EndyGel is an Electroconductive gel media used with external electrode to reduce the impedance (resistance to alternating current) of the contact between the electrode surface and the skin.
Device Story
EndyGel is an electroconductive gel applied to intact skin prior to electrode placement. It functions as a conductive interface to reduce impedance (resistance to alternating current) between the electrode surface and the patient's skin, facilitating the transfer of radiofrequency energy to tissue. The device is used in a clinical setting by healthcare professionals. It benefits the patient by ensuring efficient energy delivery during procedures requiring external electrodes.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility testing (ISO 10993-5 cytotoxicity, ISO 10993-10 irritation and sensitization), electrical performance testing (impedance measurement), and stability testing. No clinical data presented.
Technological Characteristics
Electroconductive gel composed of water, carbomer, triethanolamine, and preservatives. Non-sterile. Shelf life 2 years. Complies with ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 14971, and ISO 13485.
Indications for Use
Indicated for use as an electroconductive gel media with external electrodes to reduce impedance between the electrode surface and intact skin.
Regulatory Classification
Identification
Electroconductive media are the conductive creams or gels used with external electrodes to reduce the impedance (resistance to alternating current) of the contact between the electrode surface and the skin.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol of three stylized human profiles facing right, with flowing lines beneath them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the circular seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 7, 2017
EndyMed Medical Ltd. % Yoram Levy Osite General Manager Qsite 31 Haavoda St. Binyamina, 30500 Israel
Re: K161715
Trade/Device Name: EndyGel Regulation Number: 21 CFR 882.1275 Regulation Name: Electroconductive Media Regulatory Class: Class II Product Code: GYB Dated: December 29, 2016 Received: January 4, 2017
Dear Yoram Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michaelリ. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161715
Device Name EndyGel
Indications for Use (Describe)
Endy Gel is an Electroconductive gel media used with external electrode to reduce the impedance to alternating current) of the contact between the electrode surface and the skin.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Endymed 3DEEP Skin Science. The logo is purple and features the word "ENDYMED" in large, bold letters. Below the word "ENDYMED" is the phrase "3DEEP SKIN SCIENCE" in smaller letters. The letters are all capitalized.
# 510(K) SUMMARY
# EndyMed's EndyGel™ 510(k) Number K161715
| Applicant's Name: | EndyMed Medical Ltd. |
|-------------------|--------------------------|
| | 12 Leshem Street |
| | North Industrial Park |
| | Caesarea, 3088900 Israel |
| Tel: | (972)4-630-9100 |
| Fax: | (972)4-630-9101 |
- Contact Person: Yoram Levy, Qsite 31 Haavoda Street Binyamina, 30500 Israel Tel (972)4-638-8837; Fax (972)4-638-0510 Yoram@qsitemed.com
EndyGel™ Trade Name:
510(k) Summary Preparation Date: February 2, 2017
- Classification: Name: Electroconductive media Product Code: GYB Regulation No: 21 CFR 882.1275 Class: II Panel: Neurology
Indications for Use (IFU) Statement:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Endymed 3DEEP Skin Science. The word "ENDYMED" is in large, bold, purple letters. Below that, in smaller purple letters, are the words "3DEEP SKIN SCIENCE".
EndyGel™ is an Electroconductive gel media used with external electrode to reduce the impedance (resistance to alternating current) of the contact between the electrode surface and the skin.
## Device Description:
EndyMed's EndyGel™ is an Electroconductive gel media used with external electrode to enhance electrical conductivity by reducing the impedance (resistance to alternating current) of the radiofrequency signals path from the electrode surface between the electrode surface and the skin.
