Medline Powder-Free Nitrile Exam Glove Extended Cuff (Tested for Use with Chemotherapy Dr)

K161473 · Medline Industries, Inc. · LZA · Oct 19, 2016 · General Hospital

Device Facts

Record IDK161473
Device NameMedline Powder-Free Nitrile Exam Glove Extended Cuff (Tested for Use with Chemotherapy Dr)
ApplicantMedline Industries, Inc.
Product CodeLZA · General Hospital
Decision DateOct 19, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Story

Disposable, non-sterile, powder-free nitrile examination glove; extended cuff design. Used by healthcare professionals in clinical settings to provide a barrier against contamination between patient and examiner. Tested for resistance to chemotherapy drug permeation per ASTM D6978-05. Provides protection during handling of specific chemotherapy agents; breakthrough times vary by agent (e.g., Carmustine 40.4 min, Thiotepa 60.6 min). Single-use only; available in multiple sizes (S, M, L, XL).

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility evaluated per ISO 10993-10. Physical properties tested per ASTM D6319-10. Chemotherapy permeation tested per ASTM D6978-05.

Technological Characteristics

Material: Nitrile. Form factor: Extended cuff, blue color, non-sterile, single-use. Dimensions and physical properties (tensile strength, elongation) comply with ASTM D6319-10. Biocompatibility per ISO 10993-10. Permeation resistance tested per ASTM D6978-05.

