software Verification and Validation
- Simulated Use Validation
The test results were found to be acceptable by the respective test plans and protocols.
Sterilization testing was not required as there are no sterile components or accessories for the modified SyncVision System which consists of hardware and software.
Biocompatibility is not applicable to this submission as there are no materials in the modified SyncVision System that come into direct or indirect contact with the patient. Contact with the user involves computer hardware accessories only, i.e., keyboard, mouse.
# CONCLUSION:
Completion of these tests concluded that the proposed SyncVision System is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.