K143060 · Air Liquide Healthcare · ECX · May 21, 2015 · Anesthesiology
Device Facts
Record ID
K143060
Device Name
Intelli-Ox
Applicant
Air Liquide Healthcare
Product Code
ECX · Anesthesiology
Decision Date
May 21, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2700
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Intelli-Ox is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. When administered by properly trained personnel for oxygen deficiency and resuscitation, the Intelli-Ox is for emergency use only. For all other medical applications, the device is Rx only.
Device Story
Portable oxygen delivery system; integrates valve-regulator, flow meter, and aluminum oxygen cylinder into single unit. Input: compressed medical oxygen. Operation: single-stage piston-style regulator controls flow; user-selectable flow settings; DISS-1240 connection for high-flow delivery. Output: supplemental oxygen. Used in clinical or emergency settings by trained personnel or physicians. Digital content gauge provides real-time volume information to the operator. Device facilitates rapid patient care by eliminating need to mount separate regulators. Benefits: portable, integrated design for efficient oxygen administration.
Clinical Evidence
Bench testing only. Device performance verified through compliance with ISO 14971, ISO 13485, IEC 60513, IEC 61326-2-6, ISO 21730, IEC 61000-6-4, IEC 60601-1-2, and ISO 10651-3. Testing included electromagnetic compatibility and lithium battery safety (UN Manual of Tests and Criteria).
Technological Characteristics
Integrated valve-regulator, flow meter, and aluminum cylinder. Single-stage piston-style regulator. Max pressure 3,000 psi. Digital content gauge. Hose barb and DISS-1240 connections. IP55 rating. Materials: Aluminum. Standards: ISO 14971, ISO 13485, IEC 60601-1-2, ISO 10651-3.
Indications for Use
Indicated for adults requiring supplemental oxygen, including emergency use for oxygen deficiency and resuscitation by trained personnel.
Regulatory Classification
Identification
A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three profiles of human faces incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 21, 2015
Air Liquide Healthcare Ms. Angie Beyer Compliance Specialist 2700 Post Oak Blvd., Suite 325 Houston, Texas 77056
Re: K143060
Trade/Device Name: Intelli-Ox Regulation Number: Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Product Code: ECX Dated: April 20, 2015 Received: April 21, 2015
## Dear Ms. Beyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Beyer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Intelli-Ox
Indications For Use:
The Intelli-Ox is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. When administered by properly trained personnel for oxygen deficiency and resuscitation, the Intelli-Ox is for emergency use only. For all other medical applications, the device is Rx only.
Prescription Use ____X___ AND/OR Over-The-Counter Use ________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Air Liquide. The logo consists of a teal square with a stylized "AL" inside, followed by the words "AIR LIQUIDE" in red. The logo is enclosed in a light blue rectangle.
# 510 (k) Summary Intelli-Ox Portable Oxygen System 510(k) Number: _K143060
Traditional 510(K): Premarket Notification: Intelli-Ox Portable Oxygen System
#### APPLICANT'S/SUBMITTER'S INFORMATION 1.
Air Liquide Healthcare America Corporation 12800 West Little York Road Houston, TX 77041
| Direct Phone: | 800-624-8000 |
|--------------------------------|-----------------------|
| Internet: | www.us.airliquide.com |
| Establishment Registration No: | 3003764448 |
| Contact: | Angie Beyer |
| Contact's Phone: | 713-624-8268 |
| Contact's Fax: | 713-803-1246 |
#### 2. DATE
May 12, 2015
#### DEVICE INFORMATION 3.
Trade/Proprietary Name: Intelli-Ox
| Common Name: | Portable Oxygen Delivery System |
|------------------------|--------------------------------------------------------------|
| Device Name: | Cylinder, Compressed Gas, and Integrated Valve-<br>Regulator |
| Classification Panel: | Anesthesiology |
| Classification Number: | Unclassified |
| Product Nomenclature: | Cylinder, Compressed Gas, and Integrated Valve-<br>Regulator |
| Product Code(s): | ECX |
| Prior Submission : | No prior submission for the subject device. |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of a blue square with a white lambda symbol inside, followed by the words "AIR LIQUIDE" in red, sans-serif font. The logo is enclosed in a light blue box.
#### DEVICE CLASSIFICATION 4.
Empty Compressed gas cylinders and compressed gas cylinder with valve assemblies are unclassified devices under product code ECX and reviewed by the Anesthesiology Branch, Division of Anesthesiology and Respiratory, General Hospital, Infection Control, and Dental Devices.
Individually, gas cylinder pressure regulators and gas pressure gauges are Class 1 devices and exempted from pre-market notification, but are part of the above mentioned assembly.
#### 5. PREDICATE DEVICE(s)
The predicate device for the Intelli-Ox is the EZ-OX Plus, which was cleared under 510(k) K053117.
