K161574 · Ge Medical Systems, LLC · KPS · Aug 11, 2016 · Radiology
Device Facts
Record ID
K161574
Device Name
Discovery MI
Applicant
Ge Medical Systems, LLC
Product Code
KPS · Radiology
Decision Date
Aug 11, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The GE Discovery MI is a PET/CT system for producing attenuation corrected PET images. It is intended to be used by qualified health care professionals for imaging the distribution and localization of any positron-emitting radiopharmaceutical in a patient, for the assessment of metabolic (molecular) and physiologic function in patients, with a wide range of sizes and extent of disease, of all ages. Discovery MI is intended to image the whole body, head, heart, brain, lung, breast, bone, the gastrointestinal and lymphatic systems, and other organs. The images produced by the system may be used by physicians to aid in radiotherapy treatment planning, therapy guidance and monitoring, and in interventional radiology procedures. The images may also be used for precise functional and anatomical mapping (localization, registration, and fusion). When used with radiopharmaceuticals approved by the regulatory in the country of use, the raw and image data is an aid in; detection, localization, evaluation, diagnosis, staging, monitoring, and/or follow up, of abnormalities, lesions, tumors, inflammation, infection, organ function, disorders, and/or disease, such as, but not limited to, those in oncology, cardiology, and neurology. Discovery MI is also intended for stand-alone, diagnostic CT imaging in accordance with the stand-alone CT system's cleared indications for use.
Device Story
Discovery MI is a hybrid PET/CT imaging system. PET subsystem uses Lutetium-based Scintillator (LBS) detector arrays coupled with Silicon Photo Multipliers (SiPM) to detect gamma rays from injected radiopharmaceuticals. CT subsystem is a 64-detector row Revolution EVO. System transforms raw PET/CT data into attenuation-corrected PET images and fused anatomical/functional maps. Operated by qualified healthcare professionals in clinical settings. Output viewed on console for radiotherapy planning, therapy guidance, and diagnostic assessment of tumors, inflammation, and organ disorders. Benefits include precise localization and functional mapping. System supports static, gated, dynamic, and whole-body scanning modes.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, engineering performance evaluation, and phantom studies. Metrics included sensitivity, NECR, resolution, and lesion detectability (model observer study).
Technological Characteristics
Hybrid PET/CT. PET: Lutetium-based Scintillator (LBS) with SiPM detectors. CT: 64-detector row Revolution EVO. Scalable axial FOV architecture. Connectivity: DICOM. Standards: IEC 60601-1, IEC 60601-2-44, NEMA NU2-2012, NEMA XR-25, NEMA XR-28. Software: Moderate level of concern.
Indications for Use
Indicated for patients of all ages, sizes, and disease states for PET imaging of radiopharmaceutical distribution/localization to assess metabolic and physiologic function. Used for detection, staging, and monitoring of oncology, cardiology, and neurology conditions (e.g., tumors, inflammation, cardiac disease, dementia, TBI). Also indicated for stand-alone diagnostic CT imaging.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design of three human profiles facing to the right, with flowing lines extending from their heads.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 11, 2016
GE Medical Systems, L.L.C. % Ms. Helen Peng Regulatory Affairs Manager 3000 N Grandview Blvd. WAUKESHA WI 53188
Re: K161574
Trade/Device Name: Discovery MI Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS, JAK Dated: June 6, 2016 Received: June 7, 2016
Dear Ms. Peng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161574
Device Name Discovery MI
#### Indications for Use (Describe)
The GE Discovery MI is a PET/CT system for producing attenuation corrected PET images. It is intended to be used by qualified health care professionals for imaging the distribution and localization of any positron-emitting radiopharmaceutical in a patient, for the assessment of metabolic (molecular) and physiologic function in patients, with a wide range of sizes and extent of disease, of all ages.
Discovery MI is intended to image the whole body, head, heart, brain, lung, breast, bone, the gastrointestinal and lymphatic systems, and other organs. The images produced by the system may be used by physicians to aid in radiotherapy treatment planning, therapy guidance and monitoring, and in interventional radiology procedures. The images may also be used for precise functional and anatomical mapping (localization, registration, and fusion).