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Also indicated for use during handling of chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three human profiles facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 19, 2016 Medline, Industries, Inc. Jennifer Mason Senior Regulatory Affairs Specialist One Medline Place Mundelein, Illinois 60060 Re: K161473 Trade/Device Name: Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: September 14, 2016 Received: September 16, 2016 Dear Jennifer Mason: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K161473 Device Name Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) Indications for Use (Describe) A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The following chemicals have been tested with these gloves. Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes Carmustine 3.3mg/ml - 40.4 min. Cisplatin 1.0mg/ml - No breakthrough up to 240 min. Cyclophosphamide 20.0mg.ml - No breakthrough up to 240 min. Dacarbazine 2.0mg/ml - No breakthrough up to 240 min. Doxorubicin Hydrochloride 2/0mg.ml - No breakthrough up to 240 min. Etoposide 20.0mg/ml - No breakthrough up to 240 min. Fluoruouracil 50mg/ml - No breakthrough up to 240 min. Methotrexate 25mg - No breakthrough up to 240 min. Mitoxantrone 2.0mg/ml - No breakthrough up to 240 min. Paclitaxel 5.0mg/ml - No breakthrough up to 240 min. Thiotepa 10mg/ml - 60.6 min. Please note that the following drugs have extremely low permeation times: Carmustine (3.3 mg/ml) has a minimum breakthrough time of 40.4 minutes; Thiotepa (10.0 mg/ml) has a minimum breakthrough time of 60.6 minutes. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image features the Medline logo, set against a blue background. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a stylized star or cross symbol. The symbol is white, contrasting with the blue background, and appears to be composed of four elongated triangles converging at a central point. A small gray rectangle is visible in the upper right corner of the image. Aedline Industries, Inc. One Medline Place Mundelein, IL 60060 # K161473 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)] #### Submitter / 510(k) Sponsor Medline Industries, Inc. 1 Medline Place Mundelein, IL 60060 Registration Number: 1417592 #### Contact Person Jennifer Mason Senior Regulatory Affairs Specialist Phone: 847-643-3652 Email: JAMason@medline.com Summary Preparation Date September 14, 2016 Type of 510(k) Submission Traditional #### Device Name / Classification Name of Device: Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) Proprietary Name: Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) Common Name: Patient Examination Glove Classification Name: Patient Examination Glove Product Code: LZA, LZC Classification Panel: General Hospital Regulatory Class: I Regulation #: 21 CFR 880.6250 #### Predicate Device EMG Blue Nitrile Examination Gloves Powder Free with Tested for Use with Chemotherapy Drugs K141623 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is in white, bold, sans-serif font, and to the right of the word is a white starburst symbol. There is a gray bar on the right side of the image. ledline Industries, Inc. One Medline Place Mundelein, IL 60060 #### Device Description The Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) are single use only, disposable gloves intended for medical purposes to be worn on the hands or fingers of examiners. The gloves are powder-free and are made of nitrile with a blue colorant. The gloves are offered non-sterile and are available in small, medium, large and extra-large sizes. Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. The subject device was tested according to ASTM D6978, Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Minimum breakthrough times were determined for a wide range of chemotherapy agents. #### Indications for Use A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs. | Chemotherapy Drugs | Breakthrough Times | |------------------------------------|--------------------------------| | Carmustine 3.3mg/ml | 40.4 min. | | Cisplatin 1.0mg/ml | No breakthrough up to 240 min. | | Cyclophosphamide 20.0mg/ml | No breakthrough up to 240 min. | | Dacarbazine 2.0mg/ml | No breakthrough up to 240 min. | | Doxorubicin Hydrochloride 2.0mg/ml | No breakthrough up to 240 min. | | Etoposide 20.0mg/ml | No breakthrough up to 240 min. | | Fluorouracil 50mg/ml | No breakthrough up to 240 min. | | Methotrexate 25mg/ml | No breakthrough up to 240 min. | | Mitoxantrone, 2.0mg/ml | No breakthrough up to 240 min. | | Paclitaxel 5.0mg/ml | No breakthrough up to 240 min. | | Thiotepa 10mg/ml | 60.6 min. | The following chemicals have been tested with these gloves Please note that the following drugs have extremely low permeation times: Carmustine (3.3 mg/ml) has a minimum breakthrough time of 40.4 minutes; Thiotepa (10.0 mg/ml) has a minimum breakthrough time of 60.6 minutes. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in a sans-serif font, with a stylized white starburst symbol to the right of the text. The background is a solid dark blue color. The logo is simple and modern, and the colors are clean and professional. Medline Industries, Inc. One Medline Place Mundelein, IL 60060 # Summary of Technological Characteristics # TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES | Device Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis | |-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------| | Product Name | Medline Powder-Free Blue<br>Nitrile Exam Glove Extended<br>Cuff (Tested for Use with<br>Chemotherapy Drugs) | EMG Blue Nitrile<br>Examination Gloves Powder<br>Free with tested for use with<br>chemotherapy drugs | Different | | 510(k) Reference | K161473 | K141623 | N/A | | Product Owner | Medline Industries, Inc. | Eco Medi Glove SDN BHD | Different | | Product Code | LZA, LZC | LZA, LZC | Same | | Intended Use | A powder-free patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or<br>finger