SpotLight CT

K161066 · Arineta , Ltd. · JAK · Aug 10, 2016 · Radiology

Device Facts

Record IDK161066
Device NameSpotLight CT
ApplicantArineta , Ltd.
Product CodeJAK · Radiology
Decision DateAug 10, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2

Intended Use

The Spotlight CT Computed Tomography X-Ray system is intended for head, body, cardiac and vascular X-ray Computed Tomography applications. The SpotLight CT Computed Tomography X-ray is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission projection data taken at different angles. The system has the capability to image whole organs, including the heart, in a single rotation. The system may acquire data using Axial, Cine and Cardiac scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. The system is indicated for X-ray Computed Tomography imaging of organs that fit in a 25cm field of view, including cardiac and vascular CT imaging. The device output is useful for diagnosis of disease or abnormality and for planning of therapy procedures.

Device Story

SpotLight CT is a third-generation, multi-slice (192 detector rows), dual-tube rotate-rotate CT scanner. Input: X-ray transmission data acquired at different angles via two alternating X-ray sources (Gemini tubes) displaced in the z-axis. Operation: Electrostatic grid control alternates X-ray beams; gantry rotates at up to 0.24s/rotation; data acquired by single array detector. Output: Cross-sectional images (trans-axial and reformatted planes) produced via Stereo CT reconstruction algorithm. Used in clinical settings by radiologists/cardiologists for whole-organ imaging (including heart), cardiac/vascular CT, and therapy planning. Benefits: 140mm detector coverage, 120msec temporal resolution, 0.31mm spatial resolution, and reduced cone beam artifacts. Healthcare providers use images for diagnostic assessment and procedural planning.

Clinical Evidence

Clinical study on 38 subjects evaluated cardiac (coronary CTA, function, calcium, TAVI prep), abdominal, and head/inner ear images. Four US-certified radiologists/cardiologists assessed diagnostic value and image quality. Results confirmed diagnostic image quality, safety, and efficacy. Animal testing (porcine model) and bench testing (image quality, temporal resolution, dose performance) also performed.

Technological Characteristics

Third-generation rotate-rotate CT; 192 detector rows; dual-tube (Gemini) X-ray sources with electrostatic grid control; 450mm SAD; 140mm z-axis coverage; 250mm diagnostic FOV. Complies with AAMI/ES 60601-1, IEC 60601-1 Ed. 3, NEMA XR-25, XR-28, XR-29. Includes antiscatter collimator. Software performs Stereo CT reconstruction.