#### DEVICE DESCRIPTION 6.
The Intelli-Ox portable oxygen system is a solution for supplying Oxygen USP using a device comprised of an integrated valve-regulator, flow meter and medical Oxygen aluminum cylinder with handle and shroud all integrated into a single unit. A range of user-selectable flow setting is available with the user being able to control the flow rate, including low flows that may be clinically appropriate for certain classes of patients. An additional DISS-1240 connection provides high flow oxygen delivery. When administered by properly trained personnel for oxygen deficiency and resuscitation, the Intelli-Ox is for emergency use only. For all other medical applications, the device is Rx only.
Key specifications include hose barb connection, protective shroud, carrying handle, easy to read digital content gauge, indexed flow meter, integrated valveregulator and user instructions. This design allows medical personnel the ability to provide patient care and treatment sooner without delays caused by the need to mount a conventional regulator.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of a blue square with a white lambda symbol inside, followed by the words "AIR LIQUIDE" in red. The logo is surrounded by a thin blue border.
#### 7. INDICATIONS FOR USE
The intended use for the modified device, as described in its labeling, has not changed as a result of the modification.
The Intelli-OX is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. When administered by properly trained personnel for oxygen deficiency and resuscitation, the Intelli-Ox is for emergency use only. For all other medical applications, the device is Rx only.
#### COMPARISON WITH PREDICATE DEVICE 8.
Very similar to the original EZ-OX Plus, the newly designed Intelli-Ox has been modified by replacing the analog gauge with a digital gauge. The gauge in both devices does not control the administration of flow, but rather is for information purposes. Both the Indications for Use and the Fundamental scientific technology of the proposed device are unchanged from the legally marketed device (predicate). A summary comparison of technological characteristics, including design and materials is provided in the table below:
| Parameter | EZ-OX Plus | Intelli-Ox |
|-----------------------------|-----------------|-----------------|
| Valve/Regulator | | |
| Low Flow Settings | Yes (≥ .5L) | Yes (≥ .5L) |
| Flow Between Settings | No | No |
| Cylinder on/off | No | No |
| Filling Port | Active | Active |
| Contents Gauge | Active | Active |
| Filters | 3 | 3 |
| Pressure Design | 3,000 psi (max) | 3,000 psi (max) |
| Single stage piston style | Yes | Yes |
| Guard | | |
| Hand Grip | 1 grip | 1 grip |
| Access Ports | Yes | Yes |
| Flow selector/ hose barb/ | Yes | Yes |
| gauge aligned | | |
| Color | Green | Green |
| Height (guard and | 7" | 7" |
| integrated valve-regulator) | | |
| Cylinder | | |
| Cylinder Sizes | E | E |
| Weight (E) (product) | 950 gr | 950gr |
| Materials/construction | Aluminum | Aluminum |
The technological characteristics of the EZ-OX Plus (predicate device) and the Intelli-Ox (modified device) are the same.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Air Liquide. The logo consists of a blue square with a white lambda symbol inside, followed by the words "AIR LIQUIDE" in red, uppercase letters. The logo is enclosed in a thin blue rectangle.
### SUMMARY OF NON-CLINICAL TESTING AS BASIS FOR 9. SUBSTANTIAL EQUIVALENCE
The Intelli-Ox system complies with the following industry and international standards that are applicable to medical gas regulators:
- · ISO 14971: Medical Device Risk Management
- · ISO 13485: Medical Device Quality Management System
- · IEC 60513 ed. 2.0 (1994-01): Fundamental Aspects of Safety Standards for Medical Electrical Equipment
- · United Nations Manual of tests and Criteria, 4th Edition revised with Amendment 2, Part III, Classification Procedures, Test Methods, and Criteria Relating to Class 3, Class 4 Division 5.1 and Class 9, Section 38.3, Lithium Batteries
- · International Protection Rating (IP) of IP55
- · IEC 61326-2-6-2 2nd Ed. 2012, EMC requirements, Medical Equipment
- · ISO 21730 2nd Ed. 2007, Recommendations for Electromagnetic Compatiblity
- · IEC 61000-6-4 2nd Ed. 2006-07, Electromagnetic Compatibility
- · IEC 60601-1-2 Electrical Equipment Healthcare Application
- · IEC 60601-1-2 Ed. 4.0. Medical Electrical Equipment
- · ISO 10651-3 Electromagnetic Compatibility
- · Title 21 of the U.S. Code of Federal Regulations (21 CFR) Parts 210 and 211
#### 10. STATEMENT OF SUBSTANTIAL EQUIVALENCE
Based upon the performance testing and compliance with voluntary standards, the manufacturer believes that the Intelli-OX portable oxygen delivery system is substantially equivalent to the predicate device, and does not raise any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.