to prevent<br>contamination between patient<br>and examiner and for use with<br>chemotherapy drugs. In<br>addition these gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM<br>D6978-05 Standard Practice<br>for Assessment of Medical<br>Glove to Permeation by<br>Chemotherapy Drugs. | A powder-free patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or<br>finger to prevent<br>contamination between patient<br>and examiner and for use with<br>chemotherapy drugs. In<br>additional these gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM<br>D6978-05 standards Practice<br>for assessment of Medical<br>Glove to Permeation by<br>Chemotherapy drugs. | Same | | Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same | | Materials | Nitrile | Nitrile | Same | | Color | Blue | Blue | Same | | Design Configurations | Small<br>Medium<br>Large<br>Extra-Large | Small<br>Medium<br>Large<br>Extra-Large | Same | | Dimensions - Length | Complies with ASTM D6319-<br>10<br>270mm min. | Complies with ASTM D6319-<br>10<br>270mm min. | Same | | Dimensions - Width | Complies with ASTM D6319-<br>10<br>Small - 80 ±10mm min.<br>Medium - 95 ±10mm min. | Complies with ASTM D6319-<br>10<br>Small - 85 ±5mm min.<br>Medium - 95 ±5mm min. | Same | | | | | | | | Large - 110±10mm min. | Large - 105±5mm min. | | | | Extra-Large 120±10mm min. | Extra-Large 115±5mm min. | | | Dimensions - Thickness | Complies with ASTM D6319-<br>10 | Complies with ASTM D6319-<br>10 | Same | | | Palm - 0.10mm min. | Palm - 0.10mm min. | | | | Finger - 0.10mm min. | Finger - 0.10mm min. | | | Physical Properties -<br>Tensile | Complies with ASTM D6319-<br>10 | Complies with ASTM D6319-<br>10 | Same | | | Before Aging - 14 MPa, min. | Before Aging - 14 MPa, min. | | | | After Aging - 14 MPa, min. | After Aging - 14 MPa, min. | | | Physical Properties -<br>Elongation | Complies with ASTM D6319-<br>10 | Complies with ASTM D6319-<br>10 | Same | | | Before Aging – 500% min. | Before Aging – 500% | | | | After Aging - 400% min. | After Aging - 400% | | | Freedom from Holes | AQL 1.5 Inspection level G-2<br>Meets ASTM D6319-10 | AQL 2.5 Inspection level G-1<br>Meets ASTM D6319-00a | Similar | | Residual Powder | Complies with ASTM D6319-<br>10 < 2.0 mg | Complies with ASTM D6319-<br>00a <2.0 mg | Same | | Biocompatibility | ISO 10993-10:2010/(R) 2014<br>Under the conditions of the<br>test, not an irritant | Under the conditions of the<br>test, not an irritant | Same | | | ISO 10993-10:2010/(R) 2014<br>Under the conditions of the<br>test, not a skin sensitizer | Under the conditions of the<br>test, not a skin sensitizer | | | Prescription vs. OTC | OTC | OTC | Same | | Sterile vs. Non-Sterile | Non-Sterile | Non-Sterile | Same | | Single Use vs. Reusable | Single use | Single use | Same | | Tested for Use with<br>Chemotherapy Drugs | Yes ASTM 6978-05 | Yes ASTM D6978-05 | Same | | Chemotherapy Drugs<br>Tested | Carmustine 3.3mg/ml<br>Cisplatin 1.0mg/ml | Carmustine 3.3mg/ml | Similar | | | Cyclophosphamide 20.0mg/ml<br>Dacarbazine 2.0mg/ml | Cyclophosphamide 20.0mg/ml | | | | Doxorubicin Hydrochloride<br>2mg/ml | Doxorubicin Hydrochloride<br>2mg/ml | | | | Etoposide 20.0mg.ml | Etoposide 20.0mg.ml | | | | Fluorouracil 50 mg/ml<br>Methotrexate 25mg/ml<br>Mitoxantrone 2.0mg/ml | Fluorouracil 50 mg/ml<br>Methotrexate 25mg/ml | | | | Paclitaxel 5.0mg/ml<br>Thiotepa 10mg/ml | Paclitaxel 5.0mg/ml<br>Thiotepa 10mg/ml | | | | | Cytarabine 100mg.ml | | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image features the Medline logo against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a stylized starburst symbol. The starburst is white and appears to be composed of four elongated triangles converging at a central point. The overall design is clean and corporate, suggesting a professional and established brand identity. Medline Industries, Inc. One Medline Place Mundelein, IL 60060 {7}------------------------------------------------ Image /page/7/Picture/0 description: The image features the logo of Medline, a healthcare company. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a stylized star-like symbol. The star is white and has four points, with the bottom point extending further down than the others. The background is a solid dark blue color, providing a strong contrast to the white text and symbol. Medline Industries, Inc. One Medline Place Mundelein, IL 60060 #### Summary of Non-Clinical Testing The biocompatibility evaluation for the Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) was conducted in accordance with ANSVAAMI/ISO 10993-1:2009 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA. The Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) is classified as a surface contacting device with a limited contact duration of less than 24 hours. The following tests were performed to evaluate the biocompatibility of the Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs): - ISO 10993-10: Primary Skin Irritation Test ● - ISO 10993-10: Closed Patch Sensitization Test #### Performance Testing (Bench) Physical performance qualities were evaluated per ASTM D6319, Standard Specification for Nitrile Examination Gloves for Medical Application. Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. The gloves were tested according to ASTM D6978, Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Minimum breakthrough times were determined for a wide range of chemotherapy drugs. ## Summary of Clinical Testing Not applicable. #### Conclusion Based on the non-clinical performance testing Medline Industries, Inc. concludes that the Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) are as safe, as effective, and perform as well as the predicate device, EMG Blue Nitrile Medical Examination Gloves Powder Free with Tested for Use with Chemotherapy Drugs Labeling Claim (K141623). Therefore, the Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) are substantially equivalent to the predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...