Indications for Use

Indicated for X-ray CT imaging of organs fitting a 25cm field of view, including cardiac and vascular applications, in patients of all ages. Used for diagnosis of disease/abnormality and therapy planning.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the logo is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Auqust 10, 2016 Arineta Ltd. % Mr. Ehud Dafni CEO 15 Halamish Street Caesarea, 3088900 ISRAEL Re: K161066 Trade/Device Name: SpotLight CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: July 11, 2016 Received: July 14, 2016 Dear Mr. Dafni: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161066 Device Name SpotLight CT #### Indications for Use (Describe) The SpotLight CT Computed Tomography X-ray is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles. The system has the capability to image whole organs, including the heart, in a single rotation. The system may acquire data using Axial, Cine and Cardiac CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. The system is indicated for X-ray Computed Tomography imaging of organs that fit in a 25cm field of view, including cardiac and vascular CT imaging. The device output is useful for diagnosis of disease or abnormality and for planning of therapy procedures. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY ### Arineta Ltd.'s SpotLight CT This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92 - Date prepared: April 12, 2016 - Submitted by: Arineta Ltd. 15 Halamish Street Caesarea 3088900 Israel Tel: +972-4-6374000 Fax: +972-4-6277006 ### PRODUCT IDENTIFICATION | Device Trade Name: | SpotLight CT | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Computed Tomography X-ray System | | Classification Name: | Computed Tomography X-ray System<br>(21CFR 892.1750) | | Product Code: | 90-JAK | | Predicate Device: | GE Healthcare Revolution CT K133705 | | References Devices: | Siemens Somatom Definition Flash (K121072),<br>Elscint Excel CT Twin (K915738), Philips Mx8000<br>(K010817), Toshiba Equilion One Vision (K142465)<br>Philips Brilliance iCT (K131773) | #### INTENDED USE The Spotlight CT Computed Tomography X-Ray system is intended for head, body, cardiac and vascular X-ray Computed Tomography applications. # INDICATIONS FOR USE The SpotLight CT Computed Tomography X-ray is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission projection data taken at different angles. The system has the capability to image whole organs, {4}------------------------------------------------ including the heart, in a single rotation. The system may acquire data using Axial, Cine and Cardiac scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. The system is indicated for X-ray Computed Tomography imaging of organs that fit in a 25cm field of view, including cardiac and vascular CT imaging. The device output is useful for diagnosis of disease or abnormality and for planning of therapy procedures. # DEVICE DESCRIPTION The SpotLight CT is a third generation rotate-rotate CT scanner, designed and built based on technologies and principles of operation of the predicate device and other legally marketed CT scanners. The SpotLight CT is a multi-slice (192 detector rows), dual tube CT scanner consisting of a gantry, patient table, operator console, power distribution unit (PDU) and interconnecting cables. The system includes image acquisition hardware, image acquisition and reconstruction software for operator interface and image handling. # Technological Characteristics / Principles of Operation The system generates images through computed reconstruction of data acquired at different view angles of the rotating gantry, while irradiating the scanned subject by two alternating X-ray sources. The gantry rotates at up to 0.24 seconds per rotation and can acquire up to 240 slices of image data in a single axial scan with a maximum total coverage of 140 mm in the z direction. While the system covers field of view (FOV) of 450 mm. the radiation outside 250 mm (or 160 mm) FOV is attenuated, providing diagnostic image quality up to 250 mm FOV. The system can be operated in axial (partial or full scan), cine, cardiac and ECG gated modes. The SpotLight CT is a powerful, volumetric, high resolution CT that is designed to provide whole organ coverage, high image quality and uncompromised dose performance with the following features: - 140 mm detector coverage; - 120 msec temporal resolution (at 0.24 second rotation speed); ● - 0.31 mm spatial resolution: . - 250 mm diagnostic FOV: ● - Reduced cone beam artifacts. 0 The SpotLight CT features two "Gemini" X-ray tubes to provide two X-ray sources displaced in the z axis (parallel to rotation axis). The sources are alternating using electrostatic grid control, such that the scanned subject is irradiated alternatively by two {5}------------------------------------------------ overlapping cone X-ray beams while the gantry rotates and attenuation data is acquired by a single array detector. Image reconstruction is accomplished by a Stereo CT reconstruction algorithm based on common algorithms used in single source scanners that are modified to combine the data acquired from the two sources. The detector array comprises of three regions, center part covering FOV 250 mm with high pitch detector array and two peripheral parts with detector arrays. The data acquired in the peripheral parts is useful for patient positioning (scout scan) and for elimination of truncation artifacts from the images. The detector array features antiscatter collimator (post patient) to reduce scatter artifacts, ensure CT number uniformity and reduce beam hardening artifacts. The gantry features a short source to axis distance (SAD) of 450 mm, resulting in higher usable X-ray flux at a given source voltage and current as compared to scanners with longer SAD. ### DEVICE TESTING The SpotLight CT has completed testing and is in compliance with AAMI/ES 60601-1. IEC 60601-1 Ed. 3 and its associated collateral standards and particular standards, 21CFR Subchapter J and NEMA XR-25, XR-28 and XR-29. Testing included: - Testing on unit level; ● - Integration testing (system verification); 0 - Performance testing (verification): o - Safety testing (verification); ● - o Simulated use testing (validation); - . Animal testing (validation); - o Clinical testing (validation). The device was developed under a rigorous quality system and has successfully completed design control activities, including risk management, verification and validation. # Non-clinical Performance Testing The performance evaluation used a variety of test methods, phantoms and scan conditions. Various mathematical, physics and statistical analyses were performed to demonstrate that performance specifications are met. The tests evaluated image quality, temporal resolution and dose performance. Image quality evaluation included evaluation of artifacts, spatial resolution, low contrast detectability, noise, and uniformity and CT number accuracy. {6}------------------------------------------------ # Animal Testing The scanner was tested on a porcine model with and without contrast injection, with and without ECG gating at different heart rates. The trial included three cardiac CTA studies under different conditions, abdomen (renal) scan and neck (carotids) CTA. The images were evaluated for diagnostic quality with positive results. ### Clinical Testing Clinical data were collected on 38 subjects at one site with the approval of appropriate ethics committee and in accordance with applicable regulatory requirements as well as Arineta quality system procedures. The study protocol was designed to test the scanner across different patient populations, clinical scenarios and scan techniques. Image data sets included: - . Cardiac - Coronary CTA, cardiac function, calcium burden, preparation for TAVI - Body Abdomen (at a limited field of view), spine ● - Head brain, inner ear 0 The images were evaluated and rated by four US certified readers who are qualified radiologists or cardiologists for clinical diagnostic value and image quality. The results of this clinical testing demonstrate the diagnostic image quality performance of the SpotLight CT, as well as the safety and efficacy of the device. # SUBSTANTIAL EQUIVALENCE Based on the extensive testing as described above, the SpotLight CT is as safe and effective as the Revolution CT. The SpotLight CT has the same intended uses and similar indications, technological characteristics, energy type and principles of operation as its predicate device. The technological differences between the SpotLight CT and its predicate devices raise no new issues of safety or effectiveness. Thus, the SpotLight CT is substantially equivalent. #### CONCLUSION Based on the conformance to standards, as well as the bench and clinical testing provided, Arineta Ltd. believes that the SpotLight CT is substantially equivalent to the predicate device, GEHC Revolution CT (K133705).
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%