The UC-CARE NaviGo Workstation is an adjunctive tool for ultrasound guided procedures and is intended to be used by physicians in the clinic or hospital for 2-D and 3-D visualization of ultrasound images of the prostate gland. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the other. Additional software features include patient data management, multi-planar reconstruction, segmentation, image measurement and 3-D image registration. The device is specifically indicated to provide information within the prostate to assist needle targeted procedures (e.g. biopsy) regions of interest display, procedures planning and, reconstruction of a 3D rendered surface model of the prostate display needle locations that have been selected by the physician, as well as storage and future retrieval of this information.
Device Story
NaviGo Workstation is an adjunctive tool for prostate diagnostic and interventional procedures. It inputs 2D ultrasound video signals and pre-operative DICOM images (e.g., MRI). The system tracks the ultrasound probe position (via electromagnetic tracking) to reconstruct 3D prostate models and fuse MRI data with real-time ultrasound. It displays needle trajectories, regions of interest (ROI), and 3D models on a workstation screen. Used by physicians in clinics or hospitals, the device assists in procedure planning and needle targeting. It provides real-time visual feedback of needle advancement relative to ROIs. Benefits include improved regional orientation, accurate needle placement, and integrated data management for biopsy procedures.
Clinical Evidence
Bench testing only. Fusion accuracy evaluated using 4 phantoms with 12 isoechoic lesions; 34 ultrasound-guided biopsies performed. Results showed 100% hit rate with average error of 1.32±0.84mm. Illustration model performance compared 10 prostates (Fast vs. full contour models), showing average RMS error of 3.7±0.66mm and volume difference of 10.4%±5.15%.
Technological Characteristics
Workstation-based system; connects to standard trans-rectal ultrasound video output. Uses electromagnetic tracking (cube sensor) for probe localization. Supports side-fire and end-fire ultrasound probes. Software performs DICOM fusion, 3D surface reconstruction, segmentation, and image measurement. Connectivity via DICOM interface. No specific materials or sterilization methods listed as the device is a workstation.
Indications for Use
Indicated for physicians in clinical or hospital settings to assist in needle-targeted prostate procedures (e.g., biopsy) via 2D/3D ultrasound visualization, DICOM/MRI fusion, and 3D prostate modeling.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 6, 2017
UC-CARE Ltd. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW WASHINGTON DC 20004
Re: K160934
Trade/Device Name: NaviGo™ Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 10, 2017 Received: January 10, 2017
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160934
Device Name NaviGo™ Workstation
#### Indications for Use (Describe)
The UC-CARE NaviGo Workstation is an adjunctive tool for ultrasound guided procedures and is intended to be used by physicians in the clinic or hospital for 2-D and 3-D visualization of ultrasound images of the prostate gland. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the other. Additional software features include patient data management, multi-planar reconstruction, segmentation, image measurement and 3-D image registration.
The device is specifically indicated to provide information within the prostate to assist needle targeted procedures (e.g. biopsy) regions of interest display, procedures planning and, reconstruction of a 3D rendered surface model of the prostate display needle locations that have been selected by the physician, as well as storage and future retrieval of this information.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:flex; align-items:center;"><span><svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"><path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path><path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.75.75 0 0 1 1.08-.022z" fill-rule="evenodd"></path></svg></span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | |
| <div style="display:flex; align-items:center;"><span><svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"><path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1H2zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path></svg></span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo is in blue, with the exception of the "A" in CARE, which is in green. Above the text are five blue ovals that are progressively lighter in color.
# 510(k) SUMMARY
NaviGo™ Workstation
# 510(k) Number K160934
# Applicant's Name:
UC-CARE Ltd. Apollo Bldg., New Industrial Zone P.O.B 67 Yokneam 20692, Israel Tel: +972-4-909-7427 ext. 656 Fax: +972-4-993-7323 Email: tomers@uc-care.com Contact Person: Tomer Schatzberger
# Contact Person:
Stephen Bellomo H.M.Da Consulting and Development Ltd. 40 Geulim Stree Zichron-Yaakov 3091847 Israel Tel: +972-54-227-4831 Email: stephen.bellomo@gmail.com
# Date Prepared:
January 10, 2017
# Trade Name:
NaviGo™ Workstation
# Classification Name:
Picture archiving and communications system.