When used with radiopharmaceuticals approved by the regulatory in the country of use, the raw and image data is an aid in; detection, localization, evaluation, diagnosis, staging, monitoring, and/or follow up, of abnormalities, lesions, tumors, inflammation, infection, organ function, disorders, and/or disease, such as, but not limited to, those in oncology, cardiology, and neurology. Examples of which are:
Cardiology:
- Cardiovascular disease
- Myocardial perfusion
- Myocardial viability
- Cardiac inflammation
- Coronary artery disease
Neurology:
- Epilepsy
- Dementia, such as Alzheimer's disease, Lewy body dementia, Parkinson's disease with dementia, and frontotemporal dementia.
- Movement disorders, such as Parkinson's and Huntington's disease
- Tumors
- Inflammation
- Cerebrovascular disease such as acute stroke, chronic and acute ischemia
- Traumatic Brain Injury (TBI)
Oncology/Cancer:
- Non-Small Cell Lung Cancer
- Small Cell Lung Cancer
- Breast Cancer
- Prostate Cancer
- Hodgkin disease
- Non-Hodgkin lymphoma
- Colorectal Cancer
- Melanoma
{4}------------------------------------------------
Discovery MI is also intended for stand-alone, diagnostic CT imaging in accordance with the stand-alone CT system's cleared indications for use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are also some white swirls around the letters. The logo is simple and recognizable.
# Section 5: 510(k) Summary
Discovery MI
Aug 7, 2016 Update
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are three white, stylized water droplets or splashes surrounding the circle, positioned at the top and sides. The logo is simple, iconic, and easily recognizable as the brand symbol for General Electric.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVNESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):
| Date: | June 6, 2016 |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems, LLC<br>3000 North Grandview Blvd<br>Waukesha, WI 53188 |
| Primary Contact: | Helen Peng<br>Requlatory Affairs Manager<br>Phone 262-424-8222<br>Email: Hong.Peng@ge.com |
| Secondary Contacts: | John Jaeckle<br>Chief Requlatory Affairs Strategist<br>GE Healthcare<br>Tel: 262-424-9547<br>Email: john.jaeckle@ge.com |
| Device Trade Name: | Discovery MI |
| Common / Usual<br>Name: | PET/CT Imaging System |
| Classification Names | Emission Computed Tomography System per 21CFR 892.1200<br>Computed Tomography X-ray System per 21 CFR892.1750 |
| Device Classification | Class II |
| Product Code: | 90 KPS and 90 JAK |
| Predicate Device(s): | GE Discovery PET/CT 710 Clarity Edition (K133657) |
| Reference Devices: | GE Signa PET/MR (K142098)<br>GE Discovery IQ (K141477) |
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The frame has four water droplet-like shapes evenly spaced around the circle. The logo is presented in a blue color.
## Device Description:
The Discovery MI system is a PET/CT diagnostic imaging system combining a GE Positron Emission Tomography (PET) System and a GE Computed Tomography (CT) System.
The PET portion of the system uses a Lutetium-based Scintillator (LBS) detector. Scintillator crystal arrays are attached to Silicon Photo Multipliers (SiPM) to form detector units. The detector units are inherited from the reference device Signa PET/MR. Detector units are attached on a common support to form detector modules. The detector modules are arranged in a ring around the patient positioned inside of the PET gantry for detection of gamma rays generated as a result of PET radiopharmaceuticals injected into the patient.
The PET/CT system uses the full-featured multi-slice diagnostic CT subsystem with PET/CT post processing software to generate a map of the non-uniform attenuation in the patient. This attenuation map is then used for attenuation correction of the PET data. The CT image is also used for localization of the PET image in the patient anatomy by means of fusing the PET and CT images.
