K161179 · Air Liquide Healthcare America Corporation · ECX · Apr 13, 2017 · Anesthesiology
Device Facts
Record ID
K161179
Device Name
Intelli-Ox
Applicant
Air Liquide Healthcare America Corporation
Product Code
ECX · Anesthesiology
Decision Date
Apr 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2700
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Intelli-OX is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MR Conditional and suitable for use during MR imaging for MRI systems up to 3.0 Tesla. When administered by properly trained personnel for oxygen deficiency and resuscitation, the Intelli-Ox is for emergency use only. For all other medical applications, the device is Rx only.
Device Story
Intelli-OX is a portable oxygen delivery system supplying USP oxygen; consists of integrated valve-regulator, flow meter, medical aluminum cylinder, handle, and protective shroud. Device provides user-selectable flow rates via indexed flow meter; includes DISS-1240 connection for high-flow delivery. Used in clinical settings by medical personnel to provide oxygen therapy; MR Conditional design allows use during MRI procedures up to 3.0 Tesla. Integrated unit eliminates need to mount separate regulators, facilitating faster patient care. Benefits include portability and compatibility with MRI environments.
Clinical Evidence
Bench testing only. MRI performance testing conducted per ASTM F2052-15 and ISO TIR 10974 (2012) to validate MR Conditional status.
Technological Characteristics
Integrated valve-regulator, flow meter, and medical aluminum cylinder. Features include hose barb connection, protective shroud, carrying handle, and content gauge. MR Conditional up to 3.0 Tesla. Complies with ASTM F2052-15 and ISO TIR 10974.
Indications for Use
Indicated for adults requiring supplemental oxygen; suitable for use during MR imaging up to 3.0 Tesla; emergency use for oxygen deficiency and resuscitation by trained personnel; Rx only for other medical applications.
Regulatory Classification
Identification
A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a wing-like shape above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2017
Air Liquide Healthcare America Corporation Steve Miller Director, Quality and Regulatory Affairs 12800 West Little York Road Houston, Texas 77041-2280
Re: K161179
Trade/Device Name: Intelli-OX Regulation Number: 21 CFR 868.2700 Regulation Name: Pressure Regulator Regulatory Class: Class I Product Code: ECX Dated: February 13, 2017 Received: February 21, 2017
Dear Steve Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161179
Device Name Intelli-OX
Indications for Use (Describe)
The Intelli-OX is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MR Conditional and suitable for use during MR imaging for MRI systems up to 3.0 Tesla. When administered by properly trained personnel for oxygen deficiency and resuscitation, the Intelli-Ox is for all other medical applications, the device is Rx only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of a teal square with a stylized white "A" inside. To the right of the square is the company name "AIR LIQUIDE" in red, with a thin blue line above and below the text. The logo is simple and modern, with a focus on the company's name and a clean design.
# 510 (k) Summary Intelli-Ox 510(k) Number:
The following summary is provided in accordance with 21 CFR 807.92:
#### APPLICANT'S/SUBMITTER'S INFORMATION 1.
Air Liquide Healthcare America Corporation 12800 West Little York Road Houston, TX 77041
| Direct Phone: | 800-624-8000 |
|--------------------------------|-----------------------|
| Internet: | www.us.airliquide.com |
| Establishment Registration No: | 3003764448 |
| Contact: | Steven Miller |
| Contact's Phone: | 713-896-2280 |
| Contact's Fax: | 713-803-7427 |
- 2. DATE April 20, 2016
#### 3. DEVICE INFORMATION
Trade/Proprietary Name: Intelli-Ox (E cylinders)
| Common Name: | Portable Oxygen Delivery System |
|------------------------|--------------------------------------------------------------|
| Device Name: | Cylinder, Compressed Gas, and Integrated Valve-<br>Regulator |
| Classification Panel: | Anesthesiology |
| Classification Number: | Class 1 |
| Regulation: | 868.2700 |
| Product Code(s): | ECX |
Prior Submission: No prior submission for the subject device
{5}------------------------------------------------
AIR LIQUIDE
## DEVICE CLASSIFICATION
Empty Compressed gas cylinders and compressed gas cylinder with valve assemblies are under product code ECX and reviewed by the Anesthesiology Devices Branch, Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices.
#### 5. PREDICATE DEVICE(S)
The following device is the legally marketed device to which equivalence is being claimed:
- Intelli-Ox, K143060, Air Liquide Healthcare
#### DEVICE DESCRIPTION 6.
The Intelli-OX portable oxygen delivery system that supplies Oxygen USP using a device comprised of an integrated valve-regulator, flow meter and medical aluminum cylinder with handle and shroud all integrated into a single unit. A range of user-selectable flow setting is available with the user being able to control the flow rate, including low flows that may be clinically appropriate for certain classes of patients. An additional DISS-1240 connection provides standard high flow oxygen delivery. When administered by properly trained personnel for oxygen deficiency and resuscitation. Intelli-OX is for emergency use only. For all other medical applications, the device is Rx only.
Key specifications include hose barb connection, protective shroud, carrying handle, easy to read content gauge, indexed flow meter, and integrated valveregulator. This design allows medical personnel the ability to provide patient care and treatment sooner without delays caused by the need to mount a conventional regulator. The device is MR Conditional and suitable for use during MR imaging for MRI systems up to 3.0 Tesla.
#### 7. INDICATIONS FOR USE
The intended use for the proposed device, as described in its labeling, will include a statement that the device is MR Conditional and suitable for use during MR imaging for MRI systems up to 3.0 Tesla.
The Intelli-OX is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MR Conditional and suitable for use during MR imaging for MRI systems up to 3.0 Tesla. When administered by properly trained personnel for oxygen deficiency and resuscitation, the Intelli-Ox is for emergency use only. For all other medical applications, the device is Rx only.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of a blue square with a stylized "AL" in white, followed by the words "AIR LIQUIDE" in red. There are thin blue lines above and below the text.
#### 8. COMPARISION WITH PREDICATE DEVICE
The Intelli-OX with MR Conditional Indication is the same unit as the predicate device. The fundamental scientific technology of the proposed device is unchanged from the legally marketed predicate devices. The Intelli-OX with MR Conditional Indication adds a new indication for use. The new indication for use reflects the device being MR Conditional and suitable for use during MR imaging for MRI systems up to 3.0 Tesla. The device labeling for the Intelli-OX with MR Conditional Indication has been revised to reflect the additional indication for use. The change to the indications for use and device labeling does not raise any new questions of safety and effectiveness.
#### 9. TESTING DATA
MRI performance testing was completed and documentation was provided to FDA in support of the substantial equivalence determination.
### 10. SUMMARY OF NON-CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
The Intelli-Ox systems are MR Conditional and comply with:
- ASTM F2052-15, "Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment"
- ISO TIR 10974, 2012, Clause 21, "Assessment of the Safety Magnetic ● Resonance Imaging for Patients with an Active Implantable Medical Device"
#### STATEMENT OF SUBSTANTIAL EQUIVALENCE 11.
Based upon the MRI performance testing that is documented in this submission, the manufacturer believes that the proposed Intelli-Ox device is substantially equivalent to the predicate device, and does not raise any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.