# Classification:
Picture archiving and communications systems have been classified as Class II devices, product code: LLZ, Regulation Number: 892.2050, reviewed by the Division of Radiology.
# Predicate Devices:
Primary: NaviGo™ Workstation (UC-CARE Ltd.), cleared under K100784, concurrence date November 1, 2011.
Secondary: Urostation - 3D Prostate Suite (Koelis), cleared under K131448, concurrence date July 26, 2013.
UC-CARE Ltd.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo is in blue, with the exception of the 'A' in CARE, which is in green. Above the text are five blue ovals that are progressively lighter in color from left to right.
### Performance Standards:
No performance standards have been established for such a device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
# Device Description:
The NaviGo™ Workstation is an adjunctive tool in the management of prostate diagnostic and interventional procedures. The NaviGo™ Workstation allows prostate needle tracking, recording, and management solution. The NaviGo™ Workstation is designed to assist the physician in performing prostate diagnostic and interventional procedures by providing regional orientation information, displaying a 3D model with real-time tracking and recording of the needle location. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the . The device includes means to compensate for patient body and prostate motion at any time during the procedure.
The NaviGo™ Workstation is designed to work with standard trans-rectal ultrasound systems and biopsy setup without changing or interfering with the physician's existing workflow. The NaviGo™ Workstation connects to the video output of the ultrasound system and by tracking the ultrasound probe's position, the recorded 2D ultrasound images are transferred to the NaviGo™ Workstation for viewing and creation of a 3D model. As with any other procedure, the TRUS probe is used together with standard disposable cover sheath supplied by the user.
Two dimensional (2D) images and the 3D model of the prostate are displayed on the NaviGo™ Workstation's screen. The NaviGo™ Workstation is equipped with tools to manipulate (rotate, pan, zoom) the model, and to archive and retrieve the information for further use.
The tracking and recording enables display of an accurate 3D model of the prostate and to record needle locations on the model. Pathology diagnosis results may be updated on the 3D model and a color display representation provides a visual display of the biopsy results.
The NaviGo™ Workstation supports the display of ROIs on the 3D model and displays visual indication when the needle trajectory intersects with an ROI (region of interest) is defined by the physician by segmenting a portion of the prostate on a 2D image and displaying its location on the 3D model, thereby defining a portion of the prostate as a target to direct a needle within. The 2D model for segmentation of the ROI can be either an Ultrasound image frozen during the procedure or a DICOM complaint image from another imaging study completed prior to the Ultrasound procedure (e.g. MR) when selecting the fusion procedure to display the planned targets. The ROIs are clearly numbered and labeled by a letter symbolizing it origin (Ultrasound or DICOM), On each image, more than one ROI may be defined (segmented).
The live 2D ultrasound image is superimposed in real-time with the 2D ROI boundaries thus allowing the physician to see the real-time advancement of the needle on the live ultrasound image with reference to ROI boundaries. The 3D model is displayed on a separate window alongside the ultrasound's live continuous images display, allowing prostate 3D orientation.
Reqardless of the type of ultrasound probe used for the procedure (side-fire or end-fire) a cannula for a needle is built within the outer shell of the probe and therefore a needle can be inserted through the designated cannula. The designated cannula defines the needle path along its trajectory. The needle trajectory is displayed on the 3D model representing the potential needle route to allow the physician to direct the needle in real time into the target. When the projected path
#### UC-CARE Ltd.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo is primarily blue, with the letters "UC" and "CARE" in a bold, sans-serif font. The "A" in "CARE" is colored green. Above the text are a series of blue ovals that appear to be fading from dark to light.
intersects with an RQL the color of the ROI will change indicating the needle can be directed inside the ROI if the physician ensures that the needle is inserted deep enough.