The Discovery MI system's major components are the PET gantry/detector, Revolution EVO CT system, patient table, operator console/workspace, computing hardware, power distribution unit, system software, and reconstruction software. The operator console and software provide control of the imaging (i.e. setting and confirming conditions of operation), image acquisition, dose display, reconstruction, viewing, post processing analysis, patient management, networking, and filmina. The system may include respiratory and cardiac gating capabilities, signal analysis and display equipment, patient and equipment supports, components and accessories. In addition to being installed as a complete PET/CT system, the Discovery MI may result from an upgrade to a Revolution EVO- based Discovery PET/CT 710.
The Discovery MI system provides scalable axial coverage for the PET detector. All configurations offer reference adult and pediatric protocols for both hybrid PET/CT and CT applications. The PET 3D data acquisition modes include Static, Gated, Dynamic, and Whole Body scanning. All of which can be acquired with List mode data. The system includes standard PET iterative reconstruction alqorithms. Q.Clear full-convergence, reqularized reconstruction is optionally available. Time of Flight (ToF) may be used for all PET reconstruction types.
The CT system is the commercially available 64-detector row Revolution Evo, which may also be used for stand-alone, diagnostic CT imaging. The CT system's acquisition modes include Axial, Cine, Helical (Volumetric), Cardiac, and Gated, for head, whole body, trauma, cardiac and vascular applications.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are white swirls around the letters. The logo is simple and recognizable.
### Intended Use:
The Discovery MI PET/CT system is intended for CT attenuation corrected, anatomically localized PET imaging of the distribution of positron-emitting radiopharmaceuticals. It is intended to image the whole body, head, heart, brain, lung, breast, bone, the gastrointestinal and lymphatic systems, and other organs. The system is also intended for stand-alone, diagnostic CT imaging.
#### Indications for Use:
The GE Discovery MI is a PET/CT system for producing attenuation corrected PET images. It is intended to be used by qualified health care professionals for imaging the distribution and localization of any positron-emitting radiopharmaceutical in a patient, for the assessment of metabolic (molecular) and physiologic function in patients, with a wide range of sizes and extent of disease, of all ages.
Discovery MI is intended to image the whole body, head, heart, brain, lung, breast, bone, the gastrointestinal and lymphatic systems, and other organs. The images produced by the system may be used by physicians to aid in radiotherapy treatment planning, therapy guidance and monitoring, and in interventional radiology procedures. The images may also be used for precise functional and anatomical mapping (localization, registration, and fusion).
When used with radiopharmaceuticals approved by the regulatory authority in the country of use, the raw and image data is an aid in; detection, localization, evaluation, diagnosis, staging, restaging, monitoring, and/or follow up, of abnormalities, lesions, tumors, inflammation, infection, organ function, disorders, and/or disease, such as, but not limited to, those in oncology, cardiology, and neurology. Examples of which are:
Cardiology:
- . Cardiovascular disease
- . Mvocardial perfusion
- . Myocardial viability
- . Cardiac inflammation
- " Coronary artery disease
#### Neurology:
- " Epileosv
- . Dementia, such as Alzheimer's disease, Lewy body dementia, Parkinson's disease with dementia, and frontotemporal dementia.
- י Movement disorders, such as Parkinson's and Huntington's disease
- . Tumors
- Inflammation י
- Cerebrovascular disease such as acute stroke, chronic and acute ischemia י
- י Traumatic Brain Injury (TBI)
r1
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script in the center. There are decorative swirls around the edge of the circle. The logo is simple and recognizable.
Oncology/Cancer:
- . Non-Small Cell Lung Cancer
- . Small Cell Luna Cancer
- . Breast Cancer
- . Prostate Cancer
- י Hodgkin disease
- . Non-Hodgkin lymphoma
- . Colorectal Cancer
- . Melanoma
Discovery MI is also intended for stand-alone, diagnostic CT imaging in accordance with the stand-alone CT system's cleared indications for use.
### Technological Characteristics
Discovery MI employs the same fundamental scientific technology, basic design, construction, materials, energy source, control mechanism, operating principles as the predicate device and the reference devices.
PET Detectors: The Discovery MI detector is comprised of detector units that include a LYSO Scintillator and a SiPM. These detectors have the same operating principle as the predicates and are the same as those on the Signa PET/MR reference device.