#### Predicate Devices:
Substantial equivalence to the following predicate device is claimed:
| Predicate/<br>Reference | Device Name | Manufacturer | 510k No. | Clearance<br>Date |
|-------------------------|-------------|-----------------------------------|----------|-------------------|
| Predicate | Electro-Gel | Electro-Cap International<br>Inc. | K111717 | Jan 05 2012 |
## Comparison table with the predicate
| | EndyMed EndyGel<br>(proposed) | Predicate device:<br>Electro-Gel<br>(K111717) |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class | Class II | same |
| Classification<br>Panel | Neurology | same |
| Product Code | GYB | Same |
| Regulation<br>Description | Electroconductive media | Same |
| Regulation<br>number | 21 CFR 882.1275 | Same |
| | EndyMed EndyGel<br>(proposed) | Predicate device:<br>Electro-Gel<br>(K111717) |
| Indications for<br>Use (IFU)<br>statement | EndyMed's <i>EndyGel</i> TM is an<br>Electroconductive gel media<br>used with external electrode<br>to reduce the impedance<br>(resistance to alternating<br>current) of the contact<br>between the electrode<br>surface and the skin. | The <b>Electro-Gel</b> device is<br>intended for use with<br>external electrodes as the<br>conductor between the scalp<br>and the electrodes. It also<br>reduces impedance<br>(resistance to alternating<br>current) between the<br>electrode surface and the<br>skin. |
| Principles of<br>Operation | EndyMed's <i>EndyGel</i> TM<br>functions as a conductor<br>between the skin and the<br>electrodes. EndyMed's<br><i>EndyGel</i> TM is intended to<br>used prior to Handpiece's<br>electrodes placement to<br>reduce the impedance<br>(resistance to alternating<br>current) of the contact<br>between the electrode<br>surface and the skin. The<br>energy is then transferred via<br>the gel to the patient's tissue. | Same |
| Body contact | Intact Skin | Same |
| | EndyMed EndyGel<br>(proposed) | Predicate device:<br>Electro-Gel<br>(K111717) |
| Biocompatibility | Complies with requirements of ISO 10993-1, including:<br>Cytotoxicity (ISO 10993-5),<br>Irritation and Sensitization<br>(ISO 10993-10) | Same |
| Sterilization | Provided non-sterile | Same |
| Shelf life | 2 years | 3 years |
| Impedance (at 1<br>MHz) | 527.68 | 549.71 |
| Conductive<br>material | Water (Aqua) with<br>immersed Triethanolamine | Water (Aqua) with<br>immersed Sodium chloride |
| Composition | • Water (solvent)<br>• Carbomer (gel forming)<br>• Triethanolamine<br>(Buffering)<br>• Methylisothiazolinone and<br>Methylcloroisothiazolinone (preservative) | • Water (solvent)<br>• Sodium Chloride<br>(preservative)<br>• Aragum T-1998<br>(thickner)<br>• Potassium Bitartrate<br>(Stabilizer)<br>• Glycerin (lubricant)<br>• Methylparaben<br>(preservative)<br>• Propylparaben<br>(preservative) |
| pH | 7.15-7.33 | 6.99 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the brand name ENDYMED in large purple font. Below the brand name is the text 3DEEP SKIN SCIENCE, also in purple font, but smaller than the brand name. The text is centered and the brand name has a trademark symbol on the upper right.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for EndyMed. The logo is purple and consists of the word "ENDYMED" in large, bold letters, with the letters slightly overlapping. Below the main logo, in smaller font, are the words "3DEEP SKIN SCIENCE".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ENDYMED. The logo is purple and consists of the word "ENDYMED" in large, bold letters. Below the word "ENDYMED" is the phrase "3DEEP SKIN SCIENCE" in smaller letters. There is a trademark symbol in the upper right corner of the logo.
#### Substantial Equivalent Discussion
The EndyGel, like its predicate device, is indicated for use as an Electroconductive gel media used with external electrode to reduce the impedance (resistance to alternating current) of the contract between the electrode surface and the skin.
The EndyGel and the predicate device have similar technological Electroconductive gel media.
The impedance of the EndyGel is within 4% and the pH within 2.2% compared to the predicate device. The difference in total impedance and in pH between the two gels is negligible and do not impact the safety and effectiveness of the EndyGel.
### Performance
The safety and efficacy of the EndyGel™ were established by a series of performance tests, including biocompatibility testing, electrical performance and stability.
The design of EndyGel™ was done in accordance with EndyMed Medical Ltd. quality management system and design controls per 21CFR820 and ISO 13485. V&V and compliance testing, including: Biocompatibility testing, Electrical Performance testing and Stability testing, were successfully conducted and did not raise any new safety questions or identify any new risks.
### Standards:
The EndyGel™ complies with the following standards
- ISO 14971-1:2009 Risk management for medical devices 1.
- 2. ISO13485:2003 Medical devices -- Quality management systems -- Requirements for 0regulatory purposes.
- ISO 10993-1:2009 Biological Evaluation of Medical Devices -- Part 1: Evaluation and 3. testing within a risk management process.
- ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests for in vitro 4. cytotoxicity
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for ENDYMED, a company specializing in 3DEEP skin science. The word "ENDYMED" is displayed in large, bold, purple letters, with a small trademark symbol in the upper right corner. Below the company name, the words "3DEEP SKIN SCIENCE" are written in a smaller, sans-serif font, also in purple. The logo is clean and modern, reflecting the company's focus on advanced skin technology.
- ISO 10993-10 Biological evaluation of medical devices -- Part 10: Tests for irritation న్. and skin sensitization
- EN ISO 15223-1 Medical devices -- Symbols to be used with medical device labels, 6. labelling and information to be supplied -- Part 1: General requirements
# Conclusion:
The EndyGel™, like its predicate devices, is indicated for use as an Electroconductive gel media used with external electrode to reduce the impedance (resistance to alternating current) of the contact between the electrode surface and the skin.
The EndyGel™, the predicate device and reference device have similar technological features of an Electroconductive gel media.
Results of tests, performed with the proposed EndyGel™ demonstrate that it is as safe and effective as its predicate device, without raising any new safety and/or effectiveness concerns.
Therefore the EndyGel™ is substantially equivalent to its predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.