During the procedure, the physician may change some of the parameters on the US system or perform different actions that require the NaviGo™ Workstation to adjust. A physician may wish to freeze an image and perform measurements, label the image and save it to the report, switch between transversal and longitudinal view, or change the view of the ultrasound image. The NaviGo™ Workstation is equipped with image state algorithms to automatically detect the change in parameters and adjust itself to the new parameters. For example, if the physician freezes the US image, the NaviGo will automatically present additional menu option for a frozen image like measurements tools and labeling. The image state algorithm was developed to support generic ultrasound parameters and is specifically tested for each new ultrasound the NaviGo™ is required to support.
In off-line mode, the NaviGo™ software further enables the physician to analyze previous procedures using the prostate model, update the 3D model if necessary, update the recorded biopsies' locations, generate report, and provides a DICOM (e.g. MRI) interface for ROI definition, 3D model display, and data communication.
# Intended Use:
The UC-CARE NaviGo Workstation is an adjunctive tool for ultrasound guided procedures and is intended to be used by physicians in the clinic or hospital for 2-D and 3-D visualization of ultrasound images of the prostate gland. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the other. Additional software features include patient data management, multiplanar reconstruction, segmentation, image measurement and 3-D image registration.
The device is specifically indicated to provide information on regional orientation within the prostate to assist needle targeted procedures (e.g. biopsy) regions of interest display, procedures planning and, reconstruction of a 3D rendered surface model of the prostate display needle locations that have been selected by the physician, as well as storage and future retrieval of this information.
# Comparison of Technological Characteristics:
The subject NaviGo Workstation device and the primary predicate NaviGo Workstation device have the same principles of operation and basic design. Both devices have the same following technological characteristics:
- assist the physician by transfer and display of ultrasound images on the workstation screen
- provide regional orientation information during prostate biopsy procedures ●
- build display and manipulate a 3D model of the prostate on screen ●
- archive procedure data and report generating
- provide data management solutions ●
- track, display and record the biopsy needle trajectory location retrieved from the ultrasound probe
- . display the scanning history, including pathology analyses
Several modifications have been made from this predicate device, including the following: addition of illustration model (side-fire first procedure only), needle-fire recognition, image-state correction offline configuration, and end-fire ultrasound probe support. These modifications are considered minor and do not raise new issues of safety and efficacy.
#### UC-CARE Ltd.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo has a series of blue ovals in the top portion of the image. The text "UC-CARE" is in blue, with the "C" in "CARE" being green. Below this text is the text "Medical Systems" in blue.
Fusion functionality has been added to the subject NaviGo device, including fusion of DICOM originated information, ROI display, and procedure planning. The indications for use statement of the subject device has been modified from the predicate device to include the addition of fusion functionality. The fusion functionality of the subject NaviGo device is substantially equivalent to the fusion functionality of the secondary predicate Urostation device, including the stated accuracy of the system. Therefore, the addition of fusion functionality does not raise any safety or effectiveness concern in comparison with the cleared device.