Scalable axial field of view (FOV) architecture: The Discovery MI has a scalable architecture that supports multiple axial FOVs using the same fundamental design found on the Discovery IQ reference device.
Image reconstruction: Discovery MI's image reconstruction subsystem includes updated, more powerful computational hardware to support faster reconstruction speeds and the increased datasets from larger axial FOVs and the ability for larger image pixel matrixes. There are not any new reconstruction algorithms.
## Determination of Substantial Equivalence:
Discovery MI has been tested and certified to comply with IEC 60601-1 Ed. 3.0 and applicable collateral and particular standards (e.g. IEC 60601-2-44), 21CFR 1020.30 and 1020.33, DICOM, NEMA NU2-2012, NEMA XR-25, and NEMA XR-28.
Discovery MI has successfully completed the required design control testing per our quality system. No new hazards were identified and no unexpected test results were obtained. The Discovery MI was designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. The following quality assurance measures have been applied to the development of the system:
- Risk Analysis
- Required Reviews ●
- Design Reviews
r1
{10}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are also some white swirls around the circle.
- . Software Development Lifecycle
- Testing on unit level (Module verification) ●
- . Integration testing (System verification)
- Performance testing (Verification) ●
- Safety testing (Verification) ●
- Simulated use testing (Validation)
GE believes Discovery MI system is of comparable type and substantially equivalent to Discovery PET/CT 710. The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.
The submission follow's FDA's December 3, 1998 Guidance for pre-market submission of NM, SPECT, and PET 510(k)s.
### Summary of Additional Testing
In addition to standards certification testing and quality system verification testing, additional engineering testing (i.e. non-clinical testing) was performed to provide the data to: substantiate product performance and claims; establish that the device is substantially equivalent to the predicate device; and demonstrate that the device is as safe and as effective as legally marketed devices.
The testing and results do not raise different questions of safety and effectiveness than the predicate device.
#### Non-Clinical Testing
The additional engineering performance evaluation testing used a variety of test methods and phantoms appropriate for the performance metric/claim to be tested and evaluated. Mathematical and physics analysis were performed to demonstrate that each performance metric/claim was successfully verified and substantiated.
The items additionally evaluated for the non-clinical testing included sensitivity, NECR, resolution, and lesion detectability. The lesion detectability evaluation included a model observer study.
Product performance and claims have been substantiated via scientific, established / standardized, engineering and physics-based performance testing.
#### Clinical Testing
Discovery MI is designed and built from existing and cleared systems, sub-system, components, and technologies of its Predicate Device (Discovery PET/CT 710) and those found on its reference devices, (Signa PET/MR and Discovery IQ).
The difference between various ring configurations of Discovery MI all constitute only an increase in size of the PET detector, while keeping the identical detector material, construction, performance, resolution, pitch, etc. This type of change supports using scientific, established/standardized, engineering/physics-based performance testing,
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script in the center. There are decorative swirls around the edge of the circle. The logo is simple and recognizable.
without inclusion of clinical images, to demonstrate that the device is as safe and as effective as the predicate and reference devices.
Given the above information and the type and scope of the changes, particularly that the new system uses PET detector modules from the cleared Signa PET/MR, and that its 510k included numerous clinical images, clinical testing is not required to demonstrate that the Discovery MI is as safe and as effective as the legally marketed predicate and reference devices.
### Conclusion:
Given that:
- The updated Indications for Use fall within the predicate device's intended use, do not raise new safety and effectiveness questions or issues, and does not represent any new intended uses.
- . Discovery MI has been tested and certified to comply with the standards above identified.
- . Verification testing along with additional advanced engineering testing demonstrated the Discovery MI's equivalent or superior performance to currently marketed PET/CT devices and is therefore as safe and effective.
- . The changes and testing did not raise new or different questions of safety and effectiveness, and no new hazards were created.
GE Healthcare believes that the Discovery MI is substantially equivalent to the predicate device, Discovery PET/CT 710, with supporting evidence and comparisons with its reference systems; Discovery IQ and Signa PET/MR, and hence is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.