| Product | NaviGo (K100784)<br>[Primary Predicate<br>Device] | Urostation<br>(K131448)<br>[Secondary Predicate<br>Device] | Subject NaviGo<br>Device<br>(K160934) |
|----------------------------|-------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Product code | LLZ | LLZ | LLZ |
| Class | II | II | II |
| Device Classification Name | System, Image<br>Processing,<br>Radiological | System, Image<br>Processing,<br>Radiological | System, Image<br>Processing,<br>Radiological |
| Visualization Means | Side-fire Ultrasound | End-fire Ultrasound | Side -fire & End-fire<br>Ultrasound |
| Supported Modalities | Ultrasound | Ultrasound, DICOM<br>images (e.g. MRI) | Ultrasound, DICOM<br>images (e.g. MRI) |
| Accuracy | +/-5mm | +/-3mm | +/-3mm |
| Segmentation (Contouring) | Yes | Yes, only for fusion | Yes |
| 3D model reconstruction | Yes, contouring only | Yes, only for fusion | Yes, illustration &<br>contouring |
| Needle trajectory display | Real-time | Upon request | Real-time |
| Tracking technology | Electro-magnetic<br>(mattress and cube) | Image processing | Electro-magnetic<br>(cube only) |
| Offline | Post procedure data<br>analysis, pathology<br>results | Post procedure data<br>analysis, pathology<br>results, planning (on<br>DICOM images) | Post procedure data<br>analysis, pathology<br>results, planning (on<br>DICOM images) |
| Regions of interest | No | Originated from<br>DICOM images (e.g.<br>MRI) | Originated from US<br>images + DICOM<br>images (e.g. MRI) |
| Registration | US - US | US - US, and MR - US<br>(fusion) | US - US, and MR - US<br>(fusion) |
| DICOM connectivity | No | Yes | Yes |
# Substantial Equivalence Comparison Table:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for UC-CARE Medical Systems. The logo consists of the text "UC-CARE" in a bold, blue font, with the "A" in "CARE" being green. Above the text are five blue ovals, arranged horizontally and decreasing in size from left to right. Below the main text, "Medical Systems" is written in a smaller, blue, italicized font.
### Non-clinical Performance Data:
The performance characteristics and operation/usability of the modified NaviGo™ Workstation were evaluated in the following non-clinical (bench) testing.
# Fusion accuracy performance test:
The accuracy of targeting DICOM (MRI) derived regions of interest (ROI) by using the NaviGo Workstation was evaluated. . Four phantoms each with 3 isoechoic lesions (CIRS Model 778-05, containing 3 randomly located isoechoic lesions of 0.5 cc) were targeted 2-3 times each using the NaviGo MRI fusion algorithm. Two phantoms were targeted using side fire ultrasound probes and two phantoms were targeted using end fire ultrasound probes. Needle location were defined relative to the ROI center line and accuracy is defined as the difference in needle location between the needle trajectory (targeted location in lesion) as marked on the real time US image, to the actual result as seen in post procedure CT scan. A total of 34 ultrasound guided biopsies were targeted into 12 isoechoic lesions. All 34 biopsies (100%) successfully hit the target lesion with an average error of 1.32±0.84mm. 32.3% of the biopsies had accuracy of <1mm, 47% of the biopsies had an accuracy of 1 to 2mm, 20.6% of the biopsies had an accuracy of 2 to 3mm, and no biopsies had an error larger than 3mm. .
# Illustration model performance test:
The accuracy of the illustrated (Fast) 3D prostate model in comparison to the full contours 3D prostate model was evaluated. 10 prostates were scanned using the NaviGo Workstation, and for each prostate a Fast 3D model and a full contours 3D model was created. Each pair of 3D models was compared for volume difference and root mean square error (representing the average distance between the surfaces). The prostate shape of the Fast model was found to be very similar to the prostate shape of the full contour model, with an average RMS error of 3.7±0.66mm between the surfaces. Volume calculations were also similar, with an average difference of 10.4%±5.15%. These results demonstrate the ability of the Fast model to provide the physician with a tool which requires minimum effort but still maintains good similarity to the full contour model.
# Substantial Equivalence Conclusion:
In summary, the NaviGo™ Workstation is substantially equivalent to the previously cleared NaviGo™ Workstation (K100784) primary predicate device in all functionality except for fusion functionality. Modifications from the primary predicate device are minor and do not raise new issues of safety and efficacy. The fusion functionality of the subject NaviGo device is substantially equivalent to the fusion functionality of the secondary predicate Urostation device, including the stated accuracy of the system. Therefore, the addition of fusion functionality does not raise any safety or effectiveness concern in comparison with the cleared device. In summary, the subject NaviGo™ Workstation described in this submission is substantially equivalent to the NaviGo™ Workstation (K100784) predicate device and Urostation (K131